DIGITAL RADIOGRAPHY SYSTEM, MODELS GALAXY 900 AND 1600
K093816 · Medien International Co, Ltd. · KPR · Oct 5, 2010 · Radiology
Device Facts
Record ID
K093816
Device Name
DIGITAL RADIOGRAPHY SYSTEM, MODELS GALAXY 900 AND 1600
Applicant
Medien International Co, Ltd.
Product Code
KPR · Radiology
Decision Date
Oct 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Galaxy 900 is integrated into the user's stationary radiography system. This typical configuration permits a qualified/trained doctor or technologist to take a range of head-to-toe diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions in an integrated system. The Galaxy 900 is not intended for mammography.
Device Story
Galaxy 900/1600 are optical-based digital X-ray imagers. System integrates with an operating PC; visible light is deflected by a mirror to a high-resolution CCD camera to produce digital images. BLADE software acquires, processes, saves, and manages images. Operated by qualified doctors or technologists in clinical settings. Provides digital X-ray images for diagnostic radiography; replaces film/screen-based systems. Benefits include wide dynamic range, 16-bit digitization, and selectable image processing parameters tailored to specific body parts to optimize diagnostic quality.
Clinical Evidence
No clinical data provided. Safety and effectiveness established through bench testing, including electrical, mechanical, and environmental safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-1).
Technological Characteristics
Optical-based digital X-ray imager using a Charge Coupled Device (CCD) and lens. 17"x17" imaging area; 16-bit digitization. Galaxy 900: 3,056 x 3,056 format, 3.5 lp/mm. Galaxy 1600: 4,096 x 4,096 format, 4.6 lp/mm. DICOM 3.0 compliant. Operates on MS Windows XP. Sterilization not applicable.
Indications for Use
Indicated for general diagnostic radiographic exposures of skull, spine, chest, abdomen, and extremities in adult and pediatric patients. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# Medien International Co., Ltd.
OCT 5 2010
First of the start of the state the state of the many of the may be the state of the states of the the states of the states of
## 510(k) Summary
140938/6
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: _May. 20, 2009
1. Company and Correspondent making the submission:
Name - Medien International Co., Ltd.
Address - 1-40, Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si,
Gyeonggi-do, 435-862, Republic of Korea
Telephone - +82-31-451-9466
Fax - +82-31-451-9467
Contact - Jae-Hyun, Lee / RA Manager
E-mail – jhlee@medien.co.kr
- 2. Device :
| Trade/proprietary name | : Galaxy 900 |
|------------------------|------------------------------|
| Common Name | : Digital Radiography System |
| Classification Name | : Stationary X-ray System |
- 3. Predicate Devices :
| Manufacturer | : Imaging Dynamics Co., Ltd. |
|---------------|----------------------------------------|
| Device | : Xaminer |
| 510(k) Number | : K061595(Decision Date - 8. 17. 2006) |
- 4. Classifications Names & Citations :
21CFR 892.1680, KPR - Stationary X-ray System, Class 2
5. Description :
5.1 General
1-40, Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea
{1}------------------------------------------------
# Medien International Co., L
The Digital Radiography System, Galaxy is an optical based digital x-ray imager. This device should be integrated with an operating PC. The visible light is deflected by a mirror to a high-resolution CCD camera that produces a digital image. It can do to utilize as digitalizing x-ray images and transfer for radiography diagnostic.
The BLADE software acquires X-ray image and viewing the image.
Get image from detector, process it to ease the diagnostic, save it in database and manage it.
### 5.3 Product features
- One Charge Coupled Device (CCD) armed with one lens.
- 17" x17" imaging area.
- Wide dynamic range with 16-bit digitization
- SW is designed to be operated on MS Windows XP operating system
- Image process parameters are selectable according to the body part to make best images for diagnosis
- DICOM3.0 standard compliance
- Image Format : 3,056 x 3,056
- High Resolution image : with 3.5 lp/mm
#### 6. Indication for use :
The Galaxy 900 is integrated into the user's stationary radiography system. This typical configuration permits a qualified/trained doctor or technologist to take a range of head-to-toe diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions in an integrated system.
The Galaxy 900 is not intended for mammography.
- 7. Comparison with predicate device :
Medien International Co., Ltd. believes that the Digital Radiography System (Galaxy 900) is substantially equivalent to to Xaminer of Imaging Dynamics Co., Ltd..
ירי וגבער האמרים מעשים שניי
8. Safety, EMC, Biocompatibility and Performance Data :
<sup>1-40,</sup> Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea
{2}------------------------------------------------
# Medien International Co., L
Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 was performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2(2001). Biocompatibility testing was conducted in accordance with Standard ISO 10993-1.
Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.
- 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Medien International Co., Ltd. concludes that the Digital Radiography System(Galaxy 900) is safe and effective and substantially equivalent to predicate devices as described herein.
- 10. Medien International Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
1-40, Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea
100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
Page 23 of 926
{3}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: May. 20, 2009
9. Company and Correspondent making the submission:
Name - Medien International Co., Ltd.
Address - 1-40, Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si,
Gyeonggi-do, 435-862, Republic of Korea
Telephone - +82-31-451-9466
Fax - +82-31-451-9467
Contact - Jae-Hyun, Lee / RA Manager
E-mail - jhlee@medien.co.kr
- 10. Device :
| Trade/proprietary name | : Galaxy 1600 |
|------------------------|------------------------------|
| Common Name | : Digital Radiography System |
| Classification Name | : Stationary X-ray System |
- Predicate Devices : 11. : Imaging Dynamics Co., Ltd. Manufacturer : Xplorer 1000 Device : K992955 (Decision Date - 2. 6. 2000) 510(k) Number .
Classifications Names & Citations : 12. 21CFR 892.1680, KPR - Stationary X-ray System, Class 2
13. Description :
5.1 General
Controller Comments of the comments of the comments of
1-40, Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea
יי יראש משמשתמש
Page 24 of 926
{4}------------------------------------------------
# Medien International Co., L
The Digital Radiography System, Galaxy is an optical based digital x-ray imager. This device should be integrated with an operating PC. The visible light is deflected by a mirror to a high resolution CCD camera that produces a digital image. It can do to utilize as digitalizing x-ray images and transfer for radiography diagnostic.
The BLADE software acquires X-ray image and viewing the image.
Get image from detector, process it to ease the diagnostic, save it in database and manage it.
## 5.3 Product features
- One Charge Coupled Device (CCD) armed with one lens.
- 17" x17" imaging area.
- Wide dynamic range with 16-bit digitization
- SW is designed to be operated on MS Windows XP operating system
- Image process parameters are selectable according to the body part to make best images for diagnosis
- DICOM3.0 standard compliance
- Image Format : 4,096 x 4,096
- High Resolution image : with 4.6 lp/mm
#### Indication for use : 14.
The Galaxy 1600 Digital CCD X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. The Galaxy 1600 is not intended for mammography.
- Comparison with predicate device : 15. Medien International Co., Ltd. believes that the Digital Radiography System (Galaxy 1600) is substantially equivalent to to Xaminer of Imaging Dynamics Co., Ltd. and Xplorer 1000 of Imaging Dynamics Co., Ltd.,
- Safety, EMC, Biocompatibility and Performance Data : 16. Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1 was performed, and EMC testing was conducted in accordance
<sup>1-40,</sup> Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea
{5}------------------------------------------------
with standard IEC 60601-1-2(2001).
Biocompatibility testing was conducted in accordance with Standard ISO 10993-1. Non-clinical & Clinical considerations according to FDA Guidance "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.
- 9. Conclusions :
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Medien International Co., Ltd. concludes that the Digital Radiography System(Galaxy 1600) is safe and effective and substantially equivalent to predicate devices as described herein.
- 10. Medien International Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
END
1-40, Hanlim Human Tower #501, Geumjeong-dong, Gunpo-si, Gyeonggi-do, 435-862, Korea
: 12 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 1
Page 26 of 926
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol, possibly representing human figures or abstract shapes, and the overall design is in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
Mr. Charles Mack Principal Engineer Medien International Co., Ltd. 77325 Joyce Way ECHO OR 97826
OCT 5 2010
Re: K093816
Trade/Device Name: Digital Radiography System/Galaxy 900 and Galaxy 1600 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: September 13, 2010 Received: September, 2010
#### Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{7}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use
510(k) Number(if known):
KOORZIO
Device Name: Digital Radiography System/ Galaxy 900
Indications for Use:
The Galaxy 900 is integrated into the user's stationary radiography system. This typical configuration permits a qualified/trained doctor or technologist to take a range of head-to-toe diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts on both adult and pediatric patients. Applications can be performed with patient sitting, standing or lying in the prone or supine positions in an integrated system.
The Galaxy 900 is not intended for mammography.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation(ODE)
NARR
(Division Sign-Off) Division of Radiological De Office of In Vitro Diagnostic Device Evaluation
K093816
510K
Page 1 of
Page 19 of 926
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.