VLOC9ABSORBABLE WOUND CLOSURE DEVICE

K100257 · Surgical Devices, A Division of Tyco Healthcare · GAM · May 10, 2010 · General, Plastic Surgery

Device Facts

Record IDK100257
Device NameVLOC9ABSORBABLE WOUND CLOSURE DEVICE
ApplicantSurgical Devices, A Division of Tyco Healthcare
Product CodeGAM · General, Plastic Surgery
Decision DateMay 10, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4493
Device ClassClass 2
AttributesTherapeutic

Indications for Use

V-Loc™ 90 Absorbable Wound Closure Device is indicated for soft tissue approximation where use of an absorbable suture is appropriate.

Device Story

V-Loc™ 90 is a synthetic absorbable wound closure device; composed of Glycomer™ 631 (glycolide, dioxanone, trimethylene carbonate). Features unidirectional barbs along monofilament axis to facilitate tissue approximation without knots. Supplied in various USP sizes, dyed or undyed, affixed to needles. Used by clinicians for soft tissue approximation. Device functions by anchoring tissue via barbs; provides mechanical closure while material undergoes absorption. Benefits include simplified wound closure procedure by eliminating knot-tying.

Clinical Evidence

Bench testing included needle attachment, diameter, and tensile strength evaluations. In-vivo porcine model (21-day study) compared pull force of sutured incisions against Quill™ SRS and Biosyn™; no statistical difference in strength observed between test device and barbed control. Rat model strength loss evaluation confirmed degradation profile met label claims (75% USP minimum knot pull at 2 weeks; 90% at 1 week).

Technological Characteristics

Material: Glycomer™ 631 (glycolide, dioxanone, trimethylene carbonate) compliant with ISO 10993-1. Form factor: Monofilament with unidirectional barbs. Sizes: USP 4-0 to 0. Energy source: None (mechanical). Sterilization: Not specified. Connectivity: None.

Indications for Use

Indicated for soft tissue approximation in patients where use of an absorbable suture is appropriate.

Regulatory Classification

Identification

An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K1002571 pg 1 of 2 ## 510(k) Summary of Safety and Effectiveness SUBMITTER: Surgical Devices, a global business unit of Tyco Healthcare Group LP (d/b/a Covidien) 60 Middletown Avenue North Haven, CT 06473 MAY 1 0 2010 CONTACT PERSON: Jennifer Brennan Manager, Regulatory Affairs Tel. No .: (203) 492-5346 DATE PREPARED: April 13, 2010 TRADE/PROPRIETARY NAME: V-Loc™ 90 Absorbable Wound Closure Device COMMON/USUAL NAME: Synthetic Absorbable Suture CLASSIFICATION NAME: Absorbable poly(glycolide/l-lactide) surgical suture V-loc™ 180 Absorbable Wound Closure Device (K091087) PREDICATE DEVICE(S): Syneture™ Biosyn™ Synthetic Absorbable Suture (K000037) Quill™ Self-Retaining System (SRS) comprised of MONODERM™ (K072028) DEVICE DESCRIPTION: The V-Loc™ 90 Absorbable Wound Closure Device is a suture prepared from a synthetic polyester (Glycomer™ 631) composed of glycolide, dioxanone and trimethylene carbonate. Each device has unidirectional barbs along the axis of the monofilament. > The V-Loc™ 90 Absorbable Wound Closure Device will be offered both Undyed (clear) and Dyed with D&C Violet No. 2 at a level not exceeding 0.2% by weight of the suture in sizes U.S.P. (EP) 4-0 (Metric 2.0) 3-0 (Metric 3), 2-0 (Metric 3.5) and 0 (Metric 4). They will be supplied in pre-cut lengths affixed to various needle types. > V-Loc™ 90 Absorbable Wound Closure Devices are indicated for soft tissue INDICATIONS: 이러한 그래서 . TECHNOLOGICAL CHARACTERISTICS: MATERIALS: PERFORMANCE DATA: approximation where use of an absorbable suture is appropriate. V-Loc™ 90 Absorbable Wound Closure Device is substantially equivalent to the predicate devices with regards to use in soft tissue approximation. All components of the V-Loc™ 90 Absorbable Wound Closure Device are comprised of materials that are in compliance with ISO standard 10993-1. Performance testing was conducted to verify that the V-Loc™ 90 Absorbable Wound Closure Device is safe and effective and performs as intended. The following is a description of tests performed and associated conclusions: 0 - In-vitro performance evaluation (bench testing) - Needle Attachment meets USP/EP specification. 비 I - Diameter (non-barbed suture) Maximum Overage of USP/EP specification as stated in the Instructions for Use. . - Tensile Strength (T=0, T=1 and T=2 straight pull evaluation)- Pass - I Barb Holding Strength (T=0, T=1, and T=2 simulated barb holding of felt medium) - Pass - In-vivo performance evaluation O - Product safety and efficacy in porcine model. Chronic 21 day study 마 with biomechanical testing at T=0, T=3, T=10, and T=21 days to {1}------------------------------------------------ ## K100257: 19292 compare the pull force of the sutured incision of the control (predicate - Quill™ SRS MONODERM and Biosyn™ suture) devices and test device. Results: No statistical differences in strength between the barbed control (Quill™ SRS MONODERM) and test devices. The non-barbed control (Biosyn™) was stronger than both the barbed control and test device at time points T=0 and T=3 days. No statistical differences in strength were observed between the non-barbed control and barbed devices at T=10 and T=21 days. - Strength Loss evaluation in the rat model. Comparison of the degradation over 2 weeks of the non-barbed control (Biosyn ™ suture) and the test device. The degradation of the V-loc™ 90 met the label claim of the Biosyn™ suture at T= 2 weeks, 75% of USP minimum knot pull specification. The additional data obtained at T=1 week was used to establish a strength loss specification of 90% USP minimum knot pull. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular fashion around the eagle. MAY 1 0 2010 Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Surgical Devices, d/b/a Covidien % Ms. Jennifer Brennan Manager, Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06473 Re: K100257 Trade/Device Name: v-Loc™ 90 Absorbable Wound Closure Device Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable poly(glycolide/L-lactide) surgical suture Product Code: GAM Dated: April 6, 2010 Received: April 15, 2010 Dear Ms. Brennan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Jennifer Brennan forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark H. Mikkelson Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications For Use 510(k) Number (if known): Device Name: V-Loc™ 90 Absorbable Wound Closure Device Indications For Use: V-Loc™ 90 Absorbable Wound Closure Device is indicated for soft tissue approximation where use of an absorbable suture is appropriate. Prescription Use × (Part 21 CFR 801 Subpart D) ﺎ ﻣﺴ AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daúd Krone for JXW (Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K100257
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...