ACTIWAVE 1E, ACTIWAVE2E, ACTIWAVE2M, ACTIWAVE4E, ACTIWAVE CARDIO MODEL 08-514, 08-521, 08-536, 08-557, 08-603

K100266 · Camntech, Ltd. · MWJ · Apr 29, 2010 · Cardiovascular

Device Facts

Record IDK100266
Device NameACTIWAVE 1E, ACTIWAVE2E, ACTIWAVE2M, ACTIWAVE4E, ACTIWAVE CARDIO MODEL 08-514, 08-521, 08-536, 08-557, 08-603
ApplicantCamntech, Ltd.
Product CodeMWJ · Cardiovascular
Decision DateApr 29, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.2800
Device ClassClass 2

Indications for Use

The Actiwave Recorder is an ambulatory, body-worn, dermally-affixed, physiological signal and activity recorder. Actiwave is indicated for use in recording physiological waveforms and activity as part of a physiological recording system. The subject is not restricted to a medical environment. The subject may participate in normal activities while the Actiwave recorder is in place. Actiwave is a multi-use recorder. Actiwave may be used wherever quantifiable measurement of human physiological signal waveforms and/or activity are needed. Actiwave may be used in a clinical setting or in basic scientific research under the supervision of a trained clinician.

Device Story

Actiwave is a compact, battery-operated, ambulatory physiological signal recorder. It captures physiological waveform voltages via skin-surface gel electrodes and records activity counts via an internal accelerometer (Cardio model). The device logs digital data to internal storage for later transfer to a PC; it performs no onboard data analysis or real-time viewing. Used in clinical or research settings, it is worn by patients during normal daily activities. Clinicians retrieve the stored data from the device to a computer for subsequent review and decision-making. The device provides a portable, unobtrusive method for long-term physiological monitoring, enabling data collection outside of traditional medical environments.

Clinical Evidence

Bench testing only. Performance evaluated against Special Controls for ambulatory electrocardiographs, including input dynamic range, input impedance, gain stability, system noise, multichannel crosstalk, frequency response, timing accuracy, and temporal alignment. Electrical safety and electromagnetic interference testing performed per international standards.

Technological Characteristics

Polycarbonate/ABS case; stainless steel snap or no-touch connectors; 3.0V lithium rechargeable battery. Sensing via gel-type electrodes (voltage) and accelerometer (activity). 1, 2, or 4-channel configurations. Dimensions range from 19x27x9 mm to 37x27x9 mm; weight 5-10g. Data stored digitally for PC transfer. No onboard analysis or display.

Indications for Use

Indicated for patients requiring ambulatory recording of physiological waveforms and activity. Suitable for use in clinical settings or scientific research under clinician supervision; patients may perform normal daily activities while wearing the device.

Regulatory Classification

Identification

A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) Premarket Notification Actiwave™M #### 5.0 510(K) SUMMARY ## APR 2 9 2010 #### 5.1 Summary information #### 5.1.1 Submitter's name and address Howard Smith, Engineering Director CamNtech, UK, Ltd. Upper Pendrill Court Ermine Street North Papworth Everard Cambridge, United Kingdom CB233UY Tel: 011-44-1480-831223 011-44-1480-831733 Fax: Date summary was prepared: 01/25/2010 #### 5.1.2 Name of device | Trade Name: | ActiwaveTM | |----------------------|-----------------------------------------| | Common Name: | Digital Ambulatory Monitor | | Classification Name: | Electrocardiograph, Ambulatory (Without | | Analysis) | | | Product Regulation: | 21 CFR 870.2800 | Product Code: MWJ #### 5.1.3 Identification of predicate device SEER Light Compact Digital Holter Recorder, 510(k) Number : K021470 {1}------------------------------------------------ K100266 P2/4 510(k) Premarket Notification Actiwave™M #### 5.2 Device description #### 5.2.1.1 Functions of the device Activave is a compact, ambulatory, battery-operated physiological signal recorder. Actiwave is commonly known as an Ambulatory Electrocardiograph. Because of its small size and battery operation, Actiwave may be used in a clinical or non-clinical setting to record physiological signal waveforms while the patient is unattended. The device obtains its signal from the skin surface through the use of standard gel-type signal electrodes. Actiwave may be worn during daily activities and is fully ambulatory due to its small size and its battery-operated power source. Recorded data may be transferred later to a personal computer for storage and archiving. Actiwave does not provide any data viewing or data analysis functions. #### 5.2.1.2 Basic scientific concepts The Actiwave device acquires and logs digital data whose values represent the amplitude-varying physiological waveform signal voltages at progressive time intervals. The scientific concepts and technologies that are used to sense the signals are summarized in Table 5.1. | Physiological Parameter | Recorder Used | Technology | Value Obtained | |-------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------| | Physiological waveform | Actiwave1-, 2-, and 4-<br>channel recorders and<br>Actiwave Cardio | Amplifier senses and<br>amplifies voltage signal. | Digital value<br>corresponding to<br>instantaneous signal<br>voltage. | | Activity | Actiwave Cardio | Signal from accelerometer<br>amplifier is digitized and<br>recorded. | Digital value<br>corresponding to Activity<br>Counts or acceleration. | # TABLE 5.1: BASIC TECHNOLOGIES USED FOR PHYSIOLOGICAL SIGNAL RECORDING IN ACTIWAVE RECORDERS {2}------------------------------------------------ 510(k) Premarket Notification Actiwave™e™ #### 5.2.1.3 Physical characteristics of Recorder Pertinent physical characteristics of the Actiwave Recorder are shown in Table 5.2. # TABLE 5.2: PHYSICAL CHARACTERISTICS OF ACTIWAVE ### RECORDER | Parameter | Value | Condition/Note | | |------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------| | Size | Actiwave 1-channel: 19x27x9 mm<br>Actiwave 2-channel: 25x27x9 mm<br>Actiwave 4-channel: 37x27x9 mm<br>Actiwave Cardio: 32diam x 10 mm | Outer dimensions | | | Weight | 1-channel: 5 gm<br>2-channel: 6 gm<br>4-channel: 9 gm<br>Cardio: 10 gm | With no electrodes<br>attached | | | | Case material | Polycarbonate/ABS | | | | Attachment type | Stainless steel snap connector or<br>standard no-touch connectors | | | | Battery type | 3.0 volt lithium rechargeable | Not user replaceable | | Indicators | None | | | #### 5.3 Statement of the intended use of the device The Actiwave Recorder is an ambulatory, body-worn, dermally-affixed, physiological signal and activity recorder having 1, 2, or 4 channels of physiological waveform storage. Actiwave is indicated for use in recording physiological waveforms and activity as part of a physiological recording system. The subject is not restricted to a medical environment. The subject may participate in normal activities while the Actiwave recorder is in place. Actiwave is a multi-use recorder. Actiwave may be used wherever quantifiable measurement of human physiological signal waveforms and/or activity are needed. Actiwave may be used in a clinical setting or in basic scientific research under the supervision of a trained clinician. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the alphanumeric strings "K100266" and "P4/4". The strings are written in a handwritten style with black ink on a white background. The strings appear to be some kind of identifier or code. 510(k) Premarket Notification Actiwave™ ### Technological characteristics of the device compared to predicate 5.4 device Actiwave and the SEER Light Compact Digital Holter Recorder (FDA 510(k) Number: K021470) are each physiological recording systems based upon the concept of an ambulatory, unattended physiological recorder that logs physiological data to the recorder. Each of these devices is a solid-state recorder with electronic amplifiers, data collection means, and with the ability to store data until it is transferred to a personal computer. Actiwave and the SEER Light Compact Digital Holter Recorder are of similar mechanical materials, construction, size, and human interface characteristics. Activave and the SEER Light Compact Digital Holter Recorder are of similar electronic design and their operational characteristics are functionally equivalent. #### 5.5 Assessment of non-clinical performance data and safety testing The performance of the Actiwave Recorder has been tested in accordance with a Special Control specific to this device type. The categories of tests applied to this device are entirely contained within the Special Control. These categories are: Input dynamic range, Input Impedance, Gain Stability, System noise, Multichannel crosstalk, Frequency response, Timing accuracy, and Temporal alignment. For all the categories, we have determined that Actiwave performance meets the test criteria in the Special Control. In addition to performance testing, the device has been tested for electrical safety and electromagnetic interference according to internationally recognized standards. The testing according to these standards has raised no issues as to the safety and effectiveness of the present device or the present device compared to the predicate device. #### 5.6 Conclusion The results of the tests conducted in accordance with the Special Control demonstrate that the Actiwave is as safe, as effective, and performs as well as the predicate device. {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling a stylized bird or human figure, with flowing lines forming the shape of a wing or hair. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002 CamNtech, UK, Ltd. c/o Mr. Howard Smith Engineering Director Upper Pendrill Court, Ermine Street North Papworth Everard, Cambridge, UK CB233UY APR 2 9 2010 Re: K100266 Device Name: Actiwave™ Regulation Number: 21 CFR 870.2800 Regulation Name: Ambulatory Electrocardiograph (Without Analysis) Regulatory Class: Class II (Two) Product Code: MWJ Dated: January 25, 2010 Received: January 29, 2010 Dear Mr. Smith: . We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {5}------------------------------------------------ Page 2 - Mr. Howard Smith comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRF/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours Brom D. Zuckerman, M.D. Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### 1.0 INDICATIONS FOR USE STATEMENT ## Indications for Use 510(k) Number (if known): K100266 Device Name: Actiwave™ Indications For Use: The Actiwave Recorder is an ambulatory, body-worn, dermally-affixed, physiological signal and activity recorder. Actiwave is indicated for use in recording physiological waveforms and activity as part of a physiological recording system. The subject is not restricted to a medical environment. The subject may participate in normal activities while the Actiwave recorder is in place. Actiwave is a multi-use recorder. Actiwave may be used wherever quantifiable measurement of human physiological signal waveforms and/or activity are needed. Actiwave may be used in a clinical setting or in basic scientific research under the supervision of a trained clinician. Prescription Use AND/OR -Over-The-Counter Use (Part 21 CFR 801 Subpart D) . (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K100266 Page 4-1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...