K100448 · Ascension Orthopedics, Inc. · KWS · Aug 30, 2010 · Orthopedic
Device Facts
Record ID
K100448
Device Name
ASCENSION MODULAR TOTAL SHOULDER SYSTEM
Applicant
Ascension Orthopedics, Inc.
Product Code
KWS · Orthopedic
Decision Date
Aug 30, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ascension Modular Total Shoulder System is indicated for use as a hemi or total shoulder replacement for: 1. Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis. 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory. 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component). Shoulder Hemiarthroplasty is also indicated for: 1. Ununited humeral head fractures. 2. Avascular necrosis of the humeral head. 3. Rotator cuff arthropathy. 4. Deformity and/or limited motion. The humeral component is intended for cemented or uncemented use. The glenoid component is intended for cemented use only.
Device Story
Modular shoulder prosthesis system; includes proximal bodies, humeral stems, humeral heads, and all-polyethylene glenoid components. Components connect via Morse-type taper. Humeral heads (CoCrMo alloy) available in concentric/eccentric configurations; glenoid components (UHMWPE) available in keeled/pegged designs. Used by orthopedic surgeons in clinical settings for joint replacement. Provides mechanical bearing surface to replace damaged articular surfaces; restores joint function; reduces pain. Humeral component supports cemented or uncemented fixation; glenoid component requires cement.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included axial disassembly force of taper connections, bending taper fatigue endurance, and dynamic evaluation of glenoid loosening.
Indicated for patients with severely painful/disabled shoulder joints due to osteoarthritis, traumatic/rheumatoid arthritis, proximal humerus fracture-dislocations, ununited humeral head fractures, avascular necrosis, rotator cuff arthropathy, or deformity/limited motion requiring hemi or total shoulder replacement.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Smith & Nephew/Plus Orthopedics PROMOS Modular Shoulder System (K032126/K063578)
Osteonics' All Polyethylene Glenoid Shoulder Keeled Components (K962082)
Submission Summary (Full Text)
{0}------------------------------------------------
K100448
510(k) Summary
| SUBMITTER NAME: | Ascension Orthopedics, Inc.<br>8700 Cameron Road<br>Austin, TX 78754-3832 | AUG 30 2010 |
|-----------------|------------------------------------------------------------------------------------------|-------------|
| 510(k) CONTACT: | Stanley J. Harris<br>Phone: (512) 836-5001 x1545 | |
| DATE: | August 26, 2010 | |
| TRADE NAME: | Ascension® Modular Total Shoulder System | |
| COMMON NAME: | Hemi- or Total shoulder | |
| CLASSIFICATION: | 21 CFR 888.3690 Shoulder joint humeral (hemi-shoulder)<br>metallic uncemented prosthesis | |
| | 21 CFR 888.3660, Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis | |
| PRODUCT CODE: | HSD and KWS | |
| PANEL: | Orthopedic | |
#### PREDICATE DEVICES:
K032126/K063578 - Smith & Nephew/Plus Orthopedics PROMOS Modular Shoulder System
K962082: Osteonics' All Polyethylene Glenoid Shoulder Keeled Components
# DEVICE DESCRIPTION:
The Modular Total Shoulder System consists of a line of proximal bodies, humeral stems, humeral heads and all polyethylene glenoid components. The body, stem and humeral head may be used by themselves, as a hemiarthroplasty, if the natural numeral nead may befficient bearing surface, or in conjunction with the glenoid, as a gronold proximal bodies and humeral stems are manufactured from total replacemont: "The and connect together via a Morse type taper. The humeral theads are manufactured from cobalt-chrome-molybdenum (CoCrMo) alloy and are offered in both concentric and eccentric configurations. The humeral head may onered in both contral glenoid bone if it is of sufficient quality, or against the all antionate against the naturanid. The glenoid is manufactured from ultra high polycthylone oomonton gronone (UHMVPE) and is offered in a keeled and pegged molecular weight porychylons (Ornaare designed to function with both the concentric and eccentric heads.
{1}------------------------------------------------
# INTENDED USE:
The Ascension Modular Total Shoulder System is indicated for use as a hemi or total shoulder replacement for:
- 1. Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis.
- 2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory.
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. - revision of a failed primary component).
Shoulder Hemiarthroplasty is also indicated for:
- 1. Ununited humeral head fractures.
- 2. Avascular necrosis of the humeral head.
- 3. Rotator cuff arthropathy.
- 4. Deformity and/or limited motion.
The humeral component is intended for cemented or uncemented use. The glenoid component is intended for cemented use only.
### SUMMARY OF TECHNOLOGIES:
There are no significant differences between the Ascension® Modular Total Shoulder System, to the Smith & Nephew/Plus Orthopedics PROMOS Modular Shoulder System (K032126/K063578) currently being marketed which would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and intended use.
#### SUBSTANTIAL EQUIVALENCE:
Substantial equivalence was based upon testing and a geometrical comparison of the subject and predicate devices. Tests performed to substantiate equivalence were: axial disassembly force of taper connections, bending taper fatigue endurance, and dynamic evaluation of glenoid loosening. Geometrical comparisons were between the subject device, Ascension TITAN Modular Total Shoulder System and predicate devices: Smith & Nephew/PLUS PROMOS Modular Shoulder System (K032126, uons578) and the Osteonics' All Polyethylene Glenoid Shoulder Keeled Components (K962082).
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a stylized serpent winding around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES" are arranged in a circular fashion around the caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ascension Orthopedics, Inc. % Mr. Stanley J. Harris Vice President, Clinical/Regulatory Affairs 8700 Cameron Road, Suite 100 Austin, Texas 78754
AUG 30 2010
Re: K100448
Trade/Device Name: Ascencion® Modular Total Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: KWS. HSD Dated: August 10, 2010 Received: August 17, 2010
Dear Mr. Harris:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
#### Page 2 - Mr. Stanley J. Harris
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Sincerely yours,
Carbare Buere
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
K100448 510(k) Number.
Ascension® Modular Total Shoulder System Device Name:
## Indications for Use:
The Ascension Modular Total Shoulder System is indicated for use as a hemi or total shoulder replacement for:
- 1. Severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis.
2. Fracture-dislocations of the proximal humerus where the articular surface is severely comminuted, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory.
- 3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. – revision of a failed primary component).
Shoulder Hemiarthroplasty is also indicated for:
- 1. Ununited humeral head fractures.
- 2. Avascular necrosis of the humeral head.
- 3. Rotator cuff arthropathy.
- 4. Deformity and/or limited motion.
The humeral component is intended for cemented or un-cemented use. The glenoid component is intended for cemented use only.
Prescription Use x (Part 21 CFR 801Subpart B)
Over-The-Counter Use OR (Part 21 CFR 801Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Knutar der Mom
(Division Sign-(Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number KJ 00448
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.