← Product Code [CBF](/productcode/CBF) · K100268

# SECHRIST MONOPLACE HYPERBARIC OXYGEN CHAMBER (K100268)

_Sechrist Industries, Inc. · CBF · Apr 26, 2010 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100268

## Device Facts

- **Applicant:** Sechrist Industries, Inc.
- **Product Code:** [CBF](/productcode/CBF.md)
- **Decision Date:** Apr 26, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5470
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The intended use of the hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society (founded in 1967 to foster exchange of data on the physiology and medicine of commercial and military diving). The committee is comprised of practitioners and scientific investigators in the fields of internal medicine, infectious diseases, pharmacology, emergency medicine, general surgery, orthopedic surgery, trauma surgery, thoracic surgery, otolaryngology, oral and maxillofacial surgery and aerospace medicine. The committee is responsible for continually reviewing research and clinical data in determining the safety and efficacy of hyperbaric oxygen. Currently, there are thirteen indications that are approved by the committees these thirteen indications were accepted based on sound physiologic rationale, in vivo or in vitro studies that demonstrate effectiveness, controlled animal studies, prospective controlled clinical studies and extensive clinical experience from multiple hyperbaric medicine centers. These thirteen indications have been recommended for third-party reimbursement and most insurance carriers have established reimhursement policy based on the recommendations. The thirteen indications are: - 1. Air or Gas embolism - Carbon Monoxide Poisoning 2. - Carbon Monoxide Poisoning Complicated by Cyanide Poisoning - 3. Clostridial Myonecrosis and Myonecrosis (Gas Gangrene) - Crush Injury, Compartment Syndrome and Other Acute Traumatic Ischemias 4. - న. Decompression Sickness - Arterial Insufficiences 6. Central Retinal Artery Occlusion - Enhancement of Healing In Selected Problem Wounds - 7. Severe Anemia - 8. Intracranial Abscess - ರಿ. Necrotizing Soft Tissue Infections - 10. Osteomyelitis (Refractory) - 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis) - 12. Comprised Grafts and Flaps - 13. Thermal Burns

## Device Story

Monoplace hyperbaric chamber; administers 100% oxygen at pressures up to 3 atmospheres absolute (3 ATA). Mechanical/pneumatic system; operator-controlled pressurization/depressurization profiles to minimize patient discomfort. Used in clinical settings by trained personnel. Provides high-concentration oxygen environment for therapeutic treatment of specified conditions. Patient interface ports allow for monitoring and IV fluid delivery. Output is the controlled hyperbaric environment; clinical decision-making based on established hyperbaric medicine protocols.

## Clinical Evidence

No clinical data. Performance testing confirms device operates as designed per approved specifications. Device designed/manufactured in accordance with ASME Boiler and Pressure Vessel Code, ASME PVHO-1, and NFPA 99.

## Technological Characteristics

Monoplace pressure chamber; 41-inch internal diameter seamless acrylic cylinder; hard anodized aluminum cover and door assembly; stainless steel tie rods/nuts. Operates up to 3 ATA (29.4 psig). Pressurization rates 0.4-5.0 psig/minute. Mechanical/pneumatic control system. No software/algorithm described.

## Regulatory Identification

A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).

## Predicate Devices

- Sechrist Industries, Inc., Model 3200P/PR Hyperbaric Chamber ([K950386](/device/K950386.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ki00268

## APR 2 6 2010

### 510(K) SUMMARY

| Sechrist Industries, Inc., 4100H/HR Hyperbaric Chamber (K100268)                   |                                                                                                            |                |               |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------|---------------|
| Submitter: Sechrist Industries, Inc.<br>4225 E. La Palma Ave.<br>Anaheim, CA 92807 |                                                                                                            | Date Prepared: | March 3, 2010 |
| Contact:                                                                           | Victor Arellano, Director, Quality Assurance & Regulatory Affairs<br>Phone: 714-579-8344 Fax: 714-579-8424 |                |               |
| Trade Name                                                                         | : Model 4100H/HR Hyperbaric Chamber                                                                        |                |               |
| Common Name                                                                        | : Hyperbaric Chamber                                                                                       |                |               |
| Classification Name                                                                | : Chamber, Hyperbaric (Title 21, C.F.R. § 868.5470)                                                        |                |               |
| Product Code                                                                       | : CBF                                                                                                      |                |               |
| Predicate Device                                                                   | : Sechrist Industries, Inc., Model 3200P/PR Hyperbaric Chamber (K950386)                                   |                |               |

#### Intended Use

The intended use of the hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society (founded in 1967 to foster exchange of data on the physiology and medicine of commercial and military diving). The committee is comprised of practitioners and scientific investigators in the fields of internal medicine, infectious diseases, pharmacology, emergency medicine, general surgery, orthopedic surgery, trauma surgery, thoracic surgery, otolaryngology, oral and maxillofacial surgery and aerospace medicine. The committee is responsible for continually reviewing research and clinical data in determining the safety and efficacy of hyperbaric oxygen. Currently, there are thirteen indications that are approved by the committees these thirteen indications were accepted based on sound physiologic rationale, in vivo or in vitro studies that demonstrate effectiveness, controlled animal studies, prospective controlled clinical studies and extensive clinical experience from multiple hyperbaric medicine centers. These thirteen indications have been recommended for third-party reimbursement and most insurance carriers have established reimhursement policy based on the recommendations.

The thirteen indications are: -

- 1. Air or Gas embolism
- Carbon Monoxide Poisoning 2.
- Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
- 3. Clostridial Myonecrosis and Myonecrosis (Gas Gangrene)
- Crush Injury, Compartment Syndrome and Other Acute Traumatic Ischemias 4.
- న. Decompression Sickness
- Arterial Insufficiences 6.

Central Retinal Artery Occlusion

Enhancement of Healing In Selected Problem Wounds

- 7. Severe Anemia
- 8. Intracranial Abscess
- ರಿ. Necrotizing Soft Tissue Infections
- 10. Osteomyelitis (Refractory)
- 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
- 12. Comprised Grafts and Flaps
- 13. Thermal Burns

{1}------------------------------------------------

### Technological Characteristics

The Sechrist Model 4100H/HR Hyperbaric Chamber is a monoplace pressure chamber designed to administer 100% oxygen at pressure greater than ambient, up to 3 atmospheres absolute (30 ps) of pressure. The hyperbaric chamber is a mechanical and pneumatic system capable of controlling the operator defined pressurization profiles.

A hyperbaric oxygen chamber is a pressure vessel and control system that is designed to provide patient exposure to a very high oxygen concentration at higher than normal atmospheric pressure. Titration of the oxygen exposure is controlled by selecting the pressure achieved within the pressure vessel. Pressurization and de-pressurization rates are selected to minimize patient discomfort while increasing and decreasing the chamber pressure. Typical monoplace chambers are capable of pressurizing to 3 ATA (29.4 psig above atmospheric pressure). Typical pressurization and de-pressurization rates are in the range of 0.4 to 5.0 psig/minute.

The Model 4100H/HR is constructed with a horizontal 41 inch internal diameter seamless acrylic cylinder. One end is equipped with a hard anodized aluminum cover and the other end with a hard anodized door assembly, with access ports available for patient interface (such as patient monitoring, delivery of intravenous fluids, etc.) locking mechanism and interlocking safety device. The cylinder, end cover and the door assembly are assembled together with stainless steel tie rods/nuts and hinge assembly. The Model 4100 comes in two configurations (4100H and 4100HR). In the Standard configuration-4100H-the control panel is located on the left of the chamber when facing the door. In the Reverse configuration-4100HR-the control panel is located on the right side of the chamber.

### Performance Data

Performance testing conducted confirms the device operates as designed. The Model 4100H/HR Hyperbaric Chamber functioned as intended per its approved specifications.

A performance standard regulation Section 514 of the Food, Drug and Cosmetic Act has not been promulgated for this device type.

However, the device is designed and manufactured in accordance with the following industry standards:

- Section VIII, Division 1 of the American Society of Mechanical Engineers (ASME) Boiler and ● Pressure Vessel Code
- ASME PVHO-1 Pressure Vessels for Human Occupancy Standard .
- NFPA 99 Safety Standard for Health Care Facilities, Chapter 20 D

The following quality assurance measures were applied to the development of the 4100H/HR Hyperbaric Chamber - Requirements specifications reviews, performance testing and validation, environmental testing, electrical safety/EMC testing manufacturing process control procedures, process validation, and design controls.

### Substantial Equivalence

The Model 4100H/HR Hyperbaric Chamber is as safe and effective as the Model 3200P/PR Hyperbaric Chamber. The Model 4100H/HR Hyperbaric Chamber has the same intended uses, similar technological characteristics, and similar principles of operation as its predicate device. The minor technological differences between the Model 4100H/HR Hyperbaric Chamber Neptune and its predicate denvioler raise no new questions of safety or effectiveness. Thus, the Model 4100H/HR Hyperbaric Chamber Veotilator is substantially equivalent.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three overlapping, curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR 2 3 2010

Mr. Victor Arellano Director Quality Assurance & Regulatory Affairs -Sechrist Industries, Incorporated 4225 East La Palma Avenue Anaheim, California 92807

Re: K100268

Trade/Device Name: Sechrist Model 4100H/HR Hyperbaric Oxygen Chamber Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: II Product Code: CBF Dated: April 9, 2010 Received: April 12, 2010

Dear Mr. Arellano:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2 - Mr. Arellano

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Nh for

Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

### 510(k) Number (if known): K100268

### Device Name: Sechrist Model 4100H/HR Hyperbaric Chamber

#### Indications for Use:

The intended use of the hyperbaric chambers have been established by the Committee on Hyperbaric Oxygen Therapy of the Undersea and Hyperbaric Medical Society (founded in 1967 to foster exchange of data on the physiology and medicine of commercial and military diving). The committee is comprised of practitioners and scientific investigators in the fields of internal medicine, infectious diseases, pharmacology, emergency medicine, general surgery, orthopedic surgery, trauma surgery, thoracic surgery, otolaryngology, oral and maxillofacial surgery and aerospace medicine. The committee is responsible for continually reviewing research and clinical data in determining the safety and efficacy of hyperbaric oxygen. Currently, there are thirteen indications that are approved by the committee; these thirteen indications were accepted based on sound physiologic rationale, in vivo or in vitro studies that demonstrate effectiveness, controlled animal studies, prospective controlled clinical studies and extensive clinical experience from multiple hyperbaric medicine centers. These thirteen indications have been recommended for thirdparty reimbursement and most insurance carriers have established reimbursement policy based on the recommendations.

The thirteen indications are:

1. Air or Gas embolism

- 2. Carbon Monoxide Poisoning Carbon Monoxide Poisoning Complicated by Cyanide Poisoning
- 3. Clostridial Myonecrosis and Myonecrosis (Gas Gangrene)
- Crush Injury, Compartment Syndrome and Other Acute Traumatic Ischemias এ..
- 5. Decompression Sickness

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
L. Schult

Page 1 of 2

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K100268

{5}------------------------------------------------

# Indications for Use

### 510(k) Number (if known): K100268

Device Name: Sechrist Model 4100H/HR Hyperbaric Chamber

Indications for Use:

- 6. Arterial Insufficiencies Central Retinal Artery Occlusion Enhancement of Healing In Selected Problem Wounds
- 7. Severe Anemia
- 8. Intracranial Abscess
- 9. Necrotizing Soft Tissue Infections
- 10. Osteomyelitis (Refractory)
- 11. Delayed Radiation Injury (Soft Tissue and Bony Necrosis)
- 12. Compromise Grafts and Flaps
- 13. Thermal Burns

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

L Schulther

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K100268

Page 2 of _ 2

---

**Source:** [https://fda-staging.innolitics.com/device/K100268](https://fda-staging.innolitics.com/device/K100268)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).