The Elettra Ultrasound System is intended for diagnostic imaging or fluid flow analysis of the human body including: Fetal, Abdominal, Intraoperative, Intraoperative Neurological, Pediatric, Small Organ, Neonatal Cephalic, Cardiac, Peripheral Vessel, Musculoskeletal (Conventional), and Musculoskeletal (Superficial).
Device Story
Compact diagnostic ultrasound system; includes console, display, power supply, user controls, and transducers. Transmits ultrasonic energy into patient; receives echoes; performs post-processing to generate real-time anatomic and fluid flow images. Transducers operate in cabled or wireless configurations. Used by physicians for diagnostic imaging and flow analysis. Output displayed on-screen for clinical assessment of structures and hemodynamics. Benefits include non-invasive visualization of internal anatomy and blood flow to aid clinical decision-making.
Clinical Evidence
Bench testing only. Evaluated for acoustic output, biocompatibility, cleaning, disinfection, and sterilization effectiveness. Thermal, electrical, and mechanical safety, and electromagnetic compatibility testing performed. Conforms to applicable medical device safety standards.
Technological Characteristics
Compact ultrasound system; includes system console and transducers (L8-3 Linear, L12-5 Linear, C5-2 Curvilinear). Supports B-mode, Color Doppler, and Amplitude Doppler. Operates via ultrasonic pulsed echo/Doppler. Supports cabled or wireless transducer configurations. Complies with Track 3 acoustic output methods. Materials are biocompatible. Designed to meet electrical and physical safety standards.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ (e.g., breast, testes, thyroid, penis), neonatal cephalic, cardiac, peripheral vessel, and musculoskeletal (conventional and superficial) applications. For use by qualified physicians. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K/00598
### Summary of Safety and Effectiveness
NOV 1 9 2010
This summary of 510(k) safety and effectiveness information is submitted as a part of this Premarket Notification in accordance with the requirements of the Safe Medical Devices Act of 1990 as implemented in 21 CFR 807.92.
#### 1) Submitter's information:
Penrith Corporation 5170 Campus Drive Suite 2 Plymouth Meeting, PA 19462 USA
Corresponding official: Email: Telephone: Fax:
Lawrence Engle lengle@penrithcorp.com 610 834 1220 610 834 1220
Date Summary Prepared:
February 10, 2010
### 2) Device Information
The proprietary name of the device is the Penrith Elettra Diagnostic Ultrasound System.
The Elettra is classified as follows:
90 IYN Ultrasonic Pulsed Doppler Imaging System
Ultrasonic Pulsed Echo Imaging System 90 IYO
90 ITX Diagnostic Ultrasound Transducer
The Penrith Elettra is a compact diagnostic ultrasound device. It includes a system console housing electronic circuitry, a video display, power supply, and user controls. This connects to the transducers and together these generate the ultrasound image.
The Penrith Elettra is intended to be used by a qualified physician for diagnostic imaging or fluid flow analysis of the human body including: Fetal,
{1}------------------------------------------------
Abdominal, Intraoperative, Intraoperative Neurological, Pediatric, Small Organ, Neonatal Cephalic, Cardiac, Peripheral Vessel, Musculoskeletal (Conventional), and Musculoskeletal (Superficial).
The Company believes that the Penrith Elettra is substantially equivalent to the Acuson Sequoia, K022567, the Acuson Cypress K052331, and the Philips Avalon CTS, K023931.
The Penrith Elettra system and transducers function in a manner that is substantially equivalent to the previously cleared devices: Predicate device systems and the Elettra transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. The Elettra and the predicate devices share basic scanning modalities. Some or all predicate device systems allow for measurements of structures and flow, and calculations. Some or all predicate devices and the Elettra follow the Track 3 method for acoustic output. Patient contact materials used in the Elettra are used in equivalent formulations in the predicate devices. All indications for use claimed for the Elettra are cleared indications found on some or all of the predicate devices. The Elettra ultrasound transducer may be used in a wireless configuration or in a cabled configuration. The Elettra and the predicate devices have been designed and manufactured under comparable electrical and physical safety standards.
The Penrith Elettra has been evaluated for acoustic output, biocompatibility, cleaning, disinfection, and sterilization effectiveness, as well as thermal, electrical and mechanical safety, and electromagnetic compatibility. It has been found to conform with applicable medical device safety standards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human figures.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
Penrith Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Servies, LLC 1394 25th Street NW BUFFALO MN 55313
NOV 1 9 2010
Re: K100598
Trade/Device Name: Penrith Elettra Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: October 18, 2010 Received: October 19, 2010
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Penrith Elettra Ultrasound System, as described in your premarket notification:
Transducer Model Number
L8-3 Linear L12-5 Linear C5-2 Curvilinear
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device
{3}------------------------------------------------
can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter. please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours,
David C. Rainer, Ph.D.
David G. Brown, Ph.D. Acting Director . Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
## Section 1.3 Indications for Use
K100598
# NOV 1 9 2010
The Elettra Ultrasound System is intended for diagnostic imaging or fluid flow analysis of the human body including: Fetal, Abdominal, Intraoperative, Intraoperative Neurological, Pediatric, Small Organ, Neonatal Cephalic, Cardiac, Peripheral Vessel, Musculoskeletal (Conventional), Musculoskeletal (Superficial). See Indications for Use Forms below. The following tables provide details regarding clinical applications and modes of operation for the system and individual transducers.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K100598
. .
Page 1 of 5
יי
{5}------------------------------------------------
510(k) Number (if known): K100598
Penrith Elettra Ultrasound System
Indications for Use:
Device Name:
Diagnostic Imaging or fluid flow analysis of the human body as follows:
Indications for Use.
| Clinical Application | | Mode of Operation | | | | | | | | | |
|-------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|---------------------------------|----------------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)<br>Note 2 | Other: Harmonic<br>Imaging | |
| Opthalmic | | | | | | | | | | | |
| Fetal | | N | | | | N | N | | | | |
| Abdominal | | N | | | | N | N | | | | |
| Intraoperative Note 1 | | N | | | | N | N | | | | |
| Intraoperative Neurological | | N | | | | N | N | | | | |
| Pediatric | | N | | | | N | N | | | | |
| Small Organ Note 3 | | N | | | | N | N | | | | |
| Neonatal Cephalic | | N | | | | N | N | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | | | | N | N | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vessel | | N | | | | N | N | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal Conventional | | N | | | | N | N | | | | |
| Musculo-skeletal Superficial | | N | | | | N | N | | | | |
| Other | | | | | | | | | | | |
N: New P: Previously Cleared Blank: Not Claimed
Note 1: for example: cardiac
Note 2: B mode and PWD mode or Color Doppler and PW mode
Note 3: for example: breast, testes, thyroid, penis
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-the-Counter Use _ (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K100598
Page 2 of 5
{6}------------------------------------------------
510(k) Number (if known): K100598
Device Name: L8-3 Linear Transducer for use with
Penrith Elettra Ultrasound System
Indications for Use:
Diagnostic Imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | | | |
|-------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|---------------------------------|----------------------------|--|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)<br>Note 2 | Other: Harmonic<br>Imaging | |
| Opthalmic | | | | | | | | | | | |
| Fetal | | N | | | | N | N | | | | |
| Abdominal | | N | | | | N | N | | | | |
| Intraoperative Note 1 | | N | | | | N | N | | | | |
| Intraoperative Neurological | | N | | | | N | N | | | | |
| Pediatric | | N | | | | N | N | | | | |
| Small Organ Note 3 | | N | | | | N | N | | | | |
| Neonatal Cephalic | | N | | | | N | N | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | N | | | | N | N | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vessel | | N | | | | N | N | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal Conventional | | N | | | | N | N | | | | |
| Musculo-skeletal Superficial | | N | | | | N | N | | | | |
| Other | | | | | | | | | | | |
N: New P: Previously Cleared Blank: Not Claimed
Note 1: for example: cardiac
Note 2: B mode and PWD mode or Color Doppler and PW mode
Note 3: for example: breast, testes, thyroid, penis
| Prescription Use X<br>(Part 21 CFR 801 Subpart D) | AND/OR |
|---------------------------------------------------|--------|
|---------------------------------------------------|--------|
Over-the-Counter Use (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K100598
Page 3 of 5
{7}------------------------------------------------
510(k) Number (if known): K100598
Device Name:
L12-5 Linear Transducer for use with Penrith Elettra Ultrasound System
Indications for Use:
Diagnostic Imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|---------------------------------|----------------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)<br>Note 2 | Other: Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | N | | | | N | N | | | |
| Abdominal | | N | | | | N | N | | | |
| Intraoperative Note 1 | | N | | | | N | N | | | |
| Intraoperative Neurological | | N | | | | N | N | | | |
| Pediatric | | N | | | | N | N | | | |
| Small Organ Note 3 | | N | | | | N | N | | | |
| Neonatal Cephalic | | N | | | | N | N | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | | | | N | N | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vessel | | N | | | | N | N | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | N | | | | N | N | | | |
| Musculo-skeletal Superficial | | N | | | | N | N | | | |
| Other | | | | | | | | | | |
N: New P: Previously Cleared Blank: Not Claimed
Note 1: for example: cardiac
Note 2: B mode and PWD mode or Color Doppler and PW mode
Note 3: for example: breast, testes, thyroid, penis
| Prescription Use X<br>(Part 21 CFR 801 Subpart D) | AND/OR |
|---------------------------------------------------|--------|
|---------------------------------------------------|--------|
Over-the-Counter Use (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K100598
Page 4 of 5
{8}------------------------------------------------
510(k) Number (if known): K100598
Device Name:
C5-2 Curvilinear Transducer for use with Penrith Elettra Ultrasound System
Indications for Use:
Diagnostic Imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|---------------------------------|----------------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify)<br>Note 2 | Other: Harmonic<br>Imaging |
| Opthalmic | | | | | | | | | | |
| Fetal | | N | | | | N | N | . | | |
| Abdominal | | N | | | | N | N | | | |
| Intraoperative Note 1 | | N | | | | N | N | | | |
| Intraoperative Neurological | | N | | | | N | N | | | |
| Pediatric | | N | | | | N | N | | | |
| Small Organ Note 3 | | N | | | | N | N | | | |
| Neonatal Cephalic | | N | | | | N | N | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | | | | N | N | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral Vessel | | N | | | | N | N | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal Conventional | | N | | | | N | N | | | |
| Musculo-skeletal Superficial | | N | | | | N | N | | | |
| Other | | | | | | | | | | |
N: New P: Previously Cleared Blank: Not Claimed
Note 1: for example: cardiac
Note 2: B mode and PWD mode or Color Doppler and PW mode
Note 3: for example: breast, testes, thyroid, penis
| Prescription Use X<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|---------------------------------------------------|--------|------------------------------------------------|
|---------------------------------------------------|--------|------------------------------------------------|
Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K100598
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.