K100272 · Philips Medical Systems · DXJ · Apr 14, 2010 · Cardiovascular
Device Facts
Record ID
K100272
Device Name
INTELLIVUE CLINICAL INFORMATION PORTFOLIO
Applicant
Philips Medical Systems
Product Code
DXJ · Cardiovascular
Decision Date
Apr 14, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2450
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Intended for use in the data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. This device is indicated for use by health care providers whenever there is a need for generation of a patient record and computation of drug dosage.
Device Story
IntelliVue Clinical Information Portfolio (ICIP) is a software-only clinical information management system; collects, stores, and manages data from bedside devices and ancillary systems via direct or network connections. Used by healthcare providers in clinical environments (e.g., critical care) to generate patient records and compute drug dosages. Enhancements in Release E.0 include updated user interface, improved usability, and hospital IT requirement compliance. Output provided to clinicians for patient management and decision-making support. Benefits include streamlined data management, improved clinical workflow, and accurate drug dosage calculations.
Clinical Evidence
Bench testing only. Verification and validation activities included system-level tests, performance tests, and safety testing based on hazard analysis. Results demonstrated that the device meets all reliability requirements and performance claims consistent with the predicate device.
Technological Characteristics
Software-only clinical information management system. Connectivity via direct or network interfaces to bedside devices and ancillary systems. Operates as a data management and computation platform. No specific hardware materials or energy sources defined as it is a software-only product.
Indications for Use
Indicated for use by health care providers for patient record generation and drug dosage computation in clinical settings where data collection, storage, and management from bedside devices and ancillary systems are required.
Regulatory Classification
Identification
A medical cathode-ray tube display is a device designed primarily to display selected biological signals. This device often incorporates special display features unique to a specific biological signal.
Predicate Devices
HP CareVue 9000 Clinical Information System (K992636)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the logo for the Department of Health & Human Services - USA, which features a stylized emblem with three figures. To the right of that is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 1, 2023
Philips Medical Systems Teresa Schmidt Quality Assurance & Regulatory Engineer 3000 Minuteman Rd. Andover, Massachusetts 01810-1099
Re: K100272
Trade/Device Name: Intellivue Clinical Information Portfolio Regulation Number: 21 CFR 870.2450 Regulation Name: Medical cathode-ray tube display Regulatory Class: Class II Product Code: DXJ
Dear Teresa Schmidt:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 14, 2010. Specifically, FDA is updating this SE Letter to remove the secondary product code NSX as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter, please contact Aneesh Deoras, OHT2: Office of Cardiovascular Devices, 240-402-4363, Aneesh.Deoras@fda.hhs.gov.
Sincerely,
Aneesh S. Deoras -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a staff entwined by two snakes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
APR 1 4 2010
Philips Medical Systems c/o Ms. Teresa Schmidt Quality Assurance & Regulatory Engineer 3000 Minuteman Rd Andover, MA 01810-1099
Re: K100272
Trade/Device Name: Intellivue Clinical Information Portfolio Regulation Number: 21 CFR 870.2450 Regulation Name: Medical Cathode-Ray Tube Display Regulatory Class: Class II (two) Product Code: DXJ, NSX Dated: March 12, 2010 Received: March 15, 2010
Dear Ms. Schmidt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Ms. Teresa Schmidt
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Pr Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation · Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K100272 pg 1 of 1
Indications for Use
K100272 510 (k) Number (if known):
Device Name: IntelliVue Clinical Information Portfolio Release E.0
Intended for use in the data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. This device is indicated for use by health care providers whenever there is a need for generation of a patient record and computation of drug dosage.
AND/OR over-the-counter Use: Prescription Use: YES NO (21 CFR 807 Subpart C) (Part 21 CFFR 801 Subpart D)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
W.M.H.
Page of
(Division Sign-Off) Division of Cardiovascular Devices
510(k) Number K1600272
{4}------------------------------------------------
K100272 pg 1 of 3
## APR 1 4 2010
## 510K Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807-92(c).
- 1. The submitter of this pre-market notification is:
Teresa Schmidt Philips Medical Systems 3000 Minuteman Road Andover. MA 01810 United States
- Tel: 978-659-2220 Fax: 978-685-5624 Email: teresa.schmidt@philips.com
This summary was prepared on December 18, 2009.
- 2. The name of the subject device is the Philips IntelliVue Clinical Information Portfolio Release E.O.
- 3. The trade name of the device the Philips IntelliVue Clinical Information Portfolio.
- 4. The common usual name is Clinical Information Management System.
- 5. Device Description: As with the predicate device, the proposed IntelliVue Clinical Information Portfolio (ICIP) is a software only product used for charting and data management. The proposed updated device targets functionality needed for critical care, enhances the user experience, and provide necessary updates to meet hospital IT requirements, as is found in the Philips IntelliVue Clinical Information Portfolio device (K992636). Enhancements for the Critical Care market derived from customer feedback will be added to device in order to meet customer requirements for imoroved usability. The purpose of this submission is to formally document the renaming of the product to the Philips IntelliVue Clinical Information Portfolio (ICIP) as well as to update the 510(k) filing to include updated technologies needed to meet current hospital IT requirements and improve usability for clinicians
| Device Panel | Classification | ProCode | Description |
|------------------|----------------|---------|-----------------------------------------------------|
| General Hospital | Not classified | LNX | Software, transmission and<br>storage, patient data |
| Cardiovascular | 870.2450, II | DXJ | Display, medical cathode ray<br>tube |
The Classification names are as follows:
{5}------------------------------------------------
K 100272 pg 2 of 3
- 7. The modified device is substantially equivalent to the previously cleared HP Care Vue 9000 Clinical Information System: K992636 HP CAREVUE 9000 Clinical Information System
- 8. The major modifications are as follows:
The purpose of this submission is to formally document the renaming of the product to the Philips IntelliVue Clinical Information Portfolio (ICIP) as well as to update the 510(k) filing to include updated technologies needed to meet current hospital IT requirements and improve usability for clinicians.
- 9. Comparison with predicate:
Predicate Device : The predicate device is the HP CareVue 9000 Clinical Information System (K992636). Similarities: The predicate device and the subject device are similar with respect to the intended use and indications for use and fundamental scientific technology: Intended for use in the data collection, storage, and management with independent bedside devices, and ancillary systems that are connected either directly or through networks. This device is indicated for use by health care providers whenever there is a need for generation of a patient record and computation of drug dosage. Additionally, ICIP Release E.0 contains all the main functionality of the predicate. Differences: There were enhancements to several of the core features such as report enhancements and simple configuration changes. There is a new user interface designed around more modern technologies, this makes the product more user friendly for clinicians, support and scalability. In conclusion, there are no changes to the intended use or in the Conclusion: fundamental scientific technology employed between the predicate device and the subject devices. Thus, in our opinion, given the substantial equivalence to the predicate devices, no new issues of safety or effectiveness are raised by this premarket notification.
- 10. The subject device has the same fundamental technological characteristics as the legally marketed predicate device.
- 11. Non-Clinical Testing: Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the subject devices with respect to the predicates. Testing involved system level tests, performance tests, and
{6}------------------------------------------------
K100272 pg 3 of 3
safety testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device, the specifications of the subject device and test results showed substantial equivalence.
- 12. Conclusions from Non-Clinical Testing: The results demonstrate that the Philips IntelliVue Clinical Information Portfolio Release E.0 meets all reliability requirements and performance claims and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.