← Product Code [IYE](/productcode/IYE) · K100096

# LINAC COLLIMATOR ASSEMBLY HOUSING (LCAH) MODEL LCAHART (K100096)

_Integra Radionics, Inc. · IYE · Apr 22, 2010 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K100096

## Device Facts

- **Applicant:** Integra Radionics, Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Apr 22, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Integra Radionics™ Linac Collimator Assembly Housing (LCAHART) is indicated to be mounted on the Siemens Artiste MV linac with MLC-160 Accessory Hollder as an Accessory tray to attach Integra Radionics collimators for treatment modalities such as stereotactic radiosurgery (SRS) and radiotherapy (SRT).

## Device Story

The Linac Collimator Assembly Housing (LCAH) is a mechanical accessory tray designed for the Siemens MLC-160 LINAC Accessory Holder. It consists of a sliding tray, baseplate, and collimator tube. The device functions as a mounting interface, allowing medical physicists, radiation oncologists, and technicians to attach Integra Radionics radiation beam shaping collimators to the Siemens ARTISTE LINAC. By positioning and aligning these collimators with the LINAC's internal radiation source, the device facilitates precise beam shaping during radiosurgery and radiation therapy procedures. It is a purely mechanical component used in clinical radiation therapy environments.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Mechanical accessory tray assembly comprising a sliding tray, baseplate, and collimator tube. Designed for physical integration with the Siemens MLC-160 LINAC Accessory Holder. No electronic, software, or energy-emitting components.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- XKnife Radiosurgery System ([K912630](/device/K912630.md))
- ARTISTE MV SA ([K072485](/device/K072485.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K/00096

Confidential

Integra Radionics, Inc. Premarket Notification Traditional 510(k) Linac Collimator Assembly Housing (LCAH)

# Linac Collimator Assembly Housing (LCAH) 510(k) Summary

Submitter's name and address:

APR 2 2 2010

Integra Radionics, Inc. 22 Terry Avenue Burlington, MA 01803 USA

#### Contact person and telephone number:

Helder A. Sousa Regulatory Affairs Project Manager Integra Radionics, Inc. 22 Terry Avenue Burlington, MA 01803 USA Phone: (781) 565-1235 Fax: (781) 238-0645

## Date prepared:

December 11, 2009

#### Name of device:

| Trade Name:          | Linac Collimator Assembly Housing (LCAH) |
|----------------------|------------------------------------------|
| Common Name:         | Collimator Assembly                      |
| Classification Name: | Accelerator, Linear, Medical             |
| Regulation Number:   | 21 CFR 892.5050                          |
| Product Code:        | IYE                                      |

## Substantial Equivalence:

The Linac Collimator Assembly Housing (LCAH) is substantially equivalent in function and intended use with the XKnife Radiosurgery System (K912630) and the ARTISTE MV SA (K072485).

#### Indications Use:

The Integra Radionics™ Linac Collimator Assembly Housing (LCAH) is intended to be mounted as an Accessory tray for a LINAC to attach Integra Radionics collimators.

### Device Description:

...

The Linac Collimator Assembly Housing (LCAH) consisting of a collimator sliding tray, collimator baseplate, and collimator tube. It is designed as a slide-in attachment for the Siemens MLC-160 LINAC Accessory Holder. The LCAH positions and aligns Radionics radiation beam shaping collimators to the Siemens ARTISTE LINAC (K072485) internal collimator and radiation source during radiosurgery and radiation therapy procedures.

{1}------------------------------------------------

Intended Users of the LCAH are medical physicists, radiation oncologists, and appropriate radiation technicians.

## Conclusion:

The Linac Collimator Assembly Housing (LCAH) is substantially equivalent to the XKnife Radiosurgery System (K912630) and the Siemens ARTISTE MV SA (K072485). The Linac Collimator Assembly Housing (LCAH) is similar to the predicate devices in the intended use, the fundamental scientific technology of the device, and does not raise new issues of safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus, which is a symbol often associated with healthcare, overlaid on a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mr. Helder A. Sousa Regulatory Affairs Project Manager Integra Radionics, Inc. 22 Terry Avenue BURLINGTON MA 01803

APR 2 2 2010

Re: K100096

Trade/Device Name: Linac Collimator Assembly Housing (LCAHART) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: March 15, 2010 Received: March 16, 2010

Dear Mr. Sousa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 200). This letter will allow you to begin marketing your device as described in your Section 510(l) memarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tolloffre ne (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Donald J. Trump

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K100096

Device Name:

Linac Collimator Assembly Housing (LCAHART)

Indications For Use:

The Integra Radionics™ Linac Collimator Assembly Housing (LCAHART) is indicated to be mounted on the Siemens Artiste MV linac with MLC-160 Accessory Hollder as an Accessory tray to attach Integra Radionics collimators for treatment modalities such as stereotactic radiosurgery (SRS) and radiotherapy (SRT).

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Blod Becker

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K100096

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K100096](https://fda-staging.innolitics.com/device/K100096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
