PTA-PLUS PTA BALLOON CATHER 5 MM X 4CM, PTA-PLUS PTA BALLOON CATHETER 6MM X 4CM
Applicant
Hotspur Technologies
Product Code
DXE · Cardiovascular
Decision Date
Jul 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PTA-Plus PTA Balloon Catheter is indicated for use within synthetic arteriovenous dialysis fistulae to remove embolic material (thrombus/debris) and dilate stenosis for treatment of obstructive lesions.
Device Story
PTA-Plus PTA Balloon Catheter is a dual-function device for synthetic arteriovenous dialysis fistulae. It combines an inflatable non-compliant balloon for stenosis dilation with an integrated embolectomy coil for mechanical thrombus removal. The device is .035" guide-wire compatible and features a proprietary valve system allowing contrast injection. Operated by a clinician in a clinical setting, the device uses a proximal handle/cable to manually expand the embolectomy coil. By performing both de-clotting and angioplasty with a single catheter, the device simplifies the procedure for treating obstructive lesions, potentially reducing procedural time and patient trauma.
Clinical Evidence
No human clinical data provided. Evidence consists of biocompatibility testing (ISO MEM Elution, Rabbit Pyrogen, ASTM Hemolysis, C3a/SC5b-9, PTT, Thromboresistance, Sensitization, Acute Systemic Injection, Intracutaneous Reactivity, LAL Pyrogen) and in-vitro bench testing (balloon profile, burst pressure, compliance, fatigue, bond strength, torque, flow rate, clot removal, simulated use). In-vivo testing was conducted in an ovine model with surgically placed synthetic arteriovenous access grafts to evaluate vessel and organ response via histology compared to control devices.
Technological Characteristics
Non-compliant inflatable balloon; mechanical embolectomy coil; .035" guide-wire compatible; proprietary valve system for contrast injection; 55 cm usable length. Materials verified via biocompatibility standards (ISO 10993, ASTM). Sterilization method not specified. Device is a mechanical catheter; no software or electronic components.
Indications for Use
Indicated for patients with synthetic arteriovenous dialysis fistulae requiring removal of embolic material (thrombus/debris) and dilation of stenosis for treatment of obstructive lesions.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
K/0084/2
Hotspur Technologies, Inc. 510(k) Notification: PTA-Plus PTA Balloon Catheter CONFIDENTIAL
#### SECTION 5.0: 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
| A. Submitter Information | | |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Submitter's Name: | Hotspur Technologies, Inc. | JUL 12 2010 |
| Address: | 880 Maude Ave., Suite A | |
| | Mountain View, CA 94043 | |
| Telephone: | 650-969-3150 | |
| Fax: | 408-608-1597 | |
| Contact Person: | Eric Ankerud, Executive Vice President Clinical, Regulatory, Quality | |
| Date of Preparation: | July 6, 2010 | |
| B. Subject Device | | |
| Trade Name: | PTA-Plus PTA Balloon Catheter | |
| Common/Usual Name: | Balloon Catheter | |
| Classification Name: | Catheter, Angioplasty, Peripheral, Transluminal/Percutaneous<br>Catheter (21 CFR 870.1250, Product Code LIT)<br>Catheter, Embolectomy (21 CFR 870.5150, Product Code DXE) | |
| C. Predicate Device Name(s): | | |
| Trade Name(s): | Edwards Fogarty Adherent Clot Catheter, K901625<br>BARD Vaccess PTA Balloon Dilation Catheter, K073472 | |
### D. Device Description:
The proposed PTA-Plus PTA Balloon Catheter is designed for de-clotting and treating stenosis in synthetic dialysis grafts. It is an .035" guide-wire compatible, PTA balloon catheter with a proprietary valve system which allows injection of contrast and an embolectomy coil for mechanical removal of thrombus.
#### E. Intended Use:
The PTA-Plus PTA Balloon Catheter is indicated for use within synthetic arteriovenous dialysis fistulae to remove embolic material (thrombus/debris) and dilate stenosis for treatment of obstructive lesions.
## F. Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:
The proposed PTA-Plus PTA Balloon Catheter ("PTA-Plus") and the predicate Fogarty Adherent Clot Catheter have the same intended use. Both are indicated for removal of emboli from synthetic arteriovenous fistula. The predicate Fogarty Adherent Clot Catheter is also indicated for removal of emboli from native vessels in the arterial system. The proposed PTA-Plus and the predicate Bard Vaccess PTA Balloon Dilation Catheter have the same intended use. Both are indicated for treatment of obstructive lesions by dilation of the stenosis of synthetic arteriovenous dialysis fistula.
The proposed PTA-Plus PTA Catheter has similar technological characteristics as the predicate devices. The proposed PTA-Plus combines the features of the two predicate devices into one device. The proposed PTA-Plus and the Bard Vaccess PTA Balloon Dilation Catheter contain an inflatable non-compliant balloon for dilation of stenosis. The proposed PTA-Plus and the Fogarty Adherent Clot Catheter both have a coil which is expanded by manual control of the proximal handle/cable for embolectomy. The usable length of the proposed PTA-Plus PTA Balloon Catheter is 55 cm whereas the usable length of the
{1}------------------------------------------------
predicate Fogarty Adherent Clot Catheter is 80 cm and the predicate Bard Vaccess Balloon dilation Catheter is 50cm or 80cm. The proposed device and predicate devices are substantially equivalent in terms of intended use, fundamental scientific technology, target population, operating principles, and method of sterilization.
## G. Performance Data:
Biocompatibility testing on the proposed PTA-Plus PTA Balloon Catheter has been completed. The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and non-sensitizing to biological tissues consistent with its intended use. The following Biocompatibility tests were completed:
- ISO MEM Elution Assay .
- . Materials Mediated Rabbit Pyrogen
- . ASTM Hemolysis Assay
- Complement Activation C3a and SC5b-9 Assay .
- . Partial Thromboplastin Time
- . Four Hour Thromboresistance Evaluation
-
- ISO Guinea Pig Maximization Sensitization .
- ISO Acute Systemic Injection Test .
- ISO Intracutaneous Reactivity Test .
- Pyrogen (LAL) Chromogenic .
The proposed PTA-Plus PTA Balloon Catheter was tested in-vitro and in an in-vivo animal model to confirm the performance characteristics as compared to the predicate devices. The following in-vitro performance bench testing was completed for the PTA-Plus PTA Balloon Catheter:
- Balloon Crossing Profile ●
- . Catheter Shaft Diameter
- . Guidewire Lumen Diameter
- . Catheter Shaft Markings
- . Balloon Working Length
- . Balloon Rated Burst Pressure
- . Balloon Compliance/ Outer Diameter
- Balloon Inflation/Deflation Time .
- Balloon Fatigue �
- Catheter Body Burst Strength ●
- Catheter Bond and Tip Pull Strength .
- . Catheter Torque Strength
- Contrast Media Flow Rate .
- . Adherent Clot Removal
- Simulated Use/Flexibility/Kink .
- Radiopacity .
In-vivo testing for the PTA-Plus PTA Balloon Catheter was completed in accordance with 21 CFR Part 58 "Good Laboratory Practices for Nonclinical Laboratory Studies". Synthetic arteriovenous access grafts were surgically placed in ovine specimens and allowed to mature. Following graft maturation, a simulated procedure was performed on test and control comparator groups. Post procedure animals were survived and observed for a predetermined period. Post survival, vessel and organ histology was completed to compare vessel and organ response to the test and control device treatments.
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the PTA-Plus PTA Balloon Catheter met the established specifications necessary for consistent performance during its intended use. .
## H. Conclusions:
The PTA-Plus PTA Balloon Catheter met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance documents, test protocols, and/or customer inputs. The PTA-Plus PTA Balloon Catheter is substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Hotspur Technologies, Inc. c/o Mr. Eric Ankerud Executive Vice President, Clinical Regulatory and Quality 880 Maude Avenue, Suite A Mountain View, CA 94043
(JUL 1 2 2010
Re: K100842
PTA-Plus PTA Balloon Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE, LIT Dated: July 6, 2010 Received: July 7, 2010
Dear Mr. Ankerud:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. However, we remind you that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Eric Ankerud
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) {21 CFR 803}; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
R. beames
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### SECTION 4.0: INDICATIONS FOR USE STATEMENT
510(k) Number:
K100842
PTA-Plus PTA Balloon Catheter
Device Name:
Indication For Use:
The PTA-Plus PTA Balloon Catheter is indicated for use within synthetic arteriovenous dialysis fistulae to remove embolic material (thrombus/debris) and dilate stenosis for treatment of obstructive lesions.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Mirro ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ
(Division Sign-Off) (Division Olgh Olyn Olyn Scular Devices
510(k) Number_KIOO8AZ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.