LIMA-LIO SMR SHOULDER SYSTEM - HUMERAL STEMS, LIMA-LIO SMR SHOULDER SYSTEM - HUMERAL BODIES,
Applicant
Lima-Lto S.P.A.
Product Code
KWT · Orthopedic
Decision Date
Dec 10, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SMR Shoulder system is intended for partial or total primary shoulder joint replacement. The components are intended for use in cemented applications. Total or hemi-shoulder replacement is indicated for patients suffering from disability due to: - Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis; - Inflammatory degenerative joint disease such as rheumatoid arthritis; - Treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods
Device Story
SMR Shoulder System is a modular orthopedic prosthesis for primary shoulder joint replacement. Components include humeral stem, humeral body, adaptor tapers, humeral head, and glenoid component. Stems feature triple conicity for cemented fixation; bodies and stems couple via Morse-taper; adaptor tapers allow adjustment of rotation center and offset for joint stability. Humeral heads articulate with glenoid bone or UHMWPE glenoid component. Used in clinical settings by orthopedic surgeons for patients with degenerative joint disease or acute fractures. System provides anatomical reconstruction; modularity allows intraoperative adjustment of joint tension and alignment. Benefits include restored joint function and stability.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including static pull-out, torsional, and fatigue testing of modular connections, and static/fatigue testing of the glenoid component.
Technological Characteristics
Materials: Ti6Al4V (ISO 5832-3, ASTM F1472) for stems/bodies/adaptors; CoCrMo (ISO 5832-12, ASTM F1537) for humeral heads; UHMWPE (ISO 5834-2, ASTM F648) for glenoid components. Design: Modular humeral components with Morse-taper connections; cemented fixation. Features: Triple conicity stems, fluted proximal sections, polished surfaces for wear reduction. No software or electronic components.
Indications for Use
Indicated for patients with shoulder disability due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), or acute humeral head fractures untreatable by other fixation methods. For use in cemented partial or total primary shoulder arthroplasty.
Regulatory Classification
Identification
A shoulder joint metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings,”
(vii) F 1378-97 “Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
p 1/3
K100858
DEC 1 0 2010
# Summary of Safety and Effectiveness
Date: October 12, 2010
Manufacturer: Lima-Lto S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy
U.S. Contact Person: Cheryl Hastings Principal Consultant Phone: 574-527-4220
| Product | Product Code | Regulation and Classification Name |
|---------------------|--------------|----------------------------------------------------------------------------------------------|
| SMR Shoulder System | KWT | Shoulder joint metal/polymer non-constrained<br>cemented prosthesis per 21 CFR 888.3650 |
| | HSD | Shoulder joint humeral (hemi-shoulder) metallic<br>uncemented prosthesis per 21 CFR 888.3690 |
#### Device Description:
The SMR Shoulder System consists of a humeral stem, a humeral body, an adaptor taper, a humeral head and a glenoid component. Components are offered for cemented use for hemi or total shoulder joint arthroplasty, in primary trauma surgery.
Humeral stems are designed for cemented use and are characterized by an outline with a triple conicity to provide optimal fixation and fit independent of canal morphology. The stems are made from Ti6Al4V (ISO 5832-3, ASTM F1472). The surface in contact with cement is polished to reduce shear stresses on the cement mantle and avoid abrasion resulting from micromovements at the stem-cement interface. The distal part of the stems is characterized by a cylindrical cross-section while the proximal part is fluted to prevent rotation of the component relative to bone. The stems are provided with a male Morse taper in the proximal part for coupling with the humeral body.
Humeral bodies are made from Ti6Al4V and are coupled with the humeral stem via a female Morse-taper connection; a locking screw is provided to aid in initial mating of the stem / body assembly. Cylindrical marks are designed at the base of the Morse-taper to provide correct alignment of the eccentricity of the humeral head during surgery. A male Morsetaper connection is designed for the coupling between the humeral body and the humeral head by means of specific adaptor tapers: an angle of 45° between the axis of this Morsetaper and the axis of the stem gives the correct varus-valgus alignment to the joint. A system of holes on the external surface allows anatomical attachment of tuberosities using sutures or wires in trauma applications.
Traditional 510(k) - SMR Shoulder System October 12, 2010
Company Confidential Page 2 - Section 5
{1}------------------------------------------------
p. 2/3
K100858
Adaptor tapers (neutral and eccentrical with different heights), are made from Ti6Al4V. They allow coupling between the humeral body and the humeral head. These devices are designed to adjust the centre of rotation of the joint and provide the required offset to the humeral head to achieve the correct tensioning to the soft tissues, optimizing joint stability.
The humeral heads are made from CoCrMo (ISO 5832-12, ASTM F1537). They are intended to articulate with the glenoid bone in hemi-arthroplasty or with the glenoid component in total shoulder arthroplasty. The surface is polished to aid in the reduction of wear.
The glenoid components are manufactured from Ultra-High Molecular Weight Polyethylene (UHMWPE ISO 5834-2, ASTM F648). The articulating surface has a radius of curvature greater than the corresponding humeral head, which allows translation in the superior/inferior and anterior/posterior directions. The back surface of the component is spherical in geometry and has a single central peg which is inserted in the hole drilled in the glenoid cavity during surgery. The peg surface has three grooves to provide enhanced cement fixation.
#### Intended Use / Indications:
The SMR Shoulder system is intended for partial or total primary shoulder joint replacement. The components are intended for use in cemented applications.
Total or hemi-shoulder replacement is indicated for patients suffering from disability due to:
- . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
- Inflammatory degenerative joint disease such as rheumatoid arthritis;
- Treatment of acute fractures of the humeral head that cannot be treated with other � fracture fixation methods
#### Predicate Devices:
- Promos modular shoulder system (PlusOrthopedics, K063578) .
- . Anatomical shoulder system (Centerpulse, K030259)
- Modular shoulder system (Acumed, K992525) ●
#### Comparable Features to Predicate Device(s):
The SMR Shoulder System is similar to the predicate devices in terms of intended use, indications, design and materials. The SMR Shoulder System and the predicates are all intended for partial or total primary shoulder joint replacement. The SMR humeral stems are intended for cemented use only while some of the predicate humeral stems are intended for cemented or cementless use. The SMR and predicate glenoids are all intended for cemented
Traditional 510(k) - SMR Shoulder System October 12, 2010
Company Confidential Page 3 - Section 5
{2}------------------------------------------------
use only. The indications of all of the systems are similar with the exception that the SMR Shoulder System is not indicated for revision procedures.
Like the Promos Modular and Acumed Modular Shoulder Systems, the SMR Shoulder System provides modular humeral stem and humeral body components. The Anatomical Shoulder System is designed as a one-piece humeral stem / body component. The SMR Shoulder System and the Promos Modular Shoulder System include modular adaptor tapers for adjustment of the humeral head. The Anatomical Shoulder System and the Acumed Modular Shoulder System do not include adaptor tapers.
The components of the SMR Shoulder System are manufactured from the same or similar materials as the predicate devices. The SMR, Promos and Acumed humeral stems are manufactured from Ti6A14V while the Anatomical cemented humeral stems are manufactured from cast CoCrMo.
Non-Clinical Testing: The SMR Shoulder System has undergone static pull-out and/or torsional testing of all modular connections; fatigue testing to demonstrate both the strength of the humeral stem and the post-fatigue strength of the modular connections; static shear and torsional testing of the glenoid component; and fatigue testing of the glenoid component. All mechanical testing was done on worst case components or constructs. Where possible, standard test methods were used to allow comparison to testing of the predicate devices. The testing results demonstrated the device's ability to perform under expected clinical conditions.
Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the SMR Shoulder System to the predicate device(s).
Traditional 510(k) - SMR Shoulder System October 12, 2010
Company Confidential Page 4 - Section 5
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three swooping lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Lima-Lto- S.P.A % Hastings Regulatory Consulting, LLC. Ms. Cheryl Hastings Principal Consultant P.O. Box 696 Winona Lake, Indiana 46590
DEC 1 0 2010
Re: K100858
Trade/Device Name: SMR Shoulder System Regulation Number: 21 CFR 888.3650 Regulation Name: 21 CF R 80015860
Regulation Name: Shoulder joint metal/polymer non constrained cemented prothesis Regulatory Class: Class II Product Code: KWT, HSD Dated: November 9, 2010 Received: November 12, 2010
Dear Ms. Hastings:
We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 510(x) premaint is substantially equivalent (for the indications
referenced above and have determined the device is subsess marketed in interst for use stated in the enclosure) the arrested predicate devices marketed in interstate for use stated in the encrosuly manoned people to the Medical Device Amendments, or to commerce prior in May 28, 1970, the chance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval contraks approval provision and Cosment Act (Act) that to hot require approval controls provisions of the Act. The Act. The You may, therefore, market the device, belyev to the governments for among mishranding and general controls provisions of the frectined of the adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
and he If your device is classified (Sec aros). Existing major regulations affecting your device can be
may be subject to additional controls. may be subject to additional controls: Existing Mayor of 800 to 898. In addition, FDA may
found in the Code of Federal Regulations, Title 21, Forts 800 to 898. Register found in the Code of Federal Regarations, Four device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualite of a substitution with other requirements of the Act
that FDA has made a determination that your device and agencies. You must that FDA has made a delemination inal your are rederal agencies. You must or any Federal statutes and regulations and limited to: registration and listing (21
comply with all the Act's requirements, including, but not reporting of medical comply with all the Act Stequirements, and more reporting (teporting of medical
CFR Part 807); labeling (21 CFR Part 800); medical device requirements as CFR Part 807); labeling (21 CFR Part 807); incultur are repeative requirements as set
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as
{4}------------------------------------------------
Page 2 – Ms. Cheryl Hastings
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systems (QS) regulation (21-542 of the Act); 21 CFR 1000-1050.
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
ly of the specific advice of any and and and CDDLUGDRUOfficae(vem 115800 h go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/Offices/ucm115809.html for and go to http://www.fda.gov/AboulrDA/Cchicr.sOffices Office of Compliance. Also, please
the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. (21CFP P the Center for Devices and Radionogical realing by reference to premarket notification" (2) (CFR Patt note the regulation entitled, "Misoranding of reference to premaint of the MDR regulation (21 CFR Part 803), please go to
CFR Part 803), please go to http://www.ida.gov/wiodiatics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general information on your responsibility of the manager Division of Sman Manazones - 100 or at its Internet address (800) 038-2041 01 (301) 770-1100 et an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an an
Sincerely yours,
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): Unknown
Device Name: SMR Shoulder System
DEC 1 0 2010
Indications for Use:
## SMR Shoulder System Indications for Use
The SMR Shoulder system is intended for partial or total primary shoulder joint replacement. The components are intended for use in cemented applications.
Total or hemi-shoulder replacement is indicated for patients suffering from disability due to:
- Non-inflammatory degenerative joint disease including osteoarthritis and avascular . necrosis;
- . Inflammatory degenerative joint disease such as rheumatoid arthritis;
- . Treatment of acute fractures of the humeral head that cannot be treated with other fracture fixation methods
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for Mr. Melkerson
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
Page 1 of 1
K100858 510(k) Number
Traditional 510(k) - SMR Shoulder System March 15, 2010
Company Confidential Section 4 - Page 15 of 279
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.