K100508 · Carematix, Inc. · DRG · May 5, 2010 · Cardiovascular
Device Facts
Record ID
K100508
Device Name
CAREMATIX WELLNESS SYSTEM, MODEL CWS-5000.1-B
Applicant
Carematix, Inc.
Product Code
DRG · Cardiovascular
Decision Date
May 5, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
The Carematix™ Wellness System is a physiological monitoring system. The system collects, accumulates and transmits patient vital signs and other physiological data from a patient who may be remote from the healthcare practitioner to the practitioner. It is intended for patient home use for the following and can record physiological information such as: Non-invasive blood pressure measurement; Pulse rate measurement; Blood glucose level using a Glucometer; Blood hemoglobin oxygen saturation (%SpO2) using a digital Pulse Oximeter; Prothrombin Time (PT) and International Normalized Ratio (INR) measurement using an in-home coagulation measurement system; Peak Expiratory Flow Rate (PEFR) and Forced Expiratory Volume (FEV) measurements using an electronic peak flow meter; Patient weight using a stand-on electronic scale The results of these measurements are transmitted to a computer monitoring station in a clinical setting via common telephone lines from the patient's home setting.
Device Story
System gathers/transmits patient physiological data from home to clinical facility. Components: wireless RF transmitter adapters, communication hub, internet server. Adapters connect to existing monitoring devices (blood pressure, pulse, glucose, SpO2, PT/INR, FEV/PEFR, weight). Hub collects/stores data from adapters; transmits via PSTN, LAN, or cellular to internet server. Caregivers access server to track data, graph trends, note variances, set alert criteria, and receive notifications. Facilitates remote patient management; enables timely clinical intervention based on transmitted parameters.
Clinical Evidence
Bench testing only. Verification and validation tests conducted per Good Manufacturing Practices; system conforms to IEC 60601-1 safety requirements.
Technological Characteristics
System comprises RF transmitter adapters, communication hub, and internet server. Connectivity via PSTN, LAN, or cellular. Complies with IEC 60601-1 safety standards. Data transmission of physiological parameters from home-based monitoring devices to clinical monitoring station.
Indications for Use
Indicated for patients requiring remote monitoring of vital signs and physiological data, including blood pressure, pulse, blood glucose, SpO2, PT/INR, respiratory function (PEFR/FEV), and weight. Intended for home use with data transmission to clinical settings.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
## 510(k) Summary
## Carematix™ Wellness System
# MAY - 5 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
### SUBMITTER
Carematix™, Inc. 209 W Jackson Blvd, Suite 800 Chicago, IL 60606
## Contact Person
Sukhwant Khanuja, Ph. D. Chief Executive Officer Carematix™, Inc. 209 W Jackson Blvd, Suite 800 Chicago, IL 60606 Phone: (312) 627-9300 e-mail: skhanuja@carematix.com
### Date Prepared:
February 15, 2010
NAME OF DEVICE
Carematix™ Wellness System
CLASSIFICATION NAMES
## Radiofrequency Physiological Signal Transmitters and Receivers
DEVICE CLASSIFICATION
Regulatory Class: Product Code: Classification Panel: Regulation Number:
Class II DRG Cardiovascular Device Panel 21 CFR 870.2910
#### PREDICATE DEVICES
K040966; K073038- Carematix™ Wellness System, Carematix, Inc.
## DESCRIPTION OF DEVICE
The Carematix™ Wellness System is intended to gather and transmit patient data from the home site or remote location to caregivers at a clinical facility where it provides supplemental information for the care of the patient.
The Carematix™ Wellness System consists of wireless radiofrequency transmitter adapters, a communication hub or receiving station, and an internet server:
- The radiofrequency transmitter adapter is connected, either wired directly or to . a serial port, of monitoring devices currently in distribution having capability to monitor patient parameters for blood pressure, pulse rate, blood sugar, blood oxygen saturation, PT/INR, and FEV/PEFR.
- � The communications hub, or receiving station, collects and stores data transmitted from each of the radiofrequency transmitter adapters or wired device and transmits patient data to the internet server via PSTN. LAN. cellular, wireless etc.
- The internet server receives the patient data from the home setting or remote . location where it is made available to the caregiver to track, graph, note
{1}------------------------------------------------
Special 510(k) Notification- Response for additional Information
variances, set alert criteria, and receive notifications when parameters are outside the criteria set.
## Indications For Use
The Carematix™ Wellness System is a physiological monitoring system. The system collects, accumulates and transmits patient vital signs and other physiological data from a patient who may be remote from the healthcare practitioner to the practitioner. It is intended for patient home use for the following and can record physiological information such as:
Non-invasive blood pressure measurement;
Pulse rate measurement;
Blood glucose level using a Glucometer;
Blood hemoglobin oxygen saturation (%SpO2) using a digital Pulse Oximeter;
Prothrombin Time (PT) and International Normalized Ratio (INR) measurement using an inhome coagulation measurement system;
Peak Expiratory Flow Rate (PEFR) and Forced Expiratory Volume (FEV) measurements using an electronic peak flow meter;
Patient weight using a stand-on electronic scale
The results of these measurements are transmitted to a computer monitoring station in a clinical setting via common telephone lines from the patient's home setting.
## PERFORMANCE TESTING
Verification and Validation tests were conducted for the Carematix Wellness System in accordance with Good Manufacturing Practices and the results met acceptable criteria. The Carematix™ Wellness System also conforms to requirements of IEC 60601-1 and we consider it safe and
### CONCLUSION
The Carematix™ Wellness System is substantially equivalent to the following 510(k) cleared devices:
Carematix™ Wellness System cleared under K040966 on June 2, 2004, and Carematix™ Wellness System cleared under K073038 on Jan 11, 2008.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - GSA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - GSA" arranged around the circumference. Inside the circle is a stylized image of an abstract human figure with outstretched arms, resembling a bird in flight. The figure is composed of three curved lines that create a sense of movement and openness.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
Carematix, Inc c/o Sukhwant Khanuja, PhD CEO 209 W Jackson Blvd, Suite 800 Chicago, IL 60606
MAY - 5 2010
Re: K100508
> Trade/Device Name: Carematix Wellness System Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II (two) Product Code: DRG Dated: April 1, 2010 Received: April 5, 2010
Dear Dr. Khanuja:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Khanuja
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
W.W. Watt
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number K100508
# Device Name: CarematixTM Wellness System
# Indications for Use:
The Carematix™ Wellness System is a physiological monitoring system. The system collects, accumulates and transmits patient vital signs and other physiological data from a patient who may be remote from the healthcare practitioner. It is intended for patient home use for the following and can record physiological information such as:
Non-invasive blood pressure measurement;
Pulse rate measurement;
Blood glucose level using a Glucometer;
Blood hemoglobin oxygen saturation (%SpO2) using a digital Pulse Oximeter;
Prothrombin Time (PT) and International Normalized Ratio (INR) measurement using an in-home coagulation measurement system:
Peak Expiratory Flow Rate (PEFR) and Forced Expiratory Volume (FEV) measurements using an electronic peak flow meter;
Patient weight using a stand-on electronic scale
The results of these measurements are transmitted to a computer monitoring station in a clinical setting via common telephone lines from the patient's home setting.
Prescription Use (per 21 CFR 801.109)
and/or
Over-the-counter Use
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
ivision of Cardiovascular Devices
510(k) Number K100508
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.