K100596 · Mid-Atlantic Diagnostics, Inc. · MQK · Dec 6, 2010 · Obstetrics/Gynecology
Device Facts
Record ID
K100596
Device Name
CRYOPETTE
Applicant
Mid-Atlantic Diagnostics, Inc.
Product Code
MQK · Obstetrics/Gynecology
Decision Date
Dec 6, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6160
Device Class
Class 2
Indications for Use
The Cryopette is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human blastocyst stage embryos.
Device Story
Cryopette is a cryopreservation storage device for human blastocyst stage embryos. Device consists of a polycarbonate cryostraw with an attached silicone displacement bulb for sample aspiration and dispensing; a metering sheath for rigidity; a support ring; and an outer case for labeling. Used in clinical assisted reproduction laboratories. Operator (embryologist) aspirates sample into straw, heat-seals ends, and plunges device into liquid nitrogen for storage. Post-thaw, device is submerged in 37°C water, cut at the sealing mark, and the bulb is pressed to expel the embryo into warming media. Provides a closed system for vitrification and storage. Benefits include secure containment and maintenance of embryo viability during cryopreservation.
Clinical Evidence
Bench testing only. Mouse Embryo Assay (MEA) demonstrated >70% hatching at 96 hours. Endotoxin levels <20 EU/device. Mechanical testing confirmed aspiration, sealing, and leak-free performance. Comparative testing with predicate (CryoTip) showed equivalent cooling/warming rates and comparable mouse blastocyst survivability (96% for Cryopette vs 84% for CryoTip). Sterilization validation confirmed SAL 10^-6.
Technological Characteristics
Materials: Polycarbonate Lexan OQ-3820 (straw), stainless steel T21-RW (sheath), Black Delrin Acetal Resin (ring), Elastosil LR 3003/20 silicone rubber (bulb), Pebax 6333 (case). Principle: Manual aspiration/dispensing via displacement bulb; heat-sealed closed system. Dimensions: 3.067" overall length. Sterilization: Gamma irradiation (25-35 kGy) per ISO EN552.
Indications for Use
Indicated for use in vitrification procedures to contain and maintain human blastocyst stage embryos.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
#### 510(k) Summary
Submitted by: Mid-Atlantic Diagnostics, Inc.
77 Elbo Lane Mount Laurel, NJ 08054
DEC - 6 2010
Telephone: (856) 762-2000 Facsimile: (856) 762-2009
Contact: Susan Bush, Quality & Regulatory Affairs Manager
Date Submitted: November 8, 2010
Device Identification:
| Trade Name: | Cryopette™ |
|----------------------|-------------------------------------------------|
| Common Name: | Cryopreservation container |
| Classification Name: | Assisted Reproduction Labware (21 CFR 884.6160) |
| Product Code: | MQK |
Irvine Scientific CryoTip (K041562) Predicate Device:
Description:
The Cryopette consists of a cryostraw manufactured from polycarbonate-Lexan OQ-3820. This straw has a displacement bulb attached to one end of the cryostraw whose function is to aspirate and dispense the sample contained in the vitrification media in and out of the opposite (open) end of the cryostraw. The cryostraw is the only sample contacting portion of the Cryopette.
The metering sheath provides structural rigidity for the Cryopette. The support ring provides a backstop and seal for the displacement bulb. The outer case encases the displacement bulb to provide a surface for labeling.
The sealing mark, distal storage border and proximal storage border reference lines are found on the outside of the cryostraw.
Intended Use:
The Cryopette is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human blastocyst stage embryos.
Performance Data:
l
{1}------------------------------------------------
The following testing was performed:
- 1. Mouse Embryo Assay (MEA): >70% hatching at 96 hours MEA testing was performed to ensure that the Cryopette is nontoxic to embryos. Testing demonstrated that the hatching percentage well exceeded the specification of >70% at 96 hours, and this information will be provided on a Certificate of Analysis accompanying packages of the product.
- 2. Endotoxin (LAL) Levels: < 20 EU Endotoxin testing was performed to ensure that the Cryopette does not have unacceptable levels of endotoxin. Testing demonstrated that the endotoxin levels are <20 EU, and this information will be provided on a Certificate of Analysis accompanying packages of the product.
- 3. Mechanical Testing (Aspiration and Leakage) Mechanical testing, including evaluation of mean, minimum aspiration volumes, as well as leakage following heat sealing of the device, was performed on the Cryopette to ensure that the device functioned as intended. All testing demonstrated that the Cryopette aspirated samples appropriately and did not show evidence of leakage following heat sealing.
- 4. Blastocyst Survivability (comparative testing to CryoTip predicate device) Comparative testing was performed to demonstrate that mouse blastocysts had comparable survivability following various lengths of storage, compared to the predicate device. This testing demonstrated that mouse blastocysts vitrified and stored in Cryopettes had comparable survivability to mouse blastocysts vitrified and stored in CrvoTips.
- 5. Heating and Cooling Rates (see table below) An independent laboratory conducted a comparative evaluation of the cooling and warming rates of media in the Cryopette and CryoTip. This testing demonstrated that the cooling and warming rates were equivalent.
- 6. Sterilization Validation: SAL 10-6 Sterilization validation was performed to demonstrate that the Cryopette had a SAL of 10 t, and this information will be provided on a Certificate of Analysis accompanying packages of the product.
- 7. Shelf Life/Package Integrity Shelf life and package integrity testing was performed to demonstrate that the Cryopette maintains package integrity, sterility and functionality for the labeled shelf life of the devices.
{2}------------------------------------------------
# ADDITIONAL INFORMATION REQUESTED FOR CRYOPETTE 510(k) K100596
### Substantial Equivalence Comparison
| SUBSTANTIAL<br>EQUIVALENCE<br>COMPARISON | MidAtlantic Diagnostics<br>Cryopette (K100596) | Irvine Scientific<br>CryoTip<br>(K041562) |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Cryopette is a<br>cryopreservation storage<br>device that is intended for<br>use in vitrification<br>procedures to contain and<br>maintain human blastocyst<br>stage embryos. | The CryoTip is a<br>Cryopreservation device<br>that is intended to be used to<br>contain, freeze and maintain<br>oocytes and/or embryos. |
| Method of Sterilization | Gamma irradiation 25 -35<br>kGy according to ISO<br>EN552. | Gamma irradiation 25 -35<br>kGy according to ISO<br>EN552. |
| Sterilization | Microbiological validation of<br>25kGy Radiation<br>Sterilization by<br>ANSI/AAMI/ISO 11137-2<br>Method VD max 25 | Validation information not<br>available |
| Sterility Assurance Level<br>(SAL) | 10-6 | 10-6 |
| Testing- Endotoxin | LAL Method -<br>Each lot of Cryopettes is<br>tested for endotoxin levels.<br>The level of endotoxin units<br>per device must be less than<br>the USP limit of<br>20EU/device. | LAL Method -<br>Each lot of CryoTips tested<br>for endotoxin levels. The<br>level of endotoxin units per<br>device must be less than the<br>USP limit of 20EU/device. |
| Testing - Mouse Embryo<br>Assay | Each lot of Cryopettes is<br>tested using one-cell mouse<br>embryo bioassay. Results<br>must be equal to or greater<br>than 70% to blastocyst stage<br>within 96 hours. | Each lot of CryoTips is<br>tested using one-cell mouse<br>embryo bioassay. Results<br>must be equal to or greater<br>than 70% to blastocyst stage<br>within 96 hours. |
| Type of<br>Cryopreservation<br>system | Heat sealing capability<br>allows for a closed system. | Heat sealing capability<br>allows for a closed system. |
| | MidAtlantic Diagnostics<br>Cryopette (K100596) | Predicate Device Irvine<br>Scientific CryoTip K041562 |
| Blastocyst<br>Survivability (Refer to Feb<br>2010 Submission) | Overall survival of mouse<br>blastocysts for the Cryopette<br>was 96%. | Overall survival of mouse<br>blastocysts for the Cryotip<br>was 84%. |
| Materials | Cryo-Straw: polycarbonate<br>Lexan OQ-3820<br>Metering Sheath: stainless<br>steel full hard T21-RW<br>Support Ring: Black Delrin<br>Acetal Resin<br>Displacement Bulb: Elastosil<br>LR 3003/20 US Silicon<br>Rubber<br>Outer Case: Pebax 6333<br>Tubing<br>Adhesive: Loctite 3311<br>Ink: 18696 Ink Grade<br>(Black) | CryoTip: medical grade<br>plastic<br>Stainless steel cover sleeve |
| Mean Sample Volume | 1.08 ul | 1.50 ul |
| Loading Method | Samples loaded by gently<br>depressing the displacement<br>bulb to aspirate samples into<br>the Cryo-Straw. | Samples loaded by attaching<br>wide end of CryoTip to<br>aspiration tool such as luer<br>tip syringe using connector.<br>Samples then aspirated into<br>the CryoTip. |
| Sample Placement | Sample placement must be | Sample placement should be |
| | | between 2nd and 3rd mark of |
| | between the Proximal and<br>Distal Storage borders of the<br>Cryopette. | |
| | 100% (61/61) of Cryopettes<br>aspirated fluid as intended<br>into the device. None of the<br>Cryopettes experienced any<br>clogging or leaking<br>problems. | 100% (58/58) of Cryotips<br>aspirated fluid as intended<br>into the device. None of the<br>Cryotips experienced any<br>clogging or leaking<br>problems. |
| Mechanical Testing | The Cryopette succeeded<br>100% (61/61) of the time in<br>delivering the specimen to<br>the proper position within<br>the device. | The Cryotip succeeded<br>100% (58/58) of the time in<br>delivering the specimen to<br>the proper position within<br>the device. |
| | Visual confirmation that<br>100% (61/61) Cryopettes<br>had intact seals and no<br>leakage of fluid. | Visual confirmation that<br>100% (58/58) Cryotips had<br>intact seals and no leakage<br>of fluid. |
| | 100% (61/61) Cryopettes<br>remained intact after<br>removing from liquid<br>nitrogen. | 86% (50/58) Cryotips<br>remained intact after<br>removing from liquid<br>nitrogen; 14% (8/58)<br>Cryotips exploded<br>immediately after removing<br>from liquid nitrogen. |
| Dimensions | Cryo-Straw: 2.4"<br>Bulb: .677"<br>Distal end to sealing mark:<br>.206"<br>Sealing mark to Distal<br>Storage border: .394"<br>Distal storage border to<br>proximal storage border:<br>.394"<br>Proximal Storage border to<br>bulb: 1.406"<br>I.D. of Bulb: .028"<br>O.D of Bulb: .128"<br>I.D. of Straw: .0100"<br>O.D of Straw: .0150"<br>Overall length: 3.067" | N/A<br>N/A<br>I.D of tip: .007"<br>O.D of tip: .07"<br>Overall length: 2.99" |
| Dispensing Method Post<br>Thaw | Remove Cryopette® from<br>liquid nitrogen and<br>completely submerge entire | Remove Cryo Tip from<br>liquid nitrogen and immerse<br>into 37 degree Celsius |
| | | |
| | device into water at 37<br>degrees Celsius for 5<br>seconds.<br>Wipe excess water off.<br>Cryopette placed on cutting<br>platform and cut is made<br>just above the heat seal at<br>the Sealing Mark.<br>Displacement Bulb is then<br>pressed to expel blastocyst(s)<br>into vitrification warming<br>media | waterbath and swirl gently<br>for 3 seconds.<br>Wipe Cryo Tip with sterile<br>wipe.<br>Cut seal on wide end of Cryo<br>Tip at Mark # 4.<br>Attach wide end of Cryo Tip<br>to appropriate aspiration<br>tool.<br>Slide metal cover sleeve up<br>to expose tip end of Cryo<br>Tip.<br>Wipe tip with sterile wipe.<br>Cut seal on Mark # 2 and<br>dispense contents utilizing<br>aspiration tool into<br>vitrification warming media. |
| Cooling Rate | -23,700 °C/min | -12,000 °C/min |
| Vitrification Method | Sealed Cryopette is plunged<br>into a liquid nitrogen<br>reservoir and then placed in<br>a goblet or cryocane.<br>The CryoCane is then stored<br>in a liquid nitrogen freezer<br>for long term storage | Sealed Cryo Tip is plunged<br>into a liquid nitrogen<br>reservoir, metal covered side<br>down, then placed in a<br>goblet or CryoCane.<br>The CryoCane is then stored<br>in a liquid nitrogen freezer<br>for long term storage |
| Warming rate | 34,480 °C/min | 24,000 °C/min |
{3}------------------------------------------------
# ADDITIONAL INFORMATION REQUESTED FOR CRYOPETTE 510(k) K100596
4
{4}------------------------------------------------
•
{5}------------------------------------------------
Discussion of Similarities/Differences: Overall the Cryopette and CryoTip are very similar Divices. Each consists of a Cryo-Straw with nearly identical dimensions and volume capacity. Each has a mechanism to aspirate fluid into the Cryo-Straw, heat seal the open ends of the Cryo-Straw, plunge the closed device into liquid nitrogen, and stored onto a CryoCane. The main two differences include the aspiration methods and the presence or absence of an outer metal sheath. The Cryopette has a built-in silicone bulb used to aspirate fluid into the Cryo-Straw whereas the CryoTip uses a detachable syringe. Both aspirate nate the the of the same volume of fluid just in a slightly different manner. The acriees aspirate roughty outer metal sheath that slides over the Cryo-Straw while being handled in storage whereas the Cryopette is stored without a metal cover. Mechanical testing to include Mouse Blastocyst Survival data demonstrate that each device consistently testing to include infouse masteryst bell as survival of blastocysts following vitrification and warming.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes across its body, representing the department's commitment to promoting health and well-being. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
DEC - 6 2010
Origio MidAtlantic Devices c/o Elisa D. Harvey, DVM, PhD Senior Regulatory Consultant CardioMed Device Consultants, LLC 18905 Celebrity Lane SANDY SPRING MD 20860
Re: K100596 Trade Name: Cryopette® Regulation Number: 21 CFR \$884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MOK Dated: November 18, 2010 Received: November 23, 2010
Dear Dr. Harvey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{7}------------------------------------------------
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Hubert Lemur MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{8}------------------------------------------------
# ADDITIONAL INFORMATION REQUESTED FOR CRYOPETTE 510(k) K100596
## Indications for Use
510(k) Number (if known): K100596 Device Name: Cryopette® Indications for Use:
DEC - 6 2010
The Cryopette is a cryopreservation storage device that is intended for use in vitrification procedures to contain and maintain human blastocyst stage embryos.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-the-counter use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K100596
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.