K100481 · Smith & Nephew, Inc. · MEH · Jul 15, 2010 · Orthopedic
Device Facts
Record ID
K100481
Device Name
SMITH & NEPHEW MDF MODULAR SLEEVE
Applicant
Smith & Nephew, Inc.
Product Code
MEH · Orthopedic
Decision Date
Jul 15, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew MDF Revision Hip System components are intended for single use only and are to be implanted without bone cement.
Device Story
MDF Modular Sleeve is a component of the MDF Revision Hip System; functions as a modular adapter between femoral stem and neck. Manufactured from Ti-6Al-4V titanium alloy; features grit-blasted surface with plasma-sprayed hydroxylapatite (HA) coating to facilitate cementless fixation. Implanted by orthopedic surgeons during primary or revision hip arthroplasty. Device locks onto proximal end of femoral stem to restore hip joint function. Benefits include modularity for revision surgery and biological fixation via HA coating. No software or electronic components.
Clinical Evidence
Bench testing only. Performance evaluated per FDA guidance for femoral stem prostheses and mechanically locked modular components. Testing included environmental corrosion fatigue, pre-fatigue disassociation, and post-fatigue disassociation. No clinical data provided.
Technological Characteristics
Material: Ti-6Al-4V (ASTM F1472). Surface: Grit-blasted with plasma-sprayed hydroxylapatite (HA) coating (ASTM F1185). Form factor: Modular sleeve component for hip revision system. Fixation: Cementless. Energy source: None (mechanical).
Indications for Use
Indicated for primary and revision hip surgery in patients with hip damage due to trauma, inflammatory joint disease (e.g., rheumatoid arthritis), or noninflammatory degenerative joint disease (e.g., osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, nonunion, femoral neck/trochanteric fractures, femoral osteotomy, Girdlestone resection, fracture dislocation, or deformity). For uncemented use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
K100481 *"1/2
Smith & Nephew, Inc. Summary of Safety and Effectiveness MDF Revision Hip System Line Additions JUL 1 5 2010
Date of Summary: 07/14/2010
### Contact Person and Address
Natalie P. Williams Regulatory Affairs Specialist Smith & Nephew, Inc. 1450 Brooks Road Memphis, TN 38116 (901)399-5161
Name of Device: Smith & Nephew MDF Modular Sleeve Common Name: Modular Sleeve Device Classification Name and Reference: 21 CFR 888.3353 Hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis Device Class: II Panel Code: Orthopaedics/87 MEH
### Device Description
The MDF Revision Hip System is comprised of a stem, modular neck, and modular sleeve component. The components of the revision hip system modularly connect together to form the complete MDF Revision Hip System construct. The Extra Small (XSM) Modular Sleeve line additions mate with similar size femoral stems in the MDF Revision Hip System and lock onto the proximal end of the stem. The XSM Modular Sleeves are manufactured from titanium alloy (Ti-6Al-4V). The XSM MDF Modular Sleeve line additions are grit-blasted and contain a threaded surface with a hydroxylapatite (HA) coating applied by a plasma spray technique.
### Indications for Use
Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew MDF Revision Hip System components are intended for single use only and are to be implanted without bone cement.
#### Performance Data
Performance testing has been conducted for the subject devices in accordance with the following guidance documents:
- . Non-Clinical Information for Femoral Stem Prostheses, dated September 2007
- Draft Guidance Document for Testing Non-Articulating, "Mechanically Locked," Modular Implant Components, dated May 1995
{1}------------------------------------------------
- Calcium Phosphate (Ca-P) Coating Draft Guidance Document for Preparation of FDA Submissions for Orthopaedic and Dental Endosseous Implants, dated February 1997
Environmental corrosion fatigue, pre-fatigue disassociation, and post-fatigue disassociation have been evaluated. A review of testing has demonstrated that there are no new issues related to the safety or effectiveness of the subject devices.
# Substantial Equivalence Information
The overall design of the Smith & Nephew Extra Small MDF Modular Sleeve line additions is substantially equivalent to the modular sleeves that currently exist as part of the MDF Revision Hip System cleared via K081124. The XSM Modular Sleeves utilize an HA coating on a grit blasted surface to achieve comed hiss fixation and are substantially equivalent to the Synergy Press Fit Stem (K970337) also featuring USA coating on a roughened titanium substrate to achieve cementless fixation. Giving coosideration to the evice modifications described in the Device Description section, no changes have been made to the overall design philosophy, intended use, and material choices when compared to the predicate devices. Additional design detail is provided in Table 1 below.
| Device<br>Comparison | Modular Sleeves<br>Subject of K100481 | Predicate MDF Modular<br>Sleeves Subject of<br>K081124 | Predicate Synergy<br>Press Fit Stems<br>Subject of K970337 |
|----------------------|-----------------------------------------------------------|--------------------------------------------------------|------------------------------------------------------------|
| | HA on Grit Blast<br>Modular Sleeves | Stiktite plus HA Coated<br>Modular Sleeves | HA on Grit Blast<br>Press Fit Stems |
| Size Offering | Sizes 11 - 26/27; XSM | Sizes 11-25; S, M, L | Sizes 9-18 |
| Materials | | | |
| Substrate | Ti-6Al-4V per ASTM<br>F1472 | Ti-6Al-4V per ASTM<br>F1472 | Ti-6Al-4V per ASTM<br>F1472 |
| Roughened<br>Coating | N/A; substrate is grit-<br>blasted to roughen<br>surface. | CPTi per ASTM F67<br>(StikTite) | N/A; substrate is grit-<br>blasted to roughen<br>surface. |
| HA Coating | HA Coated per ASTM<br>F1185 | HA Coated per ASTM<br>F1185 | HA Coated per ASTM<br>F1185 |
Table 1: Comparison of Subject K100481 Modular Sleeves to Predicate Devices
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
# JUL 1 5 2010
Smith & Nephew, Inc. % Ms. Natalie P. Williams 1450 Brooks Road Memphis, Tenessee 38116
Re: K100481
Trade/Device Name: Smith & Nephew MDF Modular Sleeve Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.
Regulatory Class: Class II Product Code: MEH Dated: July 8, 2010 Received: July 9, 2010
Dear Ms. Williams;
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
# Page 2 - Ms. Natalie P. Williams
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htmr.
Sincerely vours.
Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
### KI00481 510(k) Number (if known):
Device Name: Smith & Nephew MDF Modular Sleeve
### Indications for Use:
Total hip components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteolomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. Smith & Nephew MDF Revision Hip System components are intended for single use only and are to be implanted without bone cement.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Omtu for mxn
(Division Sign-Off)
(Division Sign rthopedic, Division of Surgical, O and Restorative Devices
Page 1 of
510(k) Number K100481
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.