K100487 · Eclipsemed Global, Inc. · ONG · Nov 29, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100487
Device Name
EQUINOX CO2 LASER SYSTEM
Applicant
Eclipsemed Global, Inc.
Product Code
ONG · General, Plastic Surgery
Decision Date
Nov 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Equinox CO 2 laser when used in traditional non-fractionated scanner mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues. The Equinox CO 2 laser when used in fractionated mode is indicated for ablative skin resurfacing.
Device Story
Equinox CO2 Laser System (10,600nm wavelength) delivers thermal energy to epidermis and upper dermis. System comprises main body, color touch screen, arm, hand-piece, and foot switch. In non-fractionated mode, used for soft tissue surgery (incision, excision, ablation, vaporization, coagulation). In fractionated mode, uses scanning optics to deliver patterned thermal energy for skin resurfacing. Operated by clinicians in clinical settings. Output is laser energy; healthcare providers use device to perform surgical or dermatological procedures. Benefits include precise tissue interaction and controlled thermal damage for resurfacing.
Clinical Evidence
Bench testing only. Histology study performed on human arm tissue to evaluate thermal damage zones (depth/width) at energy levels up to 70mJ per microbeam. Measurements recorded at day 0, day 3, and day 14 post-treatment to assess healing response. No clinical study data provided.
Technological Characteristics
10,600nm CO2 laser; scanning optics for fractional delivery; system includes main body, color touch screen, arm, hand-piece, and foot switch. Class II device (21 CFR 878.4810).
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues (non-fractionated mode) and ablative skin resurfacing (fractionated mode).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Attachment 5 510(K) Summary Equinox CO2 Laser System
Pg 1 of 2
This 510(K) Summary of safety and effectiveness for the Equinox CO2 laser System is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant: | Eclipsemed Global, Inc |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 16850 Dallas Parkway<br>Dallas, TX 75248<br>972-380-2911 - phone<br>972-380-2953 - fax |
| Contact Person: | Mr. Tom O'Brien |
| Telephone: | 972-380-2911 - Phone |
| Fax: | 972-380-2953 - Fax |
| Email: | tobrien@eclipsemed.com |
| Preparation Date: | February 15, 2010 |
| Device Trade Name: | Equinox CO2 Laser System |
| Common Name: | CO2 Laser |
| Classification Name: | Instrument, Surgical, Powered, laser<br>79-ONG, 21 CFR 878-4810 |
| Legally Marketed Device: | Predicate Cynosure Affirm CO2 Laser<br>K081424 |
| Description of the Equinox CO2 laser | The Equinox CO2 laser has a wavelength of 10,600nm. CO2<br>fractional laser uses scanning optics to deliver a pattern of thermal<br>energy to the epidermis and upper dermis. Device accessories<br>include tip attachments. This system consists of main body, color<br>touch screen, Arm, hand-piece and Foot switch. |
| Intended use of the Equinox CO2 laser | The Equinox CO 2 laser when used in traditional non-fractionated<br>scanner mode is indicated for incision, excision, ablation,<br>vaporization, and coagulation of body soft tissues.<br><br>The Equinox CO 2 laser when used in fractionated mode is<br>indicated for ablative skin resurfacing. |
| Performance Data: | Histology data was submitted to support clearance of the device in<br>fractionated mode. The device was used on a human arm with<br>energy up to 70mJ per microbeam and the target area was<br>biopsied to evaluate the effect. The data was to show the depth<br>and width of thermal damage zones and healing response over<br>time. The following table shows the width and depth of<br>penetration on the day of treatment, day 3 post treatment and day<br>14 post treatment. |
{1}------------------------------------------------
K100487 pg 2 of 2
## Attachment 5 510(K) Summary Equinox CO2 Laser System
| | 10 mJ | | 40 mJ | | 70 mJ | |
|-----|-----------|-----------|-----------|-----------|-----------|-----------|
| Day | Depth | Width | Depth | Width | Depth | Width |
| 0 | 115.01 µm | 347.30 µm | 200.24 µm | 358.75 µm | 272.56 µm | 362.13 µm |
| 3 | 94.78 µm | 293.71 µm | 164.50 µm | 313.20 µm | 217.22 µm | 334.49 µm |
| 14 | X | X | X | X | 43.87 µm | 95.28 µm |
Results of Clinical Study:
None
Conclusion:
The Equinox CO2 Laser System is substantially equivalent to the Cynosure Affirm CO2 Laser cleared in K081424. The Equinox CO2 Laser System in non-fractionated scanner mode is substantially equivalent in terms of indication for use and technology based on similar technical characteristics. The Equinox CO2 Laser System in fractionated mode is substantially equivalent to the predicate device in terms of indications for use and technology based on similar technology characteristics as well as the provided histology data.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all capital letters and is evenly spaced around the circle.
"NOV 2 9 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Eclipsemed Global, Inc. % Ms. Connie Hoy 16850 Dallas Parkway Dallas, Texas 75248
Re: K100487
Trade/Device Name: Equinox CO2 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG, GEX Dated: November 19, 2010 Received: November 22, 2010
Dear Ms. Hoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 - Ms. Connie Hoy
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours
Sincerely, yours,
Mark A. Millen
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
pg 1 of (
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
NOV 2 9 2010
Device Name : Equinox CO2 Laser System
The Equinox CO 2 laser when used in traditional non-fractionated scanner mode is indicated for incision, excision, ablation, vaporization, and coagulation of body soft tissues.
The Equinox CO 2 laser when used in fractionated mode is indicated for ablative skin resurfacing.
Prescription Use XX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurr e of Device Evaluation (ODE) (Division Sign-Off) Page 1 of 1 Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.