K100390 · Elliquence, LLC · GEI · Aug 25, 2010 · General, Plastic Surgery
Device Facts
Record ID
K100390
Device Name
SURGIMAX/ SURGIMAX PLUS
Applicant
Elliquence, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 25, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Orthopedic, arthroscopic, spinal, and neurosurgical For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist. Cutting Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty. Blended Cutting and Coagulation Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal CellCarcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps. Hemostasis and Nonablative Coagulation Control of bleeding, Epilation, Telangiectasia Bipolar Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.
Device Story
Compact electrosurgical generator providing high radiofrequency (RF) energy for cutting, coagulation, and hemostasis; operates in monopolar or bipolar modes. Input: user-selected waveforms and power levels via front panel push buttons. Output: 4.0 MHz (monopolar) or 1.7 MHz (bipolar) RF energy delivered to tissue via electrodes. Used in OR/clinic settings by physicians. Operator receives feedback via digital displays and lamps; output controlled by foot/hand switches. Includes self-test and monitoring features; interlocked to prevent operation during failure. Benefits patient by providing precise tissue management and hemorrhage control during surgical procedures.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-2-2 demonstrated. No clinical data provided.
Technological Characteristics
Electrosurgical generator; 4.0 MHz (monopolar) and 1.7 MHz (bipolar) output frequency. Modes: CW Cut, Cut/Coag, Hemo, Bipolar Hemo, Bipolar Turbo. Dimensions: 23.5cm x 1.7cm x 33.7cm; Weight: 20 lbs. Complies with IEC 60601-2-2. Software/electronics identical to predicate.
Indications for Use
Indicated for soft tissue resection, ablation, coagulation, and hemostasis in orthopedic, arthroscopic, spinal, and neurological procedures; also indicated for various ENT, dermatological, and cosmetic procedures including snoring, uvulopalatoplasty, myringotomy, epistaxis, turbinate shrinkage, skin lesions, and cosmetic repairs.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
# K100390
## 510 (k) Summary
#### Date Prepared
AUG 2 5 2010
Revised August 19, 2010
#### Submitter's Information
Joseph Azary c/o Orchid Design 80 Shelton Technology Center Shelton, CT 06484 Tel: (203) 922-0105 Fax: (203) 922-0130 Email: joseph.azary@orchid-orthopedics.com
Manufacturer/Sponsor: Elliquence LLC 3333 Royal Avenue Oceanside, NY 11572
Establishment Registration for Elliquence LLC Inc. is 3007024186.
#### Device Trade Or Proprietary Names
Possible device trade names are:
- . SurgiMax
- . SurgiMax Plus
#### Device Common, Usual, or Classification Names
Electrosurgical Unit and Accessories, Electrosurgical Cutting and Coagulation and Accessories
#### Classification Panel
Classification of this device would fall under the responsibility of the Division of General, Restorative, and Neurological Devices.
#### Class
Classification: Class 2 Product Code: GEI, 21 CFR 878.4400
{1}------------------------------------------------
#### Description of the Device
The SurgiMax / SurgiMax Plus is a compact source of high radiofrequency energy employed for a variety of procedures. This action is achieved by front panel selection of waveforms and power levels. The subject device can be used with monopolar or bipolar devices. The energy is used for cutting, coagulation, and hemostasis.
The SurgiMax / SurgiMax Plus Electrosurgery Generator is a compact source of high radio-frequency RF energy to be employed for a variety of radiosurgery procedures. This action is achieved by front panel selection of waveforms and power level. All selection is effected through push buttons and lamps, which give the operator feedback of status.
Power level for each mode is indicated by front panel digital displays; which also show the status of self-test and monitoring. The display is interlocked with controls to prevent operation when FAIL is displayed. The final output power control is made through foot and/or hand switches. Both Monopolar and Bipolar electrodes are offered.
This device is designed to comply with international safety standards including applicable IEC series electrical safety standards.
The power output in the CW (Cut) mode is 120 watts into (500 ohms) a matched load. The output frequency is maintained at 4.0 MHz +/- 400Hz over all service and loading conditions including short and open for monopolar mode.
Three output waveforms are provided:
CW CUT - Continuous wave output with average power equal to the maximum with no deliberate modulation.
CUT / COAG – Deeply modulated envelope with average to peak power ratio approximately 50%. Modulation occurring at 75 Hz rate.
HEMO - Deeply modulated wave with average to peak ratio approximately 50%. Modulation occurring at 75 Hz rate.
{2}------------------------------------------------
Specifications:
o
Nominal Line Voltage:
240/220/120/100 V
Input Current at Max Output Power:
240/230/220V: 1.5A 120/100V: 3.3A
Width and Height:
।
9.25" x 5" x 13.25" (23.5cm x 1.7cm x 33.7cm)
Weight:
20 lbs (9.07 kg)
Output Characteristics:
| Output Characteristics: | | | |
|-------------------------|------------------------------------------------------|----------------------|-------------------------------|
| Mode | Output Wave-<br>form | Max Power | Activation |
| CUT | 4.0 MHz CW<br>sinusoid | 120 W at 500<br>Ohms | Footswitch or<br>fingerswitch |
| CUT/COAG | 4.0 MHz with<br>rectified full<br>wave envelope | 90 W at 500<br>Ohms | Footswitch or<br>fingerswitch |
| HEMO | 4.0 MHz with<br>square wave<br>rectified<br>envelope | 60 W at 500<br>Ohms | Footswitch or<br>fingerswitch |
| BIPOLAR<br>HEMO | 1.7 MHz with<br>square wave<br>rectified<br>envelope | 40W at 200<br>Ohms | Footswitch |
| BIPOLAR<br>TURBO | 1.7 MHz with<br>CW sinusoid | 120W at 200<br>Ohms | Footswitch |
Intended Use [21 CFR 807.92(a)(5)]
{3}------------------------------------------------
### Orthopedic, arthroscopic, spinal, and neurosurgical
For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist.
#### Cutting
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, turnors, cosmetic repairs, development of skin flaps, Blepharoplasty.
#### Blended Cutting and Coagulation
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal CellCarcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps.
#### Hemostasis and Nonablative Coagulation
Control of bleeding, Epilation, Telangiectasia
#### Bipolar
Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.
#### Technological Characteristics [21 CFR 807.92(a)(6)]
Elliquence LLC believes that the subject device has the identical technological characteristics of the predicate device. This include outputs, performance, software, electronics, and indications for use. We believe the subject device is substantially equivalent to the predicate device.
#### Performance Data [21 CFR 807.92(b)(1)]
The Ellman SurgiMax / SurgiMax Plus complies with IEC 60601-2-2. Non-clinical testing and comparison between the subject device and predicate device demonstrate equivalent performance.
#### Predicate Device [21 CFR 807.92(a)(3)]
The predicate devices are listed as follows:
{4}------------------------------------------------
#### Ellman Surgimax - K061174 .
The subject device and predicate device have identical indications for use, waveforms, functionality with bipolar and monopolar devices, frequency, maximum power, weight, and power adjustment. The electronics and software in the subject device is the same identical electronics and software in the predicate device. The outputs and performance characteristics are identical as the electronics and software have not changed.
The differences are:
- . Cosmetic differences to the outer case and faceplate
- . Difference in product name (Elliquence Surgimax or Elliquence Surgimax Plus versus Ellman Surgimax)
- Labeling changed to reflect the company name and the image to reflect the cosmetic differences. .
These differences do not affect safety or performance as they are cosmetic or aesthetic in nature.
#### Conclusion [21 CFR 807.92(b)(3)]
We believe the differences between the subject device and predicate device are minor and conclude that the subject devices are as safe and effective as the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Elliquence, LLC. % Orchid Design Mr. Joseph Azary 80 Shelton Technology Center Shelton, Connecticut 06484
AUG 2 5 2010
Re: K100390
Trade Name: Eliquence Surgimax / Surgimax Plus Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: July 20, 2010 Received: July 22, 2010
Dear Mr. Azary: .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Joseph Azary
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 100-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm/15809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Coopliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffred number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/medicaldevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark M Millerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
K100390
#### 510(k) Number (if known):
#### Device Name: Elliquence Surgimax / Surgimax Plus
#### Indications For Use:
#### Orthopedic, arthroscopic, spinal, and neurosurgical
For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist.
### Cutting
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty.
#### Blended Cutting and Coagulation
Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal CellCarcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps.
Hemostasis and Nonablative Coagulation Control of bleeding, Epilation, Telangiectasia
#### Bipolar
Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, submucosal palatal shrinkage, traditional uvulopalatop!asty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
K100390
1
(Please Do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of *Device Evaluation* (ODE)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.