← Product Code [GEX](/productcode/GEX) · K100194

# OPTICAL INTEGRITY LASERGUARD FIBERS (K100194)

_Optical Integrity, Inc. · GEX · May 20, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100194

## Device Facts

- **Applicant:** Optical Integrity, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** May 20, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Optical Integrity General Shaped Fibers with LaserGuard™ are intended to be used as an accessory to a medical laser. For specific applications, refer to the Instructions for Use documentation supplied with the laser system being used with the LaserGuard™ system.

## Device Story

Device is a surgical laser fiber waveguide with integrated LaserGuard™ thermal monitoring accessory. Fiber transmits laser light (532-2100 nm) from laser source to surgical site for tissue ablation. LaserGuard™ detects heat generated by fiber overheating (e.g., due to fiber breakage or tip contamination); upon detection, it automatically disengages the laser foot switch, placing the laser in standby mode, and triggers audible/visual alarms. Used in clinical settings by surgeons. Benefits include limiting collateral damage to endoscopes/handpieces and alerting surgeons to thermal runaway events, allowing for informed clinical intervention.

## Clinical Evidence

Bench testing only. Testing compared energy delivery of the subject device to the predicate device, confirming no difference in performance. Additional bench tests verified the LaserGuard™ functionality, specifically the ability to detect fiber breaks/overheating and trigger the laser standby mode and alarms.

## Technological Characteristics

Waveguide for laser light (532-2100 nm). Silica cores (145-940 micrometers); cladding of doped silica or polymer; outer buffer of Nylon or Tefzel. Distal tips: flat, orbed, or conical. Lengths: 2.5-5.0 meters. Accessory: electronic thermal detection circuit with audible/visual alarm and laser foot switch interlock. Sterile/non-sterile and single-use/reusable versions.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Optical Integrity, Inc. General Shaped Fiber ([K022338](/device/K022338.md))
- Dornier Midelas laser EPS mode

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a handwritten alphanumeric string. The string starts with the letter 'K', followed by the numbers '100194'. The numbers are written in a cursive style, and the overall image has a slightly rough, handwritten appearance.

# MAY 2 0 2010

## 510(k) Summary Prepared on 20 July 2009

| Submitted By: Optical Integrity, Inc.,<br>Location: 7500 McElvey Road, Panama City Beach, Florida, Zip 32408.<br>Contact: Ron Bowman at (850) 233-5312 ext. 209 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:<br>Common Name:<br>Classification name:                                                                                                             | Optical Integrity General Shaped Fiber with LaserGuard(tm)<br>surgical laser fiber<br>Laser surgical instrument for use in general and plastic surgery and in<br>dermatology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code:<br>CFR reference:                                                                                                                                 | GEX<br>21 CFR 878.4810                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate device:                                                                                                                                               | Optical Integrity, Inc. General Shaped Fiber, 510(k) number K022338<br>and Dornier Midelas laser EPS mode.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Description:<br>Technological<br>Characteristics:                                                                                                       | The product submitted herein is, in fact, the predicate device fitted with<br>a LaserGuard™ device as an accessory. The basic device is a wave<br>guide for laser light sources and will transmit light between 532 and<br>2100 nanometer wavelengths. The cladding of the fibers is either doped<br>silica or a polymer. The outer buffer material is either Nylon or Tefzel.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                                                                                 | The LASERGUARD™ accessory included with the submitted product is an<br>electronic device that will detect heat generated as a result of the fiber<br>overheating anywhere along the light path and will disengage the foot<br>switch of the laser thereby putting the laser in standby mode. The<br>LASERGUARD™ will also indicate by an audible alarm and a light that the<br>fiber has overheated.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                 | Both the predicate device and the submitted products are capable of<br>delivering laser radiation from a laser instrument to the surgical site with<br>minimal loss. Both deliver wavelengths from 532 to 2100 nanometers<br>as, for example, is generated by Ho:YAG or Nd:YAG lasers. The laser<br>energy is emitted through a flat, orbed or conical tip located at the distal<br>end of the fiber. Both the predicate and submitted products are offered<br>in a range of core sizes from 145 to 940 micrometer silica cores and are<br>from 2.5 to 5.0 meters in length. The products are supplied in sterile<br>and non-sterile as well as single-use and re-usable versions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indications for Use:<br>for both the<br>submitted product<br>and the predicate<br>device.                                                                       | The Optical Integrity General Shaped Fibers with LaserGuard™ are intended to<br>be used as an accessory to a medical laser. For specific applications, refer to<br>the Instructions for Use documentation supplied with the laser system being<br>used with the LaserGuard™ system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Comparison of<br>submitted product to<br>predicate device,<br>technological<br>characteristics:                                                                 | The submitted product, Optical Integrity General Shaped Fiber with<br>LASERGUARD™, is the predicate device fitted with an accessory capable<br>of detecting when the fiber material experiences a rapid rise in<br>temperature. This rapid rise in temperature occurs, for example, when<br>the fiber breaks as a result of a small bend radius. With the predicate<br>device, an operator must detect such a break by observing the changes<br>in energy delivery at the fiber tip and then respond to them by releasing<br>the foot or hand held switch in order to stop damage to the endoscope<br>or hand piece. With the LASERGUARD™ detector, the laser is<br>automatically put in standby, thereby limiting damage to endoscopes or<br>other fiber containing devices.<br>Additionally, if the fiber is used as a bare fiber for tissue ablation and the<br>tip of the fiber becomes contaminated and as a result the fiber heats up,<br>then the LASERGUARD™ will likewise, put the laser in standby and alert<br>the operator that a thermal runaway event has occurred. The surgeon<br>can then determine what course of action to take. That way, the<br>submitted device can assist the surgeon in delivering the appropriate<br>energy to ablate the desired tissue while minimizing collateral damage. |
| Summary of<br>Performance Data                                                                                                                                  | Bench testing demonstrated that Optical Integrity's General Shaped<br>Fiber with LASERGUARD™ performed the same as the predicate device,<br>Optical Integrity's General Shaped Fiber (without the LASERGUARD™<br>accessory). I.E. there was no difference in the energy delivered by the<br>undamaged fiber with or without the LASERGUARD™ accessory in place.<br>Additionally, the test results establish that the LASERGUARD™ will shut<br>down the laser, as intended, when a fiber break occurs and indicate<br>such by an audible alarm and an indicator light. (Test report is in<br>Section 18, Bench Tests)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

{1}------------------------------------------------

---

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines representing its wings and tail. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

#### MAY 2 0 2010

Optical Integrity, Inc. % Mr. Ronald E. Bowman 7500 McElvey Road Panama City Beach, Florida 32408

Re: K100194

Trade/Device Name: Optical Integrity General ShapedFiber with LaserGuard™ Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: May 17, 2010 Received: May 18, 2010

Dear Mr. Bowman: .

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

#### Page 2 - Mr. Ronald E. Bowman

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours

Sincerely yours,

for

Mark N. Melkerson

Mark N. Mel Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### PREMARKET NOTIFICATION INDICATIONS FOR USE STATEMENT

510(K) Number: K100194

Device Name: Optical Integrity General Shaped Fiber with LaserGuard™

Indications for Use:

The Optical Integrity General Shaped Fibers with LaserGuard™ are intended to be used as an accessory to a medical laser. For specific applications, refer to the Instructions for Use documentation supplied with the laser system being used with the LaserGuard™ system.

Concurrence of CDRH, Office of Evaluation (ODE)

Prescription Use _____________________________________________________________________________________________________________________________________________________________

_or Over-the-Counter Use _

Neil K. Oden for mxm

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100194

్లో

---

**Source:** [https://fda-staging.innolitics.com/device/K100194](https://fda-staging.innolitics.com/device/K100194)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
