K093895 · Continuity Health Solutions · DRG · May 20, 2010 · Cardiovascular
Device Facts
Record ID
K093895
Device Name
ALLY PLATFORM SYSTEM
Applicant
Continuity Health Solutions
Product Code
DRG · Cardiovascular
Decision Date
May 20, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
The ALLY™ Platform System is a multi-platform medical device that is an accessory to approved 510k devices and is intended to be used as a means of receiving, storing, and transmitting patient biometric data including vital signs measurements. The ALLY™ Platform System is to be used upon prescription and under the supervision of an authorized healthcare provider. The ALLY™ Platform System will send information to a remote database to be accessed by authorized clients, such as healthcare professionals or patient family members. Electronic data consisting of patient vital sign information, educational material, reminders, questionnaires, and patient responses are exchanged over a secure network connection. Information may be communicated via standard telephone lines, high speed internet connections, serial port interfaces, or Bluetooth® radio transceivers. The ALLY™ Platform is a multi-platform application that is not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts. The ALLY™ Platform is not intended to act as an emergency response system. All patient medical diagnosis and treatment are to be performed by an appropriate health care professional.
Device Story
ALLY Platform System acts as a communication bridge for remote patient monitoring; collects biometric data (blood pressure, pulse, weight) from legally marketed external diagnostic peripherals via home unit; transmits data over telephone lines, high-speed internet, or Bluetooth to a central server; healthcare providers access historical data via secure web portal; system facilitates exchange of vital signs, educational materials, and questionnaires; not intended for real-time alerts, automated diagnostics, or emergency response; clinical decision-making remains with the healthcare professional; benefits include improved access to patient data for out-of-hospital care management.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
System consists of a home unit (RTX 337X), external diagnostic peripherals, and a server application. Connectivity via telephone lines, high-speed internet, serial ports, and Bluetooth. Software is classified as Moderate Level of Concern. No patient contact. Operates as a communication accessory for data transmission.
Indications for Use
Indicated for patients requiring remote monitoring of biometric data and vital signs, prescribed by and under the supervision of an authorized healthcare provider.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21CFR 807.92)
#### I. DATE PREPARED: 12/14/2009
MAY 2 0 2010
- II. SUBMITTER:
Continuity Health Solutions 9304 Forest Lane, Suite N272 Dallas, TX 75243 Telephone: 866.517.3135 Fax: 214.553.8946
### III. CONTACT PERSON:
Bryan Poteet, CEO Telephone: 866.517.3135
#### IV. DEVICE NAME:
Trade/Proprietary Name: Common/Usual Name: Classification Name:
ALLYTM Platform System Telehealth System Physiological Signal Transmitters and Receivers
#### V. DEVICE CLASSIFICATION
Class II CFR section: 870.2910 Physiological Signal Transmitters and Receivers Product code: DRG Panel: Cardiology
#### VI. PREDICATE DEVICES:
K090801 Electronic House Call System Class II Decision Date: 04/08/2009
K063612 Health hero Network Class II Decision Date: 12/29/2006
{1}------------------------------------------------
K093895
p2/3
#### VII. SUBSTANTIAL EQUIVALENCE CONCLUSIONS:
Continuity Health Solutions concludes that the intended use for the Continuity Health Solutions ALLYTM Platform System is the same as that of the predicate devices, and that the technological characteristics demonstrate that they are equivalent to the predicate device. A comparison of the technological characteristics of the predicate and legally marketed devices available has been performed.
Thus, this premarket notification has demonstrated substantial equivalence.
#### VIII. DEVICE DESCRIPTION AND INTENDED USE:
### Device Description:
ALLY™ is comprised of a home unit, legally marketed diagnostic peripherals, and a server. The home unit is a communications device that collects data from the diagnostic peripherals and. The patient in the home and transmits the data to the server over the home's existing telephone line or other identified method (existing high-speed connection in patients' home). The diagnostic peripherals comprise a wide range of legally marketed medical electronic devices capable of measuring patient parameters such as blood pressure, pulse rate, weight, etc. The server application responds to calls from the home units, collects the data, and then compares the data to preset thresholds. The healthcare practitioners and privileged users access the server through a secure website.
# ALLYTM Care Portal
The ALLY TM Care Portal is intended to be used as an accessory to communication tools, such as the ALLYTM Home Appliance (RTX 337X). The ALLY TM Care Portal enables healthcare providers to request, receive, and review historical patient information. It is intended to be used in combination with a variety of external devices.
# ALLYTM Home Appliance (RTX 337X)
The ALLYTM Home Appliance (RTX 337X) is for use in non-clinical settings (such as the home), as an accessory device that is intended to be a communication tool to enable healthcare providers to receive historical patient information. It is intended to be used in combination with a variety of external devices.
The ALLY™ Home Appliance (RTX 337X) serves as the remote communication link between compatible external devices, and the compatible healthcare facility at another location. The healthcare facility could be at a disease management center or with the healthcare/wellness provider or other out of hospital caregivers.
{2}------------------------------------------------
### Intended Use:
The ALLY™ Platform System is a multi-platform medical device that is an accessory to approved 510k devices and is intended to be used as a means of receiving, storing, and transmitting patient biometric data including vital signs measurements. The ALLY™ Platform System is to be used upon prescription and under the supervision of an authorized healthcare provider.
The ALL YTM Platform System will send information to a remote database to be accessed by authorized clients, such as healthcare professionals or patient family members. Electronic data consisting of patient vital sign information, educational material, reminders, questionnaires, and patient responses are exchanged over a secure network connection. Information may be communicated via standard telephone lines, high speed internet connections, serial port interfaces, or Bluetooth® radio transceivers.
The ALLYM Platform is a multi-platform application that is not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts. The ALLY™ Platform is not intended to act as an emergency response system. All patient medical diagnosis and treatment are to be performed by an appropriate health care professional.
#### IX. SAFETY INFORMATION:
The ALLYTM Platform System has no patient contact and is utilized only by trained professionals. Trained professionals allow sufficient review to afford identification and intervention in the event of a malfunction have evaluated the output of the device. This will allow sufficient review to afford identification and intervention in the event of a malfunction.
#### X. CONCLUSION:
Continuity Health Solutions believes sufficient information is included to reach a determination of substantial equivalence. We conclude that the subject device is as safe and effective including the component and accessory devices.
As a result of this analysis and the risk reduction activities, and has reviewed the guidance document for the content of premarket submissions for software contained in medical device, we conclude that there is no software or hardware component in the ALLY TM Platform System that would be expected to result in injury or death to a patient due to a failure or latent design flaw. Thus, the "Level of Concern" of the ALLY™ Platform System product is "Moderate".
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is relatively small compared to the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Continuity Health Solutions c/o Mr. Bryan Poteet CEO 9304 Forest Lane, Suite N272 Dallas. TX 75243
MAY 0 2010
K093895 Re:
Trade/Device Name: ALLYTM Platform System Regulatory Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: II (two) Product Code: 74 DRG Dated: April 13, 2010 Received: April 19, 2010
Dear Mr. Poteet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
{4}------------------------------------------------
Page 2 - Mr. Bryan Poteet
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): N/A
Device Name: ALLY™ Platform System
Intended Use:
The ALLY™ Platform System is a multi-platform medical device that is an accessory to approved 510k devices and is intended to be used as a means of receiving, storing, and transmitting patient biometric data including vital signs measurements. The ALLY™ Platform System is to be used upon prescription and under the supervision of an authorized healthcare provider.
The ALLY " Platform System will send information to a remote database to be accessed by authorized clients, such as healthcare professionals or patient family members. Electronic data consisting of patient vital sign information, educational material, reminders, questionnaires, and patient responses are exchanged over a secure network connection. Information may be communicated via standard telephone lines, high speed internet connections, serial port interfaces, or Bluetooth® radio transceivers.
The ALLY™ Platform is a multi-platform application that is not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts. The ALLY™ Platform is not intended to act as an emergency response system. All patient medical diagnosis and treatment are to be performed by an appropriate health care professional.
Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over the Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE)
W.M. Wood.
510(k) Number K0613095
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.