MODEL MC-180MA MULTI-FREQUENCY BODY COMPOSITION ANALYZER
Applicant
Tanita Corp.
Product Code
MNW · Cardiovascular
Decision Date
Apr 15, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Body Composition Estimation
Multiple-regression algorithms
—
—
—
—
—
—
Indications for Use
The Tanita Body Composition Analyzer is indicated for use in the measurement of weight and impedance, and the estimation of body mass index (BMI), total body and segmental fat percent and weight, total body water percent and weight, intracellular and extracellular water weight, total body and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), and fat free mass (FFM), using BIA (Bioelectrical Impedance Analysis). The device is indicated for use for healthy children 5-17 years old and healthy adults with active, moderately active, to inactive lifestyles. For subjects 17 years old and younger, only fat % is displayed.
Device Story
Computer-operated body composition analyzer; utilizes bioelectrical impedance analysis (BIA). Inputs: body weight and bioelectrical impedance measured via multipolar surface electrodes. Operation: applies alternating current (5-500 kHz range, 90 microamperes); uses multiple-regression algorithms to transform impedance and weight data into body composition estimates (fat, muscle, bone, water, BMR). Output: visual display of body composition metrics. Used in clinical settings; operated by healthcare providers or technicians. Provides quantitative data to assist in health assessment and monitoring of body composition.
Clinical Evidence
Clinical testing verified the accuracy of multi-frequency bioelectrical impedance measurements used for the estimation of percent fat, fat-free mass, total body water, and extracellular water. No further clinical study details provided.
Technological Characteristics
Multi-frequency BIA analyzer; uses multipolar surface electrodes. Energy source: alternating current (5-500 kHz, 90 microamperes). Features computer-operated visual display. Software utilizes multiple-regression algorithms for body composition estimation. Standalone device.
Indications for Use
Indicated for healthy children 5-17 years and healthy adults (active to inactive) for measurement of weight/impedance and estimation of body composition metrics (BMI, fat, water, muscle, bone, BMR, FFM). For subjects ≤17 years, only fat % is displayed.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Tanita Model BC-418 Segmental Body Composition Analyzer (K033157)
Biospace Model InBody 3.0 Body Composition Analyzer (K042528)
Tanita Model SC-331 Body Composition Analyzer (K090479)
Submission Summary (Full Text)
{0}------------------------------------------------
## K100183 1 OF 2 PAGE
APR 1 5 2011
#### 510(k) Summary
The following information is provided in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92
- 1. Submitter: Tanita Corporation 14-2, 1-Chome, Maeno-cho, Itabashi-ku Tokyo 174-8630 Japan Telephone: +81 3 3968 2118 Fax: +81 3 3968 1550 E-mail: takeuchi@tanita.co.jp Data of summary: April 13, 2011
| 2. Name of the Device: | MC-180MA Multi-frequency Body Composition Analyzer |
|---------------------------|---------------------------------------------------------------------------------|
| Trade / Proprietary Name: | MC-180MA Multi-frequency Body Composition Analyzer |
| Common or Usual Name: | Body Composition Analyzer |
| Classification Name: | Analyzer, Body Composition<br>21 CFR §870.2770<br>Class II<br>Product Code: MNW |
- 3. Predicate devices to claim substantial equivalence:
- Tanita Model BC-418 Segmental Body Composition Analyzer-K033157 a.
- Biospace Model InBody 3.0 Body Composition Analyzer-K042528 b.
- Tanita Model SC-331 Body Composition Analyzer-K090479 c.
- 4. Description of the Device:
The Tanita Model MC-180MA Multi-frequency Body Composition Analyzer is a computer-operated body composition analyzer that utilizes bioelectrical impedance analysis (BIA) to estimate body composition of fat, muscle and bone, water compartments, and basal metabolic rate (BMR).
Actual measurements made by the MC-180MA include body weight and bioelectrical impedance. BIA is used to make estimates based upon these measured values.
- 5. Intended Use Statement:
The Tanita Body Composition Analyzer is indicated for use in the measurement of weight and impedance, and the estimation of body mass index (BMI), total body and
l
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K100183 PAGE 2 OF 2". The text is written in a handwritten style. The text appears to be a page number or document identifier.
segmental fat percent and weight, total body water percent and weight, intracellular and extracellular water weight, total body and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), and fat free mass (FFM), using BIA (Bioelectrical Impedance Analysis).
The device is indicated for use for healthy children 5-17 years old and healthy adults with active, moderately active, to inactive lifestyles.
For subjects 17 years old and younger, only fat % is displayed.
- 6. Summary of technological characteristics compared to predicate devices:
Both the Tanita Model MC-180MA and predicate devices (Item 3 above) are intended for use to measure weight and bioelectrical impedance, and estimate body composition (fat, muscle, bone and water) by BIA in healthy children and in healthy adults.
The technology underlying BIA estimation for the Model MC-180MA and predicate devices is the same. Design considerations-computer-operation with visual-display user interface; use of multiple-regression algorithms to predict body composition based upon body weight and segmental bioelectrical impedance measurements-and energysource considerations-multipolar surface electrodes; alternating current in the 5- to 500-kilohertz range at levels of 90 microamperes (Tanita models) to 250 microamperes (Biospace model)-are the same. Multiple-frequency measurement of bioelectrical impedence values used in estimation of percent fat, fat-free mass, total body water and extracellular water was verified in clinical testing.
Comparison of these intended-use and technological characteristics demonstrate that the Model MC-180MA is substantially equivalent to predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Tanita Corporation c/o Mr. Toshihiko Ishikawa Product Manager Tanita Corporation of America, Inc. 2625 South Clearbrook Drive ARLINGTON HEIGHTS IL 60005
APR 1 5 2011
Re: K100183
Trade/Device Name: Tanita MC-180MA Multi-frequency Body Composition Analyzer Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: April 13, 2011 Received: April 14, 2011
Dear Mr. Ishikawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{3}------------------------------------------------
#### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Hubert Humphrey
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive. Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
### 510(k) Number (if known): K100183
### Device Name: Tanita MC-180MA Multi-frequency Body Composition Analyzer
#### Indications for Use:
The Tanita Body Composition Analyzer is indicated for use in the measurement of weight and impedance, and the estimation of body mass index (BMI), total body and segmental fat percent and weight, total body water percent and weight, intracellular and extracellular water weight, total body and segmental muscle mass, physique rating, bone mass, visceral fat rating with healthy range, basal metabolic rate (BMR), and fat free mass (FFM), using BIA (Bioelectrical Impedance Analysis).
The device is indicated for use for healthy children 5–17 years old and healthy adults with active, moderately active, to inactive lifestyles.
For subjects 17 years old and younger, only fat % is displayed.
## (PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR §801.109) OR Over-the-counter _____________________________________________________________________________________________________________________________________________________________ ×
hory In Whang
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.