← Product Code [GEI](/productcode/GEI) · K100224

# ESTECH COBRA BIPOLSR II (K100224)

_Endoscopic Technologies, Inc. D/B/A Estech · GEI · Apr 7, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100224

## Device Facts

- **Applicant:** Endoscopic Technologies, Inc. D/B/A Estech
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Apr 7, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ESTECH Cobra Bipolar II System is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.

## Device Story

The ESTECH Cobra Bipolar II System is a surgical clamp-style electrosurgical device used to deliver RF energy to target tissue. The device operates in a bipolar mode, utilizing one jaw as an ablative electrode and the other as an indifferent electrode. It is intended for use by surgeons in general surgery settings to coagulate soft tissue and achieve hemostasis. The system features rotatable jaws and a single-use clamp design to improve physician convenience. The device is manually operated by the surgeon during procedures.

## Clinical Evidence

Bench testing only. Data demonstrated that mechanical aspects (joints) are adequate for use and electrode performance is equivalent to the predicate. Biocompatibility was verified per ISO 10993, and electrical safety was confirmed per IEC 60601-2-2.

## Technological Characteristics

Electrosurgical cutting and coagulation device; bipolar RF energy delivery; clamp-style form factor with rotatable jaws; single-use. Biocompatibility per ISO 10993; electrical safety per IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ESTECH Cobra Bipolar System ([K053100](/device/K053100.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Endoscopic Technologies, Inc.

APR - 7 2010 ESTECH Cobra Bipolar II Special Premarket Notification

K100224

## SECTION 5: Special 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

| Category                 | Comments                                                                                   |
|--------------------------|--------------------------------------------------------------------------------------------|
| Sponsor:                 | ESTECH, Inc.<br>2603 Camino Ramon<br>Suite 100<br>San Ramon, CA 94583<br>Tel: 925-543-2110 |
| Correspondent:           | Tamer Ibrahim<br>Vice President<br>ESTECH, Inc                                             |
| Contact Information:     | Tel: 925-543-2110<br>Fax: 925-866-7117                                                     |
| Device Common Name:      | Electrosurgical cutting and coagulation<br>device and accessories                          |
| Device Proprietary Name: | ESTECH Cobra Bipolar II System                                                             |
| Device Classification:   | Class II, GEI (21 CFR 878.4400)                                                            |

#### Device Information:

ﺬ

#### Predicate Device Information:

| Predicate Devices:                      | ESTECH Cobra Bipolar System (K053100)                             |
|-----------------------------------------|-------------------------------------------------------------------|
| Predicate Device Manufacturers:         | Endoscopic Technologies, Inc (aka<br>ESTECH)                      |
| Predicate Device Common Name:           | Electrosurgical cutting and coagulation<br>device and accessories |
| Predicate Device Classification:        | 21 CFR 878.4400                                                   |
| Predicate Device Classification Number: | Class II, GEI                                                     |

#### b. Date Summary Prepared

22 January 2010

#### c. Description of Device

The ESTECH Cobra Bipolar System is designed to guide RF energy to the target tissue via electrodes in the jaws of a surgical clamp style device. RF energy is delivered in a bipolar mode, with one jaw of the device as the ablative electrode and the other jaw as the indifferent electrode.

#### d. Intended Use

The ESTECH Cobra Bipolar II System is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.

{1}------------------------------------------------

#### Endoscopic Technologies, Inc.

ESTECH Cobra Bipolar II Svstem Special Premarket Notification

### e. Comparison to Predicate Device

The ESTECH Cobra Bipolar II System is identical in indications for use, technology, manufacture, packaging and sterilization to the predicate ESTECH Cobra Bipolar System (K053100).

There are small design differences between the application and predicate devices. These design differences (rotatable jaws, single use clamp) provide a convenience for the physician, but do not alter the performance profile seen in the predicate device.

ESTECH concludes that the ESTECH Cobra Bipolar II System is substantially equivalent to the predicate ESTECH Cobra Bipolar System.

#### f. Summary of Supporting Data

The bench testing supplied in Section 18 demonstrates that the joints and other mechanical aspects of the design of the ESTECH Cobra Bipolar II System are adequate for use. Additionally, the data demonstrates that the performance of the ESTECH Cobra Bipolar II System electrodes is equivalent to that of the predicate.

The patient contacting materials of the ESTECH Cobra Bipolar II System have been demonstrated to be biocompatible, in accordance with ISO 10993: Biological Evaluation of Medical Devices, and the safety of the electrical design is in conformance with the pertinent sections of IEC 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with healthcare, to the right. Encircling the caduceus are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

APR - 7 2010

Estech, Inc. % Coombs Medical Device Consulting, Inc. Mr. Craig Coombs, President - -1193 Sherman Street Alameda, California 94501

Re: K100224

Trade/Device Name: Estech Cobra™ Bipolar II System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: March 14, 2010 Received: March 18, 2010

Dear Mr. Coombs:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Craig Coombs, President

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for-Devices and-Radiological Health's (CDRH's) Office of Compliance -- Also, pleasenote the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

ely yours,

For

N. Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

ESTECH Cobra Bipolar II System Special Premarket Notification

# K100224 Section 4: Indications for Use

510(k) Number (if known):

4.77
K100224

ESTECH Cobra Bipolar II System Device Name:

Indications For Use:

The ESTECH Cobra Bipolar II System is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.

Prescription Use X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ZuihkeQal for num Page 1 of
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100224

CONFIDENTIAL

---

**Source:** [https://fda-staging.innolitics.com/device/K100224](https://fda-staging.innolitics.com/device/K100224)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).