UROSTATION 3D PROSTATE SUITE

K100793 · Koelis · LLZ · Sep 22, 2010 · Radiology

Device Facts

Record IDK100793
Device NameUROSTATION 3D PROSTATE SUITE
ApplicantKoelis
Product CodeLLZ · Radiology
Decision DateSep 22, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

UROSTATION - 3D PROSTATE SUITE is a computer-based software application intended to process, visualize and record 3D digital ultrasound images of the prostate.

Device Story

Urostation-3D Prostate Suite is a PC-based software application for processing, visualizing, and recording 3D digital ultrasound images of the prostate. Used by physicians in clinics or hospitals during transrectal prostate biopsy procedures. Inputs include 3D digital ultrasound images transferred from commercially available 3D ultrasound scanners via Ethernet. The device uses image fusion and display algorithms to provide a 3D representation of prostate biopsies, allowing intraoperative visualization of biopsy mapping relative to a reference prostate image. It also supports a review mode for mapping histologic results onto the 3D prostate model. The system is controlled via footswitch and manual input device. Outputs include 3D visualizations, biopsy cartography, and stored/printed patient data. The device assists physicians in tracking the organ during procedures and documenting biopsy locations, potentially improving the accuracy of prostate biopsy mapping.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use comparisons.

Technological Characteristics

PC-based software application running on standard Intel PCs with Microsoft Windows XP. Interfaces with 3D ultrasound scanners via Ethernet. Features include 3D image fusion, display, multi-planar reconstruction (MPR), and 3D image registration. No measurement features included.

Indications for Use

Indicated for use by physicians in clinic or hospital settings for 2D and 3D visualization of prostate ultrasound images. Includes patient data management, multi-planar reconstruction, and 3D image registration.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K100793 - Page 1 of 3 #### 510(k) SUMMARY 5. #### SEP 2222010 510(k) Summary for UROSTATION - 3D PROSTATE SUITE The 510(k) summary is submitted in accordance with the requirements of 21 CFR 807.92. | 510(k) Owner: | KOELIS<br>5, avenue du Grand Sablon 38700 La Tronche<br>FRANCE<br>Phone: +33 476 637 588 Fax: +33 476 549.561 | |---------------|---------------------------------------------------------------------------------------------------------------| | Contact Name | Cécile DESMULIE<br>Mail: cecile.desmulie@koelis.com | | Date Prepared | 03/17/2010 | #### Proposed Device: | Trade Name: | UROSTATION - 3D PROSTATE SUITE | |----------------------|---------------------------------------------------------------------------------------------------------------| | Common Name: | Medical Image Processing Software System | | Classification Name: | System, Image processing, Radiological<br>Picture archiving and communication system,<br>21 CFR PART 892.2050 | | Device Class | II | | Product Code | LLZ | #### Predicate Device: The Urostation system is substantially equivalent to: | 510(k) Number | K081093 | |---------------|-------------------------| | Device Name | 3-D Imaging Workstation | #### Intended Use: UROSTATION - 3D PROSTATE SUITE is a computer-based software application intended to process, visualize and record 3D digital ultrasound images of the prostate. #### Indications for Use: UROSTATION - 3D PROSTATE SUITE is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of ultrasound images of the prostate gland. Additional software features include patient data management, multi-planar reconstruction and 3D image registration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . {1}------------------------------------------------ K100793 - Page 2 of 3 #### Device Description: UROSTATION - 3D PROSTATE SUITE is a computer-based software application designed to process, visualize and record 3D digital ultrasound images of the prostate, and to manage patient and clinical data in the context of transrectal prostate biopsy. #### Hardware Platform and Operating System The application runs on standard Intel PCs under Microsoft Windows® XP operating system #### Peripheral and accessories The application is controlled by a footswitch and a manual input device. It is designed to work with commercially available 3D ultrasound scanner through Ethernet connection, 3D transrectal ultrasound probe and needle guide. #### Software Features UROSTATION - 3D PROSTATE SUITE implements image fusion and display algorithms to provide 3D representation of prostate biopsies. A typical workflow enables the physician to intraoperatively visualize the 3D mapping of biopsies with respect to a reference image of the patient's prostate. For that purpose, 3D digital images may be transferred at any time from the 3D ultrasound scanner to the Urostation for process and display, while the physician keeps track of the organ using the usual 2D live ultrasound mode. Alternatively, UROSTATION - 3D PROSTATE SUTE also provides a review mode that allows the mapping of histologic results on the said 3D representation of the patient's prostate. Patient information, images and 3D biopsy cartography may be stored or printed for future retrieval and examination. #### Technological Characteristics compared with the predicate device: UROSTATION - 3D PROSTATE SUITE utilizes the same technological characteristics as the predicate device. - Both are PC-based software applications that provide 2D/3D medical image acquisition and visualization for enhanced observation and analysis of the prostate gland. - Both devices provide in addition patient and clinical data management features. - - Both deal with ultrasound images received from commercially available imaging devices. UROSTATION - 3D PROSTATE SUITE also presents differences with the predicate device. These differences do not raise any safety or effectiveness concern. - The software architecture follows a linear workflow adapted to the physician's . practice. - The system implements a dedicated algorithm to visualize biopsies with respect to . a single reference image of the prostate. - The system has no measurement features. {2}------------------------------------------------ K100793 - Page 3 of ## Substantial Equivalence Comparison Chart | | KOELIS | EIGEN LLC | |----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------| | System: | 3D PROSTATE SUITE | 3-D Imaging Workstation | | 510(k) number: | | K081093 | | Function | • 2D/3D image<br>- acquisition<br>- viewing/reviewing<br>- processing<br>- storage<br>• Multi-Planar Reformatting (MPR)<br>• Patient and clinical data management | • Volume rendering<br>• Segmentation<br>• Measurements | | Function: | • Printing | | | Intended Use: | Process, visualize and record 3D digital ultrasound<br>images of the prostate | | | Data Source: | 3D TRUS scanners | | | Physical Characterization: | Software package<br>- operates on off the shelf hardware<br>- Windows operating system | | ### Conclusion: . . The results of comparing the intended uses, functions and technological characteristics of the UROSTATION - 3D PROSTATE SUITE with its predicate devices shows that the system is as safe and as effective as its predicate device. Therefore the UROSTATION - 3D PROSTATE SUITE is substantially equivalent to existing products currently on the market. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Mr. Cecile Desmulie Quality Manager KOELIS 5, avenue du Grand Sablon La Tronche 38700 FRANCE SEP 2 2 2010 Re: K100793 Trade Name: Urostation-3D Prostate Suite Regulation Number: 21 CFR \$892.2050 . Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: September 17, 2010 Received: September 20, 2010 Dear Mr. Desmulie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Donald J. Hett Donald J. St Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device · Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use # K10793 SEP 22 2010 510(k) Number (if known): Unknown Device Name: UROSTATION 3D PROSTATE SUITE Indications For Use: UROSTATION - 3D PROSTATE SUITE is intended to be used by physicians in the clinic or hospital for 2D and 3D visualization of ultrasound images of the prostate gland. Additional software features include patient data management, multi-planar reconstruction and 3D image registration. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) --- (Division Sign-Off) Page 1 of 1 Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety K100793 519K
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...