OptiBond SE is a two-component self-etch universal adhesive system designed to be used for all direct and indirect applications including, but not limited to, the following: Direct Applications Light-cured composite and compomer restorations . Composite/ceramic/metal repairs . Cavity sealing for amalgam restorations . . Sealing of hypersensitive and/or exposed root surfaces Core build-ups (self-cured, light-cured or dual-cured) . Indirect applications . Veneers . Porcelain, composite, and metal-based (including zirconia-based and alumina-● based) inlays, onlays, crowns, bridges Endodontic posts . Cavity sealing as a pretreatment for indirect restorations ●
Device Story
OptiBond SE is a two-component self-etch universal dental adhesive system; comprises self-etch primer and universal adhesive. Primer etches uncut enamel and dentin without separate phosphoric acid etch; adhesive contains 15% filled 0.4 micron barium glass for bond strength reinforcement. Used by dentists in clinical settings for direct and indirect dental procedures. Compatible with self-cure, light-cure, or dual-cure resin cements and core build-up materials. Simplifies bonding by eliminating secondary systems. Benefits patient by providing reliable adhesion for various restorative materials including porcelain, composite, metal, zirconia, and alumina.
Clinical Evidence
Bench testing only. Performance evaluated via bond strength testing to various dental substrates including dentin, enamel, composite, porcelain, gold, and PFM. Biocompatibility testing performed to ensure safety.
Technological Characteristics
Two-component self-etch adhesive system. Contains 15% filled 0.4 micron barium glass. Compatible with self-cure, light-cure, and dual-cure resin cements. No electronic components or software.
Indications for Use
Indicated for dental patients requiring adhesive bonding for direct restorations (composite/compomer, repairs, amalgam sealing, hypersensitive/exposed root surfaces, core build-ups) and indirect restorations (veneers, inlays, onlays, crowns, bridges, endodontic posts).
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
K100347
# APR 3 0 2010
# Section III - 510(k) Summary of Safety and Effectiveness
## Submitter:
Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7472 - Facsimile Wendy Garman - Contact Person
Date Summary Prepared: April 2010
#### Device Name:
- . Trade Name - OptiBond SE
- Common Name Bonding Agent .
- . Classification Name -- Resin Tooth Bonding Agent, per 21 CFR § 872.3200
#### Devices for Which Substantial Equivalence is Claimed:
- Kerr Corporation, OptiBond All-In-One .
#### Device Description:
· OptiBond SE is a two-component self-etch universal adhesive system including a self-etch PRIMER and a universal ADHESIVE. The self-etch primer provides effective etching to uncut enamel and dentin, without the need for a separate phosphoric acid etch, thus simplifying the bonding procedure. The adhesive component is 15% filled with 0.4 micron barium glass to help reinforce bond strength. The material's chemistry allows for compatibility with all self-cure or dual-cure resin cements and core build-up materials. Dentists can therefore utilize OptiBond SE for their direct and indirect procedures without the need for a secondary bonding system.
{1}------------------------------------------------
# Intended Use of the Device:
- OptiBond SE is a two-component self-etch universal adhesive system designed ● to be used for all direct and indirect applications including, but not limited to, the following:
# Direct Applications
- Light-cured composite and compomer restorations .
- . Composite/ceramic/metal repairs
- . Cavity sealing for amalgam restorations
- . Sealing of hypersensitive and/or exposed root surfaces
- Core build-ups (self-cured, light-cured or dual-cured) .
# Indirect applications
- . Veneers
- Porcelain, composite, and metal-based (including zirconia-based and alumina-● based) inlays, onlays, crowns, bridges
- Endodontic posts
- Cavity sealing as a pretreatment for indirect restorations ●
# Substantial Equivalence:
OptiBond SE is substantially equivalent to one other legally marketed device in the United States. OptiBond SE functions in a manner similar to and is intended for the same use as OptiBond All-In-One that is currently marketed by Kerr Corporation. A biocompatibility study was completed, which demonstrates that the material is safe for its intended use.
This 510(k) submission also includes data from bench testing used to evaluate the performance characteristics of OptiBond SE compared to the predicate device, OptiBond All-In-One. The characteristics evaluated include direct and indirect bonding strength to various materials readily found in the dental industry including, but not limited to dentin, enamel, composite, porcelain, gold and PFM.
Based upon the biocompatibility test and bench testing, the clinical performance of Optibond SE is substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with a simple, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Kerr Corporation Ms. Wendy Garman Director, Regulatory Affairs Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867
APR 30 2010
Re: K100347
Trade/Device Name: OptiBond SE Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: February 8, 2010 Received: February 12, 2010
Dear Ms. Garman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
.
{3}------------------------------------------------
Page 2- Ms. Garman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Susan Burns
Anthony Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{4}------------------------------------------------
K100347
# Indications for Use
# 510(k) Number (if known):
Device Name: OptiBond SE
## Indications For Use:
OptiBond SE is a two-component self-etch universal adhesive system designed to be used for all direct and indirect applications including, but not limited to, the following:
## Direct Applications
- Light-cured composite and compomer restorations ●
- Composite/ceramic/metal repairs .
- Cavity sealing for amalgam restorations .
- . Sealing of hypersensitive and/or exposed root surfaces
- Core build-ups (self-cured, light-cured or dual-cured) .
# Indirect applications
- Veneers .
- . Porcelain, composite, and metal-based (including zirconia-based and aluminabased) inlays, onlays, crowns, bridges
- Endodontic posts ●
- Cavity sealing as a pretreatment for indirect restorations
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
| 510(k) Number: | K100347 |
|----------------|---------|
|----------------|---------|
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.