NELLCOR COMPATIBLE REUSABLE SPO2 SOFT-FINGER SENSOR, MODEL T100A-090103
K100077 · Solaris Medical Technology, Inc. · DQA · Oct 7, 2010 · Cardiovascular
Device Facts
Record ID
K100077
Device Name
NELLCOR COMPATIBLE REUSABLE SPO2 SOFT-FINGER SENSOR, MODEL T100A-090103
Applicant
Solaris Medical Technology, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Oct 7, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
When used with a compatible patient monitor or a pulse oximeter device, Solaris Medical Technology, Inc. reusable & disposable SpO2 sensors are intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO2) and pulse rate monitoring. Solaris Medical Technology, Inc. reusable multi-patient use SpO2 Soft Sensors, reusable multi-patient use SpO2 Finger Sensors, and disposable single patient use SpO2 Soft-finger Sensors are for use with adult/pediatric patients weighing greater than 40kg. Solaris Medical Technology, Inc. disposable single patient use SpO2 Adhesive Sensors are for use with adult patients weighing greater than 40kg, pediatric patients weighing 10 - 40 kg, and infant (non-neonatal) patients weighing 3 - 15 kg. Prescription device.
Device Story
Solaris SpO2 sensors are non-invasive accessories for pulse oximeters; they measure functional arterial oxygen saturation and pulse rate. The sensor housing contains two LEDs (light sources) and one photodiode (detector). The device illuminates arterially perfused tissue with two wavelengths of light; the photodiode measures time-varying absorbance to determine oxygen saturation. Sensors are available in reusable (soft finger, finger clip) and disposable (soft-digit, adhesive) configurations. Used in clinical settings by healthcare professionals; the sensor connects to a compatible patient monitor. The monitor processes the signal to display SpO2 and pulse rate, aiding clinicians in patient monitoring and decision-making.
Clinical Evidence
Clinical testing performed under an institutionally approved protocol with informed consent. Performance validated under controlled hypoxia conditions compared to arterial oxygen saturation determined by co-oximetry. Results confirm accuracy claims over the 70% - 100% SpO2 range.
Technological Characteristics
Materials include silicone rubber and medical tape laminate. Sensing principle: two-wavelength LED illumination and photodiode detection. Form factors: reusable soft finger, reusable finger clip, disposable soft-digit, and disposable adhesive. Connectivity: wired connection to compatible pulse oximeter monitors. Sterilization: not specified.
Indications for Use
Indicated for continuous, non-invasive SpO2 and pulse rate monitoring. Reusable sensors and disposable soft-finger sensors for adult/pediatric patients >40kg. Disposable adhesive sensors for adult patients >40kg, pediatric patients 10-40kg, and infants 3-15kg (non-neonatal).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Solaris Medical Technology, Inc. The logo features a stylized graphic element resembling a waveform or mountain range to the left of the word "SOLARIS", which is written in a bold, sans-serif font. A thin line underlines the word "SOLARIS". Below the line is the text "Solaris Medical Technology, Inc." in a smaller, serif font.
OCT 7 2010
12100027
Page 1 of 4
# 510(k) Summary of Safety and Effectiveness
(a)
| Date of Preparation: | | October 7, 2010 | | |
|----------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| (1) | Submitted by: | | Solaris Medical Technology, Inc. (China)<br>6/F Building 9, 30th District, Keji Road,<br>Science Park, Nanshan District Shenzhen,<br>Guangdong, 518057 CHINA | |
| | Submitter Contact: | | Rachel Cheng<br>Vice President, Business Development<br>Solaris Medical Technology Inc. (USA)<br>Phone: +1 (650) 588-3980<br>Fax: +1 (650) 588-3988<br>rcheng@solarismedtech.com | |
| (2) | Trade Name: | | Solaris Medical Technology, Inc. Reusable & Disposable<br>SpO2 Sensors | |
| | Common Name: | | Pulse Oximeter Sensor | |
| | Classification Name: | | Oximeter | |
| | Classification Regulation: | | 21 CFR §870.2700 | |
| | Product Code: | | DQA | |
| | Class: | | Class II | |
| (3) | Predicate Device(s): | | Substantial Equivalence to: | |
| | K Number | Model | Manufacturer | |
| | K993637* | Device Name MODEL 2500A<br>Palmsat Pulse Oximeter (with model<br>8000AA Finger Clip Sensor, 7000A,<br>7000P, 7000I Flexi-Form II<br>Disposable sensors specified) | Nellcor Puritan Bennett, Inc. | |
1
6/F Building 9, 30" District, Keji Road, Science Park, Nanshan District, Shenzhen Guangdong, 518057 CHINA
Tel: +86 755 26525478 Fax: +86 755 26525386 www.szsolaris.com
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Solaris Medical Technology, Inc. The logo features a stylized waveform graphic to the left of the word "SOLARIS" in bold, capitalized letters. Below the waveform and the word "SOLARIS" is the text "Solaris Medical Technology, Inc." in a smaller font size.
Page 2 of 4
| K Number | Model | Manufacturer |
|----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| K050056* | Device Name MODEL 2500A<br>Palmsat Pulse Oximeter (with model<br>8000AA Finger Clip Sensor, 7000A,<br>7000P, 70001 Flexi-Form II<br>Disposable sensors specified) | Nonin Medical, Inc. |
| K000822* | Philips/Agilent component<br>monitoring system, Agilent 24/26<br>(with model M1191A Reusable<br>SpO2 Sensor specified as accessory) | Agilent Technologies, GmbH<br>(formerly HP) |
| K042306 | Philips M1131A SpO2 Disposable<br>SpO2 Sensor | Philips Medizinsysteme<br>Boeblingen GmbH (formerly HP) |
| K083705* | BCI WW1020 SPECTRO2 Pulse<br>Oximeter (with model 3044 Reusable<br>Finger Clip Sensor and model 1300<br>Disposable Finger Sensor, Adult<br>specified as accessories) | Smiths Medical PM, Inc. (formerly<br>SIMS BCI) |
* Please note these 510(k) numbers represent monitors containing a pulse oximeter in which the predicate sensors were included as accessories.
Reason for Submission: New Device(s)
#### Device Description: (4)
Solaris Compatible Reusable and Disposable SpO2 Sensors (Solaris Sensors) are compatible sensors for use with major types of patient monitors and oximeter devices as listed above.
Solaris Sensors employ non-invasive electro-optical means to determine the light absorption of functional arterial hemoglobin. The sensor contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The LED and photodiode are contained in the sensor housing.
Four types of sensor housings are described in this submission:
- Reusable soft finger sensor comprised of an integrated silicone rubber 트 tube.
- Reusable finger clip sensor with rigid halves and silicone pads 트
- Disposable soft-digit sensor with flexible silicone housing I
- Disposable adhesive sensors constructed of a medical tape laminate E
2
<sup>6/</sup>F Building 9, 30" District, Keji Road, Science Park, Nanshan District, Shenzhen Guangdong, 518057 CHINA Tel: +86 755 26525478 Fax: +86 755 26525386 www.szsolaris.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Solaris Medical Technology, Inc. The logo features a stylized mountain range graphic connected to the word "SOLARIS" in bold, capitalized letters. Below the logo is the text "Solaris Medical Technology, Inc." in a smaller font size.
Page 3 of 4
Each sensor has unique labeling and specifications designed for compatibility with the specific monitor manufacturer (Nellcor, Nonin, HP/Philips, BCI).
### Intended use: (2)
When used with a compatible patient monitor or a pulse oximeter device, Solaris Medical Technology, Inc. reusable & disposable SpO2 sensors are intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO2) and pulse rate monitoring.
Solaris Medical Technology, Inc. reusable multi-patient use SpO2 Soft Sensors, reusable multi-patient use SpO2 Finger Sensors, and disposable single patient use SpO2 Soft-finger Sensors are for use with adult/pediatric patients weighing greater than 40kg.
Solaris Medical Technology, Inc. disposable single patient use SpO2 Adhesive Sensors are for use with adult patients weighing greater than 40kg, pediatric patients weighing 10 - 40 kg, and infant (non-neonatal) patients weighing 3 - 15 kg.
Prescription device.
#### Technology Comparison: (6)
Solaris Sensors employ the same technological characteristics as the predicate devices to determine arterial oxygen saturation: arterially perfused tissue is illuminated sequentially by two wavelengths of LEDs, and the time varying absorbance of the tissue is measured by a photodetector.
This method is characteristic of all sensors that are the subject of this submission as well as the predicate devices.
#### Non-Clinical Performance Tests: (1) (b)
### Cleaning Instruction Testing
Solaris Sensors were tested in accordance with internal protocols to ensure that the cleaning instructions do not damage sensor labeling or degrade the material.
<sup>6/</sup>F Building 9, 30" District, Keji Road, Science Park, Nanshan District, Shenzhen Guangdong, 518057 CHINA lsthol, Roj. Roud, Coac. +86 755 26525386 www.szsolaris.com
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Solaris Medical Technology, Inc. The logo features a stylized graphic element resembling a mountain range or a waveform, with the word "SOLARIS" in bold, sans-serif font to the right of the graphic. A thin line underlines the word "SOLARIS". Below the logo is the text "Solaris Medical Technology, Inc." in a smaller font size.
Page 4 of 4
### Biocompatibility
Sensor patient contact materials meet applicable standards for biocompatibility.
### Electrical Safety and EMC Testing
The sensors were tested in accordance with current applicable standards for medical device Electrical Safety and Electromagnetic Compatibility. Test results indicated that the sensors comply with the stated clauses.
### Pulse Rate Accuracy Testing
The sensors were tested for pulse rate accuracy with a listed SpO2 simulator with predicate device monitors. The sensors met their specified pulse accuracy.
- (2) Clinical Testing:
Solaris Sensors were clinically tested to validate the performance and accuracy of the Solaris Reusable and Disposable sensors under controlled hypoxia versus arterial oxygen saturation as determined by co-oximetry. All testing was performed under an institutionally approved protocol with subject informed consent.
Clinical test results indicated that the sensors meet the stated accuracy claims over the 70 % - 100 % range.
#### Conclusions: (3)
Based upon device evaluations, predicate comparisons, and performance testing results, Solaris Reusable and Disposable SpO2 Sensors are substantially equivalent to the predicate devices.
6/F Building 9, 30" District, Keji Road, Science Park, Nanshan District, Shenzhen Guangdong, 518057 CHINA Tel: +86 755 26525478 Fax: +86 755 26525386 www.szsolaris.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a bird-like figure, composed of three curved lines.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Rachel Cheng Vice President, Business Development Solaris Medical Technology, Incorporated 400 Oyster Point Boulevard, Suite 534 South San Francisco, California 94080
OCT 7 2010
Re: K100077
Trade/Device Name: Solaris Reusable & Disposable SpO2 Sensors Regulation Number: 21 CFR 870.2700 Regulation Name: Accessory to Pulse Oximeter Regulatory Class: II Product Code: DQA Dated: October 2, 2010 Received: October 4, 2010
Dear Ms. Cheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.
{5}------------------------------------------------
Page 2- Ms. Cheng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Solaris Medical Technology, Inc. Reusable & Disposable SpO2 Sensors Device Name:
Indications for use:
When used with a compatible patient monitor or a pulse oximeter device, Solaris Medical Technology, Inc. reusable & disposable SpO2 sensors are intended to be used for continuous, non-invasive functional arterial oxygen saturation (SpO2) and pulse rate monitoring.
Solaris Medical Technology, Inc. reusable multi-patient use SpO2 Soft Sensors, reusable multipatient use SpO₂ Finger Sensors, and disposable si ngle patient use SpO₂ Soft-finger Sensors are for use with adult/pediatric patients weighing greater than 40kg.
Solaris Medical Technology, Inc. disposable single patient use SpO₂ Adhesive Sensors are for use with adult patients weighing greater than 4 0kg, pediatric patients weighing 10 – 40 kg, and infant (non-neonatal) patients weighing 3 - 15 kg.
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
× Prescription Use (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANO THER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schuttertins
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control. Dental Devices
510(k) Number: k100007 Page 1 of
C-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.