ANTHOGYR IMPULSION IMPLANTOLOGY CONTRA-ANGLE

K093894 · Anthogyr · EFA · Mar 3, 2010 · Dental

Device Facts

Record IDK093894
Device NameANTHOGYR IMPULSION IMPLANTOLOGY CONTRA-ANGLE
ApplicantAnthogyr
Product CodeEFA · Dental
Decision DateMar 3, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.4200
Device ClassClass 1

Indications for Use

ANTHOGYR's fully autoclavable contra-angles Implantology « Impulsion » are devices intended for a wide range of dental procedures including: Implant surgery such as perforating the bone, tapping and threading procedures.

Device Story

Impulsion range surgical contra-angles; used in dental implantology. Device inputs: mechanical power from dental motor. Device transforms input via geared angle mechanism to drive rotary instruments for bone perforation, tapping, and threading. Used in dental clinics by dentists/surgeons. Output: mechanical rotation of surgical burs/drills. Benefits: facilitates precise bone preparation for dental implants. Fully autoclavable design.

Clinical Evidence

Bench testing only. Device performance verified through conformance to recognized consensus standards including ISO 7785-2 (handpieces), ISO 3964 (coupling), ISO 7153-1 (stainless steel), and ISO 13402 (autoclaving/corrosion resistance).

Technological Characteristics

Stainless steel construction per ISO 7153-1. Geared angle handpiece design per ISO 7785-2. Coupling dimensions per ISO 3964. Fully autoclavable. Mechanical operation; no software or electronic components.

Indications for Use

Indicated for dental implant surgery, including bone perforation, tapping, and threading procedures. Intended for use by dental professionals.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAR - 3 2010 ## SPECIAL 510(k) - MODIFICATION TO K040674 ANTHOGYR DENTAL CONTRA-ANGLES # anthogyr ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS #### 1. GENERAL INFORMATION | Submitter | ANTHOGYR (Registration number 8020776) | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 2237, avenue André Lasquin | | | 74700 SALLANCHES FRANCE | | | Phone: 33(0)4 50 58 02 37 Fax: 33(0)4 50 93 78 60 | | | Web : www.anthogyr.com | | Contacts | Sabine BRAYETTE (QULITY INGINEER IN CHARGE OF<br>REGULATORY AFFAIRS) | | | sabine.brayette.prod@anthogyr.com | | Trade Names | ANTHOGYR « Impulsion » Implantology Contra-angles | | Legally marketed<br>predicate devices | Anthogyr Implantology Contra angles K040674 | | Classification Name | Dental handpiece and accessories | | Class | I | | Product Code | EFA | | CFR section | 872.4200 | | Intended Use | ANTHOGYR'S fully autoclavable contra-angles<br>Implantology « Impulsion » are devices intended for a wide<br>range of dental procedures including:<br>Implant surgery such as perforating the bone, tapping and<br>threading procedures | ### 2. INTENDED USE ANTHOGYR's fully autoclavable contra-angles Implantology « Impulsion » are devices intended for a wide range of dental procedures including: - Implant surgery such as perforating the bone, tapping and threading procedures. > {1}------------------------------------------------ SPECIAL 510(k) - MODIFICATION TO K040674 ANTHOGYR DENTAL CONTRA-ANGLES ## anthogyr #### 3. DEVICE DESCRIPTION ANTHOGYR has developed a full range of surgical contra angle intended to be used in implantology. The name of the range is « Impulsion » . ANTHOGYR Contra angles design, size and performance conform to NF EN ISO 7785-2 "Dental Handpieces -Part 2: Straight and geared angle handpieces". #### 4. PERFORMANCE DATA ANTHOGYR Contra angles & Handpieces conform to the following FDA recognized Consensus standards: - > ISO 14971 (2001) *Medical devices - Application of risk management to medical devices" (Recognition List Number: 005 Effective Date: 05/04/2001) - ISO 15223 (2000) « Medical devices Symbols to be used with medical device labels, labeling and information to be supplied » (Recognition List Number: 008 Effective Date: 10/29/2003) - ISO 13402 (2002) « Surgical and dental hand instruments Determination of resistance against autoclaving, corrosion and thermal exposure » (recognized Recognition List Number: 001 Effective Date: 02/19/1998) - V ISO 7785-2 (1995) "Dental Handpieces Part 2: Straight and geared angle handpieces" (Recognition List Number: 003 Effective Date: 05/03/1999) - > ISO 3964 (1982) "Dental Handpieces - Coupling dimensions" (Recognition List Number: 003 Effective Date: 05/03/1999) - > ISO 7153-1 (1999) « Surgical instruments - Metallic materials - Part 1 : stainless steel » (Recognition List Number: 006 Effective Date: 10/01/2001) In addition, ANTHOGYR Contra angles & Handpieces conform to the following standards: - I ISO 13485 (1996) *Medical devices Particular requirements for the application of the ISO 9001" - > NF EN ISO 1797-1 (1995) "Dental rotatory instruments - Shanks - Par 1: Shanks made of metal" {2}------------------------------------------------ ## SPECIAL 510(k) -- MODIFICATION TO K040674 anthogyr dental contra-angles ## anthogyr - NF EN ISO 17664 (2004) « Sterilization of medical devices Information to be > provided by the manufacturer for the processing of resterilizable medical devices » ## 5. SUBSTANTIAL EQUIVALENCE The ANTHOGYR « Impulsion » Implantology Contra-angles have the same fundamental scientific technology, operating principle and intended use as predicate devices. December 15, 2009 Summary preparation date: {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. MAR - 3 2010 Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Sabine Brayette Quality Engineer in Charge of Regulatory Affairs Anthogyr 2237 Avenue André Lasquin Sallanches FRANCE 74700 Re: K093894 Trade/Device Name: Anthogyr Contra Angles and Handpieces Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EFA Dated: January 20, 2010 Received: February 1, 2010 Dear Ms. Brayette: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Bravette Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I DA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Wh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ SPECIAL 510(k) - MODiFICATION TO K040674 ANTHOGYR DENTAL CONTRA-ANGLES thogyr ## Indications for Use 510(k) Number (if known): Device Name: ANTHOGYR CONTRA ANGLES AND HANDPIECES - / Indications for Use: ANTHOGYR's fully autoclavable contra-angles Implantology « Impulsion » are devices intended for a wide range of dental procedures including: > Implant surgery such as perforating the bone, tapping and threading procedures This range can be used with special accessories like depth stop. Prescription Use_ AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Concurrence of CDRH, Office of Device Evaluation (ODE) RSher DDS for Dr. K.P. Mulvey (Division Sign-Off) Page 1 of 1 Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: Page 15/48
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...