← Product Code [GGN](/productcode/GGN) · K100716

# AUDIT MICROCV D-DIMER LINEARITY SET, MODEL K717M-5 (K100716)

_Aalto Scientific, Ltd. · GGN · Jun 23, 2011 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100716

## Device Facts

- **Applicant:** Aalto Scientific, Ltd.
- **Product Code:** [GGN](/productcode/GGN.md)
- **Decision Date:** Jun 23, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5425
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The Audit™ MicroCV™ D-Dimer Linearity Set is a quantitative, assayed quality control material consisting of five levels of D-dimer analyte. The five D-dimer levels demonstrate a linear relationship to each other for D-dimer analyte. The Audit™ MicroCV™ D-Dimer Linearity Set is intended to use with the Biomerieux miniVIDAS® analyzer and VIDAS® D-dimer assay to verify the calibration of the measuring range. Audit™ MicroCV™ D-Dimer Linearity Set is "For In Vitro Diagnostic Use Only."

## Device Story

Audit™ MicroCV™ D-Dimer Linearity Set is a human-based, lyophilized, five-level quality control material; contains D-Dimer complex. Used in clinical laboratory settings to verify calibration, linear operating range, and reportable range of Biomerieux miniVIDAS® analyzer and VIDAS® D-dimer assay. Levels A through E are related by linear dilution; Level A represents the lower limit and Level E represents the upper limit of the instrument's measuring range. Healthcare providers use the output to confirm instrument performance and ensure accuracy of patient D-Dimer test results. Benefits include improved confidence in diagnostic accuracy for coagulation testing.

## Clinical Evidence

No clinical data. Bench testing only. Stability studies performed to determine reconstituted stability (2 days at 2-8° C) and shelf life (2 years at 2-8° C).

## Technological Characteristics

Human-based, lyophilized, five-level quality control material. Analyte: D-Dimer complex. Form factor: Vials for reconstitution. Storage: 2-8° C. Intended for use with Biomerieux miniVIDAS® analyzer.

## Regulatory Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

## Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- Triage® D-dimer Calibration Verification Controls (k050799)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k100716

B. Purpose for Submission:
Clearance of a new device

C. Measurand:
D-dimer

D. Type of Test:
Quantitative

E. Applicant:
Aalto Scientific, Ltd.

F. Proprietary and Established Names:
Audit™ MicroCV™ D-Dimer Linearity Set

G. Regulatory Information:
1. Regulation section:
21 CFR 864.5425
2. Classification:
Class II
3. Product code:
GGN-Plasma, Coagulation Control
4. Panel:
81 (Hematology)

H. Intended Use:
1. Intended use(s):
The Audit™ MicroCV™ D-Dimer Linearity Set is a quantitative, assayed quality control material consisting of five levels of D-dimer analyte. The five D-dimer levels demonstrate a linear relationship to each other for D-dimer analyte. The Audit™ MicroCV™ D-Dimer Linearity Set is intended to use with the Biomerieux miniVIDAS® analyzer and VIDAS® D-dimer assay to verify the calibration of the measuring range. Audit™ MicroCV™ D-Dimer Linearity Set is "For In Vitro Diagnostic Use Only."
2. Indication(s) for use:
Same as intended use(s)
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Biomerieux miniVIDAS® analyzer and VIDAS® D-dimer assay

I. Device Description:
The Audit™ MicroCV™ D-Dimer Linearity Set is a human based, lyophilized, five level set of QC material, A, B, C, D, and E. Each level contains D-dimer complex. It is used to confirm the proper calibration of the measuring range of D-dimer on the miniVIDAS® analyzer. Level A D-dimer is near the lower limit level and level E D-dimer is near the upper limit of the measuring range. Levels B, C, and D are related by linear dilution of level A and level E. The target ranges for five levels are:

{1}

2

|  Levels | A | B | C | D | E  |
| --- | --- | --- | --- | --- | --- |
|  Target Range | <150 ng/mL | 1105-1495 ng/mL | 2010-2760 ng/mL | 2145-4255 ng/mL | 3700-6000 ng/mL  |

## J. Substantial Equivalence Information:

1. Predicate device name(s):
Triage® D-dimer Calibration Verification Controls

2. Predicate K number(s):
k050799

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device: Audit™ MicroCV™ D-Dimer Linearity Set | Predicate: Triage® D-dimer Calibration Verification Controls  |
|  Intended Use | The Audit™ MicroCV™ D-Dimer Linearity Set is a quantitative, assayed quality control material consisting of five levels of D-dimer analyte. The five D-dimer levels demonstrate a linear relationship to each other for D-dimer analyte. The Audit™ MicroCV™ D-Dimer Linearity Set is intended to use with the Biomerieux miniVIDAS® analyzer and VIDAS® D-dimer assay to verify the calibration of the measuring range. | The Triage® D-dimer Calibration Verification Controls are to be used with the Triage® D-dimer Test and Triage® MeterPlus to verify the calibration of the Triage® D-dimer test throughout the measurable range.
The Triage® D-dimer Calibration Verification Controls are assayed materials to be used with the Triage® D-dimer Test and Triage® MeterPlus to assist the laboratory in monitoring test performance.  |
|  Number of levels per set | 5 | Same  |
|  Matrix | Human based plasma | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item |  |   |
|  Matrix | Lyophilized | Liquid  |
|  Closed vial stability | 2-8°C for 24 months | -20°C for 9 months  |
|  Open vial stability | 2 days when stored at 2-8°C | 1 day when stored at 2-8°C  |

## K. Standard/Guidance Document Referenced (if applicable):

CLSI EP06-A: Evaluation of Linearity of Quantitative Measurement Procedures: A Statistical Approach.

## L. Test Principle:

The Audit™ MicroCV™ D-Dimer Linearity Set is used to verify the calibration of the D-dimer measuring range of the miniVIDAS® analyzer.

## M. Performance Characteristics:

1. Analytical performance:
a. Precision/Reproducibility:

{2}

For each level of D-Dimer Linearity Set, 15 replicates were tested on the miniVIDAS® on three lots. All levels resulted &lt;10% CV.

b. Linearity/assay reportable range:
Linearity was determined using three lots of the D-Dimer Linearity Set. The mean D-dimer concentration of each of the 5 levels was plotted vs. five assigned levels of concentration. A linear regression value was obtained for each lot, and all results were within the limit of acceptance r² &gt; 0.95.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):
- Traceability: All reagent components are obtained from commercial vendors.
- Open vial stability: A 2-day open vial stability claim was verified using three lots of the Audit™ MicroCV™ D-Dimer Linearity Set. All 5 levels were tested on day zero and day 2. The percent recovery is determined in comparison to day zero values. The percent recovery is within ±10% limit of acceptance.
- Closed vial stability: An accelerated stability is used to predict shelf life. The vials are stressed at 25°C for 64 days to predict two-year stability when stored at 2-8°C. The percent recovery of all levels is determined in comparison to day zero values and the product is considered stable when the recovered values are within ±10%. Real time studies are ongoing to support the shelf life of the product.
- Value assignment: For each level of D-Dimer Linearity Set, 15 separate vials were tested on one miniVIDAS® analyzer yielding 15 measurements. The mean of each level was used to calculate the target concentration. The target range of each level was calculated as ±1SD of the target concentration.

d. Detection limit:
Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

{3}

5. Expected values/Reference range:
The mean and expected assay range for each level are provided in the labeling.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

4

---

**Source:** [https://fda-staging.innolitics.com/device/K100716](https://fda-staging.innolitics.com/device/K100716)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
