← Product Code [FEQ](/productcode/FEQ) · K100803

# OLYMPUS FLUSHING PUMP OFP MODEL: OFP-1 (K100803)

_Keymed, Inc. · FEQ · Apr 22, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100803

## Device Facts

- **Applicant:** Keymed, Inc.
- **Product Code:** [FEQ](/productcode/FEQ.md)
- **Decision Date:** Apr 22, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Olympus Flushing Pump OFP is indicated for use during GI endoscopy, when blood, faeces or other organic matter obscure the endoscopic view, to wash the site being visualized.

## Device Story

Peristaltic pump; facilitates direct washing of debris from observation site during GI endoscopic examinations. System uses peristaltic pump and associated tubing to control and direct irrigation water to site; removes blood, feces, or organic matter obstructing endoscopic view. Used in clinical settings by medical professionals. Modification involves Auxiliary Water Irrigation Tube material change (Tygon to PVC) and transition to sterilized packaging. No changes to system hardware or other components.

## Clinical Evidence

Bench testing only. Validation included biocompatibility testing of PVC tubing, validation of gamma sterilization process, validation of irrigation tube packaging, and performance testing of the new irrigation tube.

## Technological Characteristics

Peristaltic pump system. Auxiliary Water Irrigation Tube material: Polyvinyl chloride (PVC), Class VI biocompatible. Sterilization: Gamma radiation. No changes to system hardware or energy source.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Olympus OFP ([K000948](/device/K000948.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K100803
PyelqZ

# 510(k) Summary (per 21 CFR 807.92) Olympus Flushing Pump OFP-1

#### 1. SPONSOR/APPLICANT

APR 2 2 2010

KeyMed (Medical & Industrial) Ltd. Stock Road Southend-On-Sea Essex SS2 5QH United Kingdom

## 2. CONSULTANT/CONTACT

Medical Device Consultants, Inc. 11440 West Bernardo Drive, Suite 300 San Diego, CA 92127 Telephone: 858-753-1961 Facsimile: 858-753-1962

Primary Contact: Ron Warren

## 3. DEVICE NAME

| Proprietary Name:    | Olympus Flushing Pump OFP-1        |
|----------------------|------------------------------------|
| Common/Usual Name:   | Endoscopic Flushing or Lavage Pump |
| Classification Name: | Endoscope and accessories          |

## 4. DEVICE CLASSIFICATION

An endoscope and accessories has been classified by the Gastroenterology/Urology Devices Panel as a Class II device per 21 CFR 876.1500, Product Code FEQ.

## 5. PREDICATE DEVICES

The predicate device is the Olympus OFP cleared on May 12, 2000 under K000948.

## 6. DEVICE DESCRIPTION

The Olympus Flushing Pump is peristaltic pump which facilitates direct washing of debris from the observation site during endoscopic examinations. The peristaltic pump

{1}------------------------------------------------

K100803
Page 2 of 2

and associated tubing allow irrigation water to be controlled and directed to the observation site to remove blood, faeces or other organic matter that may be obstructing the endoscopic view.

#### 7. INTENDED USE/INDICATIONS FOR USE

The Olympus Flushing Pump OFP is indicated for use during GI endoscopy, when blood, facces or other organic matter obscure the endoscopic view, to wash the site being visualized.

#### 8. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The subject of this Special 510(k) is a modification of the Olympus Flushing Pump OFP that was previously cleared on May 12, 2000 under Premarket Notification Number K000948. The subject of this Special 510(k) is a modification of the Auxiliary Water Irrigation Tube used in the OFP. The modifications are:

- A change in the component material of the tubing from Tygon (proprietary ● polyvinyl chloride) to polyvinyl chloride (PVC) - both materials are Class VI biocompatible.
- . Provision of the tubing in a sterilized package (previous tubing was supplied nonsterile)

No modification is being made to the OFP system hardware or other components. The change to the water irrigation tubing is being made to provide greater customer convenience by offering a sterilized irrigation ready for off-the-shelf use, and for manufacturing efficiencies by using a tubing material in use for other KeyMed systems.

## 9. PERFORMANCE TESTING

Validation activities to support the use of the modified Auxiliary Water Irrigation Tube consisted of four main elements:

- Biocompatibility testing of the PVC tubing o
- Validation of the Gamma Sterilization Process 0
- o Validation of the Irrigation Tube Packaging
- Performance testing of the new irrigation tube o

Testing of the modified irrigation tube has demonstrated that the OFP fulfills prospectively defined performance criteria and that the modified system meets user needs.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

KeyMed (Medical & Industrial) Ltd. % Mr. Ronald S. Warren, RAC Principal Consultant, Regulatory Services Medical Device Consultants, Inc. 11440 W. Bernardo Drive, Suite 300 SAN DIEGO CA 92127

APR 22 2010

Re: K100803

Trade/Device Name: Olympus Endoscopic Flushing Pump Model OFP-1 Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FEQ Dated: March 19, 2010 Received: March 24, 2010

Dear Mr. Warren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

100803 510(k) Number (if known):

Device Name: Olympus Endoscopic Flushing Pump Model OFP-1

Indications for Use:

The Olympus Flushing Pump OFP is indicated for use during GI endoscopy, when blood, faeces or other organic matter obscure the endoscopic view, to wash the site being visualized.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hulin

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K100803

---

**Source:** [https://fda-staging.innolitics.com/device/K100803](https://fda-staging.innolitics.com/device/K100803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).