WIRELESS BODY TEMPERATURE MONITOR MODEL BTM-DIX SERIES (BTM-DIC,D,E,F,G & H)
K100226 · Rio Flexon Technology Co., Ltd. · FLL · Mar 18, 2010 · General Hospital
Device Facts
Record ID
K100226
Device Name
WIRELESS BODY TEMPERATURE MONITOR MODEL BTM-DIX SERIES (BTM-DIC,D,E,F,G & H)
Applicant
Rio Flexon Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Mar 18, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Wireless Body Temperature Monitor, model BTM-D1X series (model BTM-D1C, D, E, F, G and H), is a battery-operated electronic device with intended use of measuring human ear temperature precisely and continuously monitoring armpit temperature via wireless signal transmission of the measuring result. This device is reusable and is intended for ear temperature measurement as well as armpit temperature monitoring for persons over two years old. Model BTM-D1C has two functions, both ear temperature measurement and armpit temperature monitoring. The rest of models, BTM-D1D, E, F, G and H, have only the function of armpit temperature monitoring.
Device Story
Battery-operated electronic thermometer system; consists of receiver unit and wearable armband. Armband uses thermo sensor to measure armpit temperature; transmits data wirelessly (2.4GHz) to receiver. Receiver displays temperature on LCD; updates every 12 seconds. Model BTM-D1C includes additional infrared tympanic ear thermometer function. User-settable high/low audible alarms alert to temperature thresholds. Used in home or clinical settings; operated by patient or caregiver. Provides continuous monitoring to track temperature trends; audible alarms facilitate timely clinical intervention or patient care adjustments.
Clinical Evidence
No new clinical data provided. Accuracy for ear temperature measurement relies on clinical data previously submitted for Taidoc/TD-1107 (K050463) and referenced in K081256. Bench testing performed for electrical safety (EN 60601-1), electromagnetic compatibility (EN 60601-1-2), and temperature measurement accuracy (ASTM E1112-00, ASTM E1965-02). Biocompatibility testing per ISO 10993 confirmed negative results for cytotoxicity, sensitization, and skin irritation.
Technological Characteristics
Battery-operated (3V DC). Components: Receiver (IC HT-49R50) and Armband (IC HT-46R52) with SENTECH sensor. Wireless 2.4GHz transmission. Ear thermometer uses infrared sensing. Dimensions/form factor: wearable armband and handheld receiver. Standards: ASTM E1112-00, ASTM E1965-02, EN 60601-1, EN 60601-1-2. Biocompatible materials used for skin contact.
Indications for Use
Indicated for persons over two years old for ear temperature measurement and/or continuous armpit temperature monitoring.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Rio Flexon Technology Co., Ltd Wireless Body Temperature Monitor, BTM-D1X series (K081256)
{0}------------------------------------------------
K100226
# 1.4 510(K) SUMMARY
# 510(K) SUMMARY
# MAR 1 8 2010
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
#### Manufacturer:
RIO FLEXON TECHNOLOGY CO., LTD 15F, No 868-2, Jhongjheng Road, Jhonghe City Taipei County, 235, Taiwan
#### Official Correspondent:
Mr. Chi-Hung, Liao / General Manager 15F, No 868-2, Jhongjheng Road, Jhonghe City Taipei County, 235, Taiwan Tel: 886-2-8221 8199 Fax: 886-2-8221 8198 E-amil: Jeff-Liao@rio-flexon.com.tw
## US agent and correspondent:
Alan Schwartz/Executive VP mdi Consultants, Inc. 55 Northern Blvd. Great Neck, NY 11021.U.S.A Tel: 516 482 9001 Fax: 516 482 0186 Alan P. Schwartz [Alan@mdiconsultants.com]
### Date of Submission:
January 25, 2010
#### Classification name:
Clinical Electronic Thermometer / Class II
#### Proprietary Name:
Wireless Body Temperature Monitor/ BTM-D1X series (BTM-D1C, D, E, F, G & H)
#### Common name:
Wireless Body Temperature Monitor
# Regulatory Reference:
21 CFR 880.2910
{1}------------------------------------------------
#### Predicate Device:
Rio Flexon Technology Co., Ltd Wireless Body Temperature Monitor, BTM-D1X series (K081256)
#### Intended Use:
The Wireless Body Temperature Monitor, model BTM-D1X series (model BTM-D1C, D, E, F. G and H), is a battery-operated electronic device with intended use of measuring human ear temperature precisely and continuously monitoring armpit temperature via wireless signal transmission of the measuring result. This device is reusable and is intended for ear temperature measurement as well as armpit temperature monitoring for persons over two vears old. Model BTM-D1C has two functions, both ear temperature measurement and armpit temperature monitoring. The rest of models, BTM-D1D, E, F, G and H, have only the function of armpit temperature monitoring.
#### Device Description:
The Wireless Body Temperature Monitor BTM-D1X series includes models BTM-DIC, D, E, F, G and H. This device is a battery-operated electronic device that is intended to be worn at left arm to monitor the armpit temperature continuously. In addition, BTM-D1C has the additional function of an ear thermometer.
The device is composed of two operational parts, the receiver and armband. An optional accessory may be included to assist in affixing the sensor of the armband more tightly. The receiver is the main operation unit on which the ear thermometer, the measuring circuit, LCD display control circuit, and the main operation keys are included. The armband was designed and constructed with a thermo sensor and signal communication unit. For the monitoring operation, both the receiver and armband must be switched on. Soon after these two parts are switched on, the wireless signal communication will set up between the receiver and the armband. The temperature monitoring signal measured at armpit will be continuously indicated on the LCD of the receiver and will update every 12 seconds.
The device also has a high/low audible alarm, which will sound if the body temperature goes above the high alarm temperature or below the low alarm temperature set by the user.
In addition to the continuous armpit temperature monitoring, the user can also operate the functional key on the receiver to take an ear temperature measurement at any time if needed. The reading on the LCD screen will return to the armpit temperature after 30 seconds.
This system uses a 3.0V DC battery for operation of the complete system. Whenever the battery is low, the MCU circuit will detect the low battery condition automatically, and will display "Low battery" on the LCD display. The BTM-D1X series was designed and verified according to the US standard ASTM E1112-00 and ASTM E1965-02.
#### Comparison to Predicate Devices:
The new BTM-D1X series is the same as the predicate BTM-D1X series under K081256, except for the addition of the high/low alarm function.
{2}------------------------------------------------
| Element of<br>comparison | Prior BTM-<br>D1C model<br>from K081256 | Prior BTM-<br>D1D/E/F/G/H<br>models from<br>K081256 | Modified BTM-<br>D1C model | Modified BTM-<br>D1D/E/F/G/H<br>models |
|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ear<br>Thermometer<br>type | Infrared<br>tympanic<br>thermometer | N/A | Infrared<br>tympanic<br>thermometer | N/A |
| Intended use | Intended use of<br>measuring<br>human ear<br>temperature<br>precisely and<br>continuously<br>monitoring<br>armpit<br>temperature via<br>wireless signal<br>transmission of<br>the measuring<br>result. This<br>device is<br>reusable and<br>intended for ear<br>temperature<br>measurement as<br>well as armpit<br>temperature<br>monitoring for<br>persons over<br>two years old | This device is<br>intended to<br>continuously<br>monitor armpit<br>temperature via<br>wireless signal<br>transmission of<br>measuring<br>result. The<br>device is<br>intended to be<br>used for persons<br>over two years<br>old. | Intended use of<br>measuring<br>human ear<br>temperature<br>precisely and<br>continuously<br>monitoring<br>armpit<br>temperature via<br>wireless signal<br>transmission of<br>the measuring<br>result. This<br>device is<br>reusable and<br>intended for ear<br>temperature<br>measurement as<br>well as armpit<br>temperature<br>monitoring for<br>persons over<br>two years old | This device is<br>intended to<br>continuously<br>monitor armpit<br>temperature via<br>wireless signal<br>transmission of<br>measuring<br>result. The<br>device is<br>intended to be<br>used for persons<br>over two years<br>old. |
| Brand Name | Rio Flexon | Rio Flexon | Rio Flexon | Rio Flexon |
| Signal<br>processing and<br>display | Wireless 2.4G<br>transmission &<br>display on LCD<br>screen of<br>monitor | Wireless 2.4G<br>transmission &<br>display on LCD<br>screen of<br>monitor | Wireless 2.4G<br>transmission &<br>display on LCD<br>screen of<br>monitor | Wireless 2.4G<br>transmission &<br>display on LCD<br>screen of<br>monitor |
| Power<br>requirements of<br>Armband | 3V / CR2032<br>Lithium | 3V / CR2032<br>Lithium | 3V / CR2032<br>Lithium | 3V / CR2032<br>Lithium |
| Power<br>requirement of<br>Receiver | AAA battery x<br>2pcs | AAA battery x<br>2pcs | AAA battery x<br>2pcs | AAA battery x<br>2pcs |
| Temperature<br>range | 25 °C - 43 °C | 25 °C - 43 °C | 25 °C - 43 °C | 25 °C - 43 °C |
| High/low alarm | N/A | N/A | Manual | Manual |
| Element of<br>comparison | Prior BTM-<br>D1C model<br>from K081256 | Prior BTM-<br>D1D/E/F/G/H<br>models from<br>K081256 | Modified BTM-<br>D1C model | Modified BTM-<br>D1D/E/F/G/H<br>models |
| High alarm<br>temperature<br>range | N/A | N/A | 36°C - 40°C | 36°C - 40°C |
| Low alarm<br>temperature<br>range | N/A | N/A | 32°C - 35.5°C | 32°C - 35.5°C |
| Ambient<br>temperature | 15 °C - 42 °C<br>(with 95% RH<br>humidity) | 15 °C - 42 °C<br>(with 95% RH<br>humidity) | 15 °C - 42 °C<br>(with 95% RH<br>humidity) | 15 °C - 42 °C<br>(with 95% RH<br>humidity) |
| Storage<br>condition | -20 °C - 50 °C<br>(with 95% RH<br>humidity) | -20 °C - 50 °C<br>(with 95% RH<br>humidity) | -20 °C - 50 °C<br>(with 95% RH<br>humidity) | -20 °C - 50 °C<br>(with 95% RH<br>humidity) |
| Accuracy for<br>armpit<br>temperature | $\pm$ 0.1 °C | $\pm$ 0.1 °C | $\pm$ 0.1 °C | $\pm$ 0.1 °C |
| Accuracy for ear<br>temperature | $\pm$ 0.2 °C for 36<br>°C - 39 °C; $\pm$ 0.3<br>°C for the others | N/A | $\pm$ 0.2 °C for 36<br>°C - 39 °C; $\pm$ 0.3<br>°C for the others | N/A |
| Monitoring time | 24 hours | 24 hours | 24 hours | 24 hours |
| Components | 1. Receiver<br>IC#: HT-<br>49R50 | 1. Receiver<br>IC#: HT-<br>49R50 | 1. Receiver<br>IC#: HT-<br>49R50. | 1. Receiver<br>IC#: HT-<br>49R50 |
| | 2. Armband<br>IC#: HT-<br>46R52 | 2. Armband<br>IC#: HT-<br>46R52 | 2. Armband<br>IC#: HT-<br>46R52 | 2. Armband<br>IC#: HT-<br>46R52 |
| | 3. Ear<br>temperature<br>measuring<br>module | 3. RF<br>transmission<br>module<br>(2.4GHz) | 3. Ear<br>temperature.<br>measuring<br>module (for<br>BTM-D1C<br>models only) | 3. RF<br>transmission<br>module<br>(2.4GHz) |
| | 4. RF<br>transmission<br>module<br>(2.4GHz) | | 4. RF<br>transmission<br>module<br>(2.4GHz) | |
| Sensor | SENTECH<br>sensor | SENTECH<br>sensor | SENTECH<br>sensor | SENTECH<br>sensor |
| Electrical safety | EN 60601-1 | EN 60601-1 | EN 60601-1 | EN 60601-1 |
| Element of<br>comparison | Prior BTM-<br>D1C model<br>from K081256 | Prior BTM-<br>D1D/E/F/G/H<br>models from<br>K081256 | Modified BTM-<br>D1C model | Modified BTM-<br>D1D/E/F/G/H<br>models |
| Bio-<br>compatibility for<br>the skin contact<br>material<br>(Armband) | Cytotoxicity /<br>Negative<br>Sensitization /<br>Negative<br>Primary skin<br>irritation /<br>Negative | Cytotoxicity /<br>Negative<br>Sensitization /<br>Negative<br>Primary skin<br>irritation /<br>Negative | Cytotoxicity /<br>Negative<br>Sensitization /<br>Negative<br>Primary skin<br>irritation /<br>Negative | Cytotoxicity /<br>Negative<br>Sensitization /<br>Negative<br>Primary skin<br>irritation /<br>Negative |
| Measuring<br>location | Armpit and ear<br>(for BTM-D1C<br>models only) | Armpit and ear<br>(for BTM-D1C<br>models only) | Armpit and ear<br>(for BTM-D1C<br>models only) | Armpit and ear<br>(for BTM-D1C<br>models only) |
:
{3}------------------------------------------------
.
{4}------------------------------------------------
## Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence
Compliance to applicable voluntary standards includes ASTM E1112:2000, ASTM E1965:2002, EN 60601-1, and EN 60601-1-2. All of the required conformity reports are included in the 510(k) submission documents.
#### Discussion of clinical report for measurement accuracy
The ear temperature measuring function of the modified BTM-D1X series (BTM-D1C model only) is the same as the predicate BTM-D1X series (BTM-D1C model only) from K081256, which involved the integration of the 510(k) cleared model, Taidoc/TD-1107 (K050463). Therefore, the clinical report for measurement accuracy (per ASTM E1965:2002), which was included in the 510(k) submission for Taidoc/TD-1107 (K050463) and referenced in K081256, is still applicable for the modified BTM-D1X series.
No additional clinical report is included in this 510(k) submission.
#### Performance Tests:
Test Performed
- 1. EN 60601-1
- 2. EN 60601-1-2/ EN 300 440-2 / EN · 3011489-17-3
- 3. ASTM/ E 1112-00
- 4. ASTM/ E 1965-02
- 5. Biocompatibility (ISO 10993)
- 6. FCC
# Laboratory
Electronics Testing Center, Taiwan Electronics Testing Center, Taiwan
Rio Flexon Technology Co., Ltd Taidoc Technology Corporation Taiwan National Chung-Hsing University Lab Electronics Testing Center, Taiwan
#### Conclusions:
{5}------------------------------------------------
The modified device has the same fundamental scientific technology and intended use as the predicate device. The only difference between the modified BTM-DIX series and the predicate device (K081256/BTM-D1X series) is the addition of the high/low alarm function.
·
、
1
:
.
-
:
. .
1. September 19.
.
.
.
.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a stylized human figure. The symbol is composed of three curved lines that converge at the top and then separate into three wavy lines at the bottom.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## MAR ] & 2010
Mr. Chi-Hung Liao General Manager RIO Flexon Technology Company, Limited 15F, 868-2, JhongJheng Road Jhonghe City, Taipei County China (TAIWAN) 235
Re: K100226
Trade/Device Name: Wireless Body Temperature Monitor/Model: BTM-D1C.D.E.F.G and H. Regulation Number: 21CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: March 4, 2010 Received: March 5, 2010
Dear Mr. Liao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2- Mr. Liao
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Danner
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name: Wireless Body Temperature Monitor / Model: BTM-D1C, D, E, F, G and H.
Indication for Use:
The Wireless Body Temperature Monitor, model BTM-D1X series (model BTM-D1C, D, E, F, G and H), is a battery-operated electronic device with intended use of measuring human ear temperature precisely and continuously monitoring armpit temperature via wireless signal transmission of the measuring result. This device is reusable and is intended for ear temperature measurement as well as armpit temperature monitoring for persons over two years old. Model BTM-D1C has two functions, both ear temperature measurement and armpit temperature monitoring. The rest of models, BTM-D1D, E, F, G and H, have only the function of ampit temperature monitoring.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
1-10
Charles R. Hoppe
(Division Sign-Off) (7/vision of Anesthesiology, General Hospital Intection Control, Dental Devices
Page 1 of 1.
510(k) Number: K100226
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.