The PMS8210A Multi Parameter Patient Monitor is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Patient Monitor has certain features and functions. The patient parameters that can be monitored by PMS8210A Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2) and Temperature (TEMP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs. The PMS8210A Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment. The PMS8210A Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.
Device Story
Battery or line-powered multi-parameter patient monitor; acquires physiological signals (ECG, respiration, NIBP, SpO2, temperature, HR, PR). Signals converted to digital data, processed for alarm conditions, and displayed on TFT LCD. Used in hospital clinical areas (ICU, CCU, OR, emergency department) by qualified medical personnel. Provides audible/visual alarms for parameter excursions. Benefits patient by providing continuous physiological status information to clinical staff for decision-making.
Clinical Evidence
Bench testing only. Device underwent safety and performance testing in accordance with IEC standards (e.g., IEC 60601-1, 60601-1-2, 60601-2-27, 60601-2-49) and AAMI/ANSI standards for NIBP and ECG. No clinical data presented.
Technological Characteristics
Battery/line-powered monitor; 3.2" TFT LCD display; 3/5-lead ECG; oscillometric NIBP; thoracic impedance respiration; Nellcor SpO2 module. Connectivity: AC power adapter (Class I). Sterilization: N/A. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, AAMI SP10, ISO 10993.
Indications for Use
Indicated for adult, pediatric, and neonatal patients in healthcare facilities for monitoring ECG, HR, PR, respiration rate, NIBP, SpO2, and temperature. Not for home use or without physician order.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: January 27, 2010
1. Company and Correspondent making the submission:
Name - Shanghai 3F Electronics Co., Ltd. Address - Building 3, No. 128, Jiujing Road, Shanghai, China Telephone -- +86-21-67696500/51695200 ext: 8022 Fax - +86-21-67696296 Contact - Mr. Lu Yulin Email - rmdlu@3fmedical.cn
# 2. Device :
Trade/proprietary name: PMS8210A (IRIS) Patient Monitor/ Multi-Parameter
Patient Monitor
Common Name : Multi-parameter Patient Monitor
Classification of the device: Class II
Panels: Cardiovascular, General Hospital, Anesthesiology
Product code: 21CFR870.2700, Oximeter, DQA
21CFR870.2300, Monitor, Physiological, MWI
21CFR880.2910, Thermometer, Electronic, Clinical, FLL
21CFR871.1130, System, Measurement, Blood Pressure,
DXN
Establishment Registration Number : None Yet Predicate Devices:
{1}------------------------------------------------
| Predicate Model | Manufacturer | K Number | Submitted Device |
|---------------------------------------|----------------------------------------------------------------|----------|------------------------------------------------------------------------|
| PM-9000<br>EXPRESS PATIENT<br>MONITOR | Shenzhen<br>Mindtay Bio-<br>medical<br>Electronics Co.,<br>LTD | K053234 | PMS8210A (Iris) Patient<br>Monitor/ Multi-Parameter<br>Patient Monitor |
## 3. Description :
### 3.1 General
PMS8210A Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG, respiration (RESP), Non-Invasive blood pressure (NIBP), Saturation of pulse oxygen (SPO2), Temperature (TEMP), Heart rate (HR) and Pulse Rate (PR), The signals are converted into digital data and processed, examines the data for alarm conditions and displays the data. The monitor also provides operating control for the user.
The patient monitor is intended to be used in a hospital clinical area such as intensive care units, cardiac care units, operation room, emergency department, to provide additional information to the medical and nursing staff about the physiological condition of the patient. The PMS8210A patient monitor is intended to be used only under reqular supervision of clinical personnel. The intended location of use is clinics.
4. Indication for use :
The PMS8210A Patient Monitor is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Patient Monitor has certain features and functions.
The patient parameters that can be monitored by PMS8210A Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP),
{2}------------------------------------------------
Arterial Hemoglobin Oxygen Saturation(SpO2) and Temperature (TEMP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PMS8210A Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The PMS8210A Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.
- 5. Comparison with predicate device: Please see next page for the comparison table.
{3}------------------------------------------------
Table of Comparison to Predicate Device 出来ないからない。 「
1. General Specifications (i.e. physical/electrical)
| Characteristics<br>(i.e. physical/electrical) | Subject Device | Claimed SE Device 510(K) No. |
|------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Name and model | PMS8210A Multi-parameter Patient Monitor | PM-9000 EXPRESS PATIENT MONITOR (K053234) |
| Physical dimension/weight | Dimensions: 250 (W)×180 (H)× 180 (D) (mm)<br>Weight: 2.0kg | Dimensions: 318(W)×270(H)×137(D) (mm)<br>Weight: <7.5kg |
| Display | 7 segment LED + 3.2" colorful TFT LCD (320×240) | 10.4" color TFT LCD(800×600) |
| Button | keys – front panel | Soft keys – front panel |
| Type, Degree of protection<br>against electric shock | AC power adapter<br>Electr. Class I and internal power supply | AC power adapter<br>Electr. Class I and internal power supply |
| Power supply | 100~240VAC(±10%),<br>50/60Hz(±3Hz),45VA | 100~240VAC(± 10%),<br>50/60Hz(±3Hz),140VA |
| Internal power source | Insetting sealed lithium batter: 2200mAh and 4400mAh | Sealed Lead acid battery(for 2 pieces)<br>And insetting li-ion battery (for 2 pieces) |
| Battery charging indicator | Yes | Yes |
| Low battery indicator | Yes | Yes |
| Battery charge time, typ. | 2200mAh: approx. 3 hours<br>4400mAh: approx. 6 hours | Acid battery: approx. 12 hours<br>And li-ion battery: approx. 6.5 hours |
| Flammable anesthetics | not suitable | not suitable |
| Operating condition | Temperature: 0°C to 40°C (32°F to 104°F)<br>Relative Humidity: ≤95% (non-condensing) | Temperature: 0°C to 40°C (32° F to 104° F)<br>Relative Humidity: 15%~95% (non-condensing) |
| Storage condition | Temperature: -40°C to 55°C (-40°F to 131°F)<br>Relative Humidity: ≤95% (non-condensing) | Temperature: -20°C to 60°C (-4°F to 140°F)<br>Relative Humidity: 10%~95% (non-condensing) |
| EMC | IEC 60601-1-2:2007 | IEC 60601-1-2:2001+A1:2004 |
| Power on self test | Yes | Yes |
| Optional printer | Yes | Yes |
| Characteristics | Subject Device | Claimed SE Device 510(K) No. |
| Name and model | PMS8210A Multi-parameter Patient Monitor | PM-9000 EXPRESS PATIENT MONITOR (K053234) |
| Lead | 3lead(RA,LA,LL);<br>5lead(RA, RL, LA, LL,V)). | 3lead(RA, LA,LL);<br>5lead(RA, RL, LA, LL,V));<br>10lead(RA, RL, LA,LL,V1-V6)). |
| Lead option | Monitor lead(3 lead) / standard lead(5 lead) | 3 lead/ 5 lead/ 12 lead |
| Gain | ×0.5; x1. | ×0.125; ×0.25; ×0.5; ×1; ×2;auto. |
| Sweep speed | 12.5mm/s, 25mm/s, 50mm/s | 12.5mm/s, 25mm/s, 50mm/s |
| Range of heart rate monitoring | Adult: 20~300 bpm;<br>Neonate/ Pediatric: 20~350 bpm | Adult: 15~300 bpm;<br>Neonate/ Pediatric: 15~350 bpm |
| Resolution | 1 bpm | 1 bpm |
| Precision | 20~200 bpm: 5% or ±5bpm;<br>201~350 bpm: 10%. | 1 bpm or ±1% |
| Alarm setting | The limit of alarm (setup range: 20~350 bpm), and leads-off alarm display. | 15~350 bpm |
| Input resistance | ≥5 MΩ | ≥5 MΩ |
| CMRR | ≥89 dB | Diagnostic mode: ≥90dB<br>Monitoring mode: ≥105dB<br>Surgical mode: ≥105dB |
| Heart disorder analysis | NO | NO |
| Anti-polarized voltage | ≤±500 mV | 3/5 lead ≤±300mV;<br>12 lead ≤±500mV. |
| Baseline renewing time | <5 s after the defibrillation | <5 s after the defibrillation |
| ECG mode | Mode 1 (Monitoring mode), mode 2(Monitoring mode), mode 3 (Surgical mode) | Diagnostic mode, Monitoring mode, Surgical mode. |
| Frequency characteristic | Mode 1: 0.1Hz-40Hz;<br>Mode 2: 0.67Hz-40Hz<br>Mode 3: 1Hz-25Hz | Diagnostic mode: 0.05Hz-100Hz<br>Monitoring mode: 0.5Hz-40Hz<br>Surgical mode: 1Hz-20Hz |
| Safeguard | 4000V high voltage isolation, anti-defibrillation | Withstand 4000VAC/50Hz voltage in isolation; |
{4}------------------------------------------------
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||\n
2. ECG
バイト (
32 of 866
K100394
{5}------------------------------------------------
| Respiration |
|-------------|
| |
| તં |
נייר ל
| Characteristics | Subject Device | Claimed SE Device 510(K) No. |
|------------------|---------------------------------------------------|----------------------------------------------------|
| Name and model | PMS8210A Multi-parameter Patient Monitor | PM-9000 EXPRESS PATIENT MONITOR (K053234) |
| Measuring method | The thorax impedance method (used with ECG lead) | Thoracic impedance |
| Measuring range | 15 ~ 120rpm | Adult: 0~120 rpm;<br>Neonate/ Pediatric: 0~150 rpm |
| Resolution | 1 rpm | 1 rpm |
| Precision | The bigger one between ±2 rpm or ±2% | 0~6 rpm: Unspecified;<br>7~120rpm: ±2 rpm or ±2% |
| Alarm setup | Yes | Yes |
| Alarm method | Audible and visual alarm, alarm events recallable | Audible and visual alarm, alarm events recallable |
# 4. Non-Invasive Blood Pressure (NIBP)
| Characteristics | Subject Device | Claimed SE Device 510(K) No. |
|-------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Name and model | PMS8210A Multi-parameter Patient Monitor | PM-9000 EXPRESS PATIENT MONITOR (K053234) |
| Method | Oscillometric | Oscillometric |
| Patient type | Neonatal, pediatric and adult patients | Neonatal, pediatric and adult patients |
| Unit of measure | mmHg & kPa | mmHg & kPa |
| Pressure measurement<br>range - Systolic | Adult: 40 ~ 260mmHg<br>pediatric: 40 ~ 160mmHg<br>Neonate: 40 ~ 130mmHg | Adult: 40 ~ 270mmHg<br>pediatric: 40~200mmHg<br>Neonate: 40 ~ 135mmHg |
| Pressure measurement<br>range - Diastolic | Adult: 20 ~ 200mmHg<br>pediatric: 20 ~ 120mmHg<br>Neonate: 20 ~ 100mmHg | Adult: 10 ~ 210mmHg<br>pediatric: 10 ~ 150mmHg<br>Neonate: 10 ~ 100mmHg |
| Pressure measurement | Adult: 26 ~ 220mmHg | Adult: 20 ~ 230mmHg |
33 of 866
ri
{6}------------------------------------------------
| Range-Dean pressure | pediatric: 26 ~ 133mmHg<br>Neonate: 26 ~ 110mmHg | pediatric: 20 ~ 165mmHg<br>Neonate: 20 ~ 110mmHg |
|-------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| BP accuracy | Arithmetic mean values:±5 mmHg;<br>Standard deviation no greater than 8 mmHg | Arithmetic mean values:±5 mmHg;<br>Standard deviation no greater than 8 mmHg |
| BP measurement accuracy | ANSI/AAMI SP10:2002;<br>EN1060-4 | Meets ANSI/AAMI SP10:1992+A1:2002 |
| Cuff pressure range | 0 to 300mmHg | 0 to 300mmHg |
| Auto zero CAL | Yes | Yes |
| Over pressure protector | Adult/ Pediatric: 300mmHg;<br>Neonate: 150mmHg | Adult: 297±3mmHg<br>Pediatric: 240±3mmHg<br>Neonate: 147±3mmHg |
| Alarm setup | The range is the same as parameter measurement<br>range of SYS、DIA、MAP | The range is the same as parameter measurement<br>range of SYS、DIA、MAP |
| Alarm method | Sound light alarm, and record the alarm status for<br>review | Sound light alarm, and record the alarm status for<br>review |
# 5. Pulse Oximetry. (SpO2)
| Characteristics | Subject Device | Claimed SE Device 510(K) No. |
|---------------------------|------------------------------------------|-------------------------------------------|
| Name and model | PMS8210A Multi-parameter Patient Monitor | PM-9000 EXPRESS PATIENT MONITOR (K053234) |
| SpO2 module | Nellcor SpO2 | Nellcor SpO2 |
| Patient type | Adult, Pediatric & Neonate | Adult, Pediatric & Neonate |
| SpO2 measurement range | 0~100% | 0~100% |
| | adult/ Pediatric: | adult/ Pediatric: |
| | 70~100%: ±2%; | 70~100%: ±2%; |
| SpO2 measurement accuracy | 0~69%: Unspecified. | 0~69%: Unspecified. |
| | neonate | neonate |
| | 70~100%: ±3%; | 70~100%: ±3%; |
| | 0~69%: Unspecified. | 0~69%: Unspecified. |
| Alarm range(%) | 0~100% | 0~100% |
| Pulse rate measurement | 20~250bpm | 20 to 250 BPM |
34 of 866
K100394
Page 4 of 5
{7}------------------------------------------------
K100394 p8/12
:
| No. | Category | Directives/Standards | Title & Comments |
|-----|--------------------|-------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | General | 93/42/EEC | Medical Device Directive |
| | | 21CFR820 | Code of Federal Regulations |
| | | 91/157/EEC | Battery Declaration Directive |
| | | 93/86/EEC | Battery Disposal Directive |
| | | IEC60601-1:1988 | General requirements for Safety and |
| | | A1:1991,+ A2:1995 | Essential Performance |
| | | IEC60601-1-1:2000 | Medical electrical equipment -- Part 1:<br>General requirements for safety- Collateral<br>standard-Safety requirements for medical<br>electrical systems |
| | | IEC60601-1-4:2000 | Programmable medical systems |
| | | IEC 60601-1-6:2006 | Medical electrical equipment - Part 1-6:<br>General requirements for safety - Collateral<br>standard: Usability |
| | | IEC 60601-1-8: 2006 | Medical electrical equipment - Part 1-8:<br>General requirements for safety - Collateral<br>standard: General requirements, tests and<br>guidance for alarm systems in medical<br>electrical equipment and medical electrical<br>systems |
| | | IEC 60601-2-25 | Medical electrical equipment, Part 2-25:<br>Particular requirements for the safety of<br>electrocardiographs |
| | | IEC 60601-2-27: 2005 | Medical electrical equipment -- Part 2-27:<br>Particular requirements for the safety,<br>including essential performance, of<br>electrocardiographic monitoring<br>equipment. (Cardiovascular) |
| | | IEC 60601-2-49:2001 | Medical electrical equipment --- Part 2-49:<br>Particular requirements for the safety of<br>multifunction patient monitoring equipment |
| | | ISO14971:2007 | Medical devices-Application of risk<br>management to medical devices |
| 2 | Alarm | IEC60601-1-8 2006 | Medical electrical equipment - Part 1-8:<br>General requirements for safety - Collateral<br>standard: General requirements, tests and<br>guidance for alarm systems in medical<br>electrical equipment and medical electrical<br>systems |
| 3 | NIBP | AAMI SP 10:2002 | Electronic or Automated<br>Sphygmomanometers |
| | | EN1060-1:1995 | Non-invasive sphygmomanometers - Part 1:<br>General requirements |
| | | EN1060-3:1997 | Non-invasive sphygmomanometers - Part 3<br>supplementary requirements for |
| | | EN1060-4: 2004 | Non-invasive sphygmomanometers - Part 4:<br>Test procedures to determine the overall<br>system accuracy of automated non-invasive<br>sphygmomanometers. |
| | | IEC60601-2-30:1999 | Medical Electrical Equipment-Part 2-30:<br>Particular requirements for safety of<br>automatic cycling indirect blood pressure<br>monitoring equipment. |
| 5 | SpO2 | EN 865:1997<br>ISO 9919:2005 | Pulse oximeters, Particular requirements<br>Medical electrical equipment -- Part 2-34:<br>Particular requirements for the basic safety<br>and essential performance of pulse oximeters<br>equipment for medical use |
| 6 | Temperature | ASTM E1112:2000<br>ASTM E1104-03<br>EN 12470-4:2000 | Electronic thermometer for intermittent<br>determination of patient temperature<br>Standard Specification for Clinical<br>Thermometer Probe Covers and Sheaths<br>Clinical thermometers-Part 4: Performance of<br>electrical thermometers for continuous<br>measurement. |
| 7 | ECG<br>Measurement | ANSI/AAMI<br>EC11:1991/(R)2001<br>AAMI/ANSI EC13:2002/(R)2007<br>ANSI/AAMI EC<br>12:2000/(R) 2005<br>AAMI EC53/(R) 2001 | Diagnostic electrocardiographic devices<br>Cardiac monitors, heart rate meters, and<br>alarms<br>Disposable ECG electrodes<br>ECG cables and leadwires. (Cardiovascular) |
| 8 | EMC | IEC60601-1-2:2007 | Medical Electrical Equipment-Part<br>1-2:General Requirements for Safety -<br>2.Collateral Standard-Electromagnetic<br>compatibility - Requirements and tests |
| | | IEC61000-3-2 | Harmonic Emission |
| | | IEC61000-3-3 | Voltage Fluctuations/Flicker Emission |
| | | IEC61000-4-2 | Electrostatic Discharge (ESD) |
| | | IEC61000-4-3 | Radiated RF electromagnetic field |
| | | IEC61000-4-4 | Electrical fast Transient/Burst (EFT) |
| | | IEC61000-4-5 | Surge current |
| | | IEC61000-4-6 | Conducted disturbances, induced by RF field |
| | | IEC61000-4-8 | Power frequency (50/60Hz) Magnetic field |
| | | IEC61000-4-11 | Voltage dips, short interruptions, and voltage<br>variation on power supply input lines |
| | | CISPR 11, EN55011 | RF emissions |
| | | | |
| 9 | Biocompatibility | ISO10993-1 | Biological evaluation of medical devices - |
| | | | Part 1: Evaluation and testing |
| | | 15010993-2 | Biological evaluation of medical devices - |
| | | | Part 5: Tests for In Vitro cytotoxicity. |
| | | ISO10993-10 | Biological evaluation of medical devices - |
| | | | Part 10: Tests for irritation and delayed-type |
| | | | hypersensitivity |
| 10 | Labeling | EN1041:1998 | Terminology, symbols and information |
| | | | provided with medical devices - information |
| | | | supplied by the manufacturer with medical |
| | | | devices. |
| 11 | Marking | IEC60878, EN980, | Graphical Symbols for use in the labeling of |
| | | ISO7000, EN60417-1, | Medical Devices |
| | | EN60417-2 | |
| 12 | Package | ISTA: Pre-Shipment Test | Pre-Shipment Test Procedures (Package) |
| | | Procedures | |
| | | (Procedure 1A, 1994 | |
| | | Rev.) | |
| 13 | Reliability<br>i | IEC60068-2-1 | Environmental testing - Part 2-1: Tests - Test |
| | | | A: Cold |
| | | IEC60068-2-2 | Environmental testing - Part 2-2: Tests - Test |
| | | | B: Dry heat |
| | | IEC 60068-2-6 | Environmental testing - Part 2-6: Tests - Test |
| | | | Fc: Vibration (sinusoidal) |
| | | IEC 60068-2-27 | Environmental testing - Part 2-27: Tests - Te |
| | | | st Ea and guidance: Shock |
| | | IEC60068-2-30 | Environmental testing- Part 2-30: Tests - Tes |
| | | | t Db: Damp heat, cyclic |
| | | IEC 60068-2-64 | Environmental testing - Part 2-64: Tests - Te |
| | | | st Fh: Vibration, broadband random and guid |
| | | | ance |
| | | | |
.
1
・
:
:
s
、、 .
35 of 866
.
Page 1 / 3
:
{8}------------------------------------------------
K100394
p9/12
··············································································································································································
・
{9}------------------------------------------------
K100394
p10/12
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.
_____________________________________________________________________________________________________________________________________________________________________________
{10}------------------------------------------------
| ·m (Geostations<br>Or ±5 bpm (Campa<br>20~250bpn<br>3bpm ({<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>ale<br>Ullee<br>ange-Pul<br>Pulse rate accurar | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| AND AND A CLASS OF A | |
| AND of the first of the country of can and other free for the first for for<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Or ±5 bpm (Campar<br>±3bpm (Geostationa<br>20~250bpm<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
Note: Because the same NELLCOR SPO2 module and the sensors and cables are used in PM-9000 EXPRESS PATIENT MONITOR (K053234 Alarm
(bpm)
the Characteristics are same
# dictive 6. Tempera
| ature (Predictive & Monitor) | | |
|--------------------------------------|------------------------------------------|-------------------------------------------|
| Characteristics | Subject Device | Claimed SE Device 510(K) No. |
| Name and model | PMS8210A Multi-parameter Patient Monitor | PM-9000 EXPRESS PATIENT MONITOR (K053234) |
| Temperature parameter | Optional parameter | Optional parameter |
| Patient type | Adult, Pediatric & Neonate | Adult, Pediatric & Neonate |
| Unit of measure | °C & °F | °C & °F |
| Measurement site | Oral, Rectal & Axillary | Oral, Rectal & Axillary |
| Temperature measurement<br>range | 0°C~50°C (32~122°F) | 0°C~50°C (32~122°F) |
| Temperature measurement<br>accuracy | ±0.1°C (±0.2°F) | ±0.1°C (±0.2°F) |
| | ASTM E1112:00 | ASTM E1112:00 |
| Probe cross contamination<br>control | Single use | Single use |
| | Disposable cover | Disposable cover |
38 of 866
Page 5 of 5
{11}------------------------------------------------
-
- 7. Safety and Performance Data :
Please refer to the Declaration of Conformity for the comprehensive list of testing performed on the PMS8210A Multi-parameter Patient Monitor. The PMS8210A has undergone Third Party safety testing in accordance with IEC standards and completed performance testing in accordance with IEC standards. In that this device has software of Moderate concern; the appropriate level of Software evaluation was performed.
# 8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Shanghai 3F Electronics Co., Ltd. concludes that the Patient Monitor, Model PMS8210A, is safe and effective and substantially equivalent to predicate devices as described herein.
- 9. Shanghai 3F Electronics Co., Ltd.will update and include in a summary any other information deemed seasonably necessary by the FDA.
END
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
APR 1 5 2010
Shanghai 3F Electronics Co., Ltd. c/o Mr. Charles Mack Principal Engineer International Regulatory Consultants (IRC) 77325 Joyce Way Echo, OR 97826
Re: K100394
> Trade/Device Name: PMS8210A (IRIS) Multi-parameter Patient Monitor Regulatory Number: 21 CFR 870.2300 Regulation Name: Physiological Patient Monitor (without arrhythmia detection or alarms) Regulatory Class: II (two) Product Code: 74 MWI Dated: February 7, 2010 Received: February 16, 2010
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
:
{13}------------------------------------------------
Page 2 - Mr. Charles Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vo
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
Indications for Use
# Indications for Use
510(k) Number (if known):
Device Name: PMS8210A (IRIS) Multi-parameter Patient Monitor
Indications for Use:
The PMS8210A Multi Parameter Patient Monitor is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Patient Monitor has certain features and functions.
になってきましたができ
The patient parameters that can be monitored by PMS8210A Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2) and Temperature (TEMP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.
The PMS8210A Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.
The PMS8210A Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.
Prescription Use __
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
W. M
510(k) Number
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|--|
| Division of Cardiovascular Devices | |
27 of 866
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.