The V-Go 20 Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 20 Units of insulin in one 24hour time period (0.83U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin. The V-Go 30 Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 30 Units of insulin in one 24hour time period (1.25U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin. The V-Go 40 Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 40 Units of insulin in one 24hour time period (1.67U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin.
Device Story
V-Go is a mechanical, non-electronic, single-use disposable insulin infusion pump. Patient fills device with insulin using EZ Fill accessory; secures to skin via adhesive-backed foam pad. Device uses spring-loaded hydraulic mechanism for continuous basal infusion at fixed rates (20, 30, or 40 Units/day). User-operated bolus button provides 2-Unit increments (up to 36 Units/day) via manual hydraulic displacement. Integrated stainless steel 'J' needle delivers insulin subcutaneously. Device includes reservoir window for monitoring. Used by patients to manage diabetes; provides continuous basal and on-demand bolus insulin delivery without external controllers or batteries.
Clinical Evidence
No clinical performance data required. Design validation completed via human factors simulated use and clinical evaluation testing. Bench testing confirmed mechanical performance, accuracy (+/- 10%), biocompatibility (ISO 10993-1), and sterility (SAL 10^-6).
Technological Characteristics
Mechanical, non-electronic, single-use infusion pump. Materials: stainless steel 'J' needle, biocompatible plastics. Energy source: internal compressed spring for basal; manual force for bolus. Dimensions: 84mm x 47mm x 13mm. Sterilization: Ethylene Oxide (ANSI/AAMI ST67). Connectivity: None. No software.
Indications for Use
Indicated for adult patients requiring insulin for management of diabetes mellitus. Provides continuous subcutaneous basal insulin infusion and on-demand bolus dosing.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
valeritas
800 Tumpike Road Shrewsbury, MA 01545
Phone 508.845.1177 508.842.8928 ww.valeritas.com
#### 510(k) Summary
Owner's Name:
Valeritas, Inc. 750 Route 202 South, Suite 100 Bridgewater, NJ 08807
Company Contact:
Scott Huie Vice President. Operations Valeritas. Inc. 800 Boston Turnpike Shrewsbury, MA 01545 (508) 845-1177 X222 shuie@valeritas.com
.DEC 1 2010
K100504
Date Summary Prepared: 16 November 2010
Trade Name:
Common Name:
Classification Name:
Classification regulation:
**Substantial Equivalence:**
Device Description:
Class II: 21 CFR 880.5725 Panel: 80, LZG External Insulin Infusion Pump FPA Intravascular Administration Set
Valeritas V-Go™ Insulin Delivery Device
Disposable Insulin Delivery Device
Substantial Equivalence: The Valeritas V-Go Disposable Insulin Delivery Device is substantially equivalent to the BioValve Insulin Delivery System (BIDS) K050971, cleared by this Center on 16 August 2005
Pump, Infusion, Insulin/Set, Administration, Intravascular
The Valeritas V-Go Disposable Insulin Delivery Device is a mechanical (no electronics), self-contained, sterile, patient fillable, single-use disposable insulin infusion device with an integrated stainless steel subcutaneous needle.
It is designed for the subcutaneous infusion of insulin for the management of diabetes mellitus in persons requiring insulin. After filling the V-Go with insulin using the EZ Fill, the device is secured to the patient's skin over the infusion site with an adhesivebacked foam pad, which is attached to the back of the pump. Once activated, the V-Go delivers a continuous infusion of insulin at a fixed rate. The device also allows the user to initiate bolus
{1}------------------------------------------------
injections to supplement their daily basal insulin requirements. Three device models (20, 30 and 40 Units/day) will be available to address the different basal and bolus requirements of each patient. A window in the top of the pump allows the user to see into the reservoir to check the drug and to monitor the progress of the infusion.
#### Intended Use:
#### V-Go 20:
The V-Go Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 20 Units of insulin in one 24hour time period (0.83U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin.
## V-Go 30:
The V-Go Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 30 Units of insulin in one 24hour time period (1.25U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin
## V-Go 40:
The V-Go Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 40 Units of insulin in one 24hour time period (1.67U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin
## Technological Characteristics:
The V-Go has two delivery mechanisms, one for basal infusion and one for bolus delivery. Both operate by providing force against a plunger, which is inserted into a cylindrical reservoir filled with high viscosity fluid. This force is directly translated against the rear of the delivery piston at the rear of the insulin reservoir.
The Basal Hydraulic Pump operates by applying a constant force by means of a compressed spring. The spring pushes against a disk inserted internally at the rear of the cylindrical fluid filled basal reservoir, forcing the fluid through a restrictive orifice. This displacement translates into displacement of the medicinal delivery piston, providing a continuous basal infusion. The basal delivery
2
{2}------------------------------------------------
rate can be varied by the substitution of a different orifice into the device during assembly.
The Bolus Hydraulic Pump operates by applying a force by means of a manually depressed actuator (Bolus Button). Manual actuation allows the displacement of a disk inserted internally at the rear of the cylindrical fluid-filled bolus reservoir in pre-set increments that will displace the volume of fluid and the medicinal delivery piston. The manual actuation requires two steps. First, the user must press the Bolus Release Button. Displacement of the Bolus Release allows the Bolus Button to "pop-out" so that it protrudes slightly from the end of the pump. The user then depresses the Bolus Button, which displaces the Bolus Piston and delivers a pre-set bolus increment. This process is repeated until the desired volume has been delivered or the labeled number of actuations are used up. The Bolus Release Button extends from the pump and locks preventing further use as there are no additional bolus actuations remaining.
#### Substantial Equivalence:
The Valeritas V-Go is substantially equivalent to the predicate device, the BioValve Insulin Delivery System (BIDS), cleared by this center on 16 August 2005 (K050971). Please note that BioValve contributed the technology to Valeritas in 2006 and that the BioValve device and the Valeritas V-Go device have the same intended use and performance characteristics. The primary differences relate to minor material modifications and an improved method and accessory (EZ Fill) for filling the V-Go pump. In addition, the V -Go comes in three set basal rates (20, 30 and 40 Units/day) and a set incremental bolus (2 Units/actuation, with a maximum of 36 Units/day) whereas the BioValve device had a specified basal range of 14.4 to 57.6 Units/day and bolus increments from 2 to 4 Units.
#### Non-Clinical Performance Data:
## Physical Specifications
Maximum Dimensions: Weight: Cannula: Pump Mechanism: Interfaces:
. 84 mm x 47 mm x 13 mm 20 - 50 grams Stainless Steel' J' shaped needle (J-needle) Spring and Manual Hydraulic Displacement Mechanical Buttons
{3}------------------------------------------------
Operational Specifications
| Operational Specifications | | | |
|----------------------------|--------------------------------|--------------------------------|--------------------------------|
| Pump Model | V-Go 20 | V-Go 30 | V-Go 40 |
| Reservoir volume | 560 ( $56 \mu L$ of insulin) | 660 ( $66 \mu L$ of insulin) | 760 ( $76 \mu L$ of insulin) |
| Basal Rate | 8.3 ( $0.6 IU/hr$ ) | 12.5 ( $1.2 IU/hr$ ) | 16.7 ( $1.8 IU/hr$ ) |
| Bolus Increments | 20 18 Actuations | 20 18 Actuations | 20 18 Actuations |
| Nominal Bolus<br>Volume | 360 (36 IU) | 360 (36 IU) | 360 (36 IU) |
| Basal Volume<br>(24 Hours) | 200 (20 IU) | 300 (30 IU) | 400 (40 IU) |
| Accuracy | +/- 10% | +/- 10% | +/- 10% |
## Environmental Specifications - Per AAMI ID26, IPX8
| Temperature: | +5 to +37°C |
|---------------------------|------------------------------------------|
| Humidity: | 20 - 90% |
| Water Ingress Protection: | IPX8 (submerged at 1 meter for 24 hours) |
| Atmospheric Testing: | 0-10,000 feet |
## Biocompatibility - Per ISO 10993-1
Drug and skin contact materials passed all testing per ISO 10993-1. Drug Contact materials were tested and found to be compatible with insulin for the duration of use. Please see the package insert for specific insulin information.
## Sterilization and Package Stability
The V-Go and EZ Fill are sterilized using Ethylene Oxide to a Sterility Assurance Level (SAL) of 10 - per ANSI/AAMI ST67:2003(R)2008, Sterilization of Health Care Products - Requirements for products labeled "Sterile".
The V-Go and EZ Fill have been tested as pyrogen free using USP <85> Bacterial Endotoxin Testing Gel Clot Method with an acceptable maximum limit of 20EU/device.
Packaging consists of a thermoformed plastic tray with a Tyvek heat sealable lidstock. Packaging validation demonstrated that the packaging can maintain sterility integrity for up to 36 months per AAMI/ISO 11607-1:2006, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems.
{4}------------------------------------------------
## Drug Stability
Humalog has been tested in the V-Go and has been demonstrated to be stable for up to 24 hours storage followed by 24 hours wear. The EZ Fill has been demonstrated to be acceptable for filling Humalog for up to 30 days use.
## Clinical Performance Data:
No clinical performance data is required to validate the intended uses and user needs of the system. Design validation is completed by human factors simulated use and clinical evaluation testing.
## Conclusion:
Results of performance testing demonstrated substantial equivalence of the V-Go Disposable Insulin Delivery System to the predicate 510(k) device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three lines representing its wings and three curved lines below representing its tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Scott Huie Vice President, Operations Valeritas, LLC 800 Boston Turnpike Shrewsbury, Massachusetts 01545
DEC 1 2010
Re: K100504
Trade/Device Name: V-Go™ 20 Disposable Insulin Delivery Device, V-Go™ 40 Disposable Insulin Delivery Device & V-Go™ 30 Disposable Insulin Delivery Device Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: November 19, 2010 Received: November 22, 2010
Dear Mr. Huie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
## Page 2- Mr. Huie
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain.other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
James Duzonsted
For
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{7}------------------------------------------------
# 4. Indications for Use Statement
#### DEC 1 2010
510(k) Number (if known): K100504
Device Name:
V-GoTM 40 Disposable Insulin Delivery Device
Indications For Use:
The V-Go 40 Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 40 Units of insulin in one 24hour time period (1.67U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aled C. Chapman 12/1/2010
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
510(k) Number: K100504
{8}------------------------------------------------
## 4. Indications for Use Statement
510(k) Number (if known): K100504
Device Name:
V-GoTM 30 Disposable Insulin Delivery Device
Indications For Use:
The V-Go 30 Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 30 Units of insulin in one 24hour time period (1.25U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
 12/1/2010
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
Page 1 of 1
510(k) Number: K100504
{9}------------------------------------------------
# 4. Indications for Use Statement
510(k) Number (if known): K100504
Device Name:
V-GoTM 20 Disposable Insulin Delivery Device
Indications For Use:
The V-Go 20 Disposable Insulin Delivery Device is indicated for continuous subcutaneous infusion of 20 Units of insulin in one 24hour time period (0.83U/hr) and on-demand bolus dosing in 2-Unit increments (up to 36 Units per one 24-hour time period) in adult patients requiring insulin
Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | 12/1/2010 |
|--|-----------|
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| Page 1 of | 1 |
|----------------|---------|
| 510(k) Number: | K100504 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.