FETAL MATERNAL MONITOR MODEL F9 EXPRESS AND EXPRESS
K100797 · Edan Instruments, Inc. · HGM · Nov 10, 2010 · Obstetrics/Gynecology
Device Facts
Record ID
K100797
Device Name
FETAL MATERNAL MONITOR MODEL F9 EXPRESS AND EXPRESS
Applicant
Edan Instruments, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Nov 10, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The F6 Express and F9 Express fetal & maternal monitor is intended for monitoring physiological parameters of pregnant women during ante-partum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms. F6 Express and F9 Express fetal & maternal monitor is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. In addition, it provides a solution for maternal vital signs monitoring.
Device Story
Fetal & Maternal Monitor (Models F6/F9 Express) monitors physiological parameters during pregnancy, labor, and delivery. Inputs include fetal heart rate (FHR) via 1 MHz PW ultrasound probes, uterine activity (TOCO), internal parameters (IUP, DECG), and maternal vital signs (ECG, SpO2, NIBP, temperature). Device processes signals to provide real-time monitoring and limit alarms. Used in clinical settings (antepartum, labor/delivery rooms) by trained personnel. Output displayed for clinician review to assess fetal well-being and maternal status, supporting clinical decision-making during labor and delivery.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including software, hardware, safety, and environmental testing, alongside risk analysis and final validation.
Technological Characteristics
Perinatal monitoring system using 1 MHz PW ultrasound transducers for FHR. Parameters include FHR, TOCO, IUP, DECG, MECG, SpO2, NIBP, and temperature. System is a multi-parameter monitor for clinical use. Software-based processing for signal analysis and alarm management.
Indications for Use
Indicated for pregnant women from 28th week of gestation for fetal monitoring, Non-Stress testing, and maternal vital signs monitoring during antepartum examination, labor, and delivery. Contraindicated for use in intensive care units, operating rooms, or home use.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
GE Healthcare GE129 Maternal/Fetal Monitor (K991739)
Submission Summary (Full Text)
{0}------------------------------------------------
#### Attachment 3
### 510(k) Summary
NOV 1 0 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### Date prepared: September 20, 2010
#### 1. Applicant:
Edan Instruments, Inc. 3/F - B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, shekou, Nanshan Shenzhen, 518067 P.R. China Contact person: Randy Jiang
#### 2. Manufacturer:
Edan Instruments, Inc. 3/F - B. Nanshan Medical Equipments Park, Nanhai Rd 1019#, shekou, Nanshan Shenzhen, 518067 P.R. China
#### 3. Submitter:
Mr. Jigar Shah Official Correspondent for Edan Instruments, Inc. mdi Consultants, Inc. 55 Northern Blvd., Suite 200 Great Neck, New York 11021 Tel: 516-482-9001 Fax: 516-482-0186 Jigar@mdiconsultants.com
#### 4. Trade/proprietary Name: Fetal & Maternal Monitor (Models F9 Express and F6 Express)
- 5. Classification name: 21 CFR 884.2740 Perinatal monitoring system and accessories.
- 6. Product Code: HGM
7. Predicate Devices:
- Edan Instruments F9/F6 Maternal/Fetal Monitor K082602 .
- GE Healthcare GE129 Maternal/Fetal Monitor K991739 .
#### 8. Device Description
The Fetal & Maternal Monitor provides the following primary features that can be available for the multiple configurations:
1
{1}------------------------------------------------
K100797 p. 2 of 3
- Basic parameters: FHR, TOCO, Event Mark, AFM .
- Dual FHR monitoring .
- Internal parameters: IUP, DECG .
- FHR limit alarm
Following facilities are also provided in addition to the above:
- Maternal ECG monitoring .
- Maternal SpO2 monitoring .
- Maternal NIBP .
- Maternal temperature monitoring .
#### 9. Intended Use:
The F9 Express and F6 Express fetal & maternal monitor is intended for monitoring physiological parameters of pregnant women during antepartum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
F9 Express and F6 Express fetal & maternal monitor is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. In addition, it provides a solution for maternal vital signs monitoring.
#### 10. Contraindications:
It is not intended for use in intensive care units, operating rooms or for home use.
| Monitoring mode | F6 & F9 Express<br>New Device | F6 & F9<br>K082602 | COROMETRICS 120<br>SERIES (129<br>configuration)K991739 |
|-----------------|-------------------------------|--------------------|---------------------------------------------------------|
| FHR/ Dual FHR | Yes | Yes | Yes |
| TOCO | Yes | Yes | Yes |
| Fetal ECG, IUP | Yes | Yes | Yes |
| MECG | Yes | | Yes |
| SpO2 | Yes | | Yes |
| NIBP | Yes | | Yes |
#### 11. Substantial Equivalence Discussion:
#### 12. Summary of Testing:
The following quality assurance measures were applied to the development of the Fetal & Maternal Monitor:
- . Software testing
- . Hardware testing
- Safety testing .
- Environment test .
- Risk analysis .
- Final validation .
2
{2}------------------------------------------------
## 13. Conclusion:
・
Verification and validation testing was done on the Fetal & Maternal Monitor. This premarket notification submission demonstrates that Fetal & Maternal Monitor is substantially equivalent to the predicate device.
KID0797
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread, facing to the right. The bird is composed of three curved lines that suggest the shape of the wings and body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Edan Instruments, Inc. c/o Mr. Jigar Shah Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 GREAT NECK NY 11021
NOV 1 0 2010
Re: K100797
Trade/Device Name: Fetal and Maternal Monitor (Models F6 Express and F9 Express) Regulation Number: 21 CFR \$884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: Class II Product Code: HGM Dated: November 5, 2010 Received: November 8, 2010
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Fetal and Maternal Monitor (Models F6 and F9), as described in your premarket notification:
> 1 MHz PW fetal probe - model F6 1 MHz PW fetal probe - model F9
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Jigah Shah
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Kathy Daws-Kopp, at (301) 796-6535.
Sincerely vours.
Hebert Semmler
Herbert P. Lerner, MD, Director (Acting) Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
# Indication for Use
NOV 1 0 2010
510(k) Number (if known):_K100797
Device Name: Fetal & Maternal Monitor (Models F6 Express and F9 Express)
F6 Express and F9 Express fetal & maternal monitor is intended for monitoring physiological parameters of pregnant women during ante-partum examination, labor and delivery. It is intended to be used only by trained and qualified personnel in antepartum examination rooms, labor and delivery rooms.
F6 Express and F9 Express fetal & maternal monitor is intended for providing Non-Stress testing or fetal monitoring for pregnant women from the 28th week of gestation. In addition, it provides a solution for maternal vital signs monitoring.
× Prescription Use (21 CFR Part 801 Subpart D)
And/Or
Over the Counter Use __ . (21 CFR Part 801 Subpart C)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Hula Leun
on of Reproductive, Gastro-Renal, and ogical Devices 0(k) Number
{6}------------------------------------------------
# Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 1 MHz PW fetal probe- model F6 Express
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | Mode Operation | | | | | | | | | |
|------------------------------|----------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | P | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E Additional Comments: _ The above is a 1 mHz PW transducer for the fetal heart rate detection.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
Hebert Senn
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices 510(k) Number .
{7}------------------------------------------------
# Diagnostic Ultrasound indications for Use Form Fill out one form for each ultrasound system and each transducer. 1 MHz PW fetal probe- model F9 Express
Intended use : Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | Mode Operation | | | | | | | | | |
|------------------------------|----------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | P | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative(specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ(specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheraln Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other | | | | | | | | | | |
N=new indication; P=previously cleared by FDA; e=ADDED UNDER appendix E
Additional Comments: __ The above is a 1 mHz PW transducer for the fetal heart rate detection.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE I FNEEDED CONCURRENCE OF cdrh, Office of Device Evaluation (ODE)
Heber Leum
( Division Sign-Off) Di ision of Reproductive, Gastro-Urological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.