K100500 · Keeler Instruments, Inc. · HJO · Mar 9, 2010 · Ophthalmic
Device Facts
Record ID
K100500
Device Name
PSL CLASSIC SLIT LAMP
Applicant
Keeler Instruments, Inc.
Product Code
HJO · Ophthalmic
Decision Date
Mar 9, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The slit lamp is an instrument consisting of a light source that can be focused to shine a thin sheet (slit) of light into the eye. It is used in conjunction with a biomicroscope. The lamp facilitates an examination of the anterior segment, or frontal structures and posterior segment, of the human eye, which includes the eyelid, sclera, conjunctiva, iris, natural crystalline lens, and cornea. The binocular slit-lamp examination provides stereoscopic magnified view of the eve structures in detail, enabling anatomical diagnoses to be made for a variety of eye conditions. The Slit Lamp should be used only by suitably trained and authorised healthcare personnel. The device and accessories are indicated as noninvasive aid in the examination and diagnosis of eye conditions and indications are the same as those claimed for predicate device
Device Story
Keeler PSL Classic Slit Lamp is a portable, hand-held illuminated biomicroscope system with a desk-mounted base charger. It uses a 15W filament lamp to project light onto the eye; light is relayed to the corneal plane via a lens and prism system. The device features adjustable magnification (10x and 16x), adjustable eyepieces, and filters/slits to vary beam color and shape. Operated by trained healthcare personnel in clinical settings, the device provides a stereoscopic magnified view of eye structures (eyelid, sclera, conjunctiva, iris, lens, cornea). The clinician uses the visual output to perform anatomical diagnoses of eye conditions or trauma. The device is powered by an internal rechargeable battery, with intensity controlled by a rheostat.
Clinical Evidence
Performance evaluated via bench testing (optical radiation safety, electrical safety, electromagnetic compatibility) and clinical field trials. Testing confirmed the device met acceptance criteria for operating range and usability. No specific clinical metrics (e.g., sensitivity/specificity) were reported; performance was judged equivalent to the predicate based on comparative evaluation.
Technological Characteristics
Portable hand-held biomicroscope; 15W filament lamp illumination; lens and prism optical relay system; 10x/16x magnification; rechargeable battery power; rheostat intensity control. Compliant with ISO 9001:2000 and BS EN 13485:2003.
Indications for Use
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting structural properties of the anterior eye segment in patients requiring ophthalmic evaluation.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
K100500
MAR - 9 2010
# 510(k) Summary
ﻲ ﺃ
## Submitter's information
The submitter of this pre-market notification is:
Mr. Eugene VanArsdale (Marketing Manager). Keeler Instruments Inc 456, Parkway, Broomall, PA 19008 100/1 antituty, Broomail, FT-10
Company Phone No:
Company Fax No: (610) 353 4350 (610) 353 7814
Date summary prepared: January 04, 2010
# Device Identification
| Device Trade Name: | Keeler PSL Classic Slit Lamp |
|-----------------------|------------------------------|
| Common Name: | Portable Slit Lamp |
| Class: | II |
| Classification Panel: | 86 |
| Product Code: | HJO |
| Regulation Number: | 886.1850 |
## Device description
The Keeler Portable Slit Lamp comprises a rechargeable hand held portable illuminated biomicroscope system and a desk mounted base charger unit that is powered from a low voltage (12V) power supply.
The hand held unit incorporates a built-in rechargeable battery powering the illumination system. The illumination system and fixation targets are activated using a single/double click trigger located on the front of the grip/handle. To increase or reduce the light intensity there is a rheostat located below the eyepieces on the rear of the grip /handle.
The 10 x and 16 x magnifications optical system is controlled using the lever located under the adjustable eyepieces.
## Basic Principle of operation
The illumination system and fixation targets are activated using a single/double click trigger located on the front of the grip/handle. The construction of this slit lamp is such that light from a 15W filament lamp is relayed to the corneal plane via a lens and prism, the corneal plane being situated at 50mm from the output face of the prism. Filters or slits can be placed in the beam to vary the output in terms of colour and illuminated shape. Pupillary distance between the eye pieces, dioptric power of each eye piece and magnification can be adjusted as per requirement.
This instrument is used for non-invasive illumination, magnification and observation of the human eye.
## Intended use
The slit lamp is an instrument consisting of a light source that can be focused to shine a thin sheet (slit) of light into the eye. It is used in conjunction with a biomicroscope. The lamp facilitates an examination of the anterior segment, or frontal structures and posterior segment, of the human eye, which includes the eyelid, sclera, conjunctiva, iris, natural crystalline lens, and cornea. The binocular slit-lamp examination provides stereoscopic magnified view of the eve structures in
{1}------------------------------------------------
detail, enabling anatomical diagnoses to be made for a variety of eye conditions. The Slit Lamp should be used only by suitably trained and authorised healthcare personnel.
The device and accessories are indicated as noninvasive aid in the examination and diagnosis of eye conditions and indications are the same as those claimed for predicate device
# Substantial equivalence
ﻟﺔ ﺍﻟ
Predicate Device: KOWA SL-15 510(k) No: K063640 Manufacturer: KOWA Company Ltd.
The Kowa Portable Slit Lamp SL-15 is chosen as a substantially equivalent device, with its similar intended use and its classification as a slit lamp. The predicate device is a hand-held slit lamp and is equipped with rechargeable battery. Similar to predicate device, Keeler Portable Slit Lamp is equipped with a rechargeable battery.
The Keeler Portable slit lamp has following similarities to those of the predicate device:
-has the same indication for use,
-uses the same operating principle,
-incorporates the same basic optical design.
#### Technological characteristics
The technological characteristics are the same or similar to those found with the predicate device where eye is examined by projecting light onto it.
#### Declaration of device design
The Keeler Portable Slit Lamp has been designed using latest methods of evaluation and validation to ensure the risk of physical injury whilst performing the intended functions has been reduced as much is practicably possible.
## Keeler Quality Approvals
:
Keeler is an ISO 9001:2000 and BS EN 13485:2003 compliant company and is obliged to ensure that our operating and design practices fully comply. As part of the design process, all necessary risk analysis, risk management, design verification and validation are conducted to ensure that the safety and effectiveness of product designs either meet or surpass the product specification and comply with the relevant standards.
## Summary of Performance Testing
The performance of Keeler PSL Classic has been evaluated in Optical Radiation Safety Test, Electrical Safety Test, Electromagnetic Compatibility Test, comparative study with its predicative devices and field trial in clinical environment. The units under test met the acceptance criteria of the operating range and usability.
Based on the results of performance testing, the performance of Keeler PSL Classic is judged to be as safe, as effective and performs as well to predicate devices.
{2}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
MAR ~ 9 2010
Keeler Instruments, Inc. c/o Mr. Jeff Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Dr. Research Triangle, NC 27709
Re: K100500
Trade Name: Keeler PSL Classic Slit Lamp Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slitlamp biomicroscope Regulatory Class: Class II Product Code: HJO Dated: February 19, 2010 Received: February 22, 2010
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
# Page 2 - Mr. Jeff Rongero
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K100500
Device Name: Keeler PSL Classic Slit Lamp
Indications For Use:
Keeler PSL Classic Slit Lamp is an AC-powered Slit lamp biomicroscope. It is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dege
3/5/2010
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K100500
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.