K100434 · Icu Medical, Inc. · FPA · Jul 9, 2010 · General Hospital
Device Facts
Record ID
K100434
Device Name
CLAVE NEUTRON
Applicant
Icu Medical, Inc.
Product Code
FPA · General Hospital
Decision Date
Jul 9, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an intravascular catheter placed in the vein or artery. The device may be used for the administration of blood and fluids to patients, including pediatrics and immunocompromised patients. The device may also be used with power injector procedures up to 10mL per second of contrast media and a maximum pressure of 350psi. The device incorporates a technology that will prevent fluid displacement resulting from the following: Connection of a luer; syringe plunger compression; patient vascular pressure changes, such as coughing or sneezing; and IV solution container run-dry. The CLAVE Neutron incorporates a pre-slit septum that does not require the use of needles and will therefore passively aid in the reduction of needlestick injuries.
Device Story
CLAVE Neutron is a needleless, closed, bidirectional connector for intravascular catheters; functions as an interface for fluid/blood administration and power injection of contrast media. Device utilizes an internal silicone valve and pre-slit septum to maintain a closed system; prevents positive and negative fluid displacement caused by luer connection/disconnection, syringe plunger rebound, patient vascular pressure changes (e.g., coughing), and IV bag run-dry. Used in clinical settings by healthcare providers; connects between IV tubing/syringes and vascular access devices. Passive design prevents needlestick injuries; internal valve mechanism mitigates reflux and air entry. Benefits include reduced risk of catheter-related bloodstream infections and maintenance of catheter patency by preventing blood reflux into the catheter tip.
Clinical Evidence
No clinical data included. Evidence relies on bench testing, including microbial ingress challenges and simulated use environment testing. Literature cited (Bouza 2003, Maragakis 2006) supports the safety of the predicate CLAVE pre-slit septum design in pediatric and immunocompromised populations.
Technological Characteristics
Materials: Polyester, silicone, ABS or polycarbonate, and nylon. Design: Needleless, pre-slit septum, internal silicone valve. Connectivity: ISO 594-1/594-2 compliant male luer interface. Sterilization: Gamma or E-beam. Performance: Supports power injection up to 10mL/sec at 350psi; residual volume 0.14mL.
Indications for Use
Indicated for use as an accessory to intravascular catheters (venous or arterial) for administration of blood and fluids. Applicable to all patient populations, including pediatrics and immunocompromised patients. Supports power injection of contrast media up to 10mL/sec at 350psi.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
#### 510(K) Summary
K100434
### ICU MEDICAL INC.
4455 Atherton Drive Salt Lake City, Utah (801) 264 - 1332, Phone (801) 264 - 1755, Fax Tracy S. Best, Sr. Regulatory Affairs Specialist Preparation Date: 06/23/2010
#### Summary of Safety and Effectiveness for the:
CLAVE® Neutron™ Trade Name: Needleless Connector, Closed Access Common Name: Classification Name: 21 CFR 880.5440, Class II Device, 80FPA & 80LHI, Accessory
#### Legally Marketed Predicate Devices for Substantial Equivalence:
*K970855 - CLAVE® Connector - ICU Medical Inc. * K021692 - CLC2000 Catheter Patency Device - ICU Medical, Inc. * K083723 - A6 Luer Access Device - B. Braun Medical, Inc.
#### Rationale for Substantial Equivalence:
The CLAVE Neutron is substantially equivalent to all three predicates by providing a normally closed, needleless connection and a route to deliver blood and fluids to a patient through a vascular access device. All three predicate devices and the proposed device are free of steel needle components and designed such that the use of a needle in the device would render them inoperable. This design feature passively aids in the prevention needles cannot be introduced into the system during use.
The anti-displacement feature of the CLAVE Neutron is substantially equivalent to both the CLC2000 and the A6 in that it prevents negative displacement.
The power injectable feature is substantially equivalent as the A6 Luer Access Device which allows the infusion of contrast media at 10mL per second at a maximum pressure of 300psi. The CLAVE Neutron allows infusion of contrast media of 10mL per second with a maximum pressure of 350psi.
Components of the CLAVE Neutron are made from materials that are substantially equivalent to the predicate devices. Comprehensive biocompatibility testing for the CLAVE Neutron is included in this Submission.
The CLAVE Neutron was sublected to a variety of functional tests including microbial ingress and performance criteria which prove it to be substantially equivalent to the predicate devices. Results of the testing demonstrate that there are no new issues of Safety and Efficacy that are raised with the introduction of the CLAVE Neutron.
#### Description of Submitted Device:
The CLAVE Neutron is a normally closed, bidirectional connector that incorporates a unique silicone valve within the fluid path. The unique valve prevents fluid displacement, both positive and negative at all times, including when the device is accessed. There are four known causes of displacement associated with needleless connectors. They are:
- 1. Connection or Disconnection of an Adaptable Luer Device. Fluid displacement occurs when a luer is inserted, or removed, from a connector by displacing a specific volume. This is also commonly known as positive or negative pressure when describing connectors.
- 2. Syringe Plunger Rebound. This occurs when a flush has been done using a standard syringe. Once the syringe is emptied, the clinician has the ability to compress the syringe plunger in an effort to infuse all available solutions which, once released by the clinician's hand, will rebound and cause negative fluid displacement.
- 3. Patient Vascular Pressure Changes. Routine events such as coughing, sneezing, crying and bearing down can cause a temporary increase in a patient's vascular pressure. This pressure can cause negative fluid displacement back through a standard needleless connector.
{1}------------------------------------------------
- 4. IV Bag Run Dry. When an IV solution bag is allowed to empty, the normal gravity infusion pressure which is caused by the weight of the fluid hung at 36" head height diminishes to little or zero positive pressure. Normal patient vascular pressure may then overcome the fluid head height pressure and cause negative fluid displacement back through a standard needleless connector.
#### Intended Uses of the CLAVE® Neutron™:
CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an intravascular catheter placed in the vein or artery. The device may be used for the administration of blood and fluids to patients, including pediatrics and immunocompromised patlents. The device may also be used with power injector procedures up to 10mL per second of contrast media and a maximum pressure of 350psi. The device incorporates a technology that will prevent fluid displacement resulting from the following: Connection of a luer; syringe plunger compression; patlent vascular pressure changes, such as coughing or sneezing; and IV solution container run-dry. The CLAVE Neutron incorporates a pre-slit septum that does not require the use of needles and will therefore passively aid in the reduction of needlestick injuries.
## Safety and Performance:
The CLAVE Neutron is intended for use on a patient's vascular access device including pediatrics and immunocompromised patients. As discussed in the Executive Summary, the CLAVE Neutron incorporates proprietary design features from the CLAVE Needleless Connector. Specifically the preslit septum and the internal cannula of the predicate CLAVE which are used to transfer fluid to the patient's catheter, are identical in design. These two components are the functional components which prohibit bacterial transfer. Microbial Ingress testing as provided in this submission demonstrates that the CLAVE Neutron in comparison to the CLAVE was equivalent in its ability to prevent bacterial ingress. Said testing shows that the device will prevent bacterial in gress. completely if used in accordance with the directions for use. The prevention of bacterial ingress is specifically important for pediatrics and immunocompromised patients that would otherwise be at risk of infection.
In Bouza (2003), the CLAVE was found in a randomized clinical trial to independently reduce catheter hub and tip bacterial colonization. The patient population for the study was immunocompromised, intensive care patients at the Department of Cardiovascular Surgery, Hospital General University of Madrid, Madrid, Madrid, Spain. The identical CLAVE pre-slit septum component used in the study, is still used in the CLAVE today and also in the proposed CLAVE Neutron. This study demonstrates that the specific CLAVE split-septum design may safely be used In immunocompromised patient populations.
In Maragakis' (2006), the CLAVE was clinically studied in pediatric and immunocompromised patient populations. In this study, pediatric ICU and pediatric oncology service patients at the John's Hopkins Hospital in Baltimore Maryland were observed for Cather Related Bloodstream Infection Rates. The identical CLAVE pre-slit septum component used in this study, is still used in the CAVE today and also in the proposed CLAVE Neutron. This study demonstrates that the CLAVE split-septum design may safely be wed in pediatric and immunocompromised patient populations.
The CLAVE Neutron has been tested in accordance with its product specifications which accommodate known functional requirements for Critical Care, Neonatal Care and Pediatrics.
The CLAVE Neutron is individually packaged and pre-sterilized in a peel type pouch. The device can be included as part of a kit or set - as the physician may prescribe. ICU Medical performs analysis and design verification testing based on predetermined criteria, which is documented in the Performance Specification contained in this submission. All testing meets themements on the la as defined for the device, including power infusion up to 1000 per second of contrast media. The design of the device allows it to accept a male luer as defined by ISO 594-1 and ISO 594-2 which is considered to be an industry standard.
ICU Medical has also performed testing that is recommended by the guidance document "Intravascular Administration Sets Premarket Notifications [510(k)/i and we have included that successful testing as part of this submission.
{2}------------------------------------------------
| Specification: | CLAVE® | CLC2000® | Braun A6 | CLAVE® Neutron™ |
|-------------------------------------------------------|---------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| 510(k) Approval | K970855 | K021692 | K083723 | This submission |
| Functional use | Needleless connector | Needleless Connector<br>with anti-reflux | Needleless connector<br>with anti-fluid<br>displacement and<br>power injectable | Needleless connector<br>with anti-fluid<br>displacement and<br>power injectable |
| Patient Populations* | Adults, Pediatrics and<br>Immunocompromised<br>patients | Adults | Adults | Adults, Pediatrics and<br>Immunocompromised<br>patients |
| Residual Volume | 0.06mL | 0.07mL | 0.30mL | 0.14mL |
| Gravity Fluid Flow | 100mL/min | 200mL/min | Unavailable for test | 100mL/minute |
| Syringe Disconnect:<br>Fluid Displacement | Negative Displacement | Positive Displacement | Positive Displacement | Integral valve prevents<br>displacement, both<br>positive and negative® |
| Syringe Plunger<br>Compression: Fluid<br>Displacement | N/A | N/A | N/A | Integral valve prevents<br>displacement, both<br>positive and negative |
| Patient Pressure: Fluid<br>Displacement | N/A | N/A | N/A | Integral valve prevents<br>displacement, both<br>positive and negative |
| Bag Run-Dry:<br>Fluid Displacement | N/A | N/A | N/A | Integral valve prevents<br>displacement, both<br>positive and negative |
| Multiple Activations | 96 intermittent<br>3 extended time | 96 intermittent<br>3 extended time | Unavailable for test | 96 Intermittent<br>3 extended time |
| Positive Leak Pressure | 60psig | 60psig | Unavailable for test | 45psig ** |
| Negative Leak Pressure | -8.5psig | -8.5psig | Unavailable for test | -8.5psig |
| Luer Retention | ISO 594-1 (5.4) | ISO 594-1 (5.4) | Unavailable for test | ISO 594-1 (5.4) |
| Component Assembly | >51bf-in | >15lbf-in | Unavailable for test | >20lbf-in |
| Chemical Compatibility | Lipids, Alcohol | Lipids, Alcohol | Unavailable for test | Lipids, Alcohol |
| Power injectable | N/A | N/A | 10ml/sec of contrast<br>media and < 300psl. | 10ml/sec of contrast<br>media and < 350psi. |
| Sterilization Method | Gamma or E-beam | Gamma or E-beam | Gamma or E-beam | Gamma or E-beam |
| Packaging | Peel pouch | Peel pouch | Peel pouch | Peel pouch |
| Materials | Polyester; Silicone; and<br>ABS or Polycarbonate | Polyester; Silicone;<br>Stainless Steel;<br>Polypropylene; and<br>Polycarbonate | Polycarbonate; rubber;<br>silicone | Polyester: Silicone: ABS<br>or Polycarbonate; and<br>Nylon |
#### Technological Characteristics and Substantial Equivalence Table:
*A single, integral valve in the CLAVE Neutron is used to prevent the four forms of fluid displacement. This fluid displacement feature is substantially equivalent to the fluid displacement feature of all three predicate devices during the disconnection of a luer. For the additional three fluid displacement features (Syringe Plunger Compression, Patient Pressure and Bag-Run Dry), the valve operates in the identical manner which renders these new features not significantly different from the predicate devices.
**45 psig is adequate to prevent positive pressure leakage from a patient's line and is therefore not significantly different from the predicate devices.
#### Conclusion:
As demonstrated by the table above, there are equivalent features and functional uses between devices. Differences are in materials; maximum pressure; residual volume; and additional types of displacement that the CLAVE Neutron is able to accommodate. Materials are tested to the latest ISO 10993 standards; the unique design allows the additional pressure up to 350ps; residual volume is less than half of the A6 predicate due to an internal fluid path versus an external fluid path; and additional displacement types are due to the manner in which the valve operates on the CLAVE Neutron. These differences do not introduce any new safety or efficacy risks to the patient.
The operational characteristics are equivalent to luer needleless connection technology and intended use when compared to the three predicate devices. The proprietary technology in the CLAVE Neutron is unique to this type of product and is confidentially discussed further in this submission.
Materials, performance, operational features, environment of use, and connection methodology of the submitted device compared to the predicate devices show that they are substantially equivalent
{3}------------------------------------------------
and that they are safe and effective for their intended use as demonstrated and supported in the performance testing according to the product specification, included in this submission.
### Clinical and Non-Clinical Data:
There is no clinical data included in this submission.
I nere is no alliner solo nether CLAVE Neutron Design Verification Report in the submission as simulated use environment testing. The testing subjected the CLAVE Neutron to a series of simulated use environment cesang. The mig oblal ingress challenges to verify the safe and performance testing, alocentify alocating demonstrates that there are no differences between the eredicate and proposed device that raise any new issues of safety or effectiveness. This proprietary data is included in this submission.
² Bouza, E, Munoz, P, Lopez-Rodriguez, J, Jesus Perez, M, & Rincon, C, et. al, (2003). A Needleless closed system device (clave) protects from intravascular catheter tip and hub colonization: a prospective randomized study. Journal of Hospital Infection, 54, 279-287.
2 Maragakis, L, Bradley, K, Song, X, Beers, C, & Miller, M, et. al, (2006). Increased catheter-related bloodstream infection rates after the introduction of a new mechanical valve intravenous access port. Infection Control ond Hospital Epidemiology, 27, 67-70.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Tracy Best Senior Regulatory Affairs Specialist ICU Medical, Incorporated 4455 Atherton Drive Salt Lake City, Utah 84123
Re: K100434
Trade/Device Name: Clave® Neutron™ Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long Term Intravascular Catheter Regulatory Class: II Product Code: FPA Dated: June 24, 2010 Received: June 25, 201
AUG 9 - 2010
Dear Mr. Best:
This letter corrects our substantially equivalent letter of July 7, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Mr. Best
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Susan Panas
Anthony D. Watson, B.S.
ny D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K100434
Device Name: Clave® Neutron™
Indications for Use:
CLAVE Neutron is a normally closed, bidirectional connector intended for use as an accessory to an intravascular catheter placed in the vein or artery. The device may be used for the administration of blood and fluids to patients, including pediatrics and immunocompromised patients. The device may also be used with power injector procedures up to 10mL per second of contrast media and a maximum pressure of 350psi. The device incorporates a technology that will prevent fluid displacement resulting from the following: Connection of a luer; syringe plunger compression; patient vascular pressure changes, such as coughing or sneezing; and IV solution container run-dry. The CLAVE Neutron incorporates a pre-slit septum that does not reguire the use of needles and will therefore passively aid in the reduction of needlestick injuries.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rill C. 8/5/10
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K100434
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.