K100569 · Merit Medical Systems, Inc. · QEZ · Mar 11, 2011 · Cardiovascular
Device Facts
Record ID
K100569
Device Name
MERIT EMBOLECTOMY CATHETER MODEL ASAP100
Applicant
Merit Medical Systems, Inc.
Product Code
QEZ · Cardiovascular
Decision Date
Mar 11, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Merit Embolectomy Catheter is intended for use for the removal of fresh, soft emboli and thrombi from vessels of the arterial system. Not for use in cerebral vasculature.
Device Story
Merit Embolectomy Catheter is a dual-lumen rapid-exchange catheter designed for aspiration of fresh, soft emboli and thrombi from arterial vessels. Operated by physicians in clinical settings, the device is compatible with 0.014-inch guidewires. It features a radiopaque marker band for visualization and positioning marks for depth tracking. The device functions via manual aspiration to remove obstructions, facilitating blood flow restoration. It is a single-use, sterile device.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included ISO 10555-1/2 compliance, surface condition, force at break, leakage, dimensional verification, radiodetectability, flow, lubricious coating effectiveness, guide wire compatibility, stiffness, kink resistance, marker band retention, and simulated use in tortuous paths (tracking, crossing stents, and soft emboli aspiration). Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, genotoxicity, and hemocompatibility. Packaging performance was validated via accelerated aging and shipping simulations.
Technological Characteristics
Dual-lumen rapid-exchange catheter; 0.068"/1.73mm max OD; 140cm working length. Compatible with 0.014" guidewires. Features radiopaque marker band and positioning marks. Standards: ISO 10555-1/2 (catheters), ISO 594-2 (Luer fittings), ISO 11135 (EtO sterilization), ISO 10993-1 (biocompatibility). Mechanical design includes lubricious coating. Non-powered, manual aspiration device.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from arterial system vessels. Contraindicated for use in cerebral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Pronto V3 Extraction Catheter, Model 5003 (K063371)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" stacked on top of the words "U.S. FOOD & DRUG ADMINISTRATION".
February 8, 2021
Merit Medical Systems, Inc. Lindsay Martin Regulatory Affairs Specialist 1600 West Merit Pkwy. South Jordan, Utah 84095
Re: K100569
Trade/Device Name: Merit Embolectomy Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: OEZ
Dear Lindsay Martin:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 11, 2011. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075. Gregory. Oconnell(@FDA.HHS.gov.
Sincerely,
Digitally signed by Gregory W. Gregory W. O'connell -S O'connell -S Date: 2021.02.08 07:54:50 -05'00'
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MI) 20993-0002
Merit Medical Systems, Inc. c/o Ms. Lindsay Martin 1600 West Merit Parkway South Jordan, UT 84095
Re: K100569
Trade Name: Merit Embolectomy Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE Dated: February 26, 2011 Received: March 1, 2011
MAR 1 1 2011
Dear Ms. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
Page 2 - Ms. Linday Martin
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
R. Vachner
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and . Radiological Health
Enclosure
{3}------------------------------------------------
| 510(k) Number (if known): | K100569 |
|---------------------------|----------------------------|
| Device Name: | Merit Embolectomy Catheter |
**Indications for Use:**
The Merit Embolectomy Catheter is intended for use for the removal of fresh, soft emboli and thrombi from vessels of the arterial system.
Not for use in cerebral vasculature.
| Prescription Use | <u>X</u> | AND/OR | Over-The-Counter Use | _________________ |
|-----------------------------|----------|--------|------------------------|-------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
}
Anna R. Vahmen
(Division Sign-Off) Division of Čardiovascular Devices
510(k) Number_K1 0 0569
{4}------------------------------------------------
# Section 5 510(k) Summary for K100569
## General Provisions
| Submitter Name: | Merit Medical Systems, Inc. |
|----------------------|-------------------------------------------------|
| Address: | 1600 West Merit Parkway, South Jordan, UT 84095 |
| Telephone Number: | (801) 208-4187 |
| Fax Number: | (801) 253-6905 |
| Contact Person: | Lindsay Martin |
| Date of Preparation: | February 26, 2010 |
| Registration Number: | 1721504 |
#### Subject Device
Trade Name: ASAP Catheter Common/Usual Name: Merit ASAP Embolectomy Catheter Classification Name: Catheter, Embolectomy
## Predicate Device
| Trade Name: | Pronto V3 Extraction Catheter, Model 5003 |
|-------------------------|-------------------------------------------|
| Classification Name: | Catheter, Embolectomy |
| Premarket Notification: | K063371 |
| Manufacturer: | Vascular Solutions, Inc. |
### Classification
Class II per 21 CFR § 870.5150 DXE Product Code: Division of Cardiovascular Devices Review Branch:
#### Intended Use
The intended use of the Merit Embolectomy Catheter is to remove or aspirate fresh, soft emboli and thrombi from vessels of the arterial system. Not for use in cerebral vasculature.
#### Device Description
The Merit Embolectomy Catheter is a dual lumen rapid exchange catheter, compatible with 0.014"/0.36 mm guide wires with related accessories. The catheter has a maximum outer diameter of 0.068"/1.73mm and a working length of 140 cm. The catheter has a radiopaque marker band located approximately 2 mm proximal to the distal tip. The catheter has three (3) non-radiopaque positioning marks located approximately 90cm, 100 cm, and 110 cm proximal of the distal tip.
#### Safety & Performance Tests
No performance standards have been established under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, a battery of tests was performed according to protocols based on the requirements of the following documents, and were shown to meet the
{5}------------------------------------------------
acceptance criteria that were determined to be applicable to the safety and efficacy of the device:
@ ISO 10555-1, Sterile, single-use intravascular catheters, Part 1. General requirements.
@ ISO 10555-2, Sterile, single-use intravascular catheters, Part 2. Angiographic catheters
@ ISO 594-2, Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings
@ ISO 11135, Medical Devices - Validation and Routine Control of Ethylene Oxide Sterilization
@ ISO 10993-1, Biological Evaluation of medical Devices Part 1: Evaluation and Testing, and the FDA Modified ISO 10993 Test Profile
[The following is a list of all significant testing that was successfully completed:
Surface Condition
Force at Break
Freedom from Leakage 1 - Leak under Pressure
Freedom from Leakage 2- Aspiration Leak
Dimensional Verification
. Outside Diameter
- . Effective Length
- . Usable Length
- . Depth Marks
- . Tip Angle
Radio-detectability
Visual
Aspiration
Flow
Lubricious Coating Effectiveness - Coating Lubricity and Durability Test and Accelerated Aging Assessment
Coating Integrity- Dye Test
Lubricious Coating Length
Guide Wire Compatibility
Stiffness- Proximal, Mid, Distal Section
Kink Test- Proximal, Mid, Distal Section
Marker Band Retention
Rapid Exchange Lumen Attachment
Simulated Use Testing:
Tortuous Path- Tracking
Tortuous Path- Guide Wire
Tortuous Path- Crossing a deployed stent
Tortuous Path- Soft emboli aspiration
Biocompatibility Tests
- . Cytotoxicity
- Sensitization
- . Irritation
- Acute Systemic Toxicity
- . Pyrogenicity
- . Genotoxicity
- Hemocompatibility .
{6}------------------------------------------------
Packaging performance before and after exposure to accelerated aging and simulated shipping and handling conditions
- . bubble emission
- . dye penetration
- . seal peel tensile strength
- . burst strength
- . visual inspection
# Summary of Substantial Equivalence
Based on the indications for use, design, and safety and performance test results, the subject Merit Embolectomy Catheter meets the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device, the Pronto V3 Embolectomy Catheter manufactured by Vascular Solutions, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.