AST/GOT 4+1 SL, AST/GOT 4+1 SL, ELICAL 2, ELITROL I, ELITROL II, MODEL ASSL-0250, ASSL-0455, CALI-0580, CONT-0080, CONT-
Applicant
Seppim S.A.S.
Product Code
CIT · Clinical Chemistry
Decision Date
Sep 20, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1100
Device Class
Class 2
Indications for Use
The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) is a discrete photometric chemistry analyzer for in vitro diagnostic use. ELITech Clinical Systems AST/GOT 4+1 SL reagent is for the quantitative in vitro diagnostic determination of the activity of the enzyme Aspartate aminotransferase in human serum and plasma on the Vital Scientific Selectra/Flexor Analyzers. Aspartate Amino Transferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart disease. ELITech Clinical Systems Elical 2 is a single parameter calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL I is a single parameter control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer. ELITech Clinical Systems ELITROL II is a single parameter control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzers.
Device Story
The AST/GOT 4+1 SL is an in vitro diagnostic reagent kit used on the Vital Scientific Selectra/Flexor Junior benchtop discrete photometric chemistry analyzer. The system measures AST activity in human serum or plasma samples using a modified IFCC method (without pyridoxal-phosphate). The reagent consists of two parts: R1 (Tris buffer, L-Aspartate, MDH, LDH) and R2 (alpha-Ketoglutarate, NADH). The analyzer performs photometric measurements to determine enzyme activity, which is then used by clinicians to diagnose and monitor liver and heart conditions. The system requires calibration using ELICAL 2 and quality control using ELITROL I and II. The device is intended for professional use in clinical laboratory settings.
Clinical Evidence
Bench testing only. Performance evaluation included precision (within-run and total CVs), method comparison against the predicate (y=1.016x - 1.86, R^2=0.9998), and stability studies (refrigerated on-board stability of 28 days). No clinical prospective or retrospective studies were required.
Technological Characteristics
Reagent kit: R1 (Tris buffer, L-Aspartate, MDH, LDH, sodium azide) and R2 (alpha-Ketoglutarate, NADH, sodium azide). Photometric chemistry analyzer (discrete, benchtop). Liquid, ready-to-use reagents. Storage at 2-8°C. Calibration via ELICAL 2 (lyophilized human serum). Quality control via ELITROL I/II (lyophilized human serum).
Indications for Use
Indicated for the quantitative determination of Aspartate amino transferase (AST) in human serum and plasma to aid in the diagnosis and treatment of liver and heart diseases.
Regulatory Classification
Identification
An aspartate amino transferase (AST/SGOT) test system is a device intended to measure the activity of the enzyme aspartate amino transferase (AST) (also known as a serum glutamic oxaloacetic transferase or SGOT) in serum and plasma. Aspartate amino transferase measurements are used in the diagnosis and treatment of certain types of liver and heart disease.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
ABX PENTRA CP AST Reagent (k060318)
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s) (k033501)
Roche Diagnostics Precinorm U and Precipath U (k041227)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k093883
B. Purpose for Submission:
New Device: Analyzer, Assay, Control material and Calibrator material
C. Measurand:
Aspartate Aminotransferase (AST), Quality control and Calibrator material
D. Type of Test:
Quantitative enzymatic
E. Applicant:
Seppim S.A.S.
F. Proprietary and Established Names:
ELITech Clinical Systems AST/GOT 4+1 SL
ELITech Clinical Systems Elical 2
ELITech Clinical Systems Elitrol I and II
Vital Scientific Selectra Junior Analyzer and Vital Scientific Flexor Junior Analyzer
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1100: Aspartate aminotransferase (AST/SGOT) Test System
21 CFR 862.1150: Calibrator
21 CFR 862.1660: Quality Control material (assayed and unassayed)
{1}
21 CFR 862.2160: Discrete Photometric Chemistry Analyzer for Clinical Use
2. Classification:
Class II, Class II, Class I, reserved and Class I, respectively
3. Product code:
CIT, JJX, JIT, and JJE, respectively
4. Panel:
75 Clinical Chemistry
H. Intended Use:
1. Intended use(s):
See indications for use below
2. Indication(s) for use:
The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) is a discrete photometric chemistry analyzer for in vitro diagnostic use.
ELITech Clinical Systems AST/GOT 4+1 SL reagent is for the quantitative in vitro diagnostic determination of the activity of the enzyme Aspartate aminotransferase in human serum and plasma on the Vital Scientific Selectra/Flexor Analyzers. Aspartate Amino Transferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart disease.
ELITech Clinical Systems Elical 2 is a single parameter calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.
ELITech Clinical Systems ELITROL I is a single parameter control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzer.
ELITech Clinical Systems ELITROL II is a single parameter control serum for in vitro diagnostic use in accuracy control of quantitative ELITech Clinical Systems methods on the Vital Scientific Selectra Junior Analyzer and the Vital Scientific Flexor Junior Analyzers.
2
{2}
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Performance was provided for Vital Scientific Selectra Junior Analyzer which is also trademarked as the Vital Scientific Flexor Junior Analyzer.
I. Device Description:
ELITech Clinical Systems AST/GOT 4+1 SL reagent is available as a kit only. It consists of 2 reagents: Reagent 1 contains Tris buffer, L-Aspartate; Lactate dehydrogenase (LDH) (microorganisms), Malate dehydrogenase (MDH) (bacterial) and sodium azide. Reagent 2 contains α-Ketoglutarate, NADH and sodium azide.
ELITech Clinical Systems Elical 2 is a Lypholized calibrator based on human serum containing constituents to ensure optimal calibration.
ELITech Clinical Systems Elitrol I and II is a two level quality control product consisting of lypholized human serum containing constituents at desired levels.
All human source materials were tested with FDA-approved methods and found to be negative for HbsAG and to antibodies to HCV and HIV 1/2.
The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) is a discrete photometric chemistry analyzer for in vitro diagnostic use.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ABX PENTRA CP AST Reagent
Roche Diagnostics Calibrator for Automated Systems (C.f.a.s)
Roche Diagnostics Precinorm U and Precipath U
2. Predicate 510(k) number(s):
k060318, k033501, k041227
3. Comparison with predicate:
{3}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device
ELITech Clinical Systems AST/GOT 4+SL | Predicate
ABX Pentra
AST CP
Reagent
k060318 |
| Intended Use/Indications for Use | The quantitative in vitro diagnostic determination of the activity of the enzyme Aspartate amino transferase in human serum and plasma.
Aspartate Aminotransferase (AST) measurements are used in the diagnosis and treatment of certain types of liver and heart disease. | Same |
| Appearance of Reagents | Liquid Form; ready to use | Same |
| Test Method | Modified IFCC without pyridoxal-phosphatase | Optimized UV test according to IFCC modified method without pyridoxal phosphatase |
| Sample type | Serum, Plasma in lithium heparin | Same |
| Reagent Storage | Store at 2-8°C and protected from light. The reagents are stable until the expiration date stated on the label. | Reagents in unopened cassette, are stable up to expiry date on the label if stored at 2-8°C, and contamination is avoided |
| Instrument | The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) | ABX Pentra 400 Analyzer |
| Calibration Frequency | 28 days | 8 days |
{4}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device
ELITech Clinical Systems Elicit I | Predicate
Roche Calibrator
(C.f.a.s)
k033501 |
| Intended Use/Indications for Use | For in vitro diagnostic use in the calibration of quantitative method | Same |
| Format | Lyophilized calibrator based on human serum with constituents added as required to obtain desired components levels. | Same |
| Level | Single level | Same |
| Stability | Lyophilized: store at 2-8°C and protect from light until the expiry date
After reconstitution*: -8 hours between 15-25°C, 2 days between 2-8°C, 4 weeks between -25 and -15°C (when frozen once) | Same |
| Instrument | The Vital Scientific Selectra Junior Analyzer (also trademarked as the Vital Scientific Flexor Junior Analyzer) | Roche Analyzers |
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device
ELITech Clinical Systems Elitrol I/Elitrol II | Predicate
Roche Precinorm U/ Precipath U
k041227 |
| Intended Use/Indications for Use | For in vitro diagnostic use in accuracy control of quantitative method | Same |
| Format | Lyophilized human sera with constituents added as required to obtain defined components levels | Same |
| Levels | Two levels Level I (33-60 U/L), Level II (85-340 U/L) | Same |
| Stability | Lyophilized: Store at 2-8°C and protected from light until the expiry date. After Reconstitution*: 12 hours between 15-25°C, 5 days between 2-8°C, 4 weeks between -25 and -15°C (when frozen once) | Same |
{5}
6
K. Standard/Guidance Document Referenced (if applicable):
EP07-A2-Interference Testing in Clinical Chemistry-Second Edition
EP09-A2-Method comparison and bias estimate using patient samples-Second Edition
EP06-A-Evaluation of the linearity of the measurement of quantitative procedures: a statistical approach-First Edition
EP05-A2-Evaluation of precision performance of quantitative measurement methods; Approved guideline Second-Edition
L. Test Principle:
AST catalyzes the transfer of amino group from aspartate to oxoglutarate during the formation of glutamate and oxaloacetate. Oxaloacetate is reduced to malate by malate dehydrogenase (MHD). During this conversion, an equivalent amount of NADH is oxidized to NAD. The resulting decrease in absorbance at 340 nm is directly proportional to the activity of AST in serum.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
ELITech Clinical Systems AST/GOT 4+1 SL
Precision was evaluated with one quality control material and two adjusted human serum pool samples. A human serum pool diluted with saline (low), a control serum (medium), and a spiked human serum pool (high) on the Vital Scientific Selectra Junior Analyzer following CLSI EP5-A2 guideline. Samples were analyzed in duplicate, twice a day for 20 days. Results are summarized in the table below:
| Material | n | Mean (U/L) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | CV (%) | SD | CV (%) |
| Diluted Serum Pool (low) | 80 | 21.2 | 0.48 | 2.3% | 0.80 | 3.8% |
| Control Serum (medium) | 80 | 46.4 | 0.36 | 0.8% | 0.56 | 1.2% |
| Spiked Serum Pool (high) | 80 | 203.4 | 1.04 | 0.5% | 5.52 | 2.7% |
{6}
b. Linearity/assay reportable range:
# ELITech Clinical Systems AST/GOT 4+1 SL
Linearity across the assay range was determined by testing two pools of patient serum, one at a low concentration (9.2 U/L) and one at a high concentration (272 U/L). The low sample was prepared by diluting the serum sample pool with saline and the high sample was prepared by spiking the serum sample pool. All samples were assayed in triplicate. Data was analyzed using 1st, 2nd, and 3rd order least square regressions according to CLSI Protocol EP6-A. The difference in predicted values between the first and second order models are as follows:
1st order y = 26.27x - 17.37
2nd order y = 0.06x² + 25.51x - 15.72
The sponsor chose the 2nd order regression because it is significant and the results are shown in the table below:
| Dilution | Mean | Predicted 1st order | Predicted 2nd order | Difference 2nd-1st | % Difference |
| --- | --- | --- | --- | --- | --- |
| 1 | 9.2 | 8.9 | 9.8 | 1.0 | & |
| 2 | 36.4 | 35.2 | 35.5 | 0.4 | 1.1% |
| 3 | 60.4 | 61.4 | 61.4 | -0.1 | -0.1% |
| 4 | 89.2 | 87.7 | 87.3 | -0.4 | -0.4% |
| 5 | 113.2 | 114.0 | 113.4 | -0.6 | -0.5% |
| 6 | 139.1 | 140.2 | 139.6 | -0.6 | -0.5% |
| 7 | 165.9 | 166.5 | 165.9 | -0.6 | -0.3% |
| 8 | 192.1 | 192.8 | 192.4 | -0.4 | -0.2% |
| 9 | 218.0 | 219.0 | 219.0 | -0.1 | 0.0% |
| 10 | 246.6 | 245.3 | 245.7 | 0.4 | 0.2% |
Based on the data, the sponsor claimed that the assay's linearity range is 10 to 250 U/L.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Elical 2, Elitrol I and II are traceable to IFCC formulation (Schulman, 2002).**
Elitrol I and II
**Elical 2, Elitrol I & II are value assigned using two Vital Scientific Flexor Junior Analyzers.** Each sample is tested in triplicate over several days. The target value of Level I and II are the median of the observed values range. After validation of the target value, a confidence range (high and low values) is then calculated.
{7}
8
# Elical 2
The calibrator is tested against predetermined intervals on multiple Vital Scientific Selectra Junior using the AST reagent and 2 levels of quality control material. The mean analyte value is calculated and a target value is assigned.
The ELITech Clinical Systems AST/GOT 4+1 SL is stable until the expiration date printed on the label when stored at 2-8°C. The on-board stability of the reagent is 28 days.
Before reconstitution, the shelf-life of the ELITech Clinical Systems Elitrol 1 and Elitrol II is 30 months at 2-8°C. After reconstitution the stability is 12 hours when stored at 15-25°C, 5 days when stored at 2-8°C and 4 weeks (when frozen once) at -25° and -15°C.
Elical 2 is stable until the expiration date printed on the label when stored at 2-8°C prior to reconstitution. After reconstitution the stability is 8 hours when stored at 15-25°C, 2 days at 2-8°C and 4 weeks (when frozen once) at -25° and -15°C. The labeling stated that the Elical 2 should be stored tightly capped and protected from light when not in use.
d. Detection limit:
## AST/GOT 4+1 SL
A detection limit study was evaluated according to CLSI EP17-A guideline. LoB was determined by running a blank sample 60 times on the Vital Scientific Selectra Junior analyzer. The limit of detection (LoD) was determined by assaying 4 low (3.6 U/L) diluted sample pools 60 times on the Selectra Junior analyzer. Limit of Quantitation (LoQ) was determined by running 4 low (5 U/L) diluted sample pools 60 times on the Vital Scientific Selectra Junior analyzer.
| | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| Limits | 0.9 U/L | 1.8 U/L | 5.0 U/L |
The claimed measuring range is 10-250 U/L.
e. Analytical specificity:
## AST/GOT 4+1 SL
Testing for interfering substances was based on CLSI EP-7A. Testing was performed on a minimum of five concentrations for each interfering substances. Samples with increasing amounts of triglycerides (Intralipid®), Bilirubin, Ditaunobilirubin, Sodium Pyruvate, and Ascorbic Acid were tested in triplicate and compared to the same sample without the interferent.
{8}
The sponsor defined non-significant interference as the highest level tested that does not cause >10% change between the tested samples and the control sample.
| | Highest concentration tested that showing no significant interference (mg/dL) (≤10%) |
| --- | --- |
| Triglyceride | 614 |
| Unconjugated Bilirubin | 30.0 |
| Conjugated Bilirubin | 29.5 |
| Sodium Pyruvate | 2.0 |
| Ascorbic Acid | 20.0 |
The labeling states that hemolyzed samples should not be used since significant hemolysis may increase AST concentration because of high levels of AST in erythrocytes.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed using the HORIBA ABX AST CP reagent on a PENTRA 400 chemistry analyzer and the ELITech AST/GOT 4+1 SL reagent on the Selectra Junior analyzer according to CLSI protocol EP9-A2. A total of 70 serum samples covering the linearity of the AST/GOT 4+1 SL assay were assayed. Of these 70 samples, 61 were native samples, and 9 were spiked samples (sample range from 15.5-244.8 U/L). The comparison resulted in a slope of 1.016, an intercept of -1.86, correlation coefficient of r²=0.9998. The standard error of the estimate is equal to 3.6 U/L
b. Matrix comparison:
Thirty paired serum and plasma (lithium heparin) samples, ranging from 16.8 to 216 U/L, were tested on the Selectra Junior analyzer. 27 were native patient samples and 3 were spiked samples. The comparison resulted in a slope of 0.992, an intercept of +0.58, correlation coefficient of r² = 0.997. The standard error of the estimate is equal to 4.2 U/L.
Based on the data, the sponsor claims that lithium heparin is an acceptable anti-coagulant for this assay.
{9}
10
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Reference range: ¹Serum, plasma (37°C): < 40 U/L
1: Tietz:N.W. Clinical Guide to Laboratory tests ³rd Ed., (WB Saunders eds. Philadelphia USA), (1995), 76.
In the labeling, the sponsor also recommends that each laboratory should establish and maintain its own reference values.
N. Instrument Name:
Vital Scientific Selectra Junior Analyzer also trademarked as the Flexor Junior Analyzer
O. System Descriptions:
1. Modes of Operation:
Continuous loading on a sample by sample basis
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
{10}
Barcoding or manual entry
4. Specimen Sampling and Handling:
Samples in sample tubes or sample cups with a tube adaptor are placed into sample tray. Specimens, calibrators and controls are pipetted by a pipettor arm into a cuvette rotor. A cuvette blank must be conducted every 24 hours. The cuvette rotor must be replaced every 10,000 tests or when the cuvette blank exceeds the maximum allowable average.
5. Calibration:
Onboard automatic and on demand calibration
6. Quality Control:
The software contains a quality control program that evaluates control results and determines if they are within specified acceptable limits.
The labeling recommends that control measurements be assayed daily and the frequency depends on the test method, as well as the legislative regulations valid in each country and in the organization that oversees the laboratory.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.