INSTANT-VIEW MILTI-DRUG OF ABUSE URINE TEST (PANEL, CUP), INSTANT-VERDICT MULTI-DRUG OF ABUSE URINE TEST (PANEL, CUP)
Device Facts
| Record ID | K100051 |
|---|---|
| Device Name | INSTANT-VIEW MILTI-DRUG OF ABUSE URINE TEST (PANEL, CUP), INSTANT-VERDICT MULTI-DRUG OF ABUSE URINE TEST (PANEL, CUP) |
| Applicant | Alfa Scientific Designs, Inc. |
| Product Code | DJG · Clinical Toxicology |
| Decision Date | Jul 6, 2010 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 862.3650 |
| Device Class | Class 2 |
Indications for Use
The proposed drug of abuse device is intended for use in an OTC setting as an IVD screening test for any single one, or combination of, the following three substances in urine: Analyte Cutoff Morphine 300 ng/ml Oxycodone 100 ng/ml Oxycodone 300 ng/ml This test provides only a preliminary result. The results provided by this device indicate whether the drug or drug metabolite may be present. A positive result from the device is considered to be a presumptive result and should never be interpreted as final without laboratory confirmation.
Device Story
Lateral flow immunoassay; detects Morphine/Opiates and Oxycodone in urine; competitive binding principle. Device format: cup or cassette containing 1-3 test strips. Each strip features colloidal gold-labeled anti-drug antibodies and nitrocellulose membrane with test/control lines. User-operated (home use); sample applied to device; visual interpretation of colored bands. Presence of drug above cutoff inhibits test line development; control line confirms migration. Provides preliminary results; requires GC/MS or LC/MS confirmation. Benefits: rapid, accessible drug screening for lay users.
Clinical Evidence
Bench testing only; no clinical diagnostic data. OTC accuracy studies conducted at three sites with 264 total participants (136 for cup, 128 for cassette). Participants (ages 18-72) tested blind-labeled urine samples spiked with drugs at various concentrations relative to cutoffs. Results showed high agreement with GC/MS (97.16%-98.78% accuracy for cup; 97.7%-98.7% for cassette). Questionnaires confirmed ease of use and interpretation.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding. Components: colloidal gold-labeled anti-drug antibodies, nitrocellulose membrane with conjugated drug antigen. Format: cup or cassette. Built-in control line (C line). Temperature strip included in cup format. No electronic components or software.
Indications for Use
Indicated for home use by lay users for qualitative detection of Morphine/Opiates (300 ng/mL), Oxycodone (100 ng/mL), and Oxycodone (300 ng/mL) in human urine.
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
- INSTANT-VIEW® Multi-Drugs Urine Test (K063545)
- INSTANT-VIEW® Oxycodone Urine Test (K063545)
- INSTANT-VIEW® Morphine (300) Urine Test (K060527)