MICROTHERMX MICROWAVE ABLATION SYSTEM, MODELS MTX-180 OR MTX
K100786 · Bsd Medical Corp. · NEY · Aug 13, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100786
Device Name
MICROTHERMX MICROWAVE ABLATION SYSTEM, MODELS MTX-180 OR MTX
Applicant
Bsd Medical Corp.
Product Code
NEY · General, Plastic Surgery
Decision Date
Aug 13, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MicroThermX® Microwave Ablation System (MTX) delivers microwave energy for coagulation (ablation) of soft tissue. The system is not intended for use in cardiac procedures. The SynchroWave antennas may be used in open surgical as well as percutaneous ablation procedures. An optional temperature sensor may be used to monitor tissue temperatures.
Device Story
MicroThermX (MTX-180) is a microwave ablation system for soft tissue coagulation. System comprises a mobile generator, microwave antennas, and a closed-circuit saline cooling system. Operator sets time and power parameters via touchscreen; generator delivers 915 MHz microwave energy to up to 3 antennas simultaneously. Optional thermistor-based temperature sensors monitor non-target tissue. Used in intraoperative or percutaneous procedures by physicians. Cooling system prevents antenna overheating. Device provides repeatable ablation zone geometries; synchronous antenna operation enables larger ablation zones. Output allows physician to achieve controlled tissue necrosis; benefits include minimally invasive treatment options for soft tissue lesions.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing and ex-vivo studies in non-perfused animal tissue to demonstrate repeatable ablation zone geometries and system performance. Conformance to safety standards (UL 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-2, IEC 60601-2-6, CISPR 11) and internal requirements for power control, alarm function, cooling circuit integrity, and temperature measurement accuracy was verified.
Technological Characteristics
915 MHz microwave energy delivery; thermistor-based temperature monitoring. System includes mobile generator, peristaltic pump for saline cooling, and microwave antennas. Power output: 180W total, 60W per antenna. Connectivity: touchscreen interface. Standards: UL 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-2, IEC 60601-2-6, CISPR 11. Single-use sterile temperature sensors. Reusable antennas with fluid pathway channels.
Indications for Use
Indicated for coagulation (ablation) of soft tissue in open surgical or percutaneous procedures. Contraindicated for cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
MicroThermX 100 Microwave Ablation System (MTX-100) (K081042)
Valleylab VivaWave Microwave Ablation System (K053535)
Vivant Medical VivaTip Microwave Ablation Probe and Accessories (K032702)
Submission Summary (Full Text)
{0}------------------------------------------------
April 29, 2010 Traditional 510(k) Premarket Notification
K100786
Page 1 of (i)
# SECTION 5: 510(k) SUMMARY
## SPONSOR
BSD Medical Corporation 2188 W 2200 S Salt Lake City UT 84119
AUG 1 3 2010
#### SUBMITTED BY
Dixie Toolson Sells, V.P., Regulatory Affairs BSD Medical Corporation 2188 West 2200 South Salt Lake City UT 84119 801 972 5555 (p) Ext 215 / 801 072 5930 (f) dsells@bsdmc.com
## CONTACT/ PREPARED BY
Phil Triolo Phil Triolo and Associates LC 148 S. 1200 E. Salt Lake City, UT 84102-1643 .801 699 9846 (p) / 801 328 2399 (f) · philt@philt.com
## DATE PREPARED
April 26, 2010
# TRADE OR PROPRIETARY NAME
MicroThermX® Microwave Ablation System (MTX-180 or MTX)
### CLASSIFICATION/ NAME
Class II, (21CFR §878.4400), Electrosurgical cutting and coagulation device and accessories Product Code- NEY
### PREDICATE DEVICES
MicroThermX 100 Microwave Ablation System (MTX-100) (K081042) Valleylab VivaWave Microwave Ablation System (K053535) Vivant Medical VivaTip Microwave Ablation Probe and Accessories (K032702)
{1}------------------------------------------------
K100786
Page 2 of 4
April 29, 2010
Traditional 510(k) Premarket Notification
# DEVICE DESCRIPTION
The MicroThermX Microwave Ablation System (MTX-180 or MTX) delivers 915 MHz microwave energy for coagulation (ablation) of soft tissue. The delivery of microwave energy is controlled by time and power parameters set by the operator. The operator can select up to 3 applicators to ablate target tissue.
The MTX-180 consists of a mobile cart with generator and closed-circuit antenna cooling system, microwave antennas, and an optional temperature sensor.
The mobile generator is comprised of a computer, microwave generator, and thermistor-based temperature monitoring system. The operator interface is via the touchscreen monitor of the mobile generator.
The cooling system consists of a bag of sterile isotonic saline (not supplied by BSD); peristaltic pump; cooling circuit tubing and connectors; and fluid pathway channels within the antenna.
The MTX-180 is designed to:
- · Deliver controlled microwave energy to induce coagulation of soft tissue;
- · Provide repeatable ablation zone geometries for a given set of ablation parameters (repeatability demonstrated in ex-vivo studies performed in nonperfused animal tissue);
- · Be used in an intraoperative or minimally invasive percutaneous procedure;
- · Utilize a single SynchroWave antenna or synchronous operation of 2 or 3 antennas during a single procedure to induce larger zones of ablation .; and
- · Utilize an optional sensor to monitor the temperature of non-target tissue during a procedure.
### INTENDED USE
The MicroThermX® Microwave Ablation System (MTX) delivers microwave energy for coagulation (ablation) of soft tissue. The system is not intended for use in cardiac procedures. The SynchroWave antennas may be used in open surgical as well as percutaneous ablation procedures. An optional temperature sensor may be used to monitor tissue temperatures.
# TECHNOLOGICAL COMPARISON
The modified device has the identical intended use and employs the same fundamental technology as the predicate devices.
{2}------------------------------------------------
K100786
April 29, 2010
Traditional 510(k) Premarket Notifica
The proposed design, material, and labeling changes that triggered the submission of this notification consist of changes in design: maximum total power output increased to 180 Watts; maximum power delivered to each antenna increased to 60W; number of antennas decreased from 4 to 3; cooling circuit added to cool antenna; reusable non-sterile temperature sensors replaced by sterile, pre-calibrated single-use temperature sensors; material changes (shaft material was changed when additional markers on antenna shaft were added to facilitate identification of insertion depth); software modifications to accommodate hardware changes); and labeling modifications to accommodate hardware changes and for clarity.
### PERFORMANCE TESTING
Verification and Validation Studies were conducted to evaluate the equivalence of the new device to the predicate device, or conformance with relevant standards. Testing performed included evaluations to determine conformance with:
- UL 60601-1:2003 R6.03, Medical Electronic Equipment: General . requirements for basic safety and essential performance
- IEC 60601:1988+A1:1991+A2:1995, Medical Electronic Equipment: Particular . requirements for Safety of Microwave Therapy Equipment
- IEC 60601-1-2 (ed 2.1), Medical Electronic Equipment: General requirements . for basic safety and essential performance
- IEC 60601-1-4, Medical Electronic Equipment: Programmable Electrical . Medical Systems
- IEC 60601-2-2 (ed 4), Particular Requirements for the Safety of High . Frequency Surqical Equipment
- . IEC 60601-2-6 (ed.1), Medical Electronic Equipment: Particular requirements for Safety of Microwave Therapy Equipment
- CISPR 11 (2007), Limits and Methods of Measurement of Electromagnetic . Disturbance Characteristics of Industrial, Scientific, and Medical (ISM) Radiofrequency Equipment
- Internal safety and performance requirements for: .
- Software control of delivered power o
- o Alarms and Shut-offs:
- Temperature of applied parts o
- o Cooling circuit function
- Ablation zone sizes o
- Accuracy of temperature measurement by TempSure Temperature o Sensors
- Usability 0
{3}------------------------------------------------
K100786
Page 4 of (4)
April 29, 2010 Traditional 510(k) Premarket Notification
The results of all testing performed demonstrated conformance with applicable, external standards or internal requirements and/ or equivalence with the predicate device.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE
Based on the results of Verification and Validation Studies, BSD concludes that the modified devices are as safe and effective as, and perform as well as, or better than, the predicate devices.
人
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three swooping lines forming its body and wings. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
BSD Medical Corporation % Phil Triolo and Associates LC Phil Triolo, Ph.D. 148 S. 1200 East Salt Lake City, Utah 84102-1643
Re: K100786
Trade/Device Name: MicroThermX® Microwave Ablation System (MTX-180 or MTX) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NEY Dated: August 06, 2010 Received: August 09, 2010
Dear Dr. Triolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
AUG 1 3 2010
{5}------------------------------------------------
### Page 2 - Dr. Phil Triolo
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm.
Sincerely yours,
Mark A. Milliken
. Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
April 29, 2010 Traditional 510(k) Premarket Notification
## SECTION 4: INDICATIONS FOR USE
510(k) Number (if known): K100786
Device Name:
MicroThermX Microwave Ablation System
### Indications for Use:
The MicroThermX Microwave Ablation System (MTX) delivers microwave energy for coagulation (ablation) of soft tissue. The system is not intended for use in cardiac procedures. The SynchroWave antennas may be used in open surgical as well as percutaneous ablation procedures. An optional temperature sensor may be used to monitor tissue temperatures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE------------------------------------------------------------------------------------------------------------------------------------------PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
(Posted November 13, 2003)
510(k) Number _
Section 4: Indications for Use Statement
Page 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.