The NON-MYDRIATIC RETINAL CAMERA TRC-NW8F is intended for use in capturing images of the retina and the anterior segment of the eye and presenting the data to the eye care professional, with the use of a mydriatic or without use of a mydriatic.
Device Story
Non-mydriatic retinal camera captures color and fluorescein angiography images of fundus oculi and anterior segment; non-contact operation. Device utilizes attached commercial digital single-lens reflex camera to record images onto memory cards, PCs, or external storage. Observation monitor displays live feed for clinician. Used in clinical settings by eye care professionals; images presented for subsequent diagnosis. Benefits include non-invasive visualization of ocular structures to aid clinical decision-making. No patient contact; no biocompatibility testing required.
Clinical Evidence
Bench testing only. Performance verified against ISO 15004-2:2007 (light hazard protection) and ISO 10940:1998 (resolving power and photographic angular field of view). Results met TOPCON self-standards and international requirements. No clinical data presented.
Technological Characteristics
Ophthalmic camera, AC-powered. Non-contact imaging system. Integrates commercial digital single-lens reflex camera. Connectivity via memory card, PC, or external storage. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 15004-1:2006, and ISO 15004-2:2007. No patient-contacting components.
Indications for Use
Indicated for capturing images of the retina and anterior segment of the eye in patients requiring ophthalmic examination. Contraindicated in patients with history of photosensitivity, those recently treated with photodynamic therapy (PDT), or those taking medications causing photosensitivity. Use meticulously in infants, patients with infectious diseases (e.g., epidemic kerato conjunctivitis), or those at high risk for optical radiation hazards (e.g., aphakic eyes, patients with diminished light sensitivity).
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
## 510(k) SUMMARY
# K100207
| 510(k) Owner: | TOPCON CORPORATION |
|---------------|--------------------------------|
| | 75-1 Hasunuma-cho, Itabashi-ku |
| | Tokyo, Japan 174-8580 |
| | Phone Number: +81-3-3558-2566 |
| | FAX Number: +81-3-3965-6532 |
FEB 12 2010
| U.S. Facility: | Topcon Medical Systems, Inc. |
|----------------|------------------------------|
| | 37 West Century Road |
| | Paramus, New Jersey 07652 |
| | Phone Number: (201) 599-5153 |
| | FAX Number: (201) 599-5240 |
| Contact person: | Takao Sugawara |
|-----------------|--------------------------------|
| | Manager |
| | TOPCON CORPORATION |
| | 75-1 Hasunuma-cho, Itabashi-ku |
| | Tokyo, Japan 174-8580 |
| | Phone Number: +81-3-3558-2566 |
| | FAX Number: +81-3-3965-6532 |
| Date: | November 2, 2009 |
|----------------------|---------------------------------------|
| Trade Name: | NON-MYDRIATIC RETINAL CAMERA TRC-NW8F |
| Common names: | Retinal Camera |
| Classification Name: | Camera, Ophthalmic, AC-Powered |
| | 21CFR886.1120 |
| Product Code: | HKI |
{1}------------------------------------------------
#### Identification of a Legally Marketed Predicate Device
The TRC-NW8F is substantially equivalent to the TRC-NW7SF MARKII (K090115). KOWA VX-10 (K043213) and TRC-NW200 (K041367). All of those predicate have same FDA Product Code HKI.
## General Description
This product is a retinal camera designed to observe, photograph or record the fundus oculi of a patient without coming into contact with the patient's eye and provide as an electronic image the obtained fundus oculi information for subsequent diagnosis. This product can take both color photography and fluorescein angiography. This product is equipped with an observation monitor used for observation purpose and display of a photographed image. This product uses attached commercial digital single-lens reflex camera to photograph or record the fundus oculi of a patient. A photographed image may be recorded on a commercial memory card built into an commercial digital single-lens reflex camera or a personal computer (hereinafter referred to as a PC) or commercial memory devices (flash memories, hard disc, etc.). A commercial digital printer is connected and can print the observed images and the photographed images of the fundus.
## Intended Use / Indication for Use
The NON-MYDRIATIC RETINAL CAMERA TRC-NW8F is intended for use in capturing images of the retina and the anterior segment of the eye and presenting the data to the eye care professional, with the use of a mydriatic or without use of a mydriatic.
#### Patient Profiles
This instrument has not been designed to apply to infants. Use this instrument meticulously for infants.
This instrument is contraindicated in the following patients.
- The patient who has a history of photosensitivity. .
- The patient who has just received the treatment of photodynamic therapy (PDT) (For . the prohibition period, refer to the package insert of the taken light-sensitive substance.
- . The patient who takes the medicine that can cause the photosensitivity as its side effect.
{2}------------------------------------------------
K106207
Use this instrument meticulously for the following patients.
- The patient who has epidemic kerato conjunctivitis, or any other infectious disease. .
- The patient who has taken the medicine that can cause the photosensitivity as its side . effect.
- The patient who is at high risk for the optical radiation hazard, such as aphakic eye, . infant, and the patient who diminishes the sensibility to light by fundus disease.
#### Performance Data
The maximum exposure has been demonstrated to be well below the accepted threshold limits set out in ISO 15004-2:2007. (See Attachment 3)
The resolving power and the photographic angular field of view as defined in ISO 10940:1998 "Ophthalmic instruments -- Fundus cameras" have been measured and the result meet the requirement value set out in TOPCON self standards. (See Attachment 5) In all instances, the NON-MYDRIATIC RETINAL CAMERA TRC-NW8F functioned as intended.
#### Basis of Substantial Equivalence
The TOPCON NON-MYDRIATIC RETINAL CAMERA TRC-NW8F is as safe and effective as TOPCON TRC-NW7SF MARKII (K090115), KOWA VX-10 (K043213) and TOPCON TRC-NW200 (K041367). The NON-MYDRIATIC RETINAL CAMERA TRC-NW8F and the predicated devices have the same intended use and similar indications, technological characteristics, and principles of operation. The minor technological differences between the NON-MYDRIATIC RETINAL CAMERA TRC-NW8F and its predicate devices raise no new issue of safety or effectiveness. Performance data demonstrate that the NON-MYDRIATIC RETINAL CAMERA TRC-NW8F is as safe and effectiveness as required in applicable international standards such as ISO 15004-2:2007 and so on. Thus, the NON-MYDRIATIC RETINAL CAMERA TRC-NW8F is substantially equivalent.
#### Standards for testing
TOPCON conducted several tests for the TOPCON Retinal Camera TRC-NW7SF MARK II to ascertain conformity to following standards.
- Medical Electrical Equipment Part 1: General Requirements for IEC 60601-1 Safety, 1988; Amendment 1, 1991; Amendment 2, 1995 Medical electrical equipment - Part 1-2: General requirements for IEC 60601-1-2 basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests, Edition 3:2007
3
، ﺋﯩ
{3}------------------------------------------------
K100207
4
- ISO 15004-1:2006 Ophthalmic instruments - Fundamental requirements and test Part 1: General requirements applicable to all methods ophthalmic instruments Ophthalmic Instruments - Fundamental requirements and test ISO 15004-2:2007 methods Part 2: Light hazard protection
No deviation or adaptation is made in the use of above mentioned standards.
See Attachment 1 for IEC 60601-1 Test Reports and IEC 60601-1-2 Test Reports.
See Attachment 2 for ISO 15004-1:2006 Test Reports.
See Attachment 3 for ISO 15004-2:2007 Test Reports.
See Attachment 4 for Form FDA 3654 of each standard.
The NON-MYDRIATIC RETINAL CAMERA TRC-NW8F has no component which contacts with blood, bodily fluid or mucous membrane. Therefore, Biocompatibility tests were not performed.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Topcon Corp. c/o Mr. Stefan Preiss TUV SUD America Inc 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891
FEB 1 2 2010
Re: K100207
Trade/Device Name: Non-Mydriatic Retinal Camera TRC-NW8F Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera, AC-Powered Regulatory Class: Class II Product Code: HKI Dated: January 22, 2010 Received: January 25, 2010
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Stefan Preiss
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
.gov/MedicalDevices/ResourcesforYourIndustry/default.htm
Sincerely yours,
Kesia Alexander for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
| 510(k) Number (if known): | K100207 |
|---------------------------|---------|
|---------------------------|---------|
NON-MYDRIATIC RETINAL CAMERA TRC-NW8F Device Name:
Indications for Use:
The NON-MYDRIATIC RETINAL CAMERA TRC-NW8F is intended for use in capturing images of the retina and the anterior segment of the eye and presenting the data to the eye care professional, with the use of a mydriatic or without use of a mydriatic.
X Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR
Over-The-Counter Use __ (21 C.F.R. 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE ---- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sen
(Division Sign-Off) Jivision of Ophthalmic, Neurological and Ear, Fiose and Throat Devices
Page __ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
) Number K100207
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.