VERSALOK PEEK ANCHOR WITH ORTHOCORD SUTURE, MODELS 210818, 210819
K100532 · Depuy Mitek, A Johnson & Johnson Company · HWC · Mar 17, 2010 · Orthopedic
Device Facts
Record ID
K100532
Device Name
VERSALOK PEEK ANCHOR WITH ORTHOCORD SUTURE, MODELS 210818, 210819
Applicant
Depuy Mitek, A Johnson & Johnson Company
Product Code
HWC · Orthopedic
Decision Date
Mar 17, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
VERSALOCK Anchor is indicated for use in the following: Shoulder: Rotor Cuff Repair, Biceps Tenodesis Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Joint Capsule Closure Elbow: Biceps Tendon Reattachment
Device Story
Versalok Peek Anchor is a bone fixation fastener system used in orthopedic surgeries. The system consists of a PEEK anchor pre-mounted on an inserter shaft with an anvil, a threader tab, and optional Orthocord #2 suture. It is deployed using a reusable deployment gun. The device is operated by surgeons in a clinical setting. The anchor allows for knotless tissue fixation; the deployment gun features a tension wheel allowing the surgeon to control tissue tension. By eliminating the need for manual knot tying and providing precise tension control, the device facilitates soft tissue-to-bone reattachment, potentially improving surgical efficiency and clinical outcomes.
Clinical Evidence
No clinical data. Safety and performance were established through bench testing, including pull-out testing, shelf life, sterilization, and biocompatibility verification.
Technological Characteristics
Material: PEEK. Form factor: Threaded bone anchor pre-mounted on inserter shaft. Deployment: Reusable deployment gun with tension wheel. Connectivity: None. Sterilization: Provided sterile.
Indications for Use
Indicated for patients requiring soft tissue-to-bone fixation in shoulder (rotator cuff repair, biceps tenodesis), knee (MCL, LCL, and posterior oblique ligament repair, iliotibial band tenodesis, joint capsule closure), and elbow (biceps tendon reattachment) procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## 510(k) SUMMARY
KI00532
ﺍﻟﺘﻲ ﺗﻌﻠﻴﻘﺎﺕ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌﺔ ﺍﻟﻤﺘﻮﻗﻌ
## Versalok Peek Anchor
| Submitter's Name and Address: | DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MAR 1 7 2010 |
| Contact Person | Kristine Christo<br>Regulatory Affairs Project Manager<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br>Telephone: 508-828-3359<br>Facsimile: 508-977-6911<br>e-mail: kchristo@its.jnj.com<br>Prepared: 2/24/10 |
| Name of Medical Device | Classification Name: Fastener, Fixation, Nondegradable, Soft<br>Tissue Smooth or threaded metallic bone<br>fixation fasteners<br>Common/Usual Name: Bone Anchor<br>Proprietary Name: Versalok Peek Anchor |
| Substantial Equivalence | Versalok Peek Anchor is substantially equivalent to:<br>Trident Anchor (K060914)<br>Versalok Ti Anchor (K063478)<br>Arthrex Pushlock Anchor (K063479) |
| Device Classification | This device carries an FDA product code MBI and HWC, and is<br>classified as Fastener, Fixation, Nondegradable, Soft Tissue Smooth or<br>threaded metallic bone fixation fasteners under 21 CFR 888.3040. |
| Device Description | The Versalok Peek Anchor System includes the Versalok Peek Anchor,<br>which will be presented sterile, pre-mounted on an inserter shaft w/anvil |
510(k) Premarket Notification: Special
Versalok Peek Anchor ConfidentialVersalok Peek Anchor
{1}------------------------------------------------
.
イ
with a threader tab and suture. The System will be deployed with the use of a reusable Deployment Gun. The system is provided with and without a Orthocord#2 suture.
The VERSALOK Anchor System offers several additional clinical benefits as the anchor allows the surgeon to fix the tissue without tying knots and the delivery system allows the surgeon to control the tension placed on the tissue with the tension wheel of the deployment gun.
Technologies characteristics including material, design, packaging and indications are the same as the predicate cleared devices and use similar or identical material and packaging as the predicates.
Indications for Use
Nonclinical testing:
Safety and Performance
Predicate Comparison
The Versalok Peek Anchor is indicated for use in the following: Shoulder: Rotator Cuff Repair, Biceps Tenodesis, Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Joint Capsule Closure Elbow: Biceps Tendon Reattachment
Verification activities were performed on the implant or its predicates. Testing includes pull out testing, shelf life, sterilization and biocompatibility.
Results of performance and safety testing have demonstrated that the modified device is substantially equivalent to the predicate devices.
Based on the indications for use, technological characteristics, and comparison to predicate devices, the Versalok Peek Anchor has been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act.
The proposed Versalok Peek Anchor is compared to the predicate Trident Anchor (K063478), Versalok Anchor (K060914) and Arthrex Pushlock (K063479) below.
510(k) Premarket Notification: Special Versalok Peek Anchor
Confidential
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
MAR 1 7 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
DePuy Mitek % Ms. Kristine Christo Regulatory Affairs Project Manager 325 Paramount Drive Raynham, Massachusetts 02767
Re: K100532 Trade/Device Name: VERSALOCK Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC, MBI Dated: February 24, 2010 Received: February 25, 2010
Dear Ms. Christo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Kristine Christo
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):
K100537
Device Name:
VERSALOCK Anchor is indicated for use in the following:
Shoulder: Rotor Cuff Repair, Biceps Tenodesis Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Joint Capsule Closure Elbow: Biceps Tendon Reattachment
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krone for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K100532
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.