K100320 · Bio-Medical Research, Ltd. · NGX · May 5, 2010 · Physical Medicine
Device Facts
Record ID
K100320
Device Name
SLENDERTONE SYSTEM ULTRA MODEL E70/X70
Applicant
Bio-Medical Research, Ltd.
Product Code
NGX · Physical Medicine
Decision Date
May 5, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The proposed device, System Ultra, Type 390, Model E70/X70 is a powered muscle stimulator intended for over-the-counter use; there is no change to the indications for use from the predicate product Slendertone System-Abs, Type 390, E10/X10. Proposed Device: "System Ultra, Type 390, Model E70/X70" is intended for the improvement of abdominal muscle tone, for the strengthening of the abdominal muscles and for the development of a firmer abdomen.
Device Story
System Ultra (Type 390, Model E70/X70) is a two-channel, battery-powered muscle stimulator for OTC use. System comprises a three-electrode abdominal belt garment and a hand-held rechargeable control unit. Device delivers electrical stimulation to abdominal muscles to improve tone and strength. User operates device via control unit, selecting from ten available programs. Power provided by internal 3.6V NiMH rechargeable battery. Internal connections over-molded to prevent moisture ingress. User cannot access wiring or alter current path. Device intended for home use to assist in muscle conditioning.
Clinical Evidence
No clinical data submitted. Bench testing only, demonstrating compliance with IEC 60601-1, IEC 60601-2-10, and IEC 60601-1-2 safety and electromagnetic compatibility standards.
Technological Characteristics
Two-channel powered muscle stimulator; 3.6V NiMH rechargeable battery; three-electrode abdominal belt garment; over-molded internal connections; ten stimulation programs. Complies with IEC 60601-1, IEC 60601-2-10, and IEC 60601-1-2. Battery charger complies with IEC 60950 and UL 1950.
Indications for Use
Indicated for improvement of abdominal muscle tone, strengthening of abdominal muscles, and development of a firmer abdomen in general adult population (OTC use).
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
Predicate Devices
Slendertone System-Abs, Type 390 E10/X10 (K070142)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letters "BMR" in bold, sans-serif font. To the left of the letters is a curved, crescent-shaped graphic that partially encloses the "B". The logo is in black and white, with the letters and graphic appearing in solid black against a white background.
# search Ltd.
0191 774300
# (( ( ( 0032 )( )( ( ( ( 0032 )( )( )( )( ( )( ( )( ( )( ( ( )( ( )( ( ( ( ( )( )( ( ( ( ( )( )( ( ( ( ( )( )( ( ( ( ( ( )( )( ( ( ( ( ( ( )( )( ( ( ( ( )( ( ( ( ( )( )( ( ( (
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Contact Details
# MAY - 5 2010
| Name: | Deirdre Barrow |
|------------|-------------------------------------------------------------------------------|
| Address: | Bio-Medical Research Ltd.,<br>Parkmore Business Park, West<br>Galway, Ireland |
| Telephone: | +353 91 774300 |
| Fax: | +353 91 774302 |
| E-Mail: | dbarrow@bmr.ie |
| Prepared: | 25th January 2010 |
| Alternative Contact Person | Michael O'Sullivan |
|----------------------------|-------------------------|
| Title: | Quality Systems Manager |
| Telephone | +353 91 774395 |
| Fax: | +353 91 774301 |
| E-Mail: | mosullivan@bmr.ie |
## 2. Device Name
| Trade Name of Device: | System Ultra, Type 390, Model E70/X70 |
|-----------------------|------------------------------------------------------|
| Common Name: | Muscle Stimulator |
| Classification Name: | Stimulator, muscle, powered, for muscle conditioning |
| Product Code: | NGX |
# 3. Identification of Equivalent Legally Marketed Device
| Device Trade Name: | Slendertone System-Abs, Type 390 E10/X10 device |
|--------------------|-------------------------------------------------|
| Manufacturer: | Bio-Medical Research Ltd |
| 510(k) Nos: | K070142 |
{1}------------------------------------------------
#### 4. Description of Device
------
-- - ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ System Ultra, Type 390, Model E70/X70 is a two-channel, battery powered, muscle stimulation system. It is supplied with a three-electrode abdominal belt garment, a hand-held rechargeable control unit, a pack of 3 adhesive backed gel based electrodes, instructions for use and a carry pouch.
The control unit is interchangeable between all the cleared System models from the Slendertone® range of garments reference Table 10.1
| Device | Type, Model No. | 510k No. |
|-----------------------------|-----------------|----------|
| System-Abs (Female Version) | Type 390, E10 | K070142 |
| System-Abs (Male Version) | Type 390, X10 | K070142 |
| System-Shorts | Type 390, E20 | K070142 |
| System-Mini | Type 390, E30 | K072294 |
| System-Arms | Type 390, E60 | K072553 |
| System-Arms | Type 390, X60 | K083164 |
Table 10.1 System from Slendertone Range of devices
There are ten programs available to users of the System Ultra, Type 390, E70/X70. Power is derived from a 3.6V NiMH rechargeable battery pack pre-installed in the unit.
All the internal connections are over-molded to prevent moisture ingress. The user has no access to the wiring or connectors within the garment and is unable to alter the current path.
For purposes of hygiene, the garment may be cleaned and instructions for belt care are included in the user manual.
#### 5. Statement of Intended Use/Indications for Use
The proposed device, System Ultra, Type 390, E70/X70 is a powered muscle stimulator intended for over-the-counter use; there is no change to the indications for use from the predicate product Slendertone System-Abs, Type 390, E10/X10.
{2}------------------------------------------------
Proposed Device: "System Ultra, Type 390, Model E70/X70" is intended for the improvement of abdominal muscle tone, for the strengthening of the abdominal muscles and for the development of a firmer abdomen.
### 6. Technological Characteristics
The System Ultra, Type 390, E70/X70 incorporates the control unit and garment technology of the existing predicate Slendertone System-Abs, Type 390, E10/X10 (K070142).
## 7. Clinical and Non-Clinical Tests
No new clinical studies have been submitted as part of this premarket notification.
System Ultra, Type 390, E70/X70 device complies with the following international safety standards:
- . IEC 60601-1:1988 & A1:1991, A2:1995 Medical electrical equipment - Part 1: General requirements for safety
- . IEC 60601-2-10:1987 & Al 2001 Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators
- . IEC 60601-1-2:2001 (EN 60601-1-2:2001) Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility -Requirements and tests
The battery charger complies to safety standards IEC 60950 and UL 1950
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Bio-Medical Research Ltd. % Ms. Deidre Barrow Quality/Regulatory Engineer Parkmore Business Park, West Galway, Ireland
Re: K100320
Trade/Device Name: Slendertone System Ultra, Type 390, Model E70/X70 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: II Product Code: NGX Dated: January 26, 2010 Received: February 4, 2010
MAY - 6 2010
Dear Ms. Barrow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Ms. Deidre Barrow
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Milliken
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use Statement
510(k) Number (if known): ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Device Name:
s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Slendertone System Ultra, Type 390, Model E70/X70
Indications for Use:
The Slendertone System Ultra is indicated for the improvement of abdominal muscle tone, for the strengthening of the abdominal muscles and for the development of a firmer abdomen.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
(Division Sign-Off) (Division Sign-on), Orthopedic, and Restorative Devices
K100320
510(k) Number *_*_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.