HHC PUSH-BUTTON REPLACEMENT TURBINE FOR MIDWEST TRADITION HANDPIECE MODEL MW TURBINE REPLACEMENT
Applicant
Handpiece Headquarters
Product Code
EFB · Dental
Decision Date
Aug 12, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
This HHC-Push button air turbine is intended for use with the Midwest Tradition Handpiece in replacement of Midwest Tradition's original turbine. The Dental Handpiece with the HHC- replacement turbine is used by authorized persons in the practice of dentistry. The HHC-replacement turbine is used in the Midwest Tradition High Speed Dental Handpiece, which is intended for removing carious material, cavity preparations, finishing tooth preparations, reducing hard tooth structures, restorations and polishing teeth.
Device Story
HHC-Push button air turbine is a replacement turbine for Midwest Tradition high-speed dental handpieces. Device operates via high-pressure air (35 psi) impacting impeller blade pockets, driving rotation at minimum 420,000 rpm. Cooling water (2 bar) is delivered via inlet tube to the cutting area. Used by dentists/authorized personnel in clinical settings for tooth preparation and restoration. Output is mechanical rotation of a dental bur. Benefits include restoration of handpiece function using a compatible replacement component.
Clinical Evidence
Bench testing only; no clinical data required. Performance testing confirmed the device functioned as intended with results comparable to the predicate device.
Technological Characteristics
Materials: 440C stainless steel (spindle, bearings), aluminum (impeller), Buna (O-ring). Operation: Air-powered turbine, 35 psi, 420,000 rpm min. Chuck: Push-button auto chuck. Dimensions: 0.125" OD, 0.425" length. Conformance: Meets ISO standard for shanks (0.0625"-0.0630").
Indications for Use
Indicated for use by dental professionals in the Midwest Tradition High Speed Dental Handpiece for removing carious material, cavity preparation, finishing tooth preparations, reducing hard tooth structures, restorations, and polishing teeth.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
NSK- Replacement Turbine for Midwest Tradition Handpiece (K971774)
Submission Summary (Full Text)
{0}------------------------------------------------
K100389
**Handpiece**
Headquarters
AUG 2 2010
34
A HENRY SCHEIN® COMPANY
## 510(K) SUMMARY:
HHC- push button replacement turbine for Midwest Tradition.
210K
a- Submitted by:
HANDPIECE HEADQUARTERS 620 S. Placentia Ave. Placentia, CA 92870 Tel. 714-579-0175 Fax. 714-579-0186
b- Contact person: ·
Tina Steffanie-Oak Tel. 717-335-7230, ext. 4150 Fax. 717-335-7240 tina.steffanie-oak@henryschein.com Email:
c- Date summary prepared:
Revised 06-10-10
d- Device Name:
HHC- push button replacement turbine for Midwest Tradition Trade or Proprietary Name:
Common Name: Classification Name: Air-Power Dental Handpiece Dental Handpiece and accessories (21CFR 872.4200) Class: I
Product Code:
EFB "Handpiece, Air-powered, Dental"
e- Substantial Equivalency is claimed against the following device:
NSK- Replacement Turbine for Midwest Tradition Handpiece Products of NSK NAKANISHI INC. (K971774)
f- Description of the device:
The HHC-Push button air turbine is a complete air turbine which will be offered as replacement turbine for Midwest Tradition original turbine.
High pressured air directly impacts the blade pockets of the impeller through an air intake tube located inside the handpiece body, which causes the air turbine to rotate at a high speed. The dental cutting bur connects to the turbine's shank (spindle),
{1}------------------------------------------------
which rotates at the same speed as the impeller. Cooling water flows through an inlet water tube at high pressure and blows out at an angle near the dental cutting bur to reduce temperature at cutting area.
#### g- Statement of Intended Use:
This HHC-Push button air turbine is intended to use with Midwest Tradition Handpiece in replacement of Midwest Tradition's original turbine.
The Midwest Tradition Handpiece with the replacement turbine is used by authorized persons in the practice of dentistry.
The replacement turbine is used in the Midwest Tradition High Speed Dental Handpiece, which is intended for removing carious material, cavity preparations, finishing tooth preparations, reducing hard tooth structures, restorations and polishing teeth.
### h- Safety and effectiveness of the device:
The HHC-Push button air turbine is as safe and effective as the predicate device as cited above.
#### 1 - Device and Predicate Comparison Table:
| Descriptive Information | Subject Device: HHC-Midwest<br>Tradition Replacement Turbine | Predicate: NSK-Midwest Tradition<br>Replacement Turbine |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication of use | This HHC-Push button air turbine is<br>intended to use in Midwest Tradition<br>Handpiece in replacement of Midwest<br>Tradition's original turbine.<br><br>The Dental Handpiece with the<br>replacement turbine is used by<br>authorized persons in the practice of<br>dentistry.<br><br>This replacement turbine used in<br>Midwest Tradition Handpiece is<br>intended for removing carious material,<br>cavity preparations, finishing tooth<br>preparations, reducing hard tooth<br>structures, restorations and polishing<br>teeth: | The devices claimed herein are the air turbines<br>only, not the handpiece itself.<br>The devices are intended for use in the<br>Midwest's Quiet-air or Tradition handpieces, in<br>place of Midwest's original turbines.<br><br>The device, therefore, is, with Midwest<br>handpiece, intended for use in general dental<br>applications, such as cutting a tooth for crown<br>preparation, cavity preparation, finishing the<br>crown, inlay or the filling, etc. |
| Device Design: | | |
| Operation mode | Intake air pressure 35 psi- speed<br>420,000 rpm. Min. | Intake air pressure 35 psi- speed 420,000 rpm<br>Min. |
| Air/water ports | N/A | N/A |
| Fiber Optics | N/A | N/A |
| Type of chuck | Push button chuck (auto chuck) | Push button chuck(auto chuck) |
{2}------------------------------------------------
| Coupling dimensions | n/a | n/a |
|---------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Accessories | n/a | n/a |
| Composition of material | | |
| Spindle | 440C stainless steel (see more details in attachment 4 material table) | Stainless steel |
| Impeller | Aluminum (see more details in attachment 4 material table) | Aluminum |
| Bearings | 440C stainless steel. (see more details in attachment 4 material table) | Stainless steel |
| O-ring | Buna (see more details in attachment 4 material table) | Buna |
| Technical specification | | |
| Chuck | Push button chuck spindle<br>Outside diameter: 0.125"<br>Chuck length: 0.425"<br>Actuator height: 0.034" | Push Button:<br>Outside diameter: 0.125"<br>Chuck length: 0.435"<br>Actuator height: 0.038" |
| Light intensity | n/a | n/a |
| Bur extraction force | 5 lbs Min. push and pull forces/ 100 inspection using ONMI gauge/ per specification indicated on drawing | 5 lbs Min. push and pull forces- tested on predicate device |
| Air pressure | 35 psi. recommended | 35 psi. |
| Water pressure | 2 bar | 2 bar |
| Speed Rpm | 420,000 rpm. Min. | 420,000 rpm. Min. |
| Hose connection | n/a | n/a |
| Conformance standard for shanks | Meet I.S.O standard Shank diameter from 0.0625" to 0.0630" | Meet I.S.O standard shank |
| Lubricant | n/a | Dental lubricated oil |
| | | Lubricant recommended but specific type and it is not indicated in labeling. |
### J - Performance Data:
The HHC-Push button air turbine performance tests were carried out to evaluate it In all instances the HHC-Push button air turbine against the predicate device. functioned as intended, results observed were as expected, and was determined to have comparable performance and safety as compared to the predicate.
No clinical data was required.
#### K -Conclusion:
Based on the information in the notification Handpiece Headquarters believes that this HHC-Push button air turbine is substantially equivalent to the claimed predicate device. (NSK- Replacement Turbine for Midwest Tradition Handpiece Products of NSK-NAKANISHI INC. (K971774))
36
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and legs. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tina Steffanie-Oak Senior Regulatory Specialist Handpiece Headquarters 620 S. Placentia Avenue Placentia, California 92870
AUG 1, 2 2010
Re: K100389
Trade/Device Name: HHC- push button replacement turbine for Midwest Tradition Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: II Product Code: EFB Dated: August 6, 2010 Received: August 6, 2010
Dear Ms. Steffanie-Oak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a preemarket approval application (PMA). You may, therefore, market the device, subject to the emeral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Steffanie-Oak
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. 10, 10, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
AUG 1 2 2010
510(k) Number (if known): Kloo389
Device Name: HHC- push button replacement turbine for Midwest Tradition
Indications for Use:
This HHC-Push button air turbine is intended for use with the Midwest Tradition Handpiece in replacement of Midwest Tradition's original turbine.
The Dental Handpiece with the HHC- replacement turbine is used by authorized persons in the practice of dentistry.
The HHC-replacement turbine is used in the Midwest Tradition High Speed Dental Handpiece, which is intended for removing carious material, cavity preparations, finishing tooth preparations, reducing hard tooth structures, restorations and polishing teeth.
Prescription Use .X AND/OR (Per 21 CFR 801 Subpart D) Subpart C)
Over-The-Counter Use
(Per 21 CFR 801
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Loe Miley for MSP Page 1 of 1
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.