K100244 · Metro Optics of Austin, Inc. · LPL · Oct 28, 2010 · Ophthalmic
Device Facts
Record ID
K100244
Device Name
METRO SOFT
Applicant
Metro Optics of Austin, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Oct 28, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The Metro Soft, sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The Metro Soft, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters. The Metro Soft, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters. The Metro Soft, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eves and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add nower of up to +4.00 diopters. The Metro Soft, irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting. Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
Device Story
Daily wear silicone hydrogel contact lens; fabricated from efrofilcon A; lathe-cut for custom prescription. Lens acts as refractive medium to focus light on retina. Used by eyecare practitioners to correct ametropia, astigmatism, presbyopia, and irregular corneal conditions. Supplied sterile in buffered saline vials. Requires patient cleaning/disinfection via chemical or hydrogen peroxide systems. Benefits include vision correction and management of corneal irregularities.
Clinical Evidence
No clinical data. Clinical performance of efrofilcon A material previously established. Safety supported by non-clinical in vitro and in vivo toxicology and biocompatibility testing conducted under GLP.
Technological Characteristics
Material: efrofilcon A (silicone hydrogel); 74% water content; non-ionic. Oxygen permeability: 59.8 x 10-11 (cm2/sec)(ml O2/ml x hPa @ 35°C). Refractive index: 1.38. Lathe-cut manufacturing. Sterile, vial-packaged. Spherical, toric, multifocal, and irregular cornea configurations.
Indications for Use
Indicated for daily wear correction of ametropia (myopia/hyperopia) in aphakic and non-aphakic persons with non-diseased eyes. Includes correction for refractive astigmatism (up to 12D depending on lens type) and presbyopia (add power up to +4.00D). Also indicated for management of irregular corneal conditions (e.g., keratoconus, post-graft).
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
PAGE 1 of 6
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
UCT 2 8 2010
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K100244 |
|--------------------------------|----------------------------------------------------------------------------------------------|
| Applicant information: | |
| Date Prepared: | September 20, 2010 |
| Name:<br>Address | Metro Optics of Austin, Inc.<br>15802 Vision Drive<br>Pflugerville, TX 78660 |
| Contact Person: | Steve Webb<br>President |
| Phone number: | (512) 251-2382 |
| Consultant: | Martin Dalsing<br>Medvice Consulting, Inc.<br>806 Kimball Avenue<br>Grand Junction, CO 81501 |
| Phone number | (970) 243-5490 |
| Device Information: | |
| Device Classification: | Class II |
| Classification Number: | LPL |
| Classification Name: | Lenses, Soft Contact, Daily Wear |
| Trade Name: | Metro Soft, Silicone Hydrogel Daily Wear Soft Contact Lens<br>(efrofilcon A) |
{1}------------------------------------------------
100244 PAGE 2 of 6
#### Equivalent Devices:
The Metro Soft, Silicone Hydrogel Daily Wear Soft Contact Lenses (efrofilcon A) are substantially equivalent to the following predicate devices:
Predicate devices:
"IntelliWave", Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A)" by Art Optical Contact Lens, Inc. 510(k) number: K100221
"RevitalEyes, Post-Surgical (hioxifilcon B) by Metro Optics of Austin, Inc. 510(k) number: K080512
"Biofinity (comfilcon A)" by Coopervision, Inc. 510(k) number; K052560
"ActiFresh 400 (lidofilcon A)" By Hydron Ltd. 510(k) number: K983637
#### Device Description:
The Metro Soft Silicone Hydrogel Soft Contact Lenses are fabricated from efrofilcon A. which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The non-ionic lens material, (efrofilcon A) is a daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers and optionally incorporates D&C Green 6 as an integrated handling tint. The lenses are made by lathe-cut for custom RX. It consists of 26% efrofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN).
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly bevond the limbus forming a colorless, transparent optical surface. The (efrofilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{2}------------------------------------------------
The Physical properties of the lens are:
| Refractive Index | 1.38 |
|---------------------|------------------------------------------------------------------------|
| Light Transmission | greater than 97% |
| Surface Character | hydrophilic |
| Water Content | 74 % |
| Specific Gravity | 1.048 (hydrated) |
| Oxygen Permeability | 59.8 x 10-11 (cm2/sec) (ml O₂/ml x hPa @ 35°C), (revised Fatt method). |
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 74% water by weight. The lenses will be manufactured in spherical, toric, multifocal, multifocal toric and irregular Cornea configurations with the following features and properties.
| • Chord Diameter | 12.0 mm to 16.00 mm |
|-------------------------------------|----------------------------------|
| • Center Thickness | 0.01 mm to 0.50 mm |
| • Base Curve | 8.0 mm to 9.5 mm |
| • Power Range | -20.00D to +20.00D in 0.25 steps |
| • Cylinder Power (Toric) | -0.25D to -12.00D |
| • Cylinder Power (Multifocal Toric) | -0.25D to -4.00D |
| • Add Power (Multifocal) | +0.50D to +4.00D |
The lens is supplied sterile in vials containing a buffered saline solution. Vial labeling is printed with appropriate lot numbering, expiration dating and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.
#### Intended Use:
The Metro Soft, sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eves. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The Metro Soft, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.
The Metro Soft, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters.
{3}------------------------------------------------
K100244
PAGE 4 OF 6
The Metro Soft, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The Metro Soft, irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
#### Testing:
| Non-clinical Testing | A series of in vitro and in vivo preclinical toxicology and biocompatibility tests<br>were performed to assess the safety and effectiveness of the Metro Soft (efrofilcon<br>A) Silicone Hydrogel Soft Contact Lenses packaged in glass vials. All non-clinical<br>toxicology tests were conducted in accordance with the GLP regulation. All other<br>testing was conducted according to valid scientific protocols. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Test results of the non-clinical testing on the Metro Soft (efrofilcon A) Silicone<br>Hydrogel Soft Contact Lenses demonstrate that: |
| | Lenses supplied in glass vials are sterile for the indicated shelf-life, The packaging material and extracts are not toxic and not irritating, and Lens physical and material properties are consistent with currently marketed lenses. |
| Clinical Data | The clinical performance of the (efrofilcon A) lens material has been previously<br>established, and therefore was not required for this 510(k). |
| | The Metro Soft, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens is<br>identical to the cleared Intelliwave3 (efrofilcon A) Silicone Hydrogel Daily Wear<br>Soft Contact Lens, cleared under K100221. |
| | The Metro Soft, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lenses<br>have the identical manufacturing process (lathe-cut versus lathe-cut) as the<br>marketed Intelliwave3 (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact<br>Lens, cleared under K100221. |
#### Substantial Equivalence:
{4}------------------------------------------------
The following matrix illustrates the production method, lens function and material characteristics of the Metro Soft, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens, as well as the predicate devices.
#### Conclusions Drawn from Studies
#### Validity of Scientific Data
Several laboratories under Good Laboratory Practice regulations conducted toxicology studies, Microbiology, chemistry, shelf-life stability studies and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7
#### Substantial Equivalence
Information presented in this Premarket Notification establishes that the Metro Soft, (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens is as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the requested indication.
#### Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of Silicone Hydrogel, Daily Wear Soft Contact Lens. The benefits to the patient are the same as those for other Silicone Hydrogel contact lenses.
{5}------------------------------------------------
| | Metro Soft, Silicone<br>Hydrogel (efrofilcon A)<br>New Device | IntelliWave3, Silicone<br>Hydrogel (efrofilcon A)<br>predicate device | Metro Optics, RevitalEyes<br>(hioxifilcon B)<br>predicate device | CooperVision Biofinity<br>(comfilcon A)<br>predicate device | Hydron Ltd. ActiFresh 400<br>(lidofilcon A)<br>predicate device |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Indicated for daily wear for the<br>correction of visual acuity in aphakic<br>and not aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia. The lens may also be<br>prescribed for management of<br>irregular corneal conditions such as<br>keratoconus and post graft fitting. | Indicated for daily wear for the<br>correction of visual acuity in aphakic<br>and not aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia. The lens may also be<br>prescribed for management of<br>irregular corneal conditions such as<br>keratoconus and post graft fitting. | Indicated for daily wear for the<br>correction of visual acuity in aphakic<br>and not aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia. The lens may also be<br>prescribed for management of<br>irregular corneal conditions such as<br>keratoconus and post graft fitting. | Indicated for daily wear for the<br>correction of visual acuity in aphakic<br>and not aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia. | Indicated for daily wear for the<br>correction of visual acuity in aphakic<br>and not aphakic persons with non-<br>diseased eyes with myopia or<br>hyperopia. |
| Functionality | same as<br>predicate device | The contact lenses act as a refractive<br>medium that focus light rays from near<br>and distant objects on the retina. | The contact lenses act as a refractive<br>medium that focus light rays from near<br>and distant objects on the retina. | The contact lenses act as a refractive<br>medium that focus light rays from near<br>and distant objects on the retina. | The contact lenses act as a refractive<br>medium that focus light rays from near<br>and distant objects on the retina. |
| Indications | Daily wear, Silicone Hydrogel Soft<br>(hydrophilic) contact lens | Daily wear, Silicone Hydrogel Soft<br>(hydrophilic) contact lens | Daily wear, Soft (hydrophilic) contact<br>lens | Daily wear, Silicone Hydrogel Soft<br>(hydrophilic) contact lens | Daily wear, Soft (hydrophilic) contact<br>lens |
| Production Method | Lathe-Cut, custom manufactured | Lathe-Cut, custom manufactured | Lathe-Cut, custom manufactured | Cast Molded, mass produced | Lathe-Cut, custom manufactured |
| USAN name | efrofilcon A | efrofilcon A | hioxifilcon B | comfilcon A | lidofilcon A |
| Water Content | 74% | 74% | 48% | 48.0% | 74.0% |
| Oxygen Permeability | 60 x 10-11 (cm2/sec)<br>(ml O2/ml x mm Hg @ 35 degrees C),<br>(revised Fatt method). | 60 x 10-11 (cm2/sec)<br>(ml O2/ml x mm Hg @ 35 degrees C),<br>(revised Fatt method). | 15.89 x 10-11 (cm2/sec)<br>(ml O2/ml x mm Hg @ 35 degrees C),<br>(revised Fatt method). | 128 x 10-11 (cm2/sec)<br>(ml O2/ml x mm Hg @ 35 degrees C),<br>(revised Fatt method). | 28 x 10-11 (cm2/sec)<br>(ml O2/ml x mm Hg @ 35 degrees C),<br>(revised Fatt method). |
| Specific Gravity | 1.139 | 1.139 | 1.142 | 1.142 | 1.060 |
人100244 PAGE 6 OF 6
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Metro Optics of Austin Inc. c/o Medvice Consulting, Inc. Mr. Martin Dalsing Official Correspondent 806 Kimball Avenue Grand Junction, CO 81501
OCT 2 8 2010
Re: K100244
Trade/Device Name: Metro Soft Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL Dated: September 20, 2010 Received: September 22, 2010
Dear Mr. Dalsing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Deborah Falls
Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
#### Device Name: Metro Soft, Silicone Hydrogel Daily Wear Soft Contact Lens (efrofilcon A)
#### INDICATIONS FOR USE:
The Metro Soft, sphere (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The Metro Soft, toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with nondiseased eyes and/or possesses refractive astigmatism not exceeding 12 diopters.
The Metro Soft, multifocal (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding .75 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The Metro Soft, multifocal toric (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons with non-diseased eves and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add nower of up to +4.00 diopters.
The Metro Soft, irregular cornea (efrofilcon A) Silicone Hydrogel Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and non-aphakic persons and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.
Evecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH. Office of Device Evaluation (ODE)
Prescription Use V (Per 21 CFR 801.109)
or
(Optional Format 1-2-96)
Over-The-Counter Use
OCT 2 8 2010
Marc Robert
(Division Sign-Off) (Division Sign-On)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
K100244
Number
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.