The SeQual Model T2000 Series, MESA II Oxygen Concentrator, is intended for the administration of supplemental oxygen up to 10 LPM. The device is not intended for life support nor does it provide any patient monitoring capabilities. The device has no contraindications.
Device Story
MESA II Oxygen Concentrator (T2000 Series) is a 0.5–10.0 LPM continuous flow pressure swing adsorption (PSA) system. Inputs: ambient air. Transformation: 12-column PSA process using synthetic zeolite molecular sieve beds; rotary valve driven by gear motor; air compressor; heat exchanger. Outputs: concentrated oxygen. Used in clinical or home settings; operated by patients or caregivers. System includes inlet/outlet filtration (HEPA, glass microfiber), electronic flow control, and audible/visual alarms for power, compressor malfunction, abnormal flow, low oxygen concentration, and over-temperature. Healthcare providers use output to deliver supplemental oxygen therapy. Benefits: provides consistent oxygen supply for patients with respiratory needs.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 8359 and ASTM 1464 standards to verify oxygen concentration output across flow rates (0.5–10 LPM) and alarm functionality. No clinical data presented.
Technological Characteristics
12-column pressure swing adsorption (PSA) system. Materials: synthetic zeolite molecular sieve beds, glass microfiber/HEPA filters. Energy: 115VAC 60Hz. Dimensions: 580x400x500mm. Connectivity: RS-232 port for technician diagnostics. Control: PCBA with discrete logic and analog circuitry. Sterilization: N/A (not sterile).
Indications for Use
Indicated for patients requiring supplemental oxygen therapy up to 10 LPM. Not for life-supporting or life-sustaining applications. No patient monitoring capabilities. No contraindications.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Predicate Devices
INTEGRA E-Z Oxygen Concentrators Model 63230M-10 (K042262)
{0}------------------------------------------------
# 510(k) Summary
#### Submitter Information:
K100920
Submitter:
SeQual Technologies Co., Ltd 6F-7, No.136, Sec.3, Ren-ai Rd. 10657, Taipei, Taiwan (R.O.C.)
NOV 2 9 2010
Contact:
Cearo Huang, Director, Engineering & Research Division
Phone: Fax:
886-2-2700-2737 886-2-2707-9253
February 5, 2010 Date of Summary:
Device Name:
| Proprietary Name: | MESA II Oxygen Concentrator, Model T2000 Series |
|---------------------------|----------------------------------------------------|
| Common Name: | Oxygen Concentrator |
| Classification of Device: | Generator, Oxygen, Portable as per 21 CFR 868.5440 |
#### Predicate Device Equivalence:
SeQual Technologies Co., Ltd is claiming substantial equivalence to the following legally marketed predicate devices: K042262 – INTEGRA E-Z Oxygen Concentrators Model 63230M-10
K043006 - Respironics Millennium M10 Oxygen Concentrator
#### Description of Device:
The SeQual Model T2000 Series, MESA II Oxygen Concentrator, is a 0.5 to 10.0 Liter per minute (LPM) continuous flow pressure swing adsorption (PSA) type system that produces oxygen.
The SeQual Model T2000 Series, MESA II Oxygen Concentrator, consists of pneumatic and electrical components. The system has inlet filtration, air compressors, heat exchanger, and Synthetic Zeolite molecular sieve beds with a rotary valve, outlet filtration, electronic flow control and audible / visual alarms.
{1}------------------------------------------------
# Predicate Product Comparison Table:
| Device Name | MESA II (10LPM) | INTEGRA E-Z (10LPM) | RESPIRONICS M10<br>Concentrator |
|----------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Oxygen<br>Generating<br>System | 12-Column Pressure<br>Swing Adsorption | 12-Column Pressure Swing<br>Adsorption | Time Cycle/Pressure Swing<br>Adsorption |
| Valve Type | Single Rotary Valve<br>Driven by Gear Motor | Single Rotary Valve Driven<br>by Gear Motor | Exclusive Rotary Valve<br>System |
| Compressor | Thomas 2660CE48 | Thomas 2660CE48 | Thomas 2660CE54 |
| Filtration | Glass Microfiber, HEPA,<br>0.3µm at Outle | Glass Microfiber, HEPA,<br>0.3µm at Outlet | Final Bacteria Filter, HEPA |
| Control and<br>Communication<br>System | PCBA/Membrane<br>Switch/RS-232*** | PCBA/Membrane Switch | Communication<br>Port/Compensated Flow<br>Meter |
| Equipment Class<br>and Type | Class II / BF | Class II / BF | Class II / B |
| EMC<br>Compliance | 60601-1-2 | 60601-1-2 | 60601-1-2 |
| Dimension | | | |
| Height (mm) | 580 | 660 | 686 |
| Width (mm) | 400 | 370 | 483 |
| Depth (mm) | 500 | 500 | 330 |
| Volume (L) | 116.0 | 122.1 | 109.3 |
| Weight (kg) | 23 | 26.0 | 24.0 |
| Oxygen Concentration (OC)* | | | |
| 8-10 LPM | 91 ± 3 | 91 ± 3 | 92 ± 4 |
| 3-7 LPM | 94 ± 3 | 94 ± 3 | 94 ± 2 |
| 1-2 LPM | 94 ± 3 | 94 ± 3 | 92 ± 4 |
| Operation Condition | | | |
| Flow Rate (LPM) | 0.5~10.0 LPM by 0.5<br>LPM Increment | 0.5~10.0 LPM by 0.5 LPM<br>Increment | 1.0~10.0 LPM |
| Outlet Pressure<br>(psig) | 5.0~10.0 | 7.5~11.0 | 2.7 |
| Electrical Input | 115VAC +10% ~ -15%<br>60Hz | 115VAC +10% ~ -15%<br>60Hz ± 3Hz | 115VAC +10% ~ -15%<br>60HZ |
| Oxygen<br>Concentration<br>Monitor | PRO-WAVE Model<br>GAS400** | PRO-WAVE Model<br>GAS400** | Oxygen Percentage<br>Indicator, OPI, |
| Alarms | Power/Compressor<br>Malfunction/Abnormal<br>Flowrate /Low OC/Over<br>Temperature | Power/Compressor<br>Malfunction/Abnormal<br>Flowrate /Low OC/Over<br>Temperature | High/low pressure, low<br>oxygen, power failure, no<br>oxygen flow |
{2}------------------------------------------------
| Manufacturer | SeQual Technologies<br>Co., Ltd. | SeQual Technologies Inc. | Respironics Inc. |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | N/A | K042262 | K043006 |
| Intended Use | The SeQual MESA II<br>Oxygen Concentrator is<br>intended to provide<br>supplemental oxygen.<br>It<br>is not intended for life<br>supporting, or life<br>sustaining applications<br>nor does it provide any<br>patient monitoring<br>capabilities. | The SeQual Integra Oxygen<br>Concentrator is intended to<br>provide supplemental<br>oxygen. It is not intended<br>for life supporting, or life<br>sustaining applications nor<br>does it provide any patient<br>monitoring capabilities. | The Respironics M10<br>Concentrator is intended to<br>provide supplemental<br>oxygen to persons requiring<br>low flow oxygen therapy.<br>The device is not intended<br>to be life supporting nor<br>life sustaining. |
| Interface | A standard discrete logic<br>and analog circuitry<br>(PCBA system) is used<br>for all monitoring and<br>alarm functions. | A standard discrete logic<br>and analog circuitry (PCBA<br>system) is used for all<br>monitoring and alarm<br>functions. | Oxygen Percentage<br>Indicator(OPI) ultronically<br>measures the oxygen output<br>as purity indication. |
*: OC represents oxygen concentration.
**: Where equipped. Measure the output as an oxygen concentration monitor.
***: Where equipped. RS-232 communication channel provide technician detail system information from PCBA.
## Intended Use:
The SeQual Model T2000 Series, MESA II Oxygen Concentrator, is intended for the administration of supplemental oxygen up to 10 LPM. The device is not intended for life support nor does it provide any patient monitoring capabilities. The device has no contraindications.
### Technological Characteristics:
The SeQual Model T2000 Series, MESA II Oxygen Concentrator, operates comparably to the listed predicate devices. The technology employed to generate the oxygen is well established, and therefore, raise no new questions of safety and effectiveness.
### Performance Data:
Results of the oxygen concentraton testing to ISO 8359 and ASTM 1464 standards confirm the device meets specifications and is substantially equivalent to the predicate devices.
### Conclusion:
Based on the design, performance specifications, tests and intended use, the SeQual Model T2000 Series, MESA II Oxygen Concentrator is substantially equivalent to the currently marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized human figure with three heads, representing the department's focus on health and human services. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" which is arranged vertically along the left side of the logo.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Cearo Huang Director, Engineering & Research Division SeQual Technologies Company, Limited 6F-7, No.136, Sec.3, Ren-ai Rd.. 10657, Taipei, Taiwan (R.O.C.)
NOV 2 9 2010
Re: K100920 Trade/Device Name: MESA II Oxygen Concentrator Regulation Number: 21 CFR 868.5440 Regulation Name: Oxygen Concentrator Regulatory Class: II Product Code: CAW Dated: November 10, 2010 Received: November 12, 2010
Dear Mr. Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Huang
Please be advised that FDA s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh foc
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for TATUNG. The logo consists of a stylized symbol on the left and the word "TATUNG" in bold, sans-serif letters on the right. The symbol appears to be a stylized representation of a water droplet or flame, with concentric circles forming its shape. The text is in all caps and has a slightly distressed or textured appearance.
Image /page/5/Picture/1 description: The image shows the text "K100920" above the word "SEQUAL" with a logo to the left of the word. The logo appears to be a stylized image of water. The text is in a simple, sans-serif font and is black and white. The image is likely a logo or branding element for a company or product called "SEQUAL".
# Indications for Use
510(k) Number (if known):
NOV 2 9 2010
Device Name: MESA II Oxygen Concentrator Indications For Use:
The MESA II Oxygen Concentrator (Model No.: T2000 Series) is intended to provide supplemental oxygen. It is not intended for life supporting, or life sustaining applications nor does it provide any patient monitoring capabilities.
AND/OR Over-The-Counter Use _ Prescription Use X (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\$
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.