← Product Code [LFL](/productcode/LFL) · K100597

# HARMONIC FOCUS SHEARS (K100597)

_Ethicon Endo-Surgery, LLC · LFL · Apr 7, 2010 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100597

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, LLC
- **Product Code:** [LFL](/productcode/LFL.md)
- **Decision Date:** Apr 7, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K100597 · Apr 7, 2010 | HARMONIC FOCUS SHEARS | Ethicon Endo-Surgery, LLC | Published clinical literature | The sponsor utilized a review of published clinical literature to demonstrate that the use of ultrasonic Harmonic technology in ENT procedures is safe and effective, supporting the expansion of the Indications for Use. | Literature review; ENT procedures; Clinical experience |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Assessment of published clinical literature; Literature review | Patients undergoing otorhinolaryngological (ENT) procedures; Number of Sites: Not applicable | Not applicable for this study | Safety and effectiveness of ultrasonic technology in ENT procedures |

## Indications for Use

The HARMONIC FOCUS™ Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space), and other open procedures.

## Device Story

Harmonic FOCUS Shears are sterile, single-patient-use ultrasonic surgical instruments. Device consists of a soft-grip scissor handle with MIN/MAX power controls, a curved blade, and a clamp arm with a Teflon pad. Operates by connecting to a Harmonic Scalpel Generator and handpiece, which converts electrical energy into 55,500 Hz ultrasonic mechanical vibration. Mechanical energy at the blade tip converts to thermal energy, enabling simultaneous cutting and coagulation of soft tissue vessels up to 5mm in diameter. Used by surgeons in open procedures (ENT, general, plastic, pediatric, gynecologic, urologic, orthopedic). Provides an alternative to electrosurgery, lasers, and steel scalpels. Benefits include precise tissue incision with reduced thermal injury and effective hemostasis.

## Clinical Evidence

No new clinical data provided. Assessment of published clinical literature confirms that ultrasonic Harmonic technology is safe and effective for otorhinolaryngological (ENT) procedures, consistent with predicate device performance.

## Technological Characteristics

Ultrasonic surgical instrument; 9cm working length; 16mm active blade length; Teflon pad on clamp arm. Energy source: Harmonic Scalpel Generator (electrical to mechanical). Frequency: 55,500 Hz. Materials: Standard surgical grade components. Connectivity: Wired connection to generator/handpiece. Sterilization: Single-patient use (sterile).

## Predicate Devices

- Harmonic FOCUS shears ([K063192](/device/K063192.md))
- Harmonic Blades ([K060245](/device/K060245.md))
- Harmonic Blades ([K012176](/device/K012176.md))
- Harmonic Blades ([K990362](/device/K990362.md))
- Olympus Ultrasonic Surgical System ([K031305](/device/K031305.md))
- Conmed Electrosurgical Pencils ([K041868](/device/K041868.md))
- ArthroCare ENT Coblator/Plasma Wands ([K070374](/device/K070374.md))
- ArthroCare ENT Coblator/Plasma Wands ([K030108](/device/K030108.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## APR - 7 2010

Ethicon Endo-Surgery, LLC Traditional 510(k) Premarket Notification Harmonic FOCUS shears

K 100595

## · 510(k) Summary

| Company:           | Ethicon Endo-Surgery, LLC<br>475 Calle C<br>Guaynabo, PR 00969                                                                                                                                                                          |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:           | Marcia Arentz, RAC<br>Regulatory Affairs Portfolio Leader<br>Ethicon Endo-Surgery<br>4545 Creek Road<br>Cincinnati, OH 45242<br>Tel: (513) 337-1066<br>Fax: (513) 337-1122<br>Email: marentz5@its.jnj.com                               |
| Date Prepared:     | March 1, 2010                                                                                                                                                                                                                           |
| Device Name:       | Trade Name: Harmonic FOCUSTM shears<br>Common or Usual Name: Ultrasonic Surgical Instrument<br>Classification Name: Electrosurgical Cutting and Coagulation<br>Device [21 CFR 878.4400 (GEI/LFL)]<br>Device Class: Class II             |
| Predicate Devices: | Harmonic FOCUSTM shears (K063192)<br>Harmonic Blades (K060245, K012176, K990362)<br>Olympus Ultrasonic Surgical System (K031305)<br>Conmed Electrosurgical Pencils (K041868)<br>ArthroCare ENT Coblator/Plasma Wands (K070374. K030108) |

Purpose of Submission: To modify the Indications for Use Statement to include the specific indication to use the device in otorhinolaryngology procedures. The device design is essentially unchanged from the previously cleared FOCUS device.

Device Description: The Harmonic FOCUS™ Curved Shears is a sterile, single patient use instrument consisting of a soft grip scissor handle housing assembly with two hand controls (MIN for minimum power level and MAX for maximum power level). The instrument has a curved blade and clamp arm with Teflon pad. The instruments working length is 9cm with a 16mm active blade length. The FOCUS instrument allows for the cutting and coagulation of vessels up to and including 5mm in diameter. The Harmonic FOCUS Shears are designed for use with a generator capable of providing electrical energy to the FOCUS device and with a handpiece that converts the electrical energy to mechanical energy.

ﺴﺘﻌ

{1}------------------------------------------------

Indications for Use: The Harmonic FOCUS™ Curved Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space), and other open procedures.

Technological Characteristics: The Harmonic FOCUSTM Curved Shears is the same device as the prodicate Harmonic FOCUS™ Curved Shears with minor changes to the design and assembly drawings. The function and intended use of the device have not been changed. The FOCUS shears are designed for use with the Harmonic® Scalpel Generator and the Harmonic Handpiece. The generator provides electrical energy to the Handpiece which creates ultrasonic vibration (55,500 Hz) in the blade. The mechanical energy of the moving blade is converted into thermal energy which is used to cut and coagulate soft tissues.

The Harmonic FOCUS 1M device is similar in design and utilizes the same ultrasonic technology as the predicate Harmonic FOCUS device, the predicate Harmonic Scalpel Blades and the predicate Olympus Ultrasonic Surgical System shears. The Harmonic FOCUS device is also similar to electrosurgical devices such as the Conmed electrosurgical Pencils and the Arthrocare ENT Coblator/Plasma wands because they all use electrical energy to cut and coagulate tissues.

Performance Data: The design of the Harmonic FOCUS Shears is essentially unchanged and the performance data provided in K063192 is still applicable to the device. The Harmonic FOCUS shears device will seal vessels up to and including 5mm in diameter.

Clinical Literature: An assessment of published clinical literature demonstrates that there is a considerable body of knowledge and experience that show the use of ultrasonic Harmonic technology in otorhinolaryngological (ENT) procedures is as safe and effective as predicate technology for use.

## Substantial Equivalence

In establishing substantial equivalence of the Harmonic FOCUS shears to the predicate devices, Ethicon Endo-Surgery evaluated the intended use, the indications for use, the technological characteristics and the energy requirements of the systems. The Harmonic FOCUS is substantially equivalent to the predicatcs because they share the same intended use and technological characteristics. The use of Harmonic FOCUS shears in otorhinolaryngologic (ENT) procedures does not raise any new types of questions of safety and effectiveness compared with predicate devices currently used for these procedures.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the top half of the circle. In the center of the seal is a stylized image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

APR - 7 2010

Ethicon Endo-Surgery, LLC % Ms. Marcia Arentz, RAC -Regulatory Affairs Portfolio Leader 4545 Creek Road Cincinnati, Ohio 45242

Re: K100597

Trade/Device Name: Harmonic FOCUS™ Shears Regulatory Class: Unclassfied Product Code: LFL, GEI Dated: March 01, 2010 Received: March 02, 2010

Dear Ms. Arentz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Marcia Arentz, RAC

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K 100597

Device Name: Harmonic FOCUSTM Shears

Indications for Use:

The HARMONIC FOCUS™ Shears are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space), and other open procedures.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R.P. Ogden for mxm

DIVISION Sion-Of Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

(Posted November 13, 2003)

510(k) Number K100597

---

**Source:** [https://fda-staging.innolitics.com/device/K100597](https://fda-staging.innolitics.com/device/K100597)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
