← Product Code [LQY](/productcode/LQY) · K100450

# I-ROOT (K100450)

_S-Denti Co., Ltd. · LQY · Nov 18, 2010 · DE · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100450

## Device Facts

- **Applicant:** S-Denti Co., Ltd.
- **Product Code:** [LQY](/productcode/LQY.md)
- **Decision Date:** Nov 18, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** DE

## Indications for Use

The i-ROOT is intended for measuring the length of the root canal for the purpose of performing root canals and related dental procedures

## Device Story

i-ROOT is a battery-operated electronic apex locator used by dentists or oral surgeons to measure root canal length. Device uses a standard dental file inserted into the root canal as an electrode. It emits small electrical currents at 500 Hz and 5 kHz frequencies; measures current between file and mouth; compares impedance at these frequencies to minimize interference from blood or conducting fluids. Proximity to apex is displayed on a stabilized LCD meter. Output assists clinicians in determining precise working length for root canal procedures, potentially improving procedural accuracy and patient outcomes.

## Clinical Evidence

Bench testing only. Device tested for compliance with IEC 60601-1 (general safety) and IEC 60601-1-2 (electromagnetic compatibility). Accuracy verified at ±0.25mm.

## Technological Characteristics

Electronic apex locator; battery-powered (Lithium-ion 8.2VDC). Sensing principle: multi-frequency impedance measurement (500 Hz, 5 kHz). Dimensions: 110mm x 100mm x 117mm; weight: ~372g. Display: LCD. Connectivity: optional USB for PC software. Complies with IEC 60601-1 and IEC 60601-1-2.

## Predicate Devices

- Navi ROOT ([K083901](/device/K083901.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kloused

#### ട്. 510(k) Summary

#### [as required by 807.92(c)]

NOV 1 8 2010

### A. Applicant:

, 2 - 8

-Company name: S-Denti Co.,Ltd

-Address: 14F NO1407, Kolon Digital Tower Aston, 505-14, Gasan-dong, Geumcheon-gu, Seoul Korea -Tel:+82-2-2082-8828Fax:+82-2-2-2082-8829 web:www.s-denti.com

-Contact person: Peter Chung 412-687-3976

-Date: Dec 28th, 2009

#### B. Proprietary and Established Names:

Trade Name: i-ROOT Common Name:. Root Apex Locator Ragulation Name: Root Apex Locator Regulatory Classification: Unclassified, Dental, Product Code: LQY

#### C. Predicate device: Navi ROOT(K083901)

#### D. Device Description

The subject device is a lightweight, fully-automatic, battery operated device that allows a dentist or oral surgeon to locate a patient's anatomical root canal apex and obtain accurate root canal length measurements

### E. Intended use

The i-ROOT is intended for measuring the length of the root canal for the purpose of performing root canals and related dental procedures

#### F. Technological Characteristics:

The device allows the relative position of a dental file and the apex to be determined electrically. Using a standard dental file inserted into the root canal as an electrode, the device emits very small electrical currents having frequencies of 500 Hz and 5 KHz. The current between the file and mouth is measured at each of these frequencies, and compared, with readout of the relative proximity to the apex appearing on a stabilized meter. The use of two frequencies minimizes errors introduced by blood or other conducting medicinal fluids in the root canal.

#### G. Performance (Safety and Effectiveness Information)

The i-ROOT has been manufactured and tested to meet the safety requirements of IEC. The i-ROOT complies with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for safety and IEC 60601-1-2:2001, Medical electrical equipment - Part 1: General requirements for safety - Collateral standard: Electromagnetic compatibility -Requirements and tests.

#### I. Conclusion:

The performance tests demonstrated that i-ROOT is as safe, as effective and performs in a substantially equivalent manner to the predicate device

{1}------------------------------------------------

# 11.1 Comparison Table ________________________________________________________________________________________________________________________________________________________

| Product                            | i-ROOT                                                                                                                                                                         | Navi ROOT                           |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| 510K No                            |                                                                                                                                                                                | K083901                             |
| Manufacturer                       | S-Denti Co.,Ltd                                                                                                                                                                | S-Denti Co.,Ltd                     |
| Intended Use                       | Measurement of the length of the<br>root canal for the purpose of<br>performing root canals and related<br>dental procedures                                                   | Identical                           |
| Method of calculating              | Comparison of impedance at multi<br>frequencies                                                                                                                                | Identical                           |
| Electric Current                   | Input Current : DC 40 mA<br>Patient auxiliary current : AC 10 mA                                                                                                               | Less than 10pA                      |
| Electrical Power                   | Lithium-ion Battery 8.2VDC(4.1VDCx2)                                                                                                                                           | DC 4.5V(AA Battery 1.5Vx3EA)        |
| Power consumption                  | 0.33VA                                                                                                                                                                         | 180mW Max(150mW)                    |
| Measuring Voltage                  | 80mV                                                                                                                                                                           | Identical                           |
| Frequencies used<br>for comparison | 500 Hz, 5 kHz                                                                                                                                                                  | Identical                           |
| Time of use                        | 2000 Hr Continuous                                                                                                                                                             | Identical                           |
| Display                            | LCD                                                                                                                                                                            | Identical                           |
| Accessory                          | Probe cord 1EA, Lip holder 5EA,<br>File holder A 1EA,<br>File holder B 2EA,<br>AC adapter 1EA,<br>User manual 1EA,<br>USB Cable 1EA(optional),<br>PC software CD(optional) 1EA | Probe cord 1EA,<br>File holder 2EA, |
| Accuracy                           | ±0.25mm                                                                                                                                                                        | Identical                           |
| Storage condition<br>- Temperature | -20 ~ 70℃                                                                                                                                                                      | -10 ~ 60 ℃                          |
| - Humidity                         | 0 ~ 95%                                                                                                                                                                        | 10 ~ 80RH%                          |
| Dimensions(LxWxH)                  | 110mm(W) x 100mm(D) x 117mm(H)                                                                                                                                                 | 153mm(W) x 138mm(D) x 44mm(H)       |
| Weight                             | Approx. 372g ± 24g                                                                                                                                                             | 420g                                |

# Conclusion

Based upon the intended use, and upon the similarity of product configuration and administration, it can be concluded the i-ROOTis substantially equivalent to the identified predicate device in terms of intended use, safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an abstract caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Tae Kye Nam S-Denti Company, Limited 505-14, Gasan-Dong, Geumcheon-Gu Seongnam-City, Seoul Republic of Korea 153-803

NOV 1 8 2010

Re: K100450

Trade/Device Name: i-ROOT Electronic Apex Locator Regulation Number: Unclassified Regulation Name: None Regulatory Class: Unclassified Product Code: LQY Dated: August 18, 2010 Received: October 25, 2010

Dear Mr. Nam:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Nam

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542. of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Wh for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and dental Devices Office of Device Evaluation Center for Devices and Radiological Health

## Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K 100 450 ------------------------------------------------------------------------------------------------------------------------------------------

NOV 1 8 2010

Device Name: i-ROOT Electronic Apex Locator

Indications For Use:

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . :

> The i-ROOT is intended for measuring the length of the root canal for the purpose of performing root canals and related dental procedures

Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1
Susan Punny

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K100460

---

**Source:** [https://fda-staging.innolitics.com/device/K100450](https://fda-staging.innolitics.com/device/K100450)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
