← Product Code [ODA](/productcode/ODA) · K100345

# SYSTEMS INTEGRATION MODEL: ENDOALPHA (K100345)

_Olympus Medical Systems Corporation · ODA · Mar 22, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100345

## Device Facts

- **Applicant:** Olympus Medical Systems Corporation
- **Product Code:** [ODA](/productcode/ODA.md)
- **Decision Date:** Mar 22, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Systems Integration ENDOALPHA has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment.

## Device Story

Systems Integration ENDOALPHA acts as a central hub for endoscopic suites; connects to Olympus endoscopes and various ancillary equipment (e.g., electrosurgical units, insufflators, SonoSurg generators). Operates via Medical Control Unit (UCES-3) and Extension Unit (MAJ-1827). Provides centralized control via touch panel (MAJ-1828), voice commands, or endoscope remote switches; enables automatic initial settings and scene-based configurations for procedure efficiency. Displays equipment status, warnings, and error messages on universal displays; supports voice feedback and still image viewing from USB drives. Facilitates automatic smoke evacuation when compatible electrosurgical/insufflation units are connected. Used in clinical settings by medical staff to streamline equipment management, reduce manual adjustments, and improve workflow during endoscopic procedures.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and design verification of system modifications.

## Technological Characteristics

System consists of Medical Control Unit (UCES-3), Extension Unit (MAJ-1827), Touch Panel (MAJ-1828), and Universal Display (MAJ-1176). Functions as a centralized controller for ancillary endoscopic equipment. Connectivity via proprietary communication interfaces. Software-based control system manages interlocking operations, GUI display, and voice feedback. No specific material standards or energy sources beyond standard electrical power for medical equipment are detailed.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 ([K051613](/device/K051613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

JUL 2 7 2015

Olympus Medical Systems Corp. % Ms. Stacy Abbatiello Kluesner, M.S., RAC Regulatory Affairs Project Manager Olympus America, Inc. 3500 Corporate Parkway Center Valley, PA 18034-0610

Re: K100345

> Trade/Device Name: SYSTEMS INTEGRATION ENDOALPHA Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODA, GCJ Dated (Date on orig SE Itr): February 5, 2010 Received (Date on orig SE ltr): February 24, 2010

Dear Ms. Kluesner,

This letter corrects our substantially equivalent letter of March 22, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known): x 100 3 4 5 Device Name: SYSTEMS INTEGRATION ENDOALPHA Indications for Use:

The Systems Integration ENDOALPHA has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment.

Prescription Use_ (Part 21 CFR 801 Subpart D)

--

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Habib Remine

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number

Page 1 of _ 1

{3}------------------------------------------------

K100345
Page 143

## 510(k) SUMMARY

# MAR 2 2 2010

# SYSTEMS INTEGRATION ENDOALPHA

January 29, 2010

#### 1 General Information

1 Applicant: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047

Official Correspondent: Stacy Abbatiello Kluesner, M.S., RAC Regulatory Affairs & Quality Assurance Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5405 FAX: 484-896-7128 Email: stacy.kluesner@olympus.com Establishment Registration No: 2429304

Manufacturer:

SHIRAKAWA OLYMPUS CO., LTD. 3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura, Nishishirakawa-gun Fukushima, JAPAN 961-8061 Establishment Registration No: 3002808148

#### 2 Device Identification

- Device Name:
- Common Name:
- Regulation No:
- 트 Regulation Name:
- 월 Regulatory Class:
- 내 Product Code:
- Classification Panel: 월출
- Prescription Status: 물

Performance Standards:

## SYSTEMS INTEGRATION ENDOALPHA

Endosurgery system 21 CFR 876.1500

Endoscope and accessories

KOG, GCJ

ll

Gastroenterology/Urology

Prescription device

None established under Section 514.

{4}------------------------------------------------

K100345
Page 2 of 3

#### 3 Predicate Device Information

| Device Name: | Olympus Integrated Endosurgery System EndoALPHA<br>(Control Unit for Endosurgery UCES-2) Software Version 3 |
|--------------|-------------------------------------------------------------------------------------------------------------|
|--------------|-------------------------------------------------------------------------------------------------------------|

- 510(k) No: K051613
Decision Date: 08/15/2005

#### Device Description ব

The Systems Integration ENDOALPHA has been designed to be used with an Olymous endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment.

The ENDOALPHA integrated electrosurgical system has the following system functions.

## 1) Centralized control from a non-sterilized area

The ancillary equipment connected to the MEDICAL CONTROL UNIT FOR ENDOSURGERY (UCES-3) and EXTENSION UNIT FOR UCES-3 (MAJ-1827) can be controlled in a centralized way using the TOUCH PANEL FOR ENDOALPHA (MAJ-1828).

2} Centralized control from a sterilized area

- Ancillary equipment can be controlled by means of voice from a sterilized area.
Ancillary equipment can be controlled utilizing remote switches of the endoscope connected to the VIDEOSYSTEM CENTER (OLYMPUS CV-180, OLYMPUS OTV-S7Pro) in a sterilized area.

#### 3) Centralized display

The settings of the functions of connected ancillary equipment can be displayed on the screen of the TOUCH PANEL FOR ENDOALPHA (MAJ-1828) and UNIVERSAL DISPLAY FOR ENDOALPHA (MAJ-1176).

## 4) Message display

Various warning and error messages can be displayed on the screen of the TOUCH PANEL FOR ENDOALPHA (MAJ-1828) and UNIVERSAL DISPLAY FOR ENDOALPHA (MAJ-1176).

- 5) Voice reply .
The ancillary equipment settings, warning messages can be fed back in voice.

## 6) Automatic settings

The default settings of all ancillary equipment can be registered on a per-procedure basis and recalled with a one-touch operation for setting the ancillary equipment to the required settings at once.

## 7) Scene settings

The operations of the ancillary equipment, performed at the beginning of each scene in the procedure, can be registered and recalled with a one-touch operation for performing them at once.

## 8) Display customization

The contents of the display of the TOUCH PANEL FOR ENDOALPHA (MAJ-1828) and UNIVERSAL

{5}------------------------------------------------

K100345
Page 3 of 3

63456

DISPLAY FOR ENDOALPHA (MAJ-11-76) can be customized.

9) Still image viewing

Still images recorded in a USB thumb drive (SDCZ8) can be viewed on the UNIVERSAL DISPLAY FOR ENDOALPHA (MAJ-1176).

## 10) Automatic smoke evacuation

When a recommended electrosurgical unit and/or SonoSurg Generator (SonoSurg-G2) and an insufflator are connected to the MEDICAL CONTROL UNIT FOR ENDOSURGERY (UCES-3) or EXTENSION UNIT FOR UCES-3 (MAJ-1827), the generated smoke or vapor can be evacuated automatically.

The modifications made to the subject device are as follows:

- A) Addition or deletion of recommended ancillary equipment
- B) A change involving the communication interface
- C) A change in the design of the Graphical User interface (GUI)
- D) Addition of a voice reply function
- E) Deletion of Surgeon's controller for EndoALPHA (MAJ-1140)
- F) Transfer of Input/Output function for video signals to the Advanced system
- G) Addition of additional connectable touch panel
- H) Addition of Scene settings function
- l) Addition of system components
- Changes to the specifications of system components S

#### ട് Intended Use

The Systems Integration ENDOALPHA has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment.

The intended use of the ENDOALPHA as stated above is to enable a central system to control various pieces of ancillary equipment. However, the previously cleared indications for use for each separate ancillary device dictate the type of procedures that may be performed. This information is included in the instruction manual for each ancillary piece of equipment.

The Intended Use is identical to the Intended Use previously cleared for the Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3, K051613.

#### 6 Conclusion

The modified SYSTEMS INTEGRATION ENDOALPHA has the following similarities to the predicate device:

- Has the same intended use, l
- -Uses the same operating principle

In summary, the modified SYSTEMS INTEGRATION ENDOALPHA described in this submission is, in our opinion, substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K100345](https://fda-staging.innolitics.com/device/K100345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
