K100762 · Fujifilm Medical Systems U.S.A, Inc. · MQB · Jul 15, 2010 · Radiology
Device Facts
Record ID
K100762
Device Name
FLAT PANEL DETECTOR, MODEL FDR D-EVO
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
MQB · Radiology
Decision Date
Jul 15, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film /screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Device Story
FDR D-EVO is a portable, indirect-conversion amorphous silicon (a-Si) flat panel detector (FPD) using Gadolinium OxySulfide (GOS) scintillator; utilizes Irradiation Side Sampling system for image quality. Device captures X-ray signals, converting them to digital images for display. Used in clinical settings as a retrofit for analog bucky systems or as an additional panel for fixed digital radiography; compatible with upright bucky, table, or free cassette positioning. Operated by radiology staff; data transmitted to Fujifilm FDX Console for display, diagnostic viewing, and network storage. Benefits include rapid conversion of analog/CR workflows to digital radiography, enabling efficient image acquisition and electronic transmission.
Clinical Evidence
Bench testing included internal and international IEC requirements. Clinical image quality study performed on 30 image pairs comparing FDR D-EVO to Fujifilm Carbon XL-2. Three board-certified radiologists reviewed images; 100% of D-EVO image scores were greater than or equal to -1, meeting the success criterion of at least 90%.
Technological Characteristics
Indirect-conversion amorphous silicon (a-Si) flat panel detector; GOS (Gadolinium OxySulfide) scintillator; 14x17" cassette form factor; portable; wired connectivity; detachable power source; digital image output.
Indications for Use
Indicated for general projection radiographic imaging of human anatomy. Not indicated for mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# 100762
Image /page/0/Picture/1 description: The image shows the logo for Fujifilm Medical Systems, USA. The logo is in black and white, with the word "FUJIFILM" in large, bold letters. Below the main logo, the text "Fujifilm Medical Systems, USA" is written in a smaller font. The logo appears to be a standard corporate logo, likely used for branding and identification purposes.
JUL 1 5 2010
## 510(k) Summary for FDR D-EVO Flat Panel Detector
## Date: March 16, 2010
## Contact Person:
| Name: | |
|------------|--|
| Title: | |
| Telephone: | |
| Facsimile: | |
Debbie Peacock Regulatory Coordinator (203) 602-3774 (203) 363-3813
## Identification of Device:
Proprietary/Trade Name: Classification Name: Classification: CFR Section: Product Codes: Common Name:
FDR D-EVO (DR-ID600) Solid State X-ray Imager (Flat Panel/Digital Imager) Panel: Radiology 21 CFR 892.1650 90 MQB Flat Panel Digital Detector
#### l. INDICATIONS FOR USE
The FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film /screen or CR systems may be used. The FDR D-EVO is not intended for mammography.
#### 11. DEVICE DESCRIPTION
FDR D-EVO flat panel detector is an indirect-conversion amorphous silicon (a-Si) portable flat panel detector (FPD) utilizing GOS (Gadolinium OxySulfide) as a schifillator. FDR D-EVO detector has Fuji's unique Irradiation Side Sampling system, delivering high image quality.
D-EVO's 14x17" standard cassette size affords it the ability for use as a retrofit in any analog bucky and/or as an additional panel with fixed digital radiography systems, allowing a quick conversion to digital X-ray technology. The D-EVO features a detachable power source that enables easy positioning within the radiographic room within an upright bucky, table or as a free cassette. Data captured via operator console is sent electronically to the Fujifilm FDX Console to be displayed on the peraitor.
#### lll. SUMMARY OF STUDIES
Fujifilm's D-EVO Flat Panel Detector successfully completed internal and international IEC testing requirements.
In addition, Fujifilm performed an image quality reader study on 30 image pairs comparing the FDR D-EVO system with the cleared Fujifilm Carbon XL-2. Three board certified radiologists reviewed the images as described below.
{1}------------------------------------------------
# Fujifilm Medical Syste
Analysis Methodology:
All images must be deemed to be of diagnostic capability. In addition, the study will be considered a success if at least 90% of the scores from the D-EVO system are greater than or equal to -1.
### Results:
All images were deemed to be of diagnostic capability, in addition, all readers satisfied the success criteria with 100% of the D-EVO image scores greater than or equal to -1. Test Protocol and Results are enclosed as Attachments H and I.
#### IV. SUBSTANTIAL EQUIVALENCE
The Fujifilm FDR D-EVO Flat Panel Detector is substantially equivalent to the following currently cleared device:
| Predicate Device | 510(k) # |
|----------------------|----------|
| Canon CXDI-55G | K091435 |
| Carestream DRX-1 | K090318 |
| Fujifilm Carbon XL-2 | K042023 |
The proposed and predicate devices utilize similar technology and materials, comparable safety and effectiveness features, and are similar in design and construction. The Indications for Use and labeling are virtually the same, or similar and our labeling contains the required Cautions, Warnings and Contraindications consistent to those required for similar cleared devices ..
The FDR D-EVO and the predicate Canon CXDI-55G detector are both portable, digital indirect conversion, amorphous silicon (a-Si), wired GOS detectors used for the same application, to image general radiographic examinations, excluding mammography. The Carestream DRX-1 detector is a portable, digital indirect conversion, amorphous silicon (a-Si), wireless GOS detectors used for the same application, to image general radiographic examinations, excluding mammography. All systems produce digital images which can be sent to hardcopy printers, softcopy diagnostic workstations and/or stored in archive.
Fujifilm Carbon XL-2 reads the x-ray image from cassette based photostimuable phosphor plates and converts it to a digital image. The FDR D-EVO FPD is used to directly capture digital X-ray images. Both systems use an acquisition workstation to display the acquired images and transmit them via network connection for diagnostic viewing and printing.
#### V. CONCLUSION
The FDR D-EVO Flat Panel Detector is substantially equivalent to the cleared predicate detectors and conforms to applicable medical device safety standards
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo includes a symbol on the left and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" on the right. The symbol appears to be a stylized representation of a human figure or a caduceus. The text is in a sans-serif font and is colored in blue.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Debra A. Peacock Senior Regulatory Affairs Specialist FUJIFILM Medical Systems, USA Inc. 419 West Avenue STAMFORD CT 06902
AUG 2 3 2013
Re: K100762
Trade/Device Name: FDR D-EVO (DR-ID 600) Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: May 17, 2010 Received: May 18, 2010
Dear Ms. Peacock:
This letter corrects our substantially equivalent letter of July 15, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Morri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K100762
## Device Name: __ FDR D-EVO (DR-ID 600) Flat Panel Detector
Indications for Use:
0
The FDR D-EVO flat panel detector system is intended to capture for display radiographic images of human anatomy. It is intended for use in general projection radiographic applications wherever conventional film /screen or CR systems may be used. The FDR D-EVO is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
hnd D. O'th
(Division Sign-Off)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K100762
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.