K100358 · Edan Instruments, Inc. · HGM · Jan 6, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K100358
Device Name
CENTRAL MONITORING SYSTEM
Applicant
Edan Instruments, Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Jan 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Maternal Fetal Monitoring – Central Nurse System (MFM-CNS) is a clinical data managing software application and is indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting. The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual input for viewing at the central nursing station. The system also produces an electronic medical record.
Device Story
MFM-CNS is a software application running on a PC station; connects up to 32 bedside fetal/maternal monitors via Ethernet. Inputs include real-time physiological data (FHR, UA, maternal HR, SpO2, NIBP, RR, TEMP, pulse rate) and manual entries. System centralizes monitoring management at a nursing station; provides visual/audible alarms for parameter limit breaches or poor signal quality. Outputs include real-time display of patient data, electronic medical records, and printed reports (CTG, vital signs). Enables 24-hour CTG and vital sign data review; supports data backup to CDs. Used by clinical staff in healthcare settings to facilitate centralized patient monitoring and clinical decision-making.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including software testing, risk analysis, safety testing, and performance testing to verify and validate system functionality.
Technological Characteristics
Software-based central monitoring system; runs on PC hardware with Microsoft Windows XP. Connectivity via Ethernet to bedside monitors. Supports up to 32 concurrent connections. Features include data acquisition, storage, alarm management, and report printing. No specific materials or sterilization methods applicable as it is a software-only device.
Indications for Use
Indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
��
# 510(K)Summary of Safety and Effectiveness
JAN - 6 2011
This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92
| Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86755 26882220<br>Fax:+86 755 26882223 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Jiang yucai<br>Edan Instruments, Inc. |
| Preparing date: | 2010-08-23 |
| Proprietary Name: | Central Monitoring System (model MFM-CNS) |
| Classification<br>information: | 21 CFR 884.2740 Perinatal monitoring system and accessories<br>Class II |
| Product code: | HGM |
| Review Panel: | Obstetrics/Gynecology |
| Predicate Devices: | CIV-ob Obstetrical Monitoring Software Application (K051175) |
| Device Description: | MFM-CNS is a software production who runs on PC station<br>with Microsoft Windows XP operating system. MFM-CNS<br>by connecting one central station with some bedside fetal /<br>maternal monitors, carries out centralized monitoring<br>management for many beds. It can monitor a pregnant<br>woman during the whole parturition process, and all the<br>monitoring information can be recorded, saved and printed,<br>and alarm when the parameter exceed the user defined limit<br>and poor signal quality. At the same time, the old records can |
Device features:
:
be searched conveniently and quickly.
{1}------------------------------------------------
- Connect maximum 32 bedside fetal / maternal . monitors with Ethernet.
- Display FHR, UA, Maternal HR, PR, SpO2, NIBP, . RR and TEMP numerics on the screen.
- The screen displays all the monitors simultaneously, . or displays one monitor in full screen.
- 24-hour CTG, 1440-group Maternal Vital Sign data, . 200-group NIBP data review.
- . Print CTG report, Maternal Vital Sign list, NIBP list on the paper
- Audible & visible alarm when FHR or maternal vital . sign exceeds the user defined limit or poor signal quality.
- Patient information, CTG, Maternal Vital Sign list . and NIBP list can be saved, and burned into CDs for backup.
- Support user accessing control. .
The Maternal Fetal Monitoring - Central Nurse System (MFM-CNS) is a clinical data managing software application and is indicated for antepartum and intrapartum monitoring of pregnant women in a healthcare setting.
The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual input for viewing at the central nursing station. The system also produces an electronic medical record.
The MFM-CNS has display fields for the following obstetric data:
- patient demographics ●
- provider notes .
- FHR
- uterine activity (via tocodynamometry or IUP) .
- maternal heart rate .
- SpO2 .
- NIBP
- respiratory rate
- temperature
- pulse rate
# 1.2 Indications
# for Use
l -2
{2}------------------------------------------------
#### Test Summary:
The following quality assurance measures were applied to the development of the MFM-CNS Electrocardiograph:
- Software testing .
- Risk analysis .
- Safety testing .
- Performance test .
### Conclusion:
Verification and validation testing was done on MFM-CNS. This premarket notification submission demonstrates that the subject device MFM-CNS is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
ここで
| 小 | 彩
Edan Instruments, Inc. c/o Mr. William Stern Official Correspondent Multigon Industries, Inc. l Odell Plaza YONKERS NY 10701
Re: K100358
> Trade Name: MFM-CNS (Central Monitoring System) Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: December 20, 2010 Received: December 23, 2010
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Stuckert Lemieux MD
lerbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
Enclosure
{5}------------------------------------------------
## Indications for Use
#### K100358 510(k) Number:
# JAN - 6 2011
.. :
Device Name: Central Monitoring System (model MFM-CNS)
The Maternal Fetal Monitoring – Central Nurse System (MFM-CNS) is a clinical data managing software application and is indicated for antepartum monitoring of pregnant women in a healthcare setting.
The MFM-CNS is intended to manage perinatal monitoring data acquired from bedside monitors or manual input for viewing at the central nursing station. The system also produces an electronic medical record.
The MFM-CNS has display fields for the following obstetric data:
- . patient demographics
- . provider notes
- . FHR
- uterine activity (via tocodynamometry or IUP) .
- . maternal heart rate
- SpO2 .
- NIBP .
- respiratory rate .
- temperature .
- pulse rate .
| Prescription Use | <span style="text-decoration: overline;">X</span> | Or | Over the Counter Use | <span style="text-decoration: overline;"></span> |
|-----------------------------|---------------------------------------------------|----|-----------------------------|--------------------------------------------------|
| (21 CFR Part 801 Subpart D) | | | (21 CFR Part 801 Subpart C) | |
Concurrence of CDRH, Office of *In Vitro* Diagnostic Device Evaluation and Safety (OIVD)
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and Urological Devices
| 510(k) Number | K100358 | 1-4 |
|---------------|---------|-----|
|---------------|---------|-----|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.