The clinical performance of the MultiCode-RTx HSV 1&2 Kit was established by comparing results from 1,041 routine clinical vaginal swab specimens against a reference cell culture-based identification and typing test.
Symptomatic female patients presenting with vaginal lesions; Sample Size: 1041; Number of Sites: 3
ELVIS® Herpes Simplex Virus identification and typing test (cell culture based)
Sensitivity and specificity for HSV-1 and HSV-2 detection
Indications for Use
The MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test for the detection and typing of herpes simplex virus (HSV 1 & 2) DNA in vaginal lesions. It is indicated for use in the detection and typing of HSV-1 or HSV-2 in vaginal lesion swab specimens from symptomatic female patients as an aid in the diagnosis of genital herpes infection. Warning: The device is not FDA cleared for use with cerebral spinal fluid CSF; for any lesions other than vaginal. The assay is not intended to be used for male penile specimens, for prenatal screening, or for females under the age of 18 years.
Device Story
MultiCode®-RTx HSV 1&2 Kit is a qualitative PCR-based in vitro diagnostic test. Input: vaginal lesion swab specimens collected in universal transport media. Process: nucleic acid extraction via Roche MagNA Pure LC; PCR amplification using specific primer pairs labeled with distinct fluorophores and isobases (2'-deoxy-5-methyl-isocytidine and 2'-deoxyisoguanosine). Incorporation of quencher-labeled nucleotides decreases fluorescence during amplification; subsequent thermal denaturation increases fluorescence at specific melting temperatures (Tm) to distinguish HSV-1 and HSV-2 amplicons. Output: real-time fluorescence monitoring, analyzed by proprietary software to generate a printed results report. Used in clinical laboratories by trained personnel. Benefits: provides rapid, specific typing of HSV-1 and HSV-2 to aid clinical diagnosis of genital herpes infection.
Clinical Evidence
Prospective study at three U.S. clinical labs (N=1041 vaginal swab specimens) compared MultiCode®-RTx to ELVIS® cell culture. HSV-1: 92.4% sensitivity (95% CI: 85.7-96.1%), 98.3% specificity (95% CI: 97.2-98.9%). HSV-2: 95.2% sensitivity (95% CI: 91.4-97.4%), 93.6% specificity (95% CI: 91.8-95.1%). Discordant samples analyzed via DNA sequencing.
Technological Characteristics
Real-time PCR assay using expanded genetic alphabet (isobases: iC, iG). Targets glycoprotein B gene. Roche LightCycler 1.2 instrument. Software-based analysis of fluorescence and melting temperature (Tm) profiles. Reagents stored at -15°C to -30°C. Includes DNA Sample Processing Control (SPC).
Indications for Use
Indicated for symptomatic female patients aged 18 years and older for the detection and typing of HSV-1 and HSV-2 DNA in vaginal lesion swab specimens as an aid in the diagnosis of genital herpes infection. Not indicated for male penile specimens, prenatal screening, or CSF.
Regulatory Classification
Identification
Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K100336
B. Purpose for Submission: Clearance of New Device
C. Measurand: HSV-1 and HSV-2 nucleic acids target sequences
D. Type of Test: Real Time Polymerase chain reaction (PCR)-based qualitative *in vitro* diagnostic test for the detection and typing of herpes simplex virus 1 & 2 (HSV 1&2) DNA using vaginal swab specimens
E. Applicant: Eragen Biosciences, Inc.
F. Proprietary and Established Names: MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit
## G. Regulatory Information:
1. Regulation section: 21 CFR 866.3305
2. Classification: Class II
3. Product code: OQO, HSV NAAT assays, HSV PCR assays
4. Panel: Microbiology (83)
## H. Intended Use:
1. Intended use(s):
The MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit is a polymerase chain reaction (PCR)-based qualitative *in vitro* diagnostic test for the detection and typing of herpes simplex virus (HSV 1 & 2) DNA in vaginal lesions. It is indicated for use in the detection and typing of HSV-1 or HSV-2 in vaginal lesion swab specimens from symptomatic female patients as an aid in the diagnosis of genital herpes infection.
> Warning: The device is not FDA cleared for use with cerebral spinal fluid CSF; for any lesions other than vaginal. The assay is not intended to be used for male penile specimens, for prenatal screening, or for females under the age of 18 years.
{1}
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Extraction: Roche MagNA Pure LC Total NA Kit and Magna Pure Instrument
PCR: Roche LightCycler 1.2
Software: Eragen MultiCode®-RTx HSV 1&2 Kit Analysis Software (V2.0 RC1)
I. Device Description:
The EraGen MultiCode®-RTx Herpes Simplex Virus 1 & 2 kit is a polymerase chain reaction (PCR)-based qualitative *in vitro* diagnostic test for the detection and typing of herpes simplex virus 1 & 2 (HSV 1 & 2) DNA using vaginal swab specimens.
Patient vaginal swab specimens are collected in Copan Universal Transport Medium, or identical Copan manufactured media formulations (Becton Dickinson Universal Viral Transport Media, Copan branded Universal Transport Medium for LabCorp, and the Quest Viral Culture Media) and transported to the laboratory. An extractable sample processing control (SPC) target is added to the specimen prior to lysis. The SPC controls for specimen lysis, for recovery of extracted nucleic acid, for inhibitory substances and for PCR reagent and instrument integrity. The specimen is lysed and nucleic acid is extracted using the Roche MagNA Pure LC instrument using the Roche MagNA Pure LC Total Nucleic Acid Isolation Kit.
A sample of the extracted nucleic acid is added to the MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit reagents that contain a primer pair specific to HSV-1 and HSV-2 and a second primer pair specific to the SPC sequence. The two specific primer pairs are labeled with distinct fluorophore labels. PCR amplification is performed and assay fluorescence is monitored using the Roche LightCycler 1.2 real-time PCR instrument. Incorporation of the quencher-labeled nucleotide causes a decrease in assay fluorescence. Following amplification, the reaction is slowly heated and fluorescence is monitored. The strands of the amplification products will separate at a specific melting temperature (Tm) that is determined by an increase in fluorescence as the strands are separated. The instrument fluorescence output is analyzed and test results are determined using the MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit Analysis Software. A printed results report is generated.
MultiCode®-RTx HSV 1&2 Kit contains HSV-1&2 Primer Mix, HSV-1&2
2
{2}
Reaction Buffer, Nuclease Free Water, DNA Sample Processing Control, HSV-1 Positive Control, and HSV-2 Positive Control, Analysis Software, Certificate of Analysis, and Package Insert.
# J. Substantial Equivalence Information:
1. Predicate device name(s): ELVIS® HSV ID/Typing Test System (Diagnostic Hybrid, Inc.)
2. Predicate Numbers (s): K971662
3. Comparison with predicate: An FDA cleared ELVIS® HSV ID/Typing Test System (Diagnostic Hybrid, Inc.) was used as the predicate device. The performance of the MultiCode®-RTx HSV 1&2 assay was compared with the predicate which is a gold standard/reference method i.e., Cell Culture using an enzyme linked virus inducible system.
Note: To establish the clinical performance, it is important to compare qualitative HSV 1 & 2 DNA detecting devices to a gold standard/reference method such as cell culture using an enzyme linked virus inducible system
Similarities
| Device Characteristic | EraGen Biosciences MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit (New Device) | Diagnostics Hybrid, Inc. ELVIS® HSV ID/Typing Test System (Predicate Device) |
| --- | --- | --- |
| Intended Use | The MultiCode®-RTx HSV 1&2 Kit is a polymerase chain reaction (PCR)-based qualitative in vitro diagnostic test for the detection and typing of herpes simplex virus (HSV 1 & 2) DNA in vaginal lesions. It is indicated for use in the detection and typing of HSV-1 or HSV-2 in vaginal lesion swab specimens from symptomatic female patients as an aid in the diagnosis of genital herpes infection.Warning: The device is not FDA cleared for the use with cerebral spinal fluid (CSF) or any lesions other than vaginal. The assay is not intended to be used for male penile specimens, for prenatal screening, or females under the age of 18 years. | The ELVIS® HSV ID Typing Test System is a qualitative test indicated for use in isolation and identification of HSV from lesions and body fluids suspected of containing viable HSV-1 and/or HSV-2. Both serotypes have been isolated in various parts of the body, particularly when HSV-associated disease is indicated. Performance of this assay has not been established for use with antiviral therapy or prenatal monitoring. |
| Identification and Typing of HSV-1 and HSV-2 | Yes | Yes |
{3}
4
| Device Characteristic | EraGen Biosciences MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit (New Device) | Diagnostics Hybrid, Inc. ELVIS® HSV ID/Typing Test System (Predicate Device) |
| --- | --- | --- |
| Assay Results | Qualitative | Qualitative |
## Differences
| Device Characteristic | EraGen Biosciences MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit (New Device) | Diagnostics Hybrid, Inc. ELVIS® HSV ID/Typing Test System (Predicate Device) |
| --- | --- | --- |
| Assay Type | Real-Time PCR | Cell Culture using an enzyme linked virus inducible system. |
| Analysis Software Provided | Yes | No |
| Printed Results Report Provided | Yes | No |
| Kit Reagent Storage Conditions | -15°C to -30°C | 2°C to 8°C and 22°C to 28°C |
## K. Standard/Guidance Document Referenced (if applicable):
1. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11, 2005.
http://www.fda.gov/cdrh/ode/guidance/337.pdf
2. CLSI EP17-A, "Protocols for Determination of Limits of Detection".
## L. Test Principle:
Following nucleic acid extraction from a patient swab specimen, a fluorophore-labeled PCR primer pair amplifies a segment of the glycoprotein B gene of HSV-1 and HSV-2. The sample processing control (SPC) will also be amplified by a distinct fluorophore-labeled PCR primer pair unless there are sample processing errors, inhibitory substances in the PCR reaction, reagent failure, or instrument malfunction.
The MultiCode®-RTx system is based on an expanded genetic alphabet technology, consisting of 2'-deoxy-5-methyl-isocytidine (iC) and 2'-deoxy-isoguanosine (iG) nucleotide bases also known as isobases. The isobases pair specifically with each other and not with natural nucleotides. In addition isobases are efficiently incorporated during PCR. The isobase pair allows site-specific incorporation of a dabcyl quencher directly adjacent to a fluorophore-labeled primer. During PCR amplification, a quencher-modified iGTP is incorporated by the polymerase opposite an iC and a fluorophore reporter attached to a PCR primer. If target is present and is amplified, assay fluorescence decreases with every cycle as amplification product accumulates. The decrease in assay fluorescence is monitored in real time using the Roche LightCycler 1.2 instrument. Following PCR, the amplification products are thermally denatured and assay fluorescence is monitored. The strands of the
{4}
amplification products are separated and assay fluorescence increases, thus determining the melting temperature $(\mathrm{T_m})$ profile of the amplicon. The sequences between the PCR primer binding sites of the HSV-1 and HSV-2 amplicons have different base compositions that are distinguished by their different melting temperatures using the MultiCode®-RTx HSV 1&2 Analysis Software.
# Interpretation of Control Results:
A. Validation of Assay Run: The MultiCode®-RTx HSV 1&2 Kit Analysis Software automatically determines results for the controls based on the amplification cycle threshold (Ct) value and the melting temperature $(\mathrm{T_m})$ value.
B. Valid Assay Run:
- For a valid run, the following conditions must be met:
| Control ID | Assay Result | F1 | | | F2 | |
| --- | --- | --- | --- | --- | --- | --- |
| | | Ct | HSV-1 Tm | HSV-2 Tm | Ct | SPC Tm |
| HSV-1 Positive Control | Pass | + | + | N/A | N/A | N/A |
| HSV-2 Positive Control | Pass | + | N/A | + | N/A | N/A |
| Negative Control | Pass | - | - | - | + | + |
| | | + | - | - | + | + |
A “+” symbol in the Ct column indicates that the amplification curve crosses the amplification threshold. A “+” symbol in the $\mathrm{T}_{\mathrm{m}}$ column indicates that a melt peak is observed in the appropriate temperature range. A “-” symbol in the Ct column indicates that the amplification curve does not cross the amplification threshold. A “-” symbol in the Tm column indicates that no melt peak was observed in the appropriate temperature range. N/A indicates that the result is not dependent on the value.
- The clinical report displays only the Assay Result column.
- If the HSV-1 Positive Control fails, no data will be available for the HSV-2 Positive Control or the Negative Control because the criteria for Pass/Fail are based on the results of the HSV-1 Positive Control.
C. Invalid Assay Run: If the conditions for a Valid Assay Run are not met, the run must be repeated. Start from the original specimen(s), repeat the target extraction using a new Negative Control, and repeat the reaction setup using new Positive Controls.
# Interpretation of Specimen Results:
The MultiCode®-RTx HSV 1&2 Kit Analysis Software automatically determines results for the specimens based on the amplification cycle threshold (Ct) and the melting temperature $(\mathrm{T_m})$ value.
{5}
Clinical Report Call Logic
| Clinical Report | Input | | | | |
| --- | --- | --- | --- | --- | --- |
| | F1 | | | F2 | |
| | Ct | HSV-1 T_{m} | HSV-2 T_{m} | Ct | SPC T_{m} |
| HSV-2 Positive | + | - | + | - | - |
| HSV-2 Positive | + | - | + | - | + |
| HSV-2 Positive | + | - | + | + | - |
| HSV-2 Positive | + | - | + | + | + |
| HSV-1 Positive | + | + | - | - | - |
| HSV-1 Positive | + | + | - | - | + |
| HSV-1 Positive | + | + | - | + | - |
| HSV-1 Positive | + | + | - | + | + |
| HSV-1&2 Positive | + | + | + | - | - |
| HSV-1&2 Positive | + | + | + | - | + |
| HSV-1&2 Positive | + | + | + | + | - |
| HSV-1&2 Positive | + | + | + | + | + |
| Negative | + | - | - | + | + |
| Negative | - | - | - | + | + |
A “+” symbol in the Ct column indicates that the amplification curve crosses the amplification threshold. A “+” symbol in the $\mathrm{T}_{\mathrm{m}}$ column indicates that a melt peak is observed in the appropriate temperature range. A “-” symbol in the Ct column indicates that the amplification curve does not cross the amplification threshold. A “-” symbol in the $\mathrm{T}_{\mathrm{m}}$ column indicates that no melt peak was observed in the appropriate temperature range. N/A indicates that the result is not dependent on the value.
- A sample is Positive if both the Ct and $\mathrm{T}_{\mathrm{m}}$ are positive in the F1 channel irrespective of the Ct and $\mathrm{T}_{\mathrm{m}}$ in the F2 channel.
- A sample is Negative if the Ct and $\mathrm{T}_{\mathrm{m}}$ are negative in the F1 channel and the Ct and $\mathrm{T}_{\mathrm{m}}$ are positive in the F2 channel.
{6}
Invalid/Fail Report Call Logic
| Invalid/Fail Report | Input | | | | |
| --- | --- | --- | --- | --- | --- |
| | F1 | | | F2 | |
| | Ct | HSV-1 T_{m} | HSV-2 T_{m} | Ct | SPC T_{m} |
| Invalid | - | - | + | - | - |
| Invalid | - | - | + | - | + |
| Invalid | - | - | + | + | - |
| Invalid | - | - | + | + | + |
| Invalid | - | + | - | - | - |
| Invalid | - | + | - | - | + |
| Invalid | - | + | - | + | - |
| Invalid | - | + | - | + | + |
| Invalid | - | + | + | - | - |
| Invalid | - | + | + | - | + |
| Invalid | - | + | + | + | - |
| Invalid | - | + | + | + | + |
| Fail | + | - | - | - | - |
| Fail | + | - | - | - | + |
| Fail | + | - | - | + | - |
| Fail | - | - | - | - | - |
| Fail | - | - | - | - | + |
| Fail | - | - | - | + | - |
A “+” symbol in the Ct column indicates that the amplification curve crosses the amplification threshold. A “+” symbol in the T<sub>m</sub> column indicates that a melt peak is observed in the appropriate temperature range. A “-” symbol in the Ct column indicates that the amplification curve does not cross the amplification threshold. A “-” symbol in the T<sub>m</sub> column indicates that no melt peak was observed in the appropriate temperature range. N/A indicates that the result is not dependent on the value.
- A sample is INVALID if the T<sub>m</sub> is positive and the Ct is negative in the F1 channel. This condition warrants repeat testing using the extracted nucleic acid. If the result is INVALID again, then the specimen needs to be re-extracted and tested.
- A sample is FAIL if the T<sub>m</sub> is negative in the F1 channel, and the Ct or T<sub>m</sub> is negative in the F2 channel. This condition warrants repeat testing following extraction of another aliquot of the specimen. If the result is again a FAIL, then the specimen needs to be re-collected and tested.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{7}
The Precision/Reproducibility of the MultiCode®-RTx HSV 1&2 Kit was evaluated at 3 U.S. clinical laboratories. A panel was prepared containing 6 simulated HSV-1 and HSV-2 samples that included High Negative, Low Positive (near the assay limit of detection) and Moderate Positive samples. The panel along with the external HSV-1 and HSV-2 positive and negative controls (Cytomegalovirus) was assayed in triplicate. The external Controls were included to monitor the study site DNA extraction and amplification processes during each assay run. Kit positive and negative controls were included with each assay run. Panels and controls were tested at each site by 2 operators, 1 time each per day for 5 days (N = 900). Results of the Precision/Reproducibility studies for the MultiCode®-RTx HSV 1&2 assay are presented from three sites in the tables below.
8
{8}
Precision/Reproducibility Study Summary for the MultiCode®-RTx HSV 1&2 PCR Assay
| Targets | Site #1 | | | | Site #2 | | | | Site #3 | | | | Total agreement with expected results (%) | 95% Confidence Interval |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Agreement with expected results-# correct / #tested | Avg Tm1 | %CV-Tm | Avg % Deflection2 | Agreement with expected results-# correct / #tested | Avg Tm1 | %CV-Tm | Avg % Deflection2 | Agreement with expected results-# correct / #tested | Avg Tm1 | %CV-Tm | Avg % Deflection2 | | |
| HSV-1 Positive Control | 10/10 | 84.3 | 0.19 | 85.0 | 10/10 | 84.7 | 0.36 | 78.3 | 10/10 | 84.7 | 0.22 | 72.8 | 30/30 (100%) | 88.4%-100% |
| HSV-2 Positive Control | 10/10 | 87.2 | 0.21 | 70.9 | 10/10 | 87.7 | 0.32 | 70.3 | 10/10 | 87.7 | 0.17 | 68.8 | 30/30 (100%) | 88.4%-100% |
| HSV-1/HSV-2 Negative Control1 | 10/10 | 77.9 | 0.18 | 1.5 | 10/10 | 78.0 | 0.28 | 1.4 | 10/10 | 78.0 | 0.21 | 1.8 | 30/30 (100%) | 88.4%-100% |
| PN 1750 HSV-1 Positive External Control | 30/30 | 84.5 | 0.37 | 79.2 | 29/30 | 84.6 | 0.27 | 68.3 | 30/30 | 84.7 | 0.24 | 67.3 | 89/90 (98.9%) | 93.9%-100% |
| PN 1751 HSV-2 Positive External Control | 30/30 | 87.4 | 0.40 | 73.0 | 29/30 | 87.6 | 0.28 | 68.2 | 30/30 | 87.7 | 0.23 | 69.3 | 89/90 (98.9%) | 93.9%-100% |
| PN 1754 HSV-1/HSV-2 Negative External Control1 | 30/30 | 77.6 | 0.40 | 1.6 | 30/30 | 77.7 | 0.43 | 1.3 | 30/30 | 77.8 | 0.29 | 1.1 | 90/90 (100%) | 95.9%-100% |
| HSV-1 High Negative1 | 30/30 | 77.6 | 0.39 | 2.2 | 30/30 | 77.7 | 0.44 | 2.3 | 30/30 | 77.8 | 0.32 | 1.4 | 90/90 (100%) | 95.9%-100% |
| HSV-1 -Low Positive | 30/30 | 84.4 | 0.37 | 75.7 | 29/30 | 84.5 | 0.26 | 64.3 | 30/30 | 84.6 | 0.21 | 65.1 | 89/90 (98.9%) | 93.9%-100% |
| HSV-1 High Positive | 30/30 | 84.5 | 0.36 | 94.4 | 30/30 | 84.6 | 0.20 | 92.8 | 30/30 | 84.8 | 0.21 | 94.3 | 90/90 (100%) | 95.9%-100% |
| HSV-2 High Negative1 | 30/30 | 77.6 | 0.41 | 1.2 | 30/30 | 77.7 | 0.28 | 1.5 | 30/30 | 77.7 | 0.27 | 1.3 | 90/90 (100%) | 95.9%-100% |
| HSV-2 Low Positive | 30/30 | 87.3 | 0.41 | 71.9 | 30/30 | 87.5 | 0.25 | 63.5 | 30/30 | 87.6 | 0.19 | 67.5 | 90/90 (100%) | 95.9%-100% |
| HSV-2 High Positive | 30/30 | 87.4 | 0.39 | 92.9 | 30/30 | 87.5 | 0.19 | 89.4 | 30/30 | 87.7 | 0.20 | 91.9 | 90/90 (100%) | 95.9%-100% |
1: For the HSV-1/HSV-2 Negative Control, the PN1754 HSV-1/HSV-2 Negative Extraction Control as well as the HSV-1 High Negative and HSV-2 High Negative, the $\mathrm{T}_{\mathrm{m}}$ value of Sample Processing Control (SPC) was used, since these targets did not generate any detectable HSV-1 or HSV-2 signal.
2: % deflection is the individual sample deflection obtained during the melt curve analysis expressed as a percent of the maximum deflection of the melt curve in that assay run. Average % deflection is the average deflection for that panel member across all the 10 runs for the site.
{9}
b. Linearity/assay reportable range: N/A
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Assay Controls: Quality control procedures are intended to monitor reagent and assay performance. The kit includes the following controls:
| Control Type | Use |
| --- | --- |
| HSV-1 Positive Control | Verify proper reagent performance and Roche LightCycler instrument setup |
| HSV-2 Positive Control | Verify proper reagent performance and Roche LightCycler instrument setup |
| Negative Control | Detect contamination and verify proper reagent performance and Roche LightCycler instrument setup |
| DNA Sample Processing Control | Verify proper specimen lysis, verify nucleic acid extraction and to verify proper reagent performance and Roche LightCycler instrument setup |
Stability Studies: were performed in-house to determine the shelf life for the multiCode®-RTx Herpes Simplex Virus 1 & 2 Kit.
- Accelerated Stability: The product is stable for a minimum of 15 months when stored at an accelerated temperature of 4°C.
- Real-Time Stability: The product is stable for a minimum of 6 months when stored the recommended storage temperature of -15°C to -30°C.
- Real time stability studies will continue to be performed to support the shelf life of this product.
d. Detection limits:
A Limit of Detection (LOD) and Limit of Blank (LOB) study was performed in-house in accordance with CLSI EP-17-A (Protocols for Determination of Limits of Detection and Limits of Quantitation: Approved Guideline) to determine the analytical LOD and LOB performance obtained with the MultiCode®-RTx Herpes Simplex Virus 1 & 2 Kit.
A Limit of Detection (LoD) and Limit of Blank (LoB) study was performed in house to determine the analytical LoD and LoB performance using quantified (TCID₅₀/mL) cultures of HSV-1 and HSV-2 serially diluted in Copan Universal Transport Media. Each viral strain was extracted using the MagNA Pure LC and tested in replicates of 60 per concentration of virus. The LOD was determined to be the lowest concentration of target that was
10
{10}
detected in at least $95\%$ of replicates. The LOB was determined to be the highest concentration of target that was detected in $\leq 5\%$ of replicates.
The LOD for HSV-1 was determined to be $2.00 \times 10^{3} \mathrm{TCID}_{50} / \mathrm{mL}$ . At this concentration $100\%$ of samples were detected with a $95\%$ Confidence Interval of $93.94 - 100\%$ . The LOD for HSV-2 was determined to be $6.40 \times 10^{1} \mathrm{TCID}_{50} / \mathrm{mL}$ . At this concentration $98.3\%$ of samples were detected with a $95\%$ Confidence Interval of $91.06 - 99.96\%$ .
The LOB for HSV-1 was determined to be $2.50 \times 10^{2} \mathrm{TCID}_{50} / \mathrm{mL}$ . At this concentration $3.33\%$ of samples were detected with a $95\%$ Confidence Interval of $0.41 - 11.53\%$ . The LOD for HSV-2 was determined to be $4.00 \mathrm{TCID}_{50} / \mathrm{mL}$ . At this concentration $0.00\%$ of samples were detected with a $95\%$ Confidence Interval of $0.00 - 5.96\%$ .
LoD for HSV 1
| TCID50/mL | Pos/Total Calls | Positivity | 95% C.I. |
| --- | --- | --- | --- |
| 4.00 x 103 | 59/59* | 100.00% | 91.06 ± 99.96% |
| 5.00 x 103 | 59/59* | 100.00% | 91.06 ± 99.96% |
| 6.00 x 103 | 17/60 | 28.33% | 17.45 ± 41.44% |
| 7.00 x 102 | 6/59* | 10.17% | 4.82 ± 22.57% |
| 8.50 x 102 | 2/60 | 3.33% | 0.41 ± 11.53% |
| 1.25 x 102 | 0/60 | 0.00% | 0.00 ± 5.96% |
*One "Invalid" call due to baselines of amp curves by newest V2.0 RC1 software
LoD for HSV 2
| TCID50/mL | Pos/Total Calls | Positivity | 95% C.I. |
| --- | --- | --- | --- |
| 1.28 x 102 | 60/60 | 100.00% | 94.04 ± 100.00% |
| 6.40 x 101 | 59/60 | 98.33% | 91.06 ± 99.96% |
| 3.20 x 101 | 56/60 | 93.33% | 83.80 ± 98.15% |
| 1.60 x 101 | 17/57* | 29.82% | 18.43 ± 43.40% |
| 8.00 | 4/60 | 6.67% | 1.85 ± 16.20% |
| 4.00 | 0/60 | 0.00% | 0.00 ± 5.96% |
*Three "Invalid" calls due to baselines of amp curves by newest V2.0 RC1 software
Note: The LOD value for HSV-1 was shown higher than HSV-2. This could be due to mismatch in the forward primer because a single set of forward and reverse HSV primers was used to amplify the target region from HSV-1 and HSV-2. The HSV Forward primer is a perfect match with the HSV-2 glycoprotein B sequence but contains a one base mismatch with the HSV-1 glycoprotein B sequence. The mismatch is in the center of the primer with 10 bases $3^{\prime}$ to the mismatch. The predicted $\mathrm{T}_{\mathrm{m}}$ for the forward primer to HSV-1 is $57.4^{\circ}\mathrm{C}$ ; the predicted $\mathrm{T}_{\mathrm{m}}$ for the forward primer to HSV-2 is
{11}
$59.8^{\circ}\mathrm{C}$ . It is likely that in the initial rounds of PCR, HSV-2 is amplified more efficiently than HSV-1 which could result in a small increase in analytical sensitivity for HSV-2.
# e. Analytical specificity:
Analytical Specificity/Cross Reactivity of the MultiCode®-RTx HSV 1&2 Kit was evaluated at EraGen Biosciences. A panel was prepared containing 22 different organisms representing near-neighbors to the HSV-1 and HSV-2 virus and organisms reasonably expected to be present in vaginal swab specimens.
The cross-reactivity panel was tested in a background of Copan Universal Transport Media at the concentration indicated in the table below. No HSV positive results were observed for any of the organisms tested and the DNA SPC was detected in all cases. The organism panel was also spiked into HSV-1 and HSV-2 near the device's Limit of Detection (LoD) and tested. No interference was observed from any of the tested organisms, and all the results were positive for HSV-1 or HSV-2 as expected.
Cross Reactivity Panel
| Organism | Concentration |
| --- | --- |
| Candida albicans | 2.8 x107CFU/mL |
| Chlamydia trachomatis | 2.5 x108EBs/mL |
| Escherichia coli | 2.0 x107CFU/mL |
| Mycoplasma hominis | 4.5 x105CFU/mL |
| Neisseria gonorrhoeae | 1.8 x106CFU/mL |
| Staphylococcus aureus | 2.4 x107CFU/mL |
| Staphylococcus saprophyticus | 1.2 x107CFU/mL |
| Streptococcus pyrogenes | 1.4 x107CFU/mL |
| Trichomonas vaginalis | 1.1 x107CFU/mL |
| Bacteroides fragilis | 3.3 x107CFU/mL |
| Gardnerella vaginalis | 1.4 x106CFU/mL |
| Mobiluncus mulieris | 1.0 x107CFU/mL |
| Toxoplasma gondii | 6.6 x105Tachyzoites/mL |
| Treponema pallidum | 1.0 x107CFU/mL |
| Cytomegalovirus (AD169 strain) | 4.2 x103TCID50/mL |
| Enterovirus (Type 71) | 1.4 x104TCID50/mL |
| Epstein-Barr virus (B95-8 strain) | 9.3 x107copies/mL |
| Varicella Zoster virus | 2.4 x107copies/mL |
| Human Herpes 6 virus (Z29 strain) | 1.9 x106TCID50/mL |
| Human Herpes 7 virus (SB strain) | 3.4 x106TCID50/mL |
| Human Papilloma virus | 5-8 x105copies/mL |
| Rubella virus | 1.7 x104TCID50/mL |
{12}
f. Interfering Substances:
An Interfering Substance study was performed at EraGen Biosciences to evaluate the effects of potential interfering substances on the MultiCode®-RTx HSV 1&2 Kit. A panel was prepared containing 6 substances and was tested at two different concentrations that could reasonably be expected to be present in vaginal swab specimens. The substance panel was tested near the device's Limit of Detection (LoD) for HSV-1 and HSV-2. No interference was observed in the presence of exogenous and endogenous substances in an extractable sample.
Interfering Substances Panel
| Substance | Concentration |
| --- | --- |
| Whole Blood (with EDTA) | 10% |
| Whole Blood (with EDTA) | 1% |
| Female Urine | 10% |
| Female Urine | 1% |
| Protein (Albumin) | 10 mg/ml |
| Protein (Albumin) | 1 mg/ml |
| Protein (Casein) | 10 mg/ml |
| Protein (Casein) | 1 mg/ml |
| K-Y Brand Jelly | 5% |
| K-Y Brand Jelly | 0.5% |
| Acyclovir (Acycloguanosine) | 2.5 mg/ml |
| Acyclovir (Acycloguanosine) | 0.25 mg/ml |
g. Carry-over/Cross-Contamination
Carry-over/Contamination studies were done only with HSV-1 target, since both HSV-1 and HSV-2 share a single set of primers. The MultiCode®-RTx HSV 1&2 Kit was evaluated internally using simulated samples at the LoB and High Positive HSV-1 (1700X LoD) samples. Ten sets of samples in a sequence pattern of High Positive-LoB-LoB were aliquoted into a 32-well sample plate and extracted using the Roche MagNA Pure LC instrument. The high positive sample was positive for HSV-1 in all cases and all LoB samples were negative for HSV-1 indicating that there was no carry-over/cross-contamination of HSV-1 in the LoB samples.
2. Comparison studies:
a. Method comparison with predicate device:
The clinical performance evaluation was done against a gold standard/reference method i.e., Cell Culture using an enzyme linked virus inducible system.
{13}
For additional details please see section 3 Clinical studies subsection c.
b. Matrix Comparison
N/A
3. Clinical studies:
a. Clinical Sensitivity: N/A
b. Clinical specificity: N/A
c. Other clinical supportive data (when a. and b. are not applicable):
An FDA cleared ELVIS® HSV ID/Typing Test System (Diagnostic Hybrid, Inc.) was used as the Predicate device. The performance of the MultiCode®-RTx HSV 1&2 assay was compared with the predicate which is a gold standard/reference method i.e., Cell Culture using an enzyme linked virus inducible system.
## Clinical Performance
Performance characteristics of the MultiCode®-RTx HSV 1&2 assay were established during a prospective study performed at 3 U.S. clinical laboratories from 2008-2009. A total of 1041 vaginal swab specimens were tested by the MultiCode®-RTx HSV 1&2 Kit and by the reference method. The swab specimens were collected from lesions in Copan Universal Transport Medium from the patient population ranged from 18 to 85 years. One swab specimen was collected from each patient in Copan Universal Transport Medium and was tested for HSV-1 and HSV-2 using both the Diagnostics Hybrids, In. ELVIS culture and typing test and the EraGen MultiCode-RTx test.
A total of 1041 swab specimens were tested using the MultiCode®-RTx HSV 1&2 assay at three sites. Results from Prospective Study are shown in the tables below.
14
{14}
15
| Herpes Simplex Virus Type 1 Comparison Results | | | | |
| --- | --- | --- | --- | --- |
| | Reference Method | | | |
| | | Positive | Negative | Total |
| MultiCode®-RTx HSV 1&2 Kit | Positive | 97 | 16^{a} | 113 |
| | Negative | 8^{b} | 920 | 928 |
| | Total | 105 | 936 | 1041 |
| | | Value | 95% Confidence Interval | |
| Sensitivity | | 92.4% | 85.7 – 96.1% | |
| Specificity | | 98.3% | 97.2 – 98.9% | |
a Sequence analysis detected HSV-1 in 12 of the 16 samples identified as HSV-1 by MultiCode®-RTx. Sequence analysis did not detect HSV-1 in 4 of the samples.
b Sequence analysis detected HSV-1 in 1 of the 8 samples identified as HSV-1 negative by MultiCode®-RTx. Sequence analysis did not detect HSV-1 in 7 of the samples. Of these 7 samples: 4 of the samples were identified as HSV-2 by both MultiCode®-RTx and sequencing 2 of the samples were negative by MultiCode®-RTx and not detected by sequencing 1 sample was negative by MultiCode®-RTx and HSV-2 positive by sequencing.
| Herpes Simplex Virus Type 2 Comparison Results | | | | |
| --- | --- | --- | --- | --- |
| | Reference Method | | | |
| | | Positive | Negative | Total |
| MultiCode®-RTx HSV 1 and 2 Kit | Positive | 198 | 53^{a} | 251 |
| | Negative | 10^{b} | 780 | 790 |
| | Total | 208 | 833 | 1041 |
| | | Value | 95% Confidence Interval | |
| Sensitivity | | 95.2% | 91.4 – 97.4% | |
| Specificity | | 93.6% | 91.8 – 95.1% | |
a Sequence analysis detected HSV-2 in 43 of the 53 samples identified as HSV-2 by MultiCode®-RTx. Sequence analysis did not detect HSV-2 in 10 of the discordant samples.
b Sequence analysis detected HSV-2 in 2 of the 10 samples identified as HSV-2 negative by MultiCode®-RTx. Sequence analysis did not detect HSV-2 in 8 of the discordant samples. These 8 samples were identified as HSV-1 by both MultiCode®-RTx and sequencing.
A total of 69 discordant specimens were reference method negative and MultiCode®-RTx HSV 1&2 Kit positive for HSV-1 or HSV-2. These samples were further evaluated by DNA sequencing. In addition 22 HSV-1 and 24 HSV-2 samples that were tested positive by both the EraGen MultiCode-RTx and the Reference Method were bidirectionally sequenced by SeqWright.
{15}
16
# Competitive Inhibition
Competitive Inhibition of the MultiCode®-RTx HSV 1&2 Kit was evaluated internally using simulated samples with varying concentrations of HSV-1 virus (1X LoD to 1000X LoD) and HSV-2 virus (1X LoD to 1000X LoD). Competitive inhibition was observed. The highest concentration of co-infecting target that can be present while maintaining 95% detection of the 3X LoD target for HSV-1 was 1X LoD for HSV-2. The highest concentration of co-infecting target that can be present while maintaining 95% detection of the 3X LoD target for HSV-2 was 1X LoD for HSV-1.
Competitive Inhibition was also evaluated internally using simulated samples with equal concentrations of HSV-1 virus and HSV-2 virus (5X LoD to 500X LoD). Competitive inhibition was not observed at any of the concentrations tested when the concentrations of HSV-1 and HSV-2 were equal.
4. Clinical cut-off: N/A
5. Expected values/Reference range:
The prevalence of HSV-1 and HSV-2 during the 2008 - 2009 multi-site clinical study (n=1041 specimens collected December 2008 - June 2009) was estimated using the MultiCode®-RTx test. Geographical distribution of study population included in the study consisted of 74 Midwest (12 States), 813 South (16 States), 153 West (13 States), and 1 Northeast (9 States). The combined prevalence was used to calculate the positive predictive values (PPV) and negative predictive values (NPV) of the MultiCode®-RTx test. The calculations are based on the sensitivity and specificity values obtained from the clinical studies: sensitivity of 92.4% and specificity of 98.3% for HSV-1; sensitivity of 95.2% and specificity of 93.6% for HSV-2. The MultiCode®-RTx HSV 1 & 2 assay results are summarized in the following tables:
MultiCode®-RTx HSV 1& 2 Assay Distribution of Prospective Population by Age Group; All testing Sites
| Age Range | MultiCode®-RTx HSV-1-Positive | MultiCode®-RTx HSV-2-Positive | Total number of Specimens |
| --- | --- | --- | --- |
| 18 to 25 years | 56 | 77 | 316 |
| 26 to 30 years | 12 | 41 | 159 |
| 31 to 35 years | 7 | 20 | 109 |
| 36 to 40 years | 14 | 29 | 128 |
| 41 to 45 years | 5 | 22 | 85 |
| 46 to 50 years | 10 | 20 | 88 |
{16}
17
| 51 to 55 years | 2 | 15 | 51 |
| --- | --- | --- | --- |
| 56 to 60 years | 3 | 12 | 36 |
| 61 to 65 years | 3 | 5 | 22 |
| 66 to 70 years | 0 | 5 | 22 |
| 71 to 75 years | 1 | 2 | 11 |
| 76 to 80 years | 0 | 1 | 11 |
| 81 to 85 years | 0 | 2 | 3 |
| 86 to 90 years | 0 | 0 | 0 |
| Total | 113 (10.85%) | 251 (24.11%) | 1041 |
Prevalence vs Hypothetical Predictive Values
| Prevalence | HSV-1 | | HSV-2 | |
| --- | --- | --- | --- | --- |
| | PPV | NPV | PPV | NPV |
| 40% | 97.5% | 95.1% | 90.8% | 96.7% |
| 30% | 96.0% | 96.8% | 86.5% | 97.8% |
| 20% | 93.3% | 98.1% | 78.9% | 98.7% |
| 10% | 85.8% | 99.1% | 62.3% | 99.4% |
| 5% | 74.0% | 99.6% | 43.9% | 99.7% |
## N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.