K100103 · Helena Laboratories · GGC · Dec 15, 2010 · Hematology
Device Facts
Record ID
K100103
Device Name
SPECIALTY ASSAYED CONTROL-2
Applicant
Helena Laboratories
Product Code
GGC · Hematology
Decision Date
Dec 15, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5425
Device Class
Class 2
Indications for Use
The S.A.C.-2 is an assayed abnormal control plasma intended for use in clinical laboratories when testing human plasma for the following coagulation deficiencies: Prothrombin time (PT), activated Partial Thromboplastin Time (aPTT), Factor II, V, VII, VIII, IX, X, XI, XII deficiencies, Fibrinogen, von Willebrand Factor antigen, Ristocetin cofactor, Protein C activity, and Protein S activity (total and free). The control material can be used with manual assays, electro-mechanical, or photo-optical clot detection methods.
Device Story
S.A.C.-2 is a lyophilized, buffered abnormal control plasma derived from human donors; used in clinical laboratories to verify performance of coagulation assays. Input: human plasma samples tested via manual, electro-mechanical, or photo-optical coagulation instruments. Device serves as a quality control material to ensure accuracy of clinical testing for coagulation deficiencies (PT, aPTT, factors, fibrinogen, vWF, Protein C/S). Healthcare providers use the control results to validate assay performance before reporting patient results. Benefits include ensuring reliability of diagnostic testing for coagulation disorders.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated per CLSI EP5-A2 using three operators, multiple reagent systems, and analyzers; all parameters achieved CV ≤ 15%. Stability testing confirmed analytes remain within assigned ranges and within 20% of 0-hour values after 4.5 hours at 2-8°C.
Technological Characteristics
Lyophilized, buffered human plasma collected in 4% sodium citrate. Reconstituted control is stable for 4 hours at 2-8°C. Compatible with manual, mechanical, and photo-optical coagulation instrumentation. No electronic or software components.
Indications for Use
Indicated for use in clinical laboratories as an abnormal control plasma for human plasma coagulation testing, including PT, aPTT, specific factor deficiencies, fibrinogen, vWF, and Protein C/S activity. Prescription use only.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
Predicate Devices
Specialty Assayed Control-2 (S.A.C.-2) (k941872)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k100103
B. Purpose for Submission:
Change of manufacturing location for a currently marketed device
C. Measurand:
Abnormal control for Prothrombin Time (PT), activated Partial Thromboplastin Time (aPTT), Factors II, V, VII, X, VIII, IX, XI, XII, Fibrinogen, von Willebrand Factor Antigen, Ristocetin Cofactor, Protein C, Protein S (total and free)
D. Type of Test:
Hemostasis quality control
E. Applicant:
Helena Laboratories Corporation
F. Proprietary and Established Names:
Specialty Assayed Control-2 (S.A.C.-2)
G. Regulatory Information:
1. Regulation section:
21 CFR 864.5425, Multipurpose system for in vitro coagulation studies
2. Classification:
II
3. Product code:
GGC, Control, plasma, abnormal
4. Panel:
81 Hematology
H. Intended Use:
1. Intended use(s):
The S.A.C.-2 is an assayed abnormal control plasma intended for use in clinical laboratories when testing human plasma for the following coagulation deficiencies: Prothrombin time (PT), activated Partial Thromboplastin Time (aPTT), Factor II, V, VII, VIII, IX, X, XI, XII deficiencies, Fibrinogen, von Willebrand Factor antigen, Ristocetin cofactor, Protein C activity, and Protein S activity (total and free). The control material can be used with manual assays, electro-mechanical, or photo-optical clot detection methods.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
Mechanical and photo-optical instrumentation
I. Device Description:
The S.A.C.-2 is an assayed abnormal control plasma from normal human donors intended for use in clinical laboratories when testing human plasma for the following coagulation deficiencies: Prothrombin time (PT), activated Partial Thromboplastin
{1}
Time (aPTT), Factor II, V, VII, VIII, IX, X, XI, XII deficiencies, Fibrinogen, von Willebrand Factor antigen, Ristocetin cofactor, Protein C activity, and Protein S activity (total and free).
The control is prepared from a normal pool of human plasma collected in $4\%$ sodium citrate. The pool has 0.03M Hepes buffer and is lyophilized to ensure stability of all plasma coagulation factors. All factors are adjusted to be in the abnormal range.
# J. Substantial Equivalence Information:
1. Predicate device name(s): Specialty Assayed Control-2 (S.A.C.-2)
2. Predicate 510(k) number(s): k941872
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | | Device | Predicate |
| Intended Use | The S.A.C.-2 is an assayed abnormal control plasma intended for use in clinical laboratories when testing human plasma for the following coagulation deficiencies: Prothrombin time (PT), activated Partial Thromboplastin Time (aPTT), Factor II, V, VII, VIII, IX, X, XI, XII deficiencies, Fibrinogen, von Willebrand Factor antigen, Ristocetin cofactor, Protein C activity, and Protein S activity (total and free). The control material can be used with manual assays, electro-mechanical, or photo-optical clot detection methods. | | The S.A.C.-2 is an assayed abnormal control plasma which may be used in many facets of testing in the coagulation laboratory |
| Analytes | Prothrombin time (PT), activated Partial Thromboplastin time (aPTT), Factor II, V, VII, VIII, IX, X, XI, XII deficiencies, fibrinogen, von Willebrand Factor antigen, Ristocetin cofactor, Protein C activity, and Protein S activity (total and free). | | Same |
| Reagent | Lyophilized, buffered human plasma collected in 4% sodium citrate | | Same |
| Stability (reconstituted) | 4 hours at 2 - 8°C | | Same |
| Packaging | 10 x 1 mL vials | | Same |
| Instrumentation | Mechanical and photo-optical instruments | | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Analytes (Kinetic) | Analytes no longer available in control F VIII, X, plasminogen, antithrombin III, protein C, a2-antiplasmin | Included in control F VIII, X, plasminogen, antithrombin III, protein C, a2-antiplasmin |
| Analytes (Immunological RID) | Analytes no longer available in control Plasminogen, antithrombin III | Included in control Plasminogen, antithrombin III |
{2}
3
K. Standard/Guidance Document Referenced. (if applicable):
CLSI, EP5-A2-Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guidelines-Second Edition
L. Test Principle:
The control plasma may be used when performing tests on manual assays, mechanical or photo- optical coagulation instruments in conjunction with suitable commercial reagents.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed at Helena Laboratories with three operators, using various reagent systems and analyzers. Duplicate determinations of all analytes were made on five vials from four lots of control, as described in CLSI, EP5-A2. All parameters fell within acceptance criteria (CV of ≤ 15%).
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Assays are acceptable if the mean values are as follows:
| Assay | Mean Value |
| --- | --- |
| Prothrombin Time (PT) | ≥14.0 seconds |
| Activated Partial Thromboplastin Time (aPTT) | >35.0 seconds |
| Factor Assays | ≤50 % |
| Fibrinogen | ≤ 140 mg/dL |
| Rocket von Willebrand Factor Antigen | ≤50% |
| Rocket Protein C | ≤50% |
| Rocket Protein S | ≤50% |
| Rocket Protein S (Free) | ≤50% |
| Ristocetin Cofactor | ≤50% |
| ELISA Protein C | ≤50% |
| ELISA Protein S, S (Free), Monoclonal Free S | ≤50% |
| ELISA von Willebrand Factor Antigen | ≤50% |
Stability studies were performed on each analyte using duplicate determinations on three vials from each of three lots of control. Each analyte was tested at 0 hours and at 4.5 hours with storage at 2 to 8°C. Acceptance criteria for the studies was defined as falling within assigned range and for the 4.5 hour data point to be within 20% of the 0 hour data point. All parameters fell within acceptable performance.
d. Detection limit
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
{3}
N. Proposed Labeling:
The labeling is sufficient and satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.