K100576 · Icu Medical, Inc. · FPA · Dec 17, 2010 · General Hospital
Device Facts
Record ID
K100576
Device Name
HIGH PRESSURE SETS-400 PSI
Applicant
Icu Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Dec 17, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The high pressure extension set is a device used as conduit tubing to deliver fluids from an electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices. Fluids may include: Contrast media (at 10mL/second) or other medications, and blood or blood products as the physician may prescribe. The fluids are administered through a cannula or catheter placed in a vein or artery. The incorporation of ICU proprietary connectors such as the CLAVE; MicroCLAVE; Y-CLAVE; and CLC2000 with the extension sets, allows for needleless connectivity to other devices via universal luer connection. The maximum pressure is 400 psi and other ancillary devices must also be rated for pressure up to 400 psi.
Device Story
High pressure extension sets serve as sterile fluid conduits between injectors and patient access devices (cannula/catheter). Input: fluids (contrast media, medications, blood) from manual or electronic injectors. Transformation: provides high-pressure-rated (up to 400 psi) pathway using integrated needleless connectors (CLAVE, MicroCLAVE, Y-CLAVE, CLC2000). Output: fluid delivery to patient. Used in clinical settings (e.g., cardiovascular procedures) by healthcare professionals. Benefits: prevents needlesticks via needleless connectivity; facilitates safe high-pressure contrast injection; allows custom lengths/configurations per facility needs.
Clinical Evidence
No clinical data. Bench testing only. Simulated use testing conducted per product specifications to validate 400 psi pressure rating and performance criteria. Biocompatibility testing performed per ISO 10993-1:2003.
Indicated for patients requiring fluid delivery (contrast media, medications, blood products) via electronic or manual injector through a venous or arterial cannula/catheter. For use by healthcare professionals. No specific age/gender restrictions.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
### 510(K) Summary
ICU MEDICAL INC.
4455 Atherton Drive Salt Lake City, Utah (801) 264 - 1332, Phone (801) 264 - 1755, Fax Tracy S. Best, Sr. Regulatory Affairs Specialist Preparation Date: 05/28/2010
DEC 1 7 2010
K100576
# Summary of Safety and Effectiveness for the:
| Trade Name: | High Pressure Sets - 400 psi |
|----------------------|------------------------------------------------------------------------------------------------------|
| Common Name: | High Pressure Infusion Devices |
| Classification Name: | Primary 21 CFR 880.5440, Class II Device, 80FPA<br>Secondary 21 CFR 870.4290, Class II Device, 74DTL |
#### Legally Marketed Predicate Devices for Substantial Equivalence:
*K061285 - Smart Site Valve Sets - Cardinal Health, Inc.
*K941978 - High Pressure Tubing, - Abbott Labs (submission acquired by ICU Medical Inc.) *K963749 - High Pressure Injection Linc, Maxim Medical, Inc.
#### Rationale for SE:
The ICU Medical High Pressure Sets are substantially equivalent to the predicate devices that offer needleless connectivity and international lucr compatibility. The use of ICU's legacy line of connectors is complimentary to these pre-assembled, individually packaged, and pre-sterilized high pressure sets. High pressure sets are used in cardiovascular procedures where contrast media is injected using control syringes or power injectors is common among all devices, which are safe for use up to 400 psi. These features are substantially equivalent to the predicate devices in that they also utilize male luers; female luers or locking luer connectors. Components of the devices are made from materials that are substantially cquivalent to the predicate devices listed above and this submission includes comprehensive biocompatibility testing for all device materials included in this submission. Additionally, all devices have a proven and invaluable history in the medical device market.
#### Description of Submitted Device:
The High Pressure Sets are a group of luer compatible connection devices that aid in the prevention of needlesticks and offer ease of use in the environment for the healthcare professional. The use of the CLAVE®; MicroCLAVE™; Y-CLAVE™; and CLC2000™ is a natural fit with the high pressure sets. Testing included with this submission, validates the 400 psi claim with these unique connectors. The functional technology of the high pressure set has not changed but is augmented by the addition of the ICU connectors. The devices are sterilized to ensure a 10ª SAL and they have been tested under the ISO 10993-1:2003 ~ Biological evaluation of medical devices ~ Part 1: Evaluation and resting and the results from these tests were found to be acceptable and are enclosed in this submission.
#### Non-Clinical Testing:
Testing (simulated use) has been conducted according to the product specification and that testing demonstrates that the device is safe and effective by meeting predefined criteria. Test data is part of this submission as evidence.
{1}------------------------------------------------
## Intended Uses of the ICU Medical High Pressure Sets:
The high pressure extension set is a device used as conduit tubing to deliver fluids from an electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices. Fluids may include: Contrast media (at 10mL/second) or other medications, and blood or blood products as the physician may prescribe. The fluids are administered through a cannula or catheter placed in a vein or artery. The incorporation of ICU proprietary connectors such as the CLAVE; MicroCLAVE; Y-CLAVE; and CLC2000 with the extension sets, allows for needleless connectivity to other devices via universal luer connection. The maximum pressure is 400 psi and other ancillary devices must also be rated for pressure up to 400 psi.
| Component | ICU Medical<br>High Pressure<br>Sets | Cardinal Health<br>Smart Site Valve<br>Sets | Abbott Labs<br>(submission<br>acquired by ICU<br>Medical) | Maxxim Medical<br>High Pressure<br>Injection Line |
|-------------------------|----------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------|---------------------------------------------------|
| Materials | Tubing-PVC<br>Luers - PC, PVC<br>& proprietary<br>needleless devices | Tubing - PVC<br>Luers - PC &<br>proprietary<br>needleless devices | Tubing - PVC<br>Luers - PC &<br>PVC | Tubing - PVC<br>Luers - Plastic (PC<br>assumed) |
| Functional<br>use | High pressure<br>infusion | High pressure<br>infusion | High pressure<br>infusion | High pressure<br>infusion |
| Sterilization<br>Method | Gamma | EtO | EtO and Gamma | Unknown |
| Packaging | Peel pouch | Peel pouch | Peel pouch | Peel pouch |
| 510(k)<br>Approval | This submission | K061285 | K941978 | K963749 |
### Technological Characteristics and Substantial Equivalence Table:
## Safety and Performance:
ICU Medical's commitment to quality and the high pressure sets that we currently make with the acquired K941978 submission are part of this submission. The use of these devices at pressures up to 400 psi has been tested and thoroughly proven to be safe and effective when coupled with the CLAVE; MicroCLAVE; Y-CLAVE; and COC2000. These high pressure sets are individually packaged and pre-sterilized in a peel type pouch. Devices are obtainable at custom lengths, according to facility needs and physician preference, including the type of luer ends used. Additionally, ICU Mcdical performs analysis and design verification icsting based on predecermined criteria, which is presented in the Performance Specifications in this submission. All testing meets these performance criteria as defined for the high-pressure. The design of the proprietary needleless connectors allow lucr connections as defined by ISO 594-1 and ISO 594-2 which is considered to be an industry standard.
## Conclusion:
The materials, performance, and operational features of the submitted devices and the predicate devices are substantially equivalent to one another and effective for their intended use.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Tracy Best Senior Regulatory Affairs Specialist ICU Medical, Incorporated 4455 Atherton Drive Salt Lake City, Utah 84123
DEC 17 2010
Re: K100576
Trade/Device Name: High Pressure Sets - 400 psi Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: October 30, 2010 Received: November 5, 2010
Dear Mr. Best:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Best
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
DEC 17 2010
#### 510(k) Number (if known): K100576
Device Name: High Pressure Sets - 400 psi
Indications for Use:
The high pressure extension set is a device used as conduit tubing to deliver fluids from an electronic or manual injector and may also be used as connector tubing to provide a sterile fluid pathway between two devices. Fluids may include: Contrast media (at 10mL/second) or other medications, and blood or blood products as the physician may prescribe. The fluids are administered through a cannula or catheter placed in a vein or artery. The incorporation of ICU proprietary connectors such as the CLAVE; MicroCLAVE; Y-CLAVE; and CLC2000 with the extension sets, allows for needleless connectivity to other devices via universal luer connection. The maximum pressure is 400 psi and other ancillary devices must also be rated for pressure up to 400 psi.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mih C. Chape 12/19/10
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K100576
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.