The Medicon Mogen Circumcision Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during circumcision of a male infant or child.
Device Story
The Medicon Mogen Circumcision Clamp is a reusable, stainless steel surgical instrument designed for male infant or child circumcision. The device consists of a clamp and a locking bar; it features a spot-welded lock screw to prevent loosening and chamfered edges to prevent injury. The clamp is manually operated by a surgeon in healthcare facilities (hospitals, clinics, surgical centers). Operation involves lifting a clamping lever and turning a flap to open the device, then placing the flap against the clamp to close and lock it. The jaws open to a maximum of 2.5 mm to avoid trapping the glans. The device is supplied non-sterile and requires autoclave steam sterilization by the user before each procedure. By compressing the foreskin, the clamp facilitates the circumcision procedure, providing a controlled mechanical method for tissue management.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical laboratory testing, including mechanical performance, material properties, and corrosion resistance testing, as the technology has a long history of clinical use.
Technological Characteristics
Stainless steel construction per ASTM F 899 and ISO 7153-1. Mechanical clamping mechanism with 2.5 mm maximum jaw opening. Corrosion resistant. Reusable; requires autoclave steam sterilization. Manual operation.
Indications for Use
Indicated for use in male infants or children undergoing circumcision to compress the foreskin.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a collection of medical instruments categorized by type. The categories include implants, instruments, endoscopes, laryngoscopes, and containers. The image also includes the text "K100916" at the top right.
MEDICON &G * Postlach 44 55 * D-78609 Tuttlinger
Medicon
Instrumente
the art of surgery
55. B-78500 Tit
# DEC - 6 2010
Traditional 510 (k) - HFX
### 2. 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
Medicon Mogen Circumcision Clamp
MEDICON eG
(As required by Section 807.92)
- 2.1 Submitter: (807.92 (a) (1))
MEDICON eG Gaensaecker 15 D-78532 Tuttlingen Germany
Tel. +49 7462 2009 0 Fax: +49 7462 200950 eMail: sales@medicon.de
- 2.2 Contact Person:
Joachim Schmid CEO +49 7462 2009 0 Tel.
- Date Summary Prepared: (807.92 (a) (1)) 2.3
December 1, 2010
Page 5 of 17
{1}------------------------------------------------
K100916
# MEDICON eG
# Traditional 510 (k) - HFX
#### 2.4 Device Names: (807.92 (a) (2))
| Proprietary Name: | Medicon Mogen Circumcision Clamp |
|----------------------|------------------------------------------------------------------------------------|
| Common name: | Mogen Clamp, Mogen Circumcision Clamp,<br>Medicon Mogen Clamp, Circumcision Clamp, |
| Classification Name: | Clamp, Circumcision |
| Medical Specialty: | Obstetrics/Gynecology |
| Product Code(s): | HFX |
| Regulation Number: | 884.4530 |
| Device Class: | 2 |
- 2.5 Reason for Submission: (807.81(2))
New Device.
- 2.6 Predicate Device: (807.92(a)(3))
- T. S. Medical Circumcision Clamp, K033403
#### 2.7 Device Description:
The Medicon Mogen Circumcision Clamp is a stainless steel, reusable circumcision clamp consisting out of multiple components, including clamp and locking bar. For safety reasons the lock screw of the clamp is fixed with a spot welding point in order not to become loose. For safety reasons the clamp opens to a maximum of 2.5 mm, in order not to trap the glans and the edges of closing area are chamfered and rounded inside and outside to avoid injuries
- 2.8 Intended Use: (807.92 (a) (5))
The Medicon Mogen Circumcision Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during circumcision of a male infant or child.
{2}------------------------------------------------
# K100916
# MEDICON eG
# Traditional 510 (k) - HFX
### 2.9 Description of Device Function:
Opening of the Medicon Mogen Circumcision Clamp:
Image /page/2/Picture/5 description: The image contains two objects labeled with numbers. Object 1 is on the left side of the image and has a curved shape with an arrow pointing to the left. Object 2 is on the right side of the image and has a more complex shape with multiple sections.
(1) By lifting the clamping lever and turning away of flap, the clamp can be opened (2);
Closing and locking of the Medicon Mogen Circumcision Clamp:
Image /page/2/Picture/8 description: The image shows a black and white picture with two numbers, 3 and 4, with arrows pointing to different parts of the image. The number 3 is on the left side of the image, and the arrow points to the left side of the object. The number 4 is on the top right side of the image, and the arrow points to the top of the object.
(2) By putting the flap against the clamp, the clamp can be closed (4) and locked with the lamping lever;
{3}------------------------------------------------
K100916
## MEDICON eG
## Traditional 510 (k) - HFX
# 2.10 Environment of Use:
The Medicon Mogen Circumcision Clamp is intended for use in healthcare facilities, including hospitals, medical clinics and surgical centers, by surgeons of proper training and experience.
#### 2.11 Sterility Information:
The Medicon Mogen Circumcision Clamp is offered non-sterile for autoclave steam sterilization by user.
# 2.12 Comparison with SE Device: (807.92 (a) (6))
Based on the information provided in the predicate 510(k) summary, the predicate device is identical to the Medicon Mogen Circumcision Clamp in indications for use, operating principle, design and material. The Medicon Mogen Circumcision Clamp technical information, intended use and performance information, as provided in this premarket notification, shows the subject device to be as safe and effective as the current legally marketed predicate device.
# 2.13 Summary of physical and performance characteristics
The Medicon Mogen Circumcision Clamp is made out of stainless steel. The stainless steel used is in accordance with the standards ASTM F 899 and ISO 7153-1. All components of the clamp are corrosion resistant.
The Medicon Mogen Circumcision Clamp is supplied non-sterile but is intended to be reused. Decontamination, cleaning and moist heat sterilization has to be performed according to the general instructions described in the package insert prior to each use.
In the open position, the jaws are designed to separate to a maximum of 2.5 mm. For safe clamping the Medicon Mogen Clamp is locked with a defined closing force.
For safety reasons the lock screw of the clamp is fixed with a spot welding point in order not to become loose.
{4}------------------------------------------------
K1008916
Traditional 510 (k) - HFX
# 2.14 Discussion of Non-Clinical Laboratory Tests
The following Non-Clinical Testing was performed:
- o Sterilization validation
- Performance testing mechanical, material, and corrosion resistance o testing
The Medicon Mogen Circumcision Clamp is identical to the Predicate Device in intended use and basic design. The materials used are the same or similar as that used to manufacture the Predicate Device. The Medicon Mogen Circumcision Clamp has been tested and complies with the Standards ASTM F 899-09 and ISO 7153-1 Standard Specification for Stainless Steels for Surgical Instruments and with ASTM F 1089-10 and ISO 13402 Standard Specifications for resistance against corrosion and autoclaving of Surgical Instruments. The test results demonstrate that mechanical performance characteristics, material properties, corrosion resistance and sterilizations of the device are substantially equivalence to the Predicate Device.
# 2.15 Discussion of Clinical Tests
Clinical Testing was not performed in order to establish substantial equivalence as this type of technology and procedure has a long history of clinical use.
# 2.16 Conclusions
The Medicon Mogen Circumcision Clamp is substantial equivalence to the Predicate Device and it is as safe and effective for its intended use as the Predicate Device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Joachim Schmid CEO Medicon eG Gansacker 15 78532 Tuttlingen GERMANY
DEC - 6 2010
Re: K100916
Trade/Device Name: Medicon Mogen Clamp
Regulation Number: 21 CFR 8884.4530
Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HFX Dated: November 8, 2010 Received: November 16, 2010
Dear Mr. Schmid:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies: You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{6}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Lehman MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
## Enclosure
{7}------------------------------------------------
# INDICATIONS FOR USE
DEC - 6 2010
510(k) Number (if known):
100916
Device Name: Medicon Mogen Circumcision Clamp
Indications For Use:
The Medicon Mogen Circumcision Clamp is an instrument used in a medical procedure to compress the foreskin of the penis during circumcision of a male infrant or child.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hula Semen
(Division Sign-Off) i wision of Reproductive, Abdominal and Faciliogical Devices 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.